Page 31
•
Low angle, one direction (30S), Light is aimed at the part at an angle of 30 ±3° from one side. The light may be incident from any of the four possible orientations. The lighting plane is aligned to be parallel to the line formed by one edge of the image sensor to within ±5°. The lighting illuminates the entire symbol area with uniform energy.
30° (+/-3) from 1 side (North, East, South, West)
•
Medium angle, four direction (45Q), Light is aimed at the part at an angle of 45 ±3° from the plane of the surface of the symbol from four sides such that the lines describing the center of the beams from opposing pairs of lights are co-planar and the planes at right angles to each other. One lighting plane is aligned to be parallel to the line formed by a horizontal edge of the image sensor to within ±5°. The lighting illuminates the entire symbol area with uniform energy.
45° (+/-3) from all 4 sides (North, East, South, West)
Positioning Data Matrix Mark for Verification Always ensure the bottom plane of the light in UID-LDP and UID-DPM is parallel to the plane of the Data Matrix symbol area. Use Live Video to ensure the Data Matrix mark is centered in the FOV with the correct orientation.
Light Control The light in UID-LDP and UID-DPM will be switched on when the UIDChecker™ program starts up and switched off when the program exits. The state of the multifunction light is saved so that the program, when restarted, will use the same light configuration as in the previous operation. The default light configuration is 90 for the UID-DPM.
v2.3.1, Dec 2008
Getting Started With Your UID Compliance Verifier
1-17
1 Getting Started w/UID Compliance Verifier
Lighting for the UID-LDP & UID-DPM
Page 32
Chapter
1
Getting Started w/UID Compliance Verifier
Triggers In UID-LDP and UID-DPM, the verification can be triggered from: •
A hardware trigger button
•
The Read UID Mark software button on the UIDChecker™ GUI
Note: The hardware trigger is disabled for the verification type MIL-STD-130L Change 1 (ISO 15415) in order to support Read and Rotate. See Appendix B, “Read And Rotate,” for details.
1-18
Getting Started With Your UID Compliance Verifier v2.3.1, Dec 2008
Page 33
2
CHAPTER 2
Quickly Using UID Compliance Verifier
2 Quickly Using UID Compliance Verifier
After loading the UIDChecker™ application software on your PC and powering up the UID Compliance Verifier, you are now ready to begin using your UID Compliance Verifier to read UID marks.
Using Your UIDChecker™ Software To launch the UIDChecker™ application, select and click on the UIDChecker™ desktop icon, as shown in Figure 2–1. FIGURE 2–1. UIDChecker™ Icon
First Time Startup The first time you launch the application, the UIDChecker software will prompt you for a Reader on the network to connect to. After that, it will attempt to reconnect to the last reader each time it is launched, as shown in Figure 2–2.
v2.3.1, Dec 2008
Getting Started With Your UID Compliance Verifier
2-1
Page 34
Chapter
2
Quickly Using UID Compliance Verifier
FIGURE 2–2. Find a Reader Menu Item
This option displays a list of readers on your network. If you have been successful in setting up the network connection to the camera, you will find the reader on the list. The name should begin with “UID”. Select your reader from the drop down list. Note: On some networks, there may be a delay of several minutes when you power on and plug the Ethernet cable into your network. If no readers show up in the drop down list (see Figure 2–3), then close the UIDChecker™ software, wait five minutes, and launch the program again. FIGURE 2–3. Drop Down List of Readers
2-2
Getting Started With Your UID Compliance Verifier v2.3.1, Dec 2008
Page 35
Testing the UID Compliance Verifier
FIGURE 2–4. Command Buttons Active
Subsequent Startups As stated above, upon first-time startup, the UIDChecker will prompt for a reader on the network to connect to. Upon subsequent restarts of UIDChecker™, the UIDChecker™ will remember its reader and automatically select it, and the UID Verifier will be ready to run.
Testing the UID Compliance Verifier After completing all the hardware setup and software loading, you will want to perform a quick check to ensure that your UID-LDP or UID-DPM verifier is working properly. You will need the Microscan Calibration Test Card (Microscan part number 98-UA10-0CC0) and the UID Compliance Card (Microscan part number A4-20814-1A) to test your UID-LDP or UID-DPM.
Perform Reflectance Calibration Note: Please refer to Reflectance Calibration in Chapter 3 for more detail. Once you have positioned your UID-LDP or UID-DPM into your work space, you need to calibrate the system using this Calibration Target. 1.
v2.3.1, Dec 2008
Ensure UIDChecker™ has connected to your verifier. If you have never calibrated the verifier from the UIDChecker™, you will see that Last Calibrated: UNKNOWN displayed in Figure 2–5:
Getting Started With Your UID Compliance Verifier
2-3
2 Quickly Using UID Compliance Verifier
Once this has been done, the system will remember the reader and will connect to it automatically. The command buttons will become active, as shown in Figure 2–4:
Page 36
Chapter
2
Quickly Using UID Compliance Verifier
FIGURE 2–5. Last Calibrated: UNKNOWN
2.
The Reflectance Calibration box should be displayed automatically. If it is not, click Reader > Reflectance Calibrate to bring up the Calibration dialog box, as shown in Figure 2–6.
FIGURE 2–6. Reflectance Calibration Dialog Box
3.
In the Calibration Target Contrast text box (see Figure 2–6), enter the Contrast value shown on the card associated with the Data Matrix (80% contrast) you are using as the target. For example, if the contrast marked for the Data Matrix on the Calibration Test Card is 82.6%, you would enter 82.6 or 83.
4.
In the Calibration Target Reflectance Max text box (see Figure 2–6), enter the Rmax value shown on the card associated with the Data Matrix (80% contrast) you are using as the target. For example, if the R_max marked for the Data Matrix on the Calibration Test Card is 87.4%, you would enter 87.4 or 87. Live Video mode is enabled.
5.
2-4
Position the Data Matrix (around 80% contrast) under the UID-LDP or UIDDPM. Center the Data Matrix in the camera's FOV, as shown in Figure 2–7.
Getting Started With Your UID Compliance Verifier v2.3.1, Dec 2008
Page 37
Testing the UID Compliance Verifier
FIGURE 2–7. Data Matrix Centered in FOV
Quickly Using UID Compliance Verifier
2
6.
Click Calibrate on the Reflectance Calibration dialog box (Figure 2–6). The UIDChecker™ will calibrate the verifier in less than 10 seconds, or post error messages if it can’t. All calibration data are saved to the verifier. Click Close to exit the dialog box. The top border of the UIDChecker™ application will display the last calibration date. Also, the calibration date will be posted in the Results display.
Test the Correct Calibration v2.3.1, Dec 2008
1.
Launch the UIDChecker™ application if it is not already running.
2.
Select Reader > Verification Type > MIL-STD-130N (ISO 15415). Note that the verification type MIL-STD-130N (ISO 15415) is the easiest one to test the correct calibration.
3.
Select Results Setup > Show Verification Details in On-Screen Results View.
Getting Started With Your UID Compliance Verifier
2-5
Page 38
Chapter
2-6
2
Quickly Using UID Compliance Verifier
4.
Click Live Video. Position the higher contrast (about 80%) Data Matrix on the Microscan Calibration Test Card under the UID Verifier. Center the Data Matrix on the screen with correct orientation, as shown in Figure 2–8. (For UID-LDP, the laser lines are at the middle of the FOV). Note that the Data Matrix can be rotated but one of the “L” borders must be parallel to one of the lines in the video overlay.
Getting Started With Your UID Compliance Verifier v2.3.1, Dec 2008
Page 39
Testing the UID Compliance Verifier
FIGURE 2–8. Data Matrix in Center with Correct Orientation
Quickly Using UID Compliance Verifier
2
5.
v2.3.1, Dec 2008
When you are finished positioning the target, click Live Video again (to exit Live Video).
Getting Started With Your UID Compliance Verifier
2-7
Page 40
Chapter
2
Quickly Using UID Compliance Verifier
6.
Click Read UID Mark to verify the correct report for contrast and cell size. The contrast should match the value (with tolerance of +/- 2%) shown on the Calibration Test Card and the cell size should be reported as 0.015 +/- 0.0001 inches, as shown in Figure 2–9.
FIGURE 2–9. Test Calibration Using Contrast & Cell Size
Test the Correct Operation Note: Cell Contrast is defined differently than the normal contrast value used in ISO/IEC 15415 and will not match the value shown on the calibration card. 1.
2-8
Select Reader > Verification Type > MIL-STD-130N (AIM DPM-12006). Ensure that Report is set to First Passing Overall Grade. Click OK.
Getting Started With Your UID Compliance Verifier v2.3.1, Dec 2008