Getting Started With Your UID Compliance Verifier, pages 101 to 110

Pages 101 to 110 of 148

The original document is the PDF (2.9 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Siemens UID LDP

Page 101

B Read And Rotate

APPENDIX B

The Read And Rotate button is visible when you select the MIL-STD-130L Change 1 (ISO/IEC 15415) Verification Type. You use Read And Rotate to test for rotational consistency of measurements. Read And Rotate allows you to position your UID Code and read/rotate it five times at 72° intervals. Use Live Video to position a mark in the FOV of the reader. Notes: The Data Matrix must be completely within the screen for a successful read.

B Read And Rotate

The hardware trigger button is disabled to support Read And Rotate.

Read And Rotate Usage Details 1.

v2.3.1, Dec 2008

Specify the Verification Type to be ISO/IEC 15415 in the Verification dialog box, as shown Figure B–1.

Getting Started With Your UID Compliance Verifier

B-1

Page 102

Appendix

B

Read And Rotate

FIGURE B–1.

B-2

Setting Verification Type to Invoke Read And Rotate

2.

Click OK. The Read And Rotate button will appear on the main interface.

3.

Click Read And Rotate. Notice the following: –

The Live Video button is activated.

–

The Read And Rotate button becomes the Cancel Read And Rotate button.

–

The chevrons displayed on the main Read And Rotate window (Figure B–2).

Getting Started With Your UID Compliance Verifier v2.3.1, Dec 2008

Page 103

Read And Rotate Usage Details

FIGURE B–2. Initial Read And Rotate Screen

Align the Data Matrix solid edges to the chevron angle. This works best when you carefully line up the “L” of the Data Matrix as close to the onscreen “Chevron Angle,” as shown in Figure B–3.

B Read And Rotate

4.

v2.3.1, Dec 2008

Getting Started With Your UID Compliance Verifier

B-3

Page 104

Appendix

B

Read And Rotate

FIGURE B–3. Read And Rotate First Read

5.

After you have the Data Matrix in position, click Live Video. If the Data Matrix is aligned correctly, the unit beeps and the chevrons are rotated 72°.

6.

Align the Data Matrix solid edges to the new chevron angle.

7.

After you have the Data Matrix in position, click Live Video.

8.

Continue this process three more times, aligning the Data Matrix solid edges to the new chevron angle. When you have aligned the Data Matrix a total of five times, the Live Video button is no longer active (depressed), and a screen similar to the one in Figure B–4 is displayed. Note: To cancel out of Read And Rotate, click the Cancel Read And Rotate button.

B-4

Getting Started With Your UID Compliance Verifier v2.3.1, Dec 2008

Page 105

Read And Rotate Usage Details

FIGURE B–4. Read And Rotate Last Read

Save your Results if you have not already done so.

10.

Click on View Results to see a Summary Report, if you want to.

B Read And Rotate

9.

v2.3.1, Dec 2008

Getting Started With Your UID Compliance Verifier

B-5

Page 106

Appendix

B

Read And Rotate

FIGURE B–5. Read And Rotate Summary Report

Clicking on any of the file names in the table will take you to the individual read results for the five successful read and verifications.

B-6

Getting Started With Your UID Compliance Verifier v2.3.1, Dec 2008

Page 107

Attaching UID-DPM to UID-DPM Stand

C

C Attaching UID-DPM to UID-DPM Stand

APPENDIX C

Setting Up the UID-DPM Stand FIGURE C–1. Dowel Pins & Thumb Screw

Dowel Pins

Thumb Screw v2.3.1, Dec 2008

1.

Remove the stand base and upright arm from the packaging.

2.

Place the base on a level flat surface.

3.

Slide the upright arm onto the two dowel pins in the back of the base.

Getting Started With Your UID Compliance Verifier

C-1

Page 108

Appendix

C

Attaching UID-DPM to UID-DPM Stand

4.

Fasten the upright arm using the large thumb-screw provided with the kit.

5.

Be careful while moving the upright arm. The camera mounting head is spring-loaded and can move while carrying the arm.

Attaching the UID-DPM Use the following procedure to attach the UID-DPM to the UID-DPM stand: 1.

Rotate the arm so that it is in the position shown in Figure C–2.

FIGURE C–2. Correct Position of Arm

2.

C-2

Move the bolt on the arm to the correct position before you try to attach the UID-DPM (see Figure C–3).

Getting Started With Your UID Compliance Verifier v2.3.1, Dec 2008

Page 109

FIGURE C–3. Move Bolt to Correct Position

Move the bolt to this position before you attach the UID-DPM

3.

Attach the UID-DPM, making sure that all four feet fully contact the base of the UID-DPM stand.

4.

Using one foot of the UID-DPM, set up the correct stand-off distance for the direct part mark you want to verify, as shown in Figure C–4.

FIGURE C–4. Setting Up Correct Stand-Off Distance

5.

v2.3.1, Dec 2008

Slide the part into the FOV, as shown in Figure C–5.

Getting Started With Your UID Compliance Verifier

C Attaching UID-DPM to UID-DPM Stand

Attaching the UID-DPM

C-3

Page 110

Appendix

C

Attaching UID-DPM to UID-DPM Stand

FIGURE C–5. Part in the Field of View

6.

Center the Data Matrix mark using Live Video, as shown in Figure C–6.

FIGURE C–6. Centering Data Matrix Using Live Video

C-4

Getting Started With Your UID Compliance Verifier v2.3.1, Dec 2008