Page 41

Operation

Operating Instructions

Chapter 7

OTS400

7.3.4

Viewing and editing operating data

This function allows you to interrogate and reset statistical operating data that the OTS400/OTC400 logs in the form of counters during the reading operation. The OTS400/ OTC400 does not output reading results in this mode. 1. Click the "operating data" icon (“jagged line”) in the toolbar. The Operating Data dialog box appears. The OTS400/ OTC400 interrupts Reading mode. 2. After reading/resetting the desired counters, confirm the dialog box with "OK". The OTS400/ OTC400 resumes Reading mode. 7.3.5

Displaying Adjusting mode and reading angle limits

These two functions are not directly available in the OTS400/ OTC400 but in the bar code scanners connected to the OPS. A detailed description is provided in Chapter 6 Operation of the CLV490 Operating Instructions. 7.3.6

Background teach-in

This function is not directly available in the OTS400/ OTC400 but in the bar code scanners connected to the OPS. A detailed description is provided in Chapter 6 Operation of the CLV490 Operating Instructions. 7.3.7

Reading diagnosis

Function of the auxiliary interface (default). In this mode, the OTS400/ OTC400 outputs the data contents of all read bar codes with the associated reading diagnosis data, including the data recorded incorrectly according to the selection criteria (errored). The number of bar codes output, therefore, may be greater than the number of bar codes that are sent in the reading result on the host interface. In the default setting, the OTS400/ OTC400 does not output any reading diagnosis data on the host interface. The reading diagnosis can be activated on the "Auxiliary Interface" tab in the CLV Setup program as follows: ¾

Choose the "Read Diagnostics" option in the "Auxiliary Interface" list box.

Download to the OTS400/ OTC400: 1. Click the "page to bar code scanner" icon in the toolbar. The CLV Setup downloads the parameter set to the OTS400/ OTC400. The "Download Parameters" dialog box appears with the storage options. 2. Click the desired storage option. The dialog box closes automatically. The terminal interface of the OTS400/ OTC400 is now working in "Reading diagnosis" mode. 7.3.8

Monitoring the host interface

Function of the auxiliary interface. In this mode, the OTS400/ OTC400 outputs the data traffic on its host interface via the auxiliary interface. Repetition requests from the protocol driver and protocol-specific data, such as start and stop characters, are suppressed. Each data string is represented on a separate line. Identification of the data direction: O data string ... = I data string ... =

OTS400/ OTC400 sending to the host (O = Output) OTS400/ OTC400 receiving from the host (I = Input)

The function for monitoring the host interface can be activated on the "Auxiliary Interface" tab in the CLV Setup program as follows: ¾

Choose the "Monitor Host Interface" option from the "Auxiliary Interface" drop-down list.

8008869/RA61/2007-07-30

© SICK AG · Auto Ident · Germany · All rights reserved

7-5

Page 42

Chapter 7

Operation

Operating Instructions OTS400 Omni Tracking System

Download to the OTS400/ OTC400: 1. Click the "page to bar code scanner" icon in the toolbar. The CLV Setup downloads the parameter set to the OTS400/ OTC400. The "Download Parameters" dialog box appears with the storage options. 2. Click the desired storage option. The dialog box closes automatically. The auxiliary interface of the OTS400/ OTC400 is now working in "Monitor Host Interface" mode. Displaying data in the Terminal Emulator (for test purposes): 1. Choose the serial interface as a trigger source for the reading pulse. See Section 7.3.1 Reading mode. 2. Click the "Terminal” icon in the toolbar. The Terminal Emulator window appears. The bar code scanner is in Reading mode. 3. Click the "Start SW Trigger" button. 4. Place the bar code in front of any bar code scanner. 5. Click the "Stop SW Trigger button. The OTS400/ OTC400 outputs the result of the host interface in the main Terminal Emulator window. In the default setting, the result does not contain any reading diagnosis data. This function requires the incremental signal. The OTS400/ OTC400 outputs the reading result in the main window of the Terminal Emulator. As a prerequisite, the OTS400/ OTC400 must be able to assign the read bar code uniquely to one object, i.e. the bar code is located on the object detected by the trigger source of the reading pulse generator. If the data traffic on the host interface is very fast or very high, it may not be displayed correctly. This is due to the slower transmission speed of the auxiliary interface, with the result that missing information appears as ".........". 7.3.9

Auxiliary input

Function of the auxiliary interface. In this mode, the OTS400/ OTC400 accepts a bar code entered manually on the auxiliary interface (via keyboard or handheld scanner with decoder). It sends the bar code to the host in a separate data string on the host interface. This function can be used to correct no reads, for example, by transferring missing bar codes subsequently. 7.3.10 Self-test During the self-test, the OTS400/OTC400 checks whether its hardware components are functioning properly and verifies the number and error status of the connected bar code scanners. A concluding message output on the auxiliary interface provides information on the test result. Each time it is switched on, the device runs a self-test before it is initialized with the valid parameter set. The test can be called up explicitly at any time. The CLV does not output any reading results during the test routine. 1. Click the "Terminal" icon in the toolbar. The Terminal Emulator window appears. The OTS400/ OTC400 is in Reading mode. 2. Click the "Self-Test" option field. The OTS400/ OTC400 terminates Reading mode and starts the test routine. After a few seconds, the OTS400/ OTC400 outputs the test result in the form of a code number. 3. To return to Reading mode, click the "Reading mode" option or close the Terminal Emulator. The OTS400/ OTC400 resumes Reading mode.

7-6

© SICK AG · Auto Ident · Germany · All rights reserved

8008869/RA61/2007-07-30

Page 43

Operating Instructions

Operation

Chapter 7

OTS400

The code number "15000" means that the test has been concluded successfully and no errors were diagnosed. Section 10.2 Error messages contains a list of error codes and associated remedies. 7.3.11 Standard OTS400 statistics (single read) Function of the auxiliary interface. In this mode, the OTS400/ OTC400 outputs an overview of the information collected for each reading pulse via the auxiliary interface. The reading interval lengths and durations, as well as velocity are output. Furthermore, the code type, length, and contents, reading rate, and code reliability are output for each code read. The standard OTS400 statistics (single read) can be activated on the "Auxiliary Interface" tab in the CLV Setup program as follows: ¾

Choose the "Standard OTS Statistics (single read)" option from the "Auxiliary Interface" drop-down list.

Download to the OTS400/ OTC400: 1. Click the "page to bar code scanner" icon in the toolbar. The CLV Setup downloads the parameter set to the OTS400/ OTC400. The "Download Parameters" dialog box appears with the storage options. 2. Click the desired storage option. The dialog box closes automatically. The terminal interface of the OTS400/ OTC400 is now working in the "Standard OTS statistics (single read)" mode. The procedure for activating the described function for 7.3.11 to 7.3.16 is always the same and only described in Section 7.3.11. 7.3.12 Extended OTS400 statistics (single read) Function of the auxiliary interface. In this mode, the OTS400/ OTC400 outputs an overview of the information collected for each reading pulse via the auxiliary interface. In addition to Section 7.3.11 Standard OTS statistics (single read), the reading results of the various bar code scanners are also transferred in detail here. 7.3.13 Summarized OTS400 statistics (single read) Function of the auxiliary interface. In this mode, the OTS400/ OTC400 outputs a summarized overview on the information collected for each reading pulse via the auxiliary interface. Only the most important items of data are summarized in one line. This compressed form is particularly useful for evaluating the results in detail (e.g. with a spreadsheet program). 7.3.14 Standard OTS400 statistics (single read) with overall statistics after 100 reading pulses Function of the auxiliary interface. In this mode, the OTS400/ OTC400 outputs an overview of the information collected for every 100th reading pulse via the auxiliary interface. No data is output between the cycles, with the exception of error messages. The information output here is the same as that described in Section 7.3.11 Standard OTS statistics (single read). 7.3.15 Summarized OTS400 statistics (single read) with overall statistics after 100 reading pulses Function of the auxiliary interface. In this mode, the OTS400/ OTC400 outputs a summarized overview of the information collected for every 100th reading pulse via the auxiliary interface. No data is output between the cycles, with the exception of error messages. Only the most important items of data are summarized in one line. This compressed form is particularly useful for evaluating the results in detail (e.g. with a spreadsheet program).

8008869/RA61/2007-07-30

© SICK AG · Auto Ident · Germany · All rights reserved

7-7

Page 44

Chapter 7

Operation

Operating Instructions OTS400 Omni Tracking System

7.3.16 I/O monitor in the incremental pulse Function of the auxiliary interface. In this mode, the OTS400/ OTC400 outputs a status line referring to the time and increment on the auxiliary interface for every status change of an input or output. One position in the status line is available for this purpose for each of the 4 outputs, the relay, and 14 inputs (without increment inputs). Depending on the status of the input or output, the position contains either a "." or a "1".

7.4 Switching off the device Any changes made to the parameter set of the OTS400/ OTC400 or the bar code scanner with the CLV Setup program do not take effect until they have been transferred to the OTS400/ OTC400 or bar code scanner. Current parameter set unchanged: 1. If no changes have been made or if these are to be rejected, choose "File" and "Exit" in the menu bar. CLV Setup closes. 2. Switch off the supply voltage to the OTS400/ OTC400 and the bar code scanners connected to the OPS. The parameter set that was last stored permanently in the OTS400/ OTC400 or bar code scanner remains valid. Current parameter set has been modified in CLV Setup: 1. Download the modified parameter set to the OTS400/OTC400: Click the "page to bar code scanner" icon in the toolbar. CLV Stup transfers the parameter set to the OTS400/ OTC400 and asks you to specify the desired storage option. 2. Confirm the "Permanent" storage option with "OK". The OTS400/ OTC400 is now working with the modified parameter set. 3. Store the modified parameter set as a new configuration file in CLV Setup on the PC or overwrite the existing file: choose "File", a "Save As" from the menu bar. 4. Enter and confirm the path and file name in the dialog box. CLV Setup stores the configuration file. 5. Choose "File" and the "End" menu option in the menu bar. CLV Setup closes. 6. Switch off the supply voltage to the OTS400/ OTC400 and bar code scanners connected in the OPS.

7-8

© SICK AG · Auto Ident · Germany · All rights reserved

8008869/RA61/2007-07-30

Page 45

Operating Instructions

Parameterization

Chapter 8

OTS400

8

Parameterization

8.1 Overview The parameter settings are used to adapt the reading, evaluation, and output properties of the bar code scanner to the application. The default settings defined at the factory or a parameter set that has already been modified is used as a starting point. The parameter settings for the bar code scanner can be defined as follows: •

Using the Windows-based CLV Setup program (via a serial interface)

•

With command strings (via a serial interface)

8.2 Quick start 8.2.1

Setting parameters with the CLV Setup program

The CLV Setup program provides a convenient means of configuring the OTS400/OTC400 via a serial interface: 1. Connect the PC to the auxiliary interface on the OTS400/ OTC400 2. Switch on the power supply to the OTS400/OTC400 and bar code scanners in the OPS. 3. Start Windows and the CLV Setup program. 4. Upload the parameter settings from the OTS400/OTC400 by clicking the ‘Bar code scanner to page’ icon in the toolbar. CLV Setup then transfers the values from the current parameter set from the OTS400/OTC400 to the PC (this is carried out automatically when the program starts, provided that the PC is connected). 5. Change the required values on the tabs. You can display an explanation of the parameters and their functions by pressing the <F1> key. 6. Download the parameters to the OTS400/ OTC400 by clicking the ‘Page to bar code scanner’ icon in the toolbar. CLV Setup then transfers the modified parameter set to the OTS400/OTC400. The ‘Download Parameters’ dialog box is then displayed with the storage options. 7. Confirm the ‘Permanent’ storage options by clicking ‘OK’. The dialog box is then closed automatically. 8. Save the modified parameter set as a configuration file ‘*.scl*’ in CLV Setup. The OTS400/OTC400 is now ready to use the new parameter settings. The individual bar code scanners used in the OPS can also be parametrized via the auxiliary interface on the OTS400/OTC400. To do so, choose the ID of the relevant bar code scanner in the menu ‘Option’ under ‘Parameter in OTS network’. 8.2.2 ¾

Setting parameters with command strings

Program the host/PC in such a way that it sends the command strings to the OTS400/OTC400.

8008869/RA61/2007-07-30

© SICK AG · Auto Ident · Germany · All rights reserved

8-1

Page 46

Chapter 8

Parameterization

Operating Instructions OTS400 Omni Tracking System

8.3 Setting parameters with CLV Setup The Windows-based CLV Setup program is used to set the OTS400/OTC400 and bar code scanner parameters manually and to operate the device in general (choose functions and operating modes). The program supports the OTS400/OTC400 and all standard SICK bar code scanner types. CLV Setup communicates with the OTS400/OTC400 and bar code scanners by uploading and downloading the parameter values via the serial port on the PC (online mode). The default settings defined at the factory or a parameter set that has already been modified are used as a starting point. All of the parameter sets can be stored in configuration files in CLV Setup. You can display the context-sensitive help, which contains a description of all the OTS400/OTC400 and bar code scanner parameters, e.g. with the ‘Intrenet Explorer‘. A terminal emulator provides direct (online) access to the OTS400/OTC400 and the bar code scanner. You can display and edit the operating data on the screen in English or German. 8.3.1

Installing the CLV Setup software

Preparations: 1. Make sure you have the CD-ROM “Manuals & Software Bar Code Scanners” (no. 2029112) containing CLV Setup at hand. 2. Connect the PC to the auxiliary interface of the OTS400/ OTC400. 3. Connect the power supply to the OTS400/ OTC400. 4. Switch on your PC and start Windows. Installing the software: You can run the software on a PC with Windows 95TM/ 98TM, Windows NTTM, Windows 2000TM or Windows XPTM. The installation program creates a main directory with a series of subdirectories and generates the necessary links. Approximately 30 MB of hard disk space is required to install CLV Setup. You can also remove the program from your PC at any time by running the uninstaller. 1. Initial installation 1. Close all Windows applications! 2. Insert the CD-ROM into the CD-ROM drive. 3. If the auto run function is enabled, the start page (table of contents) of the CD-ROM is automatically displayed in the your browser 4. Select the SOFTWARE folder on top left. 5. Select CLV, ICR, OPS SETUP SOFTWARE in the listing. A table with CLV-Setup information is displayed. 6. Select DOWNLOAD under SOFTWARE FILE. 7. In the FILE DOWNLOAD dialog box select the option to install directly from the CD-ROM. Confirm with OK. The software is automatically saved in the “Programs\CLV“ directory on your hard disk. – or – Select the option to save the software locally on your local hard disk and confirm with OK. In the FILE DOWNLOAD dialog box select the desired target directory on your hard disk. The “CLVSetupxx.exe“ file is then saved at this location (xx = version number). Select OPEN in the dialog box to start the file. 8. The installation program starts and guides you through the installation with screen messages. The program asks you for your user name and company name. This information appears as a header in printouts. “CLVSetup“ and CLV Assistant are installed. The software will be entried in the Windows start menu under “Programs\CLV-Setup“.

8-2

© SICK AG · Auto Ident · Germany · All rights reserved

8008869/RA61/2007-07-30

Page 47

Operating Instructions

Parameterization

Chapter 8

OTS400

9. Please read the Readme file that contains the latest information on CLV Setup. 10. Then confirm the installation message by clicking OK. The CLV Setup program and CLV Setup Help are installed and ready. 2. Updating the program a) Installing the new version over the old one: Before you can install the new version of the software, you must remove the files of the old version – with the exception of the configuration files “*.scl” – using the uninstaller program. There are two ways of removing CLV Setup: •

Automatic, full deinstallation (default setting): All of the files in the main directory and subdirectory, with the exception of the configuration files in the ‘data’ directory (which contain the application-specific parameter sets of the OTS400/OTC400 and bar code scanner), are deleted.

•

Custom deinstallation: The uninstaller uses the log file from the installation routine to list all of the files, with the exception of the configuration files ‘*.scl’ in the ‘data’ directory. The configuration files ‘*.scl’ only remain if you do not allow the main directory to be deleted. We recommend that you move these files to a different directory and copy them back to the “data” directory later.

Starting the uninstaller: 1. Choose the uninstaller for CLV Setup under Programs in the Start menu. The uninstaller is then started and guides you step by step through the deinstallation process. 2. Choose the deinstallation method. 3. Install the new version of CLV Setup as described in Point 1. Initial installation, making sure to choose the same directory. The new version of CLV Setup is installed. The configuration files of the old version can be used again. b) Installing the new version in parallel with the old one: 1. Start the installation program for the new CLV Setup version as described under 1. Initial installation. 2. When the program asks you for the target directory, you must specify a new directory. The new version of CLV Setup is then installed. We recommend that you do not run both versions simultaneously. You can also use your old configuration files with the new CLV Setup version. 8.3.2

Starting CLV Setup

1. Switch on your PC and start Windows. 2. Choose CLV Setup from the Start menu. The SICK logo is then followed by the initial screen. 3. Confirm the initial screen. CLV Setup then checks whether a bar code scanner is connected to the COM 1 port on the PC (default setting in CLV Setup) and whether the OTS400/OTC400 communication parameters match those in CLV Setup. 4. Once the program has successfully established communication, it enters the OTS400/OTC400 type in the ‘Scanner’ drop-down list in the tool bar and changes the status field to ‘Connected’. The program then loads the internal OTS400/OTC400 or scanner description as well as the default parameter values and displays them in the tab. Finally, CLV Setup uploads the parameter set that was last stored permanently in the OTS400/OTC400 and displays the values in the tabs.

8008869/RA61/2007-07-30

© SICK AG · Auto Ident · Germany · All rights reserved

8-3

Page 48

Chapter 8

Parameterization

Operating Instructions OTS400 Omni Tracking System

Troubleshooting If the program cannot establish communication, it outputs a timeout warning and displays ‘No connection’ in the status field. CLV Setup then enters the OTS400/OTC400 or scanner type that it last communicated with in the ‘Scanner’ drop-down list (CLV410 the first time the program is started). The program then loads the internal description of the scanner type and displays the default parameter values on the tabs. 1. OTS400/ OTC400 or bar code scanner connected but the communication parameters differ 1. Choose the connected OTS400/OTC400 or CLV type from the ‘Scanner’ drop-down list in the toolbar. CLV Setup then attempts to connect to the device. Depending on whether it was successful, it displays either ‘Connected’ or ‘No connection’ in the status field. The program then loads the internal scanner description and displays the default parameter values on the tabs. 2. If ‘No connection’ is displayed, choose ‘Options’, ‘Serial interface’. CLV Setup then displays the current communication parameter settings in the ‘COM parameters’ dialog box. If you choose the ‘Option‘‚ ‘Auto Baud Detect’, CLV Setup attempts to detect and set the communication parameters automatically. 3. Make sure that the communication parameters defined in CLV Setup are identical to those of the bar code scanner (9,600 bd, 8 data bits, 1 stop bit, no parity) and confirm the dialog box. CLV Setup attempts to communicate with the OTS400/OTC400 again. If it succeeds, it displays ‘Connected’ in the status window. You can then change the current parameter values on the tabs. 2. No OTS400/OTC400 or bar code scanner connected when CLV Setup is started 1. Connect the PC to the auxiliary interface on the OTS400/OTC400. 2. Choose the connected OTS400/OTC400 or CLV type from the ‘Scanner’ drop-down list in the toolbar. CLV Setup then attempts to connect to the device. If it was successful, it displays either ‘Connected’ in the status field. The program then loads the internal scanner description and displays the default parameter values on the tabs. 3. Click the ‘bar code scanner to page’ icon in the toolbar. CLV Setup uploads the parameter set that was last stored permanently in the OTS400/OTC400 and displays the values in the tabs. You can then change the current parameter values on the tabs.

8-4

© SICK AG · Auto Ident · Germany · All rights reserved

8008869/RA61/2007-07-30

Page 49

Parameterization

Operating Instructions

Chapter 8

OTS400

8.3.3

Functions

The graphical user interface (GUI) is largely self explanatory and comprises the following elements: •

Title bar and status display for the configuration files

•

Menu bar with pull-down menus

•

Toolbar with scanner selection field for connecting to the OTS400/OTC400 or bar code scanner

•

Max. 9 tabs containing parameters that are grouped according to their functions. Some of these parameters open further dialog boxes.

The following functions are provided in CLV Setup: •

Upload the default parameter set from the database

•

Automatic communication attempt with the OTS400/OTC400 and bar code scanner when the program is started

•

Automatic storage of last OTS400/OTC400 or bar code scanner type selected

•

The Terminal Emulator window provides direct access to the OTS400/OTC400 and can be used to select operating modes and functions. For further information, see Chapter 7 Operation.

•

A communication window for selecting the communication port on the PC and for setting the data transfer parameters

•

An “auto baud detect” function that attempts to detect the communication parameters automatically

•

A menu for selecting the units of measure

•

A menu for selecting the GUI language

•

A function for printing the parameter sets as profile bar codes (not for OTS400/ OTC400 and CLV490)

•

A function for displaying the last 10 configuration files

•

Context-sensitive help in HTML format (<F1> Help)

The online ‘CLV Setup Help’ function provides a description of how to use the program under ‘Program information’. Default settings of CLV Setup Parameter

Value

Communication

COM 1, 9,600 bd, 8 data bits, 1 stop bit, no parity

Units of measure

Metric

Browser

Not assigned

Language

German

Last type selected

CLV410

File repository

‘data’ (bar code scanner configuration files)

Table 8-1: CLV Setup – default settings

8008869/RA61/2007-07-30

© SICK AG · Auto Ident · Germany · All rights reserved

8-5

Page 50

Chapter 8

Parameterization

Operating Instructions OTS400 Omni Tracking System

8.3.4

Transferring parameter sets between CLV Setup and the OTS400/OTC400 or CLV

Uploading from the OTS400/OTC400 or CLV (displaying the current parameter set in CLV Setup): Each time you select the OTS400/OTC400 or scanner type, CLV Setup loads the internal scanner description and displays the default parameter sets on the tabs, irrespective of whether it can communicate with the connected device. To be able to edit the current OTS400/OTC400 or CLV parameter set, you must first transfer it from the device to the CLV Setup program. ¾

Click the ‘bar code scanner to page’ icon in the toolbar. CLV Setup uploads the parameter set that was last stored permanently in the OTS400/OTC400 and displays the values in the tabs. - or Open the correct configuration file ‘*.scl’ (if it exists) in CLV Setup.

Download the parameter set to the OTS400/OTC400 or CLV (transfer a new parameter set to the OTS400/OTC400 or CLV): The parameter values displayed on the tabs only affect how the data is stored in CLV Setup. Any changes to these values do not take effect until they have been transferred to the OTS400/OTC400. CLV Setup always transfers the entire parameter set, i.e. all of the parameter values are overwritten in the OTS400/OTC400. 1. Click the ‘page to bar code scanner’ icon in the toolbar. CLV Setup transfers the parameter set to the RAM of the OTC400/OTS400 or CLV. The ‘Download Parameters’ dialog box is displayed with the following storage options. ‘PERMANENT’: ‘TEMPORARY’:

CLV Setup transfers the parameters to the RAM and to the non-volatile parameter store (EEprom) of the OTS400/OTC400 or bar code scanner. CLV Setup transfers the parameters to the RAM. The changes are lost when the OTS400/OTC400 or bar code scanner is switched off.

2. Click the required storage option. The dialog box is closed automatically. The new parameter set is now stored in the OTS400/OTC400 or bar code scanner (either temporarily or permanently). 3. Recommendation: Save the changes to the parameter set as a new configuration file in CLV Setup on the PC or overwrite the existing file: choose ‘File’ -> ‘Save as’ from the menu bar. 4. Enter the path and file name in the dialog box and confirm your entries. CLV Setup then saves your configuration file. The new parameter set is then saved in CLV Setup. Unknown OTS400/OTC400 or CLV parameters in CLV Setup If CLV Setup detects unknown parameters when it uploads the parameter set to the OTS400/OTC400 or bar code scanner (CLV Setup has not been updated for handling new parameters or the OTS400/OTC400 or bar code scanner is a special model), it outputs a warning and displays the unknown parameters in the window on the ‘Extras’ tab. The parameters are displayed in the form of command strings and can be edited using the command

8-6