Operating Instructions: CLV 42x Bar Code Scanner, Standard line: 10 Appendix, pages 93 to 101

Pages 93 to 101 of 146, from Operating Instructions: CLV 42x Bar Code Scanner, Standard line

The original document is the PDF (3.7 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Sick CLV420 and Sick CLV422

Page 93

Appendix

Operating Instructions

Chapter 10

CLV 42x bar code scanner

10

Appendix 10.1

Overview

The Appendix contains the following additional information: •

Specification diagrams – Line/Raster scanner – Line/Raster scanner with 105° angle attachment

•

System messages

•

Installation and operating instructions for the PC-based "CLV Setup" program

•

Configuring the CLV with profile bar codes

•

Configuring the CLV with commands

•

Calculating parameter values for setting the CLV

•

Tables

•

Special applications and procedures

•

Replacing a CLV (copying the parameter set)

•

Accessories

•

Dimensional drawings of the accessories

•

Supplementary documentation

•

Glossary

•

Copy of the EC Declaration of Conformity

•

Index

•

Scannable sample bar codes

10.2

Specification diagrams

10.2.1

Reading conditions for all diagrams

Test code

Code 39/ITF

Print ratio

2:1

Print contrast

> 90 %

Tilt

±10°

Ambient light

< 2000 lx

Good read rate

> 75 %

Table 10-1: Reading conditions for specification diagrams

Raster scanner: raster height 15 mm (0.59 in) at 200 mm (7.88 in) reading distance Note

Line/raster scanner with 105° angle attachment: the whole reading range shifts by 15 mm (0.59 in) towards the reading window. For the CLV 420 a corresponding reading field diagram is shown in Fig. 10-2.

8 009 981/O078/16-08-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-1

Page 94

Chapter 10

Appendix

Operating Instructions CLV 42x bar code scanner

10.2.2

Reading field diagram for CLV 420 line/raster scanner (front reading window)

General reading conditions: see Table 10-1, Page 10-1

Scan frequency/reading distance characteristics: see Fig. 10-5, Page 10-6

Fig. 10-1:

10-2

CLV 420-0010/-1010: reading range (DOF) for line/raster scanner

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 981/O078/16-08-2004

Page 95

Operating Instructions

Appendix

Chapter 10

CLV 42x bar code scanner

10.2.3

Reading field diagram for CLV 420 line/raster scanner with 105° angle attachment (lateral reading window)

General reading conditions: see Table 10-1, Page 10-1

Scan frequency/reading distance characteristics: see Fig. 10-5, Page 10-6

Fig. 10-2:

CLV 420-2010/3010: reading range (DOF) for line/raster scanner with 105° angle attachment

8 009 981/O078/16-08-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-3

Page 96

Chapter 10

Appendix

Operating Instructions CLV 42x bar code scanner

10.2.4

Reading field diagram for CLV 421 line/raster scanner (front reading window)

General reading conditions: see Table 10-1, Page 10-1

Scan frequency/reading distance characteristics: see Fig. 10-5, Page 10-6

Fig. 10-3:

10-4

CLV 421-0010/1010: reading range (DOF) for line/raster scanner

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 981/O078/16-08-2004

Page 97

Operating Instructions

Appendix

Chapter 10

CLV 42x bar code scanner

10.2.5

Reading field diagram for CLV 422 line/raster scanner (front reading window)

General reading conditions: see Table 10-1, Page 10-1

Scan frequency/reading distance characteristics: see Fig. 10-5, Page 10-6

Fig. 10-4:

CLV 422-0010/-1010: reading range (DOF) for line/raster scanner

8 009 981/O078/16-08-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-5

Page 98

Chapter 10

Appendix

Operating Instructions CLV 42x bar code scanner

10.2.6

CLV 42x characteristics field of scanning frequency

Reading conditions: see Table 10-1, Page 10-1

Fig. 10-5:

CLV 42x: scanning frequency characteristic as a function of the reading distance and resolution

10.3

System messages

The CLV outputs all system messages in plain text via the terminal interface. The messages are in English and can be displayed in extended mode of the terminal emulator in the CLV Setup user interface. (See Chapter 6.6.1 Displaying messages, Page 6-25.) Message

Meaning

"CLV 42 Soft.Vers. X.XX Change No. 0000 Host: RS232"

The software version and revision is displayed when the power supply is switched on.

"no code"

In “Reading/Parameter Evaluation“ mode, the CLV indicates that no codes matching the parameterized evaluation conditions were detected during the reading interval. (error status ST = 2 output on the host interface).

"laser safety timeout"

The CLV has deactivated the laser diode 10 minutes (default setting) after a reading pulse was initiated. The reading pulse is still active, even though the CLV is no longer reading. The reading pulse is to be terminated by resetting the pulse signal. The laser diode is activated again by the next reading pulse.

"cancel AutoSetup"

The CLV has terminated the AutoSetup, no valid bar code was found.

Table 10-2: CLV system messages

10-6

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 981/O078/16-08-2004

Page 99

Appendix

Operating Instructions

Chapter 10

CLV 42x bar code scanner

10.4

Installation and operating instructions for the PC-based "CLV Setup" program

The scope of delivery of the CLV includes a CD-ROM (“Manuals & Software“) containing the "CLV Setup" program. The following is required to run the program: a PC with at least an 80486 processor running at 66 MHz, 16 MB RAM, CD-ROM drive, serial interface, mouse (recommended) and Windows 95TM/98TM, Windows NTTM, Windows 2000TM or Windows XPTM. Approximately 25 MB of hard disk space is required to install CLV Setup (V 4.10) and CLV Assistant, approximately 20 MB to install CLV Setup Help (V 4.10). For the lastest version of the software, see www.sick.com. 10.4.1

Preparations for installation

1.

Make sure you have the CD-ROM (“Manuals & Software“) at hand.

2.

Connect the CLV to the CDB 420 or CDM 420 connection module. (See Chapter 5.5.3 Connecting the power supply, Page 5-5.)

3.

Connect the PC to the terminal interface (port) of the CLV using a 3-core RS 232 data cable (null modem cable) (CDB 420 or CDM 420: connect the PC to the internal, 9-pin "Service" plug). See Chapter 5.5.6 Connecting the PC, Page 5-7.

4.

Switch on the power supply to the CDB 420 or CDM 420. Once the CLV has started, it outputs a tone to indicate that the self-test was successfully completed. Shortly afterwards, it outputs two tones to indicate that it has assumed Reading mode. The "Device Ready" LED lights up.

5.

Switch on your PC and start Windows.

10.4.2

Installation

The CLV Setup installation program creates a directory called "CLV", a series of subdirectories and the necessary links. Using the included uninstaller, CLV Setup can be completely removed from the PC. As of version V2.6, CLV Setup will only be available and further developed as a 32-bit application. Note

The following instructions describe how to install the software on Windows 95TM. These instructions may differ depending on the operating system you are using (for example, administrator rights are required to install the software on Windows NT4.0). Initial installation 1. Shut down all the applications running in Windows. 2. Insert the CD ROM into the CD ROM drive. 3. If the auto run function is enabled, the start page (table of contents) of the CD ROM is automatically displayed in the your browser. 4. Select the SOFTWARE folder on top left. 5. Select CLV, ICR, OPS SETUP SOFTWARE in the listing. A table with CLV Setup information is displayed. 6. Select DOWNLOAD under SOFTWARE FILE. 7. In the FILE DOWNLOAD dialog box select the option to install directly from the CD ROM. Confirm with OK. The software is automatically saved in the “Programs\CLV“ directory on your hard disk. – or –

8 009 981/O078/16-08-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-7

Page 100

Chapter 10

Appendix

Operating Instructions CLV 42x bar code scanner

Select the option to save the software locally on your local hard disk and confirm with OK. In the FILE DOWNLOAD dialog box select the desired target directory on your hard disk. The “CLVSetupxx.exe“ file is then saved at this location (xx = version number). Select OPEN in the dialog box to start the file. 8.

The installation program starts and guides you through the installation with screen messages. The program asks you for your user name and company name. This information appears as a header in printouts. “CLV Setup“ and CLV Assistant are installed. The software will be entried in the Windows start menu under “Programs\CLV Setup“.

9.

Please read the Readme file, as it contains the latest information and problems regarding CLV Setup.

10. Then confirm the final installation message by clicking “OK“. Online-Hilfe CLV-Setup Help 1.

Return to the start page of the CD ROM with the opened folder “Software“.

2.

Select CLV, CLV HELP SOFTWARE in the listing. A table with CLV Help information is displayed.

3.

Select DOWNLOAD under SOFTWARE FILE.

4.

In the FILE DOWNLOAD dialog box select the option to install directly from the CD ROM. Confirm with OK. The software is automatically saved in the “Programs\CLV“ directory on your hard disk. – or – Select the option to save the software locally on your local hard disk and confirm with OK. In the FILE DOWNLOAD dialog box select the same target directory as selected for CLV Setup on your hard disk. The “CLVHelp.exe“ file is then saved at this location.

5.

After downloading the software, select OPEN in the dialog box to start the software. The software is installed in the sub directory “Help“ of the main directory “CLV“. CLV Help will also be entried in the Windows start menu under “Programs\CLV Setup“. Restart the PC. Some Windows DLL files are possibly updated.

The “CLV Setup“ program, the CLV Assistant and the online “CLV Setup Help“ are installed and ready. The program logs all of the files that are installed to the “install.log“ file. This list is used by the uninstaller to remove CLV Setup and must not be deleted if you want to use the uninstaller at some time in the future.

10-8

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 981/O078/16-08-2004

Page 101

Operating Instructions

Appendix

Chapter 10

CLV 42x bar code scanner

Updating program There are two options for installing a new version of CLV Setup: •

You can install the new version and still keep the old version (parallel installation)

– or – •

You can install the new version over the old version (overwrite)

Installing new version and retaining old version If you want to install the new version and still keep the old version of CLV Setup, follow the procedure described under Initial installation. When the program asks you for the target directory, you must specify a new directory. Both versions of the software are then available but may not be started simultaneously. The configuration files "*.scl" of the old version can be used in the new version and contain the CLV parameter sets. In order to do so, copy the configuration files from the "data" directory of the old version to the "data" directory of the new version. Installing new version over old version When installing the new version over the old version, the files of the old version must first be deleted (exception: the "*.scl" configuration files). They contain the parameter sets of the CLV. The uninstaller of the old version removes the program files. You can use the uninstaller to remove all of the files (with the exception of the configuration files) or only selected files. The default setting is a complete uninstall (the configuration files are not deleted). With the custom uninstall, the files are listed and only selected files are uninstalled. If you choose this uninstall method, ensure that the main directory containing the configuration files "*.scl" is not deleted. We recommend that you move these files to a different directory and copy them back to the "data" directory later. 1.

From the Windows Start menu, choose "Programs" and the uninstaller for CLV Setup. The uninstaller starts and guides you step by step through the uninstall procedure.

2.

Select the uninstall method (complete or custom).

3.

Install the new version of CLV Setup as described under Initial installation. Select the same directory.

The new version of CLV Setup is installed. The configuration files of the old version can be used again.

8 009 981/O078/16-08-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-9