Page 61

Operator’s Guide

Appendix K

Bit The narrowest code element (bar of space) that may contain information. Character The smallest group of elements assigned by a symbology to uniquely represent one or more numbers, letters, punctuation marks or other information. Character Set The numbers and/or letters and markings included in a bar code symbol. Check Character (or Check Digit) A character included within a bar code symbol whose value is used for performing a mathematical check of the validity of the decoded data. Contact Code Reader A light pen or other scanning device that must come into physical contact with the code medium in order to read the symbol. Continuous Code A bar code or symbol wherein the space between the characters is part of the code. Decodability This parameter grade can be “A,” “B,” “C,” “D” or “F.” The Decodability grade indicates the amount of error in the width of the most deviant element in the symbol. The less deviation, the higher the grade. Decodability is a measure of print accuracy using the symbology reference decode algorithm.

002-7858

55

RJS, Minneapolis, MN

Page 62

Operator’s Guide

Appendix K

Decode Determining the information which has been encoded in a bar code symbol. Decoder The portion of a bar code reading system that performs the decode function. Defects This parameter grade can be “A,” “B,” “C,” “D” or “F.” Defects are of two types, voids and spots. Voids are light areas in bars, and spots are dark areas in spaces. The defect grade is determined by a relationship between the largest defect in the symbol and symbol contrast. The smaller the defect, the better the grade. Aperture size can affect grade; for example, using a small aperture to analyze a very wide element will permit detection of the largest defects. Defects are usually voids, and these defects can be reduced by increasing the amount of ink (or equivalent). Diffuse Reflection Reflected light which emanates uniformly in all directions from the reflecting surface. Dimensional Deviation (DD) The measured deviation of bars and/or spaces of a scanned symbol from the specification. Discrete Code A bar code or symbol wherein the spaces between the characters are not part of the data.

002-7858

56

RJS, Minneapolis, MN

Page 63

Operator’s Guide

Appendix K

Edge Contrast (EC) See Edge Contrast(min) The difference between the space reflectance (Rs) and adjoining bar reflectance (Rb). EC = Rs - Rb Edge Contrast (min) This parameter grade can be “A” or “F.” Edge contrast is the Reflectance difference between adjoining bars and spaces. The minimum edge contrast is the smallest value of EC = R space – R bar found in the scan. The grade is determined by calculating the edge contrast of every element in a symbol and then comparing the lowest value to a fixed threshold in the specification (15%). If the value is equal to or greater than the threshold, the grade is “A.” If the value is less than the threshold, the grade is “F.” Element A generic term used to refer to either a bar or space in a bar code symbol. Element Edge The location where the scan reflectance profile intersects the midpoint between the space reflectance (R space) and bar reflectance (R bar) of adjoining elements. Visual measuring techniques will generally locate the element edge closer to the center of the bar.

002-7858

57

RJS, Minneapolis, MN

Page 64

Operator’s Guide

Appendix K

Element Reflectance Non-uniformity (ERN) The reflectance difference between the highest peak and lowest valley within each individual element and quiet zone. When an element consists of a single peak or valley, its element reflectance non-uniformity is zero. Element Width The thickness or width of a bar or space as measured from its leading edge to its trailing edge. Encoded Area The total linear dimensional space taken by all characters of a code pattern including start/stop and data. First Read Rate The percentage of successful “reads” of a bar code symbol on the first attempt. Fixed Beam Scanner A bar code reading device wherein coded items pass across a stationary incandescent or LED light source. Flexo Film Master A measurement standard symbol produced by printing the symbol on clear plastic film using the Flexographic printing process. Gloss A phenomenon related to the specular reflection of incident light. The effect of gloss is to reflect more of the incident light in a specular manner, and to scatter less. This effect occurs at all angles of incidence and should not be confused with the grazing angle which is specular reflection often referred to as sheen.

002-7858

58

RJS, Minneapolis, MN

Page 65

Operator’s Guide

Appendix K

Global Threshold (GT) The global threshold is drawn through the middle of a profile, to distinguish spaces above the line and bars below. The reflectance value is determined by dividing the symbol contrast (SC) by 2 and adding the minimum reflectance, Rmin. GT = Rmin + (SC/2) GS-1 GS1 designs and implements global standards and solutions to improve the efficiency and visibility of supply and demand chains globally. Guard Bar The first and last bars of a bar code symbol usually having the pattern 101. A guard bar generally follows the leading quiet zone and precedes the trailing quiet zone. This term is used mostly for UPC/EAN symbologies. Infinite Pad Method The method for measuring reflectance in which the sample substrate being measured is backed with enough thickness of the same type of substrate so that doubling the number of sheets does not change the measured value of reflectance. Inspection Band An area of the bar code symbol where measurements shall be taken spanning from 10% to 90% of the average bar height.

002-7858

59

RJS, Minneapolis, MN

Page 66

Operator’s Guide

Appendix K

Inter-character Gap In discrete barcodes, the space that separates two adjacent characters. When present, inter-character gaps are considered spaces (elements) for purposes of edge determination and reflectance parameter grades. Interleaved A bar code in which characters are paired together using bars to represent the first character and spaces to represent the second. ISO International Organization for Standardization, organization that maintains the standards related to bar codes and bar code verifiers. Ladder Code A bar code or symbol printed vertically with the individual bars looking like the rungs of a ladder. Laminate See Over-laminate Laser Scanner A bar code reading device that uses a low energy laser light source for illumination. Magnification Factor The size of a printed bar code compared to a standard (nominal) size. Maximum Element Reflectance Non-uniformity (ERN max) The largest element reflectance non-uniformity in a scan reflectance profile. Maximum Reflectance (Rmx)

002-7858

60

RJS, Minneapolis, MN

Page 67

Operator’s Guide

Appendix K

The greatest reflectance value in a scan reflectance profile including quiet zone. (Note: eRmx is the greatest reflectance value of an element, not including quiet zone). Minimum Edge Contrast (Ecmn) The smallest edge contrast in a scan reflectance profile. Minimum Reflectance (Rmn) The smallest reflectance value in a scan reflectance profile. Misread A condition that occurs when the data output of a reader does not agree with the encoded data. Modulation (MOD) This parameter grade can be “A,” “B,” “C,” “D” or “F.” The modulation grade is based on the relationship between minimum edge contrast (Ecmin) and symbol contrast (SC). MOD = ECmin/SC The greater the difference between minimum edge contrast and symbol contrast, the lower the grade. Scanners and verifiers perceive the narrower bars and spaces to have less intensity than wider bars and spaces; the comparison of this diminished intensity of narrow elements to wide elements is called modulation. This condition is affected by aperture size. Note: Since “ink spread” will reduce the width and intensity of single module space within a symbol, this is one thing to check in seeking to correct a low modulation grade

Module

002-7858

61

RJS, Minneapolis, MN

Page 68

Operator’s Guide

Appendix K

The narrowest expected bar or space width. Moving Beam Scanner A laser device that dynamically searches for a bar code pattern by sweeping a moving optical beam through a field of view. N (wide to narrow ratio) In symbologies with two element widths, the wide to narrow ratio of elements is calculated by summing the average wide bar width and average wide space width and dividing the sum by 2 times Z. Intercharacter gaps, if applicable, are not included. N = (avg.wide bar + avg. wide space) / (2*Z) Nanometer (nm) A unit of measure used to define the wavelength of light, equal to 10-9 meter. Nominal The intended value for a specific parameter. Tolerances are generally specified as positive and negative deviations from this value. No-Read (Non-read, Non-scan) The absence of data at the scanner output after an attempted scan because of no code, defective code or operator error. Nominal Size The target size for a specific element or group of elements. Numeric A character set that contains only numbers. Omni-directional

002-7858

62

RJS, Minneapolis, MN

Page 69

Operator’s Guide

Appendix K

The ability to read a bar code symbol from any angle as long as the bar code passes across the scanner window. Opacity The property of a material that minimizes the showthrough of printing from the back side or the next sheet. Overhead The number of characters in a symbol required for start, stop and checking. Over-laminate A coating or material adhered to the scanning surface of a bar code symbol. Parity A system for encoding characters as “odd” or “even” for self checking of barcodes. Peak The graphical pattern on a scan reflectance profile which looks like an upside down “U” or “V.” Within a profile a peak represents a space. One or more peaks could also be found within an element representing a reflectance change within an element. Plaque A template used as a reflectance calibration standard (RCS). The known reflectance values are posted on the back of the plaque.

002-7858

63

RJS, Minneapolis, MN

Page 70

Operator’s Guide

Appendix K

Print Contrast Signal (PCS) A comparison between the reflectance (brightness difference) of bars and spaces in a symbol. PCS under a given set of illumination conditions is defined as follows: PCS is calculated as follows: L - D x 100% L Where: L = Lightest (highest reflectance) D = Darkest (lowest reflectance)

Profile See Scan Reflectance Profile Quiet Zone The area immediately preceding the start character and following the stop character in a bar code symbol as specified in a particular application and/or symbology specification. Reference Decode Each symbology type specifies a specific decoding method to be used in determining overall symbol grade. This parameter grade can be “A” or “F.” (“A” is pass and “F” is fail). If this parameter is “F” the overall symbol grade will also be “F” regardless of any other parameter. If this parameter is “A” the lowest of the other parameter grades determines the overall symbol grade. Reflectance A measure of the amount of light reflected from an illuminated surface. Reflectance Minumum

002-7858

64

RJS, Minneapolis, MN