Every data set and data source in the GLX has data properties, which control how the data are displayed in the Graph, Table, and other screens.
To Open the Data Properties Dialog Box To access and edit the properties of a data source, set up the Graph or Table to display the data source, then open the data source menu and select Properties.26 The Data Properties dialog box will open. You can also access the data properties of a sensor measurement from the Sensors screen. In the Sensors screen, press F2 . (See page 55 for more information.) When you are finished viewing or editing the data properties, press close the dialog box and return to the previous screen, or press F2 any changes.
(OK) to to cancel
F1
Navigating in the Data Properties Dialog Box Measurement Name The Measurement Name is the label that designates a data source on a graph axis or at the top of a table column. You may want to edit the Measurement Name to add more information about that measurement (change “Temperature” to “Water Temperature,” for instance) or to shorten it (change “Force” to “F,” for instance). To edit the Measurement Name, use the up and down arrows keys to highlight it and press (or click it if you are using a mouse). Type the new name on the keypad and press (or press Esc to cancel). See “Multipress Text Input Mode” on page 90 for instructions on using the keypad to enter text.
Measurement Unit The Measurement Unit is the name of units that are displayed on a graph axis or at the top of a table column. You many want to edit the Measurement Unit to add units to a calculation or manually entered data set, or to abbreviate or spell out the units of a sensor measurement. For instance, you might change “m” to “meters.” To edit the Measurement Unit, use the up and down arrows keys to highlight it and press (or click it if you are using a mouse). Type the new unit name on the keypad and press (or press Esc to cancel).
26 For instructions on displaying data in the Graph and selecting items from the data source menu, see pages 13 and 15. For instructions on displaying data in the Table and selecting items from the data source menu, see pages 28 and 28.
Page 76
70
D a t a
P r o p e r t i e s
See “Multipress Text Input Mode” on page 90 for instructions on using the keypad to enter text.
Note that changing the name of the units will have no effect on the actual data. For instance, if you change the name of the Temperature units from “°C” to “°F,” the numeric values of displayed temperature will not change. To change the actual unit of measure, select the units field in the Graph (see page 15), Table (page 29), or other display; or create a calculation (see page 47). Numeric Style and Number of Digits Select Numeric Style27 and Number of Digits28 to set how numbers are displayed in the Table and Digits displays.
27To select a Numeric Style
•
Fixed Decimals: in this style, the Number of Digits determines how many digits are displayed to the right of the decimal point.
Keypad 1. Use the arrow keys to highlight Numeric Style. 2. Press to open the menu. 3. Use the arrow keys to highlight the desired style and press ; or press the number on the keypad corresponding to the desired style.
•
Significant Figures: the Number of Digits determines the number of significant figures that are displayed.
Mouse 1. Click Numeric Style to open the menu. 2. Click the desired style.
•
Scientific Notation: all numbers are displayed with a single digit to the left of the decimal point, and the other digits to the right, followed by symbol E and an exponent. (The E means “times 10 raised to the power of….”) The Number of Digits is the total number of digits displayed on both sides of the decimal point (not including the exponent).
28To select the Number of Digits
•
Engineering Notation: This style is similar to scientific notation, but there are three digits to the left of the decimal place and Number of Digits determines the number of digits to the right of the decimal place.
Mouse 1. Click Number of Digits to highlight it. 2. Press + or increase or decrease the value.
•
NSEW 000°00.000: Use this style for longitude or latitude data collected with a GPS sensor. The data will be displayed in degrees and minutes.
The meaning of Number of Digits depends on the selected Numeric Style, as described below.
EE for Exponent <= This data property determines the smallest number that will be displayed without using exponential notation. A value of -3, for instance, means that anything less than or equal to 10-2 will be displayed in exponential notation. EE for Exponent >= This data property determines the largest number that will be displayed without using exponential notation. A value of 6, for instance, means that anything greater than or equal to 106 will be displayed in exponential notation. Zero |Value| < This data property is the smallest absolute value that will be displayed. Any number with a smaller absolute value will be displayed as 0.
Keypad 1. Use the arrow keys to highlight Number of Digits. 2. Press + or increase or decrease the value.
Page 77
X p l o r e r
G L X
U s e r s ’
G u i d e
Ca lib rat ion Some of the sensors that you use with the GLX can be calibrated in the Calibration dialog box, which is accessible through the Sensors screen. Calibration can make a measurement more accurate or adjust multiple measurements so that they agree with each other. Sensors designed for calibration include the temperature probes included with the GLX. Certain other PASPORT sensors are also calibratable. To find out if a sensor can be calibrated, check its documentation, or connect it to the GLX and look for it in the Sensor field of the Calibration dialog box.
To Open the Calibration Dialog Box 29See “Sensors Screen” on page 55.
From the Sensors Screen29 follow these steps. F4
1.
Press to open the Sensor Menu.
2.
Use the arrow keys to highlight Calibrate and press ber on the keypad corresponding to Calibrate.
; or press the num-
The Calibration dialog box will not open during data collection.
Theory of Calibration To skip to the practical instructions on calibrating a sensor, see “Navigating in the Calibration Dialog Box” on page 73 and “Calibration Procedures” on page 75.
One of the functions of the GLX is to take the stream of raw data from a sensor and transform it into the calibrated data that you see in the Graph, Table, and other displays. If you do not calibrate a sensor yourself, the GLX uses a default calibration that is loaded when the sensor is connected. You can think of the GLX as containing a device that takes in raw data and outputs calibrated data. Raw Input Measurement (from sensor)
Calibrated Output Measurement (to Graph, Table, etc.) Calibration
When you perform a calibration, the GLX redefines the linear equation that transforms the raw input into the calibrated output. The linear function is of the form: Raw Input = Slope × Calibrated Output + Offset Or: Calibrated Output = ( Raw Input – Offset ) ⁄ Slope This function can be graphically represented as a line.
71
Page 78
72
C a l i b r a t i o n
Input Data
Pt 2
DInput Pt 1
Slope =
DOutput
DInput DOutput
Offset Output Data
Two points, Pt 1 and Pt 2, define the line. In the two-point calibration procedure, each point is reset by associating a known standard value (for instance, the pH of a buffer solution) with a raw input measurement that the sensor sends to the GLX when it is in that standard. In a one-point calibration, only one of the points is reset by the user.
Types of Calibration There are three types of calibration: two-point, one-point slope, and one-point offset. Any of these calibrations can be performed on a single sensor, or simultaneously on multiple similar sensors; however, for any given sensor, the GLX will automatically select the most typical calibration type as the default setting. Two-Point In a two-point calibration, you reset two points to define a new line. This type of calibration affects both the slope and the offset. For step-by-step instructions on the two-point calibration procedure, see page 75.
Input Data old calibration new calibration
Pt 1 Pt 1
Pt 2
Output Data Pt 2 Two-point calibration; two points are set to redefine the line
Input Data
One-Point Slope In a one-point slope calibration, you reset only one point. The slope of the line changes so that the line intersects the new point, while the offset (or Y-intercept) does not change. For step-by-step instructions on the onepoint slope calibration procedure, see page 76. new point old point new calibration slope changes Offset does not change old calibration
Output Data One-point slope calibration; the line changes its slope to intersect the new point
Page 79
X p l o r e r
G L X
U s e r s ’
Input Data
One-Point Offset In a one-point offset calibration, you reset only one point. The line shifts so that it intersects the new point, but its slope does not change.
new calibration new point old calibration line shifts
Output Data One-point slope calibration; the line shifts to intersect the new point
Offset calibration is usually used to make one sensor agree with another sensor. The graph below on the left shows the measurements from two temperature probes in the same container of water. Due to normal variation among probes, the second probe consistently reads about 0.08 °C higher than the first probe. Normally this difference would be insignificant; however, an offset calibration can be used to bring the sensors into closer alignment. See “One-Point Offset Calibration or Multiple-Measurement Alignment Procedure” on page 77 for step-bystep instructions. One measurement is shifted to align with other measurement
Before offset calibration
After offset calibration
Navigating in the Calibration Dialog Box For step-by-step instructions on the calibration procedure, see page 75.
The Calibration dialog box contains a list of fields that are used to select the measurement (or measurements) to be calibrated, and to enter the parameters of the calibration. One of the fields is always highlighted. Use the up and down arrow keys to move the highlight to a different field. You will sometimes see an up or down arrow ( or ) near the bottom of the dialog box indicating that more fields are available, but not currently visible. When you see the up arrow displayed on the screen, you can press the up arrow key on the keypad multiple times to make more fields visible. When you see the down arrow displayed on the screen, you can press the down arrow key to make more fields visible.
G u i d e
73
Page 80
74
C a l i b r a t i o n
Sensor Field Measurement Field Single/All Field Calibration Type Arrow indicating that additional fields can be displayed with the down arrow key
Pt 1 Standard Value Pt 1 Sensor Input Value Pt 2 Standard Value Pt 2 Sensor Input Value Arrow indicating that additional fields can be displayed with the up arrow key The top and bottom portions of the Calibration dialog box
Sensor Field Highlight this field and press + or calibratable sensors connected to the GLX.
to cycle through the
Measurement Field Highlight this field and press + or the calibratable measurements of the selected sensor.
to cycle through
Single/All Field Highlight the Single/All Field and press to switch between “Calibrate selected measurement only” and “Calibrate all similar measurements.”30 Calibration Type In most cases, it is not necessary to access this field because the GLX automatically selects the most typical calibration type for the selected measurement. To override the automatic selection, highlight Calibration Type and press to open the menu. Highlight the desired calibration type and press , or press the number on the keypad corresponding to the desired type. The options are 2 Point, 1 Point Slope, and 1 Point Offset. Pt 1 Standard Value Highlight this field and press press .
. Enter a value and
Pt 1 Sensor Input Value It is usually not necessary to access this field because its value is entered automatically when you press F3 . To enter a value manually, highlight the field and press press .
. Enter a value and
Pt 2 Standard Value Highlight this field and press standard and press .
. Enter the value of the
Pt 2 Sensor Input Value It is usually not necessary to access this field because its value is entered automatically when you press F4 . To enter a value manually, highlight the field and press the standard and press .
. Enter the value of
30See “One-Point Offset Calibration or
Multiple-Measurement Alignment Procedure” on page 77 for more information.
Page 81
X p l o r e r
G L X
U s e r s ’
G u i d e
75
Because they are not needed, the fields for Pt 1 are hidden when the 1 Point Slope calibration type is selected, and the fields for Pt 2 are hidden when the 1 Point Offset calibration type is selected.
Calibrated Output and Raw Input Two live values are displayed at the bottom of the Calibration dialog box: the Calibrated Output Measurement and the Raw Input Measurement. The Raw Input Measurement is the data transferred from the sensor to the GLX. The Calibrated Output Measurement is the value resulting from the currently defined calibration. The Calibrated Output Measurement is displayed in the Graph, Table, and other displays.
Note About Raw Input Measurement The value and units of the raw input measurement data can be confusing because they do not necessarily have an obvious meaning. Depending on which sensor you are using, the raw input may represent the default calibration, the voltage produced by an electrode in the sensor, or the digital output of the sensor’s analog-to-digital converter.
Calibrated Output Raw Input Measurement Measurement Live Data
Keep in mind that it does not matter what the value is, or what its units are. It is only important that the raw sensor input varies in direct proportion to the quantity being measured.
Calibration Procedures If you don’t know whether a two-point or one-point slope calibration is best suited for a particular sensor, follow steps 1–3 of either procedure and observe the Calibration Type that is automatically selected.
Two-Point Calibration Procedure A two-point calibration requires two known standards. For instance, if you are calibrating a pH sensor, you might use two buffer solutions: one of pH 4 and the other of pH 7. The fields of the Calibration dialog box are identified on page 74.
1.
From the Sensor field, select the desired sensor.
2.
From the Measurement field, select the desired measurement.
3.
Set the Single/All field to “Calibrate selected measurement only.”
4.
From the Calibration Type menu, select 2 Point, if it is not already selected.
5.
Place the probe in the first known standard.
6.
Enter the value of the standard in the Standard Value field for Pt 1.
7.
Observe the Raw Input Measurement displayed near the bottom right corner of the screen and wait until it has stabilized.
Enter standard value here
Press F3 to copy Raw Input Measurement to Pt 1 Sensor Input Value
Page 82
76
C a l i b r a t i o n
8.
Press
F3 to read the sensor input value of Pt 1.
Note that the Raw Input Measurement has been automatically entered into the Pt 1 Sensor Input Value field, and the standard value is now displayed as the Calibrated Output Measurement. 9.
Place the probe in the second known standard.
10. Enter the value of the standard in the Standard Value field for Pt 2. 11. Observe the Raw Input Measurement displayed near the bottom right corner of the screen and wait until it has stabilized. 12. Press
F4
Enter standard value here to read the sensor input value of Pt 2.
Note that the Raw Input Measurement has been automatically entered into the Pt 2 Sensor Input Value field, and the standard value is now displayed as the Calibrated Output Measurement. 13. Press F1 to accept the new calibration and return to the Sensors screen, or press F2 to cancel the new calibration and revert to the previous or default calibration.
F1
F2
Accept new calibration or Cancel new calibration and return to Sensors and revert to previous screen calibration
One-Point Slope Calibration Procedure A one-point slope calibration requires one known standard. For instance, if you are calibrating a dissolved oxygen sensor, you might use 100% saturated water, for which you can look up the oxygen concentration based on temperature and pressure. The fields of the Calibration dialog box are identified on page 74.
1.
From the Sensor field, select the desired sensor.
2.
From the Measurement field, select the desired measurement.
3.
Set the Single/All field to “Calibrate selected measurement only.”
4.
From the Calibration Type menu, select 1 Point Slope, if it is not already selected.
5.
Place the probe in the known standard.
Press F4 to copy Raw Input Measurement to Pt 2 Sensor Input Value
Page 83
X p l o r e r
6.
Enter the value of the standard in the Standard Value field for Pt 2.
7.
Observe the Raw Input Measurement displayed near the bottom right corner of the screen and wait until it has stabilized.
8.
Press
F4
G L X
U s e r s ’
G u i d e
77
Enter standard value here to read the sensor input value.
Note that the Raw Input Measurement has been automatically entered into the Pt 2 Sensor Input Value field, and the standard value is now displayed as the Calibrated Output Measurement. 9.
F1
Press to accept the new calibration and return to the Sensors screen, or press F2 to cancel the new calibration and revert to the previous or default calibration.
Press F4 to copy Raw Input Measurement to Pt 2 Sensor Input Value
One-Point Offset Calibration or Multiple-Measurement Alignment Procedure If you are making two or more simultaneous measurements of the same type, it is sometimes desirable to slightly adjust the calibrations so that all of the measurements agree. For instance, if you are using several temperature probes, you may find that they have slightly different readings when immersed in the same container of liquid; a multiple-measurement alignment will make all of the probes read the same temperature. This procedure is a one-point offset calibration performed on two or more similar measurements simultaneously. The fields of the Calibration dialog box are identified on page 74.
1.
From the Sensor field, select the desired sensor.
2.
From the Measurement field, select the desired measurement.31
31In this case, the calibration of the
3.
Set the Single/All field to “Calibrate all similar measurements.”
4.
From the Calibration Type menu, select 1 Point Offset.
selected measurement will not change. Rather, the calibrations of all other similar measurements will be changed to match the selected measurement.
5.
Place all the probes so that they are all measuring the same thing. It is not necessary to know the value of what they are measuring at this point. (For instance, place all the temperature probes into the same container of water.)
6.
Press F3 . Note that the Calibrated Output Measurement has been automatically entered into the Pt 1 Standard Value field, and the Raw Input Measurement has been automatically entered into the Pt 1 Sensor Input Value field.
7.
Press F1 to accept the new calibration and return to the Sensors screen, or press F2 to cancel the new calibration and revert to the previous or default calibration.
To confirm that the measurements are aligned, open the Digits display and view two or more of the measurements simultaneously while recording data. They should be identical or very close. For information on the Digits display, see page 37.
Press F3 to copy Calibrated Output Measurement to Pt 1 Standard value; and Raw Input Measurement to Pt 1 Sensor Input Value
Page 84
78
D a t a
F i l e s
S c r e e n
Da ta F ile s S cre en Collected data, calculations, notes, display configurations, and sensor settings are all stored in the GLX’s memory. You can save data in a file during or after an experiment, or save a file without data to use as the starting point of an experiment. Use the Data Files screen to view, create, copy, move, and delete these files and their components. In this screen you can also transfer files to and from a second GLX or a flash drive connected to the USB port.
To Open the Data Files Screen From the Home Screen, do one of the following: use the arrow keys to highlight the Data Files icon, then press
; or click the Data Files icon. The Data Files icon on the Home Screen
Navigating in the Data Files Screen The Data Files screen is divided into two sections: the Folder Selector and the File List. The Folder Selector is used to select one of the GLX’s two data-storage locations. The File List displays the files stored in the selected folder. At any time, the focus is either on the Folder Selector or the File List, indicated by a dark box surrounding that section. The focused section also has one of its items highlighted.
Folder Selector
File List
Dark box surrounding active section
Highlighted folder
Folder Selector active
Highlighted file list entry
File List active
Press the down arrow key to move the focus from the Folder Selector to the File List section. Press the up arrow key (when the highlight is on the top entry of the File List) to move the focus from the File List to the Folder Selector. If the File List is empty, the focus cannot be moved to it.
Navigating in the Folder Selector There are two folders in the Folder Selector: RAM and Flash. When the Folder Selector has the focus, the selected folder is highlighted. Press the left and right arrow keys to switch between RAM and Flash. The size and free capacity of the selected folder are indicated below the Folder Selector. The files in the selected folder are displayed in the File List section. Press the down arrow key to move the focus to the File List of the selected folder. When the focus is not on the Folder Selector, the selected folder is indicated by a dashed box.
Selected folder; Press to change selection