Page 31
OaktonLog The other end of the serial communication cable ends with a small black flat plug. Plug this end into the socket at the back of the OaktonLog. Note that the socket has three pins, with one further away from the other two (Figure 1). Take care to plug the cable in correctly.
Chapter 1
Figure 1: Connecting OaktonLog
Note: You can also connect to the PC via the Cradle (see page 168).
31
Page 32
Chapter 1
OaktonLog
2. Connect OaktonLog to an external sensor Use the same PC connection socket to connect to an external sensor (see Figure 1). Note that the socket has three pins, with one further away from the other two (Figure 1). Take care to plug the cable in correctly. Note: You can also connect the external sensor to the Cradle (see page 168). 1.3.4. Sleep Mode The OaktonLog battery's maximum lifespan is approximately 1.5 years. This long battery life is achieved by OaktonLog putting itself "to sleep" between recordings, or after 4 minutes goes by
32
Page 33
OaktonLog
Chapter 1 without communication with the computer. While sleeping the data logger consumes a minimal amount of power. OaktonLog wakes up every second for a very short time (a few micro-seconds) to check if one of the keys has been pressed or if there is a message coming in from the computer. This is the reason you have to press and hold a key for at least two seconds before a status message appears. If you use high sampling rates, the OaktonLog battery’s lifespan will be considerably shorter (see Figure 2).
33
Page 34
Duration of working in monthes
Chapter 1
OaktonLog
30 25 24.1
23.3
21.5
20 16.4
15
24.6
24.4
EC650 EC600
12.1
10 5.9
5
0.8
0 0
5
10
15
Minutes between samples
Figure 2: OaktonLog battery life
34
20
Page 35
OaktonLog
Chapter 1
Note: The figures on the graph refer to OaktonLog usage without daily download and without viewing min/max values on OaktonLog's display 1.3.5. Replacing the Battery Warning: The back cover of the OaktonLog contains a special compartment for the humidity sensor, which is very fragile. When removing or replacing the cover, be careful not to harm the sensor
35
Page 36
Chapter 1
OaktonLog Humidity sensor
Humidity sensor compartment +
Figure 3: Replacing the battery OaktonLog’s battery is a 1/2AA 3.6V lithium battery 1.
36
Unfasten the four screws on the back of the OaktonLog.
Page 37
OaktonLog 2. 3.
4.
5.
Chapter 1
Carefully remove the back cover Replace the battery, and be sure to insert the new battery into the correct corresponding polarities (look for the “+” symbol next to the positive terminal) Carefully reposition the electronic board on OaktonLog’s back cover and make sure that the humidity sensor is placed in its compartment. Refasten the four screws on the back of the OaktonLog.
37
Page 38
Chapter 1
OaktonLog
1.4. Working with the MicroLab Software 1.4.1.
Installation
1. System Requirements To work with MicroLab, your system should be equipped with the following: Software Windows 95, or later (Windows 95 will not support USB) Internet Explorer 4.0 or later (you can install Internet Explorer 5 when you install OaktonLog Plus, since it ships with the product) Hardware Pentium 300MHz or higher 32 MB RAM (64 MB recommended) 5MB available disk space for the MicroLab application
38
Page 39
OaktonLog 2.
Chapter 1
Install the Software 1. Insert the CD into your CD drive. 2. The installation will start automatically. Once the process begins, follow the on-screen instructions. 3. In case auto run is not working, open the CD drive folder and double-click the setup icon, then follow the on-screen instructions.
To un-install the software: From the Start menu select Settings/ Control Panel, and use the Add/Remove programs function to remove the MicroLab application.
39
Page 40
Chapter 1
OaktonLog
To install the USB driver (optional for cradle communication): 1. Insert the CD into your CD drive. If Installation begins automatically (and you have already installed MicroLab), click Cancel to stop installation. 2. Connect the Cradle to a USB port on your PC. Windows will automatically detect the new device and open the Add New Hardware Wizard. 3. Select Specify the location of the driver, and then click Next. 4. Select Search for the best driver for your device, then check the Removable Media checkbox, and then click Next.
40