Page 41

Appendix

C

General Specifications

Communications, I/O, Indicators, Illumination P/N / Model

Comm.

Discrete I/O

Indicators

Illumination

GMV-6310-1100G GMV-6310-1102G GMV-6310-1104G GMV-6310-1106G GMV-6310-1110G GMV-6310-1112G GMV-6310-1114G GMV-6310-1116G GMV-6310-1172G GMV-6310-1200G GMV-6310-1202G GMV-6310-1204G GMV-6310-1206G GMV-6310-1210G

High-Output Red LEDs @ 617nm Trigger Input, Learn: Bi-directional, optoisolated, 4.5-28V rated (10mA at LEDs: Trigger, Pass, High-Output White 28VDC) Fail, Mode, Power, LEDs RS-232 and Ethernet Outputs (1, 2, 3): Link/Act; Green Flash: Pass; Blue V: Bi-directional, optoisolated, 1-28V Target rated, ICE<mA at 24VDC, current High-Output Red limited by user) LEDs @ 617nm

GMV-6310-1212G GMV-6310-1214G GMV-6310-1216G GMV-6310-1272G

C-4

High-Output White LEDs

Vision MINI Xi Smart Camera Guide

Page 42

General Specifications

C Power P/N / Model

Power

GMV-6310-1100G GMV-6310-1102G GMV-6310-1104G GMV-6310-1106G GMV-6310-1110G GMV-6310-1112G GMV-6310-1114G GMV-6310-1116G GMV-6310-1172G GMV-6310-1200G

10-30VDC, 200 mV p-p max. ripple, 132mA @ 24VDC (typ.)

GMV-6310-1202G GMV-6310-1204G GMV-6310-1206G GMV-6310-1210G GMV-6310-1212G GMV-6310-1214G GMV-6310-1216G GMV-6310-1272G

Vision MINI Xi Smart Camera Guide

C-5

Page 43

Appendix

C

General Specifications

Operating Environment; Agency Compliance P/N / Model

Operating Temperature

Storage Temperature

Humidity

Enclosure

Agency Compliance

GMV-6310-1100G GMV-6310-1102G GMV-6310-1104G GMV-6310-1106G GMV-6310-1110G GMV-6310-1112G GMV-6310-1114G GMV-6310-1116G GMV-6310-1172G GMV-6310-1200G GMV-6310-1202G GMV-6310-1204G GMV-6310-1206G GMV-6310-1210G GMV-6310-1212G

0° to 40° C (32° –50° to 75° C Up to 90% IP54 to 104° F) (–58° to 167° F) (non-condensing) (category 2)

FCC, UL/cUL, CE (General Immunity for Light Industry: EN 55024 ITE Immunity Standard; Radiated and Conducted Emissions of ITE Equipment: EN 55022 ITE Disturbances), CB, Class A, RoHS/WEEE

GMV-6310-1214G GMV-6310-1216G GMV-6310-1272G

C-6

Vision MINI Xi Smart Camera Guide

Page 44

Dimensions

Note: Nominal dimensions shown. Typical tolerances apply.

Vision MINI Xi Smart Camera Guide

C General Specifications

Dimensions

C-7

Page 45

Appendix

C

General Specifications

Field of View and Working Distance WVGA/SXGA

QXGA

C-8

Vision MINI Xi Smart Camera Guide

Page 46

D

APPENDIX D

Serial Commands

D Serial Commands

This section provides descriptions of the serial commands that can be sent to the camera via TCP (Telnet) port, AutoVISION Terminal, or HyperTerminal.

Vision MINI Xi Smart Camera Guide

D-1

Page 47

Appendix

D

Serial Commands

Serial Command Syntax < > = Required argument. Replace appropriately. For example: -u <DB_User_name> becomes -u av where av replaces DB_User_name. | = Mutually exclusive arguments. Choose one from the list. { } = Used with | to specify a list of choices for an argument. [ ] = Optional parameter. Important: Unless otherwise stated, commands will respond with !OK on success and !ERROR on failure.

GETIMAGE <-transfer=ymodem> [-format={jpg|png}] [-quality ={0-100}] [-woi=left,top,right,bottom] [-inspection=n] Initiates serial transfer of inspection image (RS-232 only). Note: This command always returns the last (most recent) image. -transfer=ymodem is currentlynot optional - only Ymodem protocol is supported. -format={jpg|png} specifies the format of the image. If omitted, the image format is JPG. -quality=n specifies a JPG compression quality of n less than or equal to 100. The default quality is 80 if not specified. Note: The PNG format provides lossless image compression. If format is set to PNG, the quality setting does not apply. woi=left,top,right,bottom specifies a rectangular area of the image to be included in the output image. If omitted, the full image buffer is returned. -inspection=n specifies the inspection from which to retrieve an image. The image will be from the first snapshot within that inspection. If not specified, the image will be from the first inspection that does contain a snapshot. The following example will retrieve an image from the camera with these settings: Protocol: ymodem; Format: png; Quality: N/A; Inspection: second inspection. GETIMAGE –transfer=ymodem –format=png –inspection=2 The following example will retrieve an image from the camera with these settings: Protocol: ymodem; Format: jpg (default); Quality: 50; Inspection: first inspection (default). GETIMAGE –transfer=ymodem –quality=50 D-2

Vision MINI Xi Smart Camera Guide

Page 48

Serial Commands

D ONLINE Starts all inspections.

OFFLINE Stops all inspections.

TRIGGER Triggers an inspection.

vt [n] Triggers an inspection by pulsing a Virtual I/O point. For example: vt 1 will return pulse VIO1. The inspection will run if it is configured to use VIO 1 as a trigger. If specified, the VIO index must be in the allowed range for Virtual I/O points within Visionscape. The virtual I/O line will be set high then low. If VIO Index is not specified, VIO1 is assumed. Fail Return: Return !ERROR followed by the reason for the failure. For example: !ERROR No such trigger when the index specified ‘n’ is out of range of virtual triggers.

REBOOT [-noload] Reboots the device. -noload = do not load BOOT job.

Vision MINI Xi Smart Camera Guide

D-3

Page 49

Appendix

D

Serial Commands

MEMAVAIL [-cp] Returns available memory for device or coprocessor.

MEMCONTIG [-cp] Returns maximum memory block for device or coprocessor.

MEMFRAGS [-cp] Returns memory fragments for device or coprocessor.

MEMINFO [-cp] [-v] Returns memory summary “avail/contig/frags” for device or coprocessor. Verbose.

VERSION Returns Visionscape software version.

JOBSAVE [-slot=]<n> Saves current job to slot n.

JOBLOAD [-slot=]<n> [-r] Loads job from slot n. -r = Start inspections.

JOBDELETE {[-slot=]n|-all} Deletes job in slot n, or all jobs if -all. Important: Does not delete the current job loaded in camera memory.

D-4

Vision MINI Xi Smart Camera Guide

Page 50

JOBINFO [[-slot=]n] [-v] Gets job summary or info about slot n. JOBINFO with no arguments returns a list of all jobs on the device. -v = Verbose n. This option shows the amount of space that would be freed if the job were deleted. It also lists the total disk space and free disk space.

JOBBOOT [-slot=]<n> Sets bootup job slot n (RS-232 only).

JOBDOWNLOAD <-transfer=ymodem> Downloads .avz job packaged via transfer method (RS-232 only).

SET <tagname> <value> Sets value of a global tag. The tagname must correspond to one of the supported tags within the device. The value can contain spaces. The command is terminated by a carriage return and/or line feed character. The value can be a list of comma-separated items to set a sequence of tags: Send SET int1 1, 2, 3 to set int1 = 1, int2 = 2, int3 = 3. The AVP service allows setting of step and datum information from the job tree using forward slash ‘/’ in the symbolic name path. SET avp/insp1/snapshot1/acq1/gain 2.0 paths are not case-sensitive and do not need to be fully qualified if unique. SET avp/acq1/gain 2.0 will set the same gain value if there is only one acquire.

Vision MINI Xi Smart Camera Guide

D-5

Serial Commands

D