Page 11
Introduction
Warning and Caution Summary WARNING LED LIGHT DO NOT VIEW DIRECTLY WITH OPTICAL INSTRUMENTS CLASS 1M LED PRODUCT Light Output: 648 cd Wavelength: 464 nm; 518 nm; 635 nm IEC 60825-1:1993+A1:1997+A2:2001
• Viewing the Quadrus Verifier’s LED output with optical instruments such as magnifiers, eye loupes, or microscopes within a distance of 100 mm could cause serious eye injury. • Maximum LED light output: 648 cd • Wavelength: 464 nm; 518 nm; 635 nm • Location of the Quadrus Verifier’s LED aperture window:
LED Aperture Window
• CAUTION: Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure.
Quadrus Verifier User’s Manual xi
Page 12
Warning and Caution Summary
Warning and Caution Summary (cont.) WARNING LED RADIATION DO NOT STARE INTO BEAM CLASS 2 LED PRODUCT Max Power: 67mW Wavelength: 660 nm IEC 60825-1:1994+A1:2002+A2:2001
• Max Power: 67 mW • Wavelength: 660 nm • Location of the Quadrus Verifier’s LED array:
LED Array
• CAUTION: Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure.
xii
Quadrus Verifier User’s Manual
Page 13
1 Quick Start Contents Step 1 Check Required Hardware................................................................................................ 1-2 Step 2 Connect the System.......................................................................................................... 1-3 Step 3 Install ESP......................................................................................................................... 1-4 Step 4 Select Model ..................................................................................................................... 1-5 Step 5 Select Communications Protocol ...................................................................................... 1-6 Step 6 Position Verifier and Symbol ............................................................................................. 1-7 Step 7 Locate Symbol .................................................................................................................. 1-8 Step 8 Calibrate Reflectance....................................................................................................... 1-9 Step 9 Set Verification Parameters ........................................................................................... 1-10 Step 10 Verifiy Symbol .............................................................................................................. 1-11 Step 11 Generate and Save Verification Report ........................................................................ 1-14
This section is designed to get your Verifier up and running quickly so you can get a sense of its capabilities and test sample symbols. Detailed setup information for installing the Verifier in your application can be obtained in the subsequent sections.
Quadrus Verifier User’s Manual
1-1
Page 14
Check Required Hardware
Step 1, Check Required Hardware
Hardware Required
Caution: Be sure that all cables are connected BEFORE applying power to the system. Always power down BEFORE disconnecting any cables.
Item 1 2 3 4 5 6 7 8
1-2
Description Quadrus Verifier IB-150 Kit Power Supply Illumination Power Supply Light Control Cable Quadrus EZ Stand (not shown) Monitor Kit Object Detector
Part Number FIS-6700-100XG Included with Verifier Included with Verifier Included with Verifier Included with Verifier Included with Verifier 98-000096-01 99-000017-01
Quadrus Verifier User’s Manual
Page 15
Quick Start
Step 2, Connect the System
Hardware Configuration
Caution: Be sure that all cables are connected BEFORE applying power to the system. Always power down BEFORE disconnecting any cables.
Connecting to a Host by RS-232 1. Connect the Verifier to the IB-150 Kit. 2. Connect the IB-150 Kit host cable to the host. 3. Connect the Lighting Power Supply to the lighting chamber. 4. Connect the main power supply and cycle power to the Verifier. 1
Note: When wiring the IB-150 Kit to a host with a 25-pin host connector, cross pins 2 and 3. When wiring the interface box to a host with a 9-pin host connector, do NOT cross pins 2 and 3.
Connecting to a Host by TCP/IP
25
Side View of IB-150 showing Host 25-pin Socket Connection
See Chapter 16, Ethernet. Quadrus Verifier User’s Manual
1-3
Page 16
Install ESP
Step 3, Install ESP Easy Setup Program (ESP) is Microscan’s proprietary setup and testing application. The purpose of ESP is to provide a quick and easy way to set up and configure Microscan products. When the Quadrus Verifier is connected to a host computer (Windows Vista, XP, or 2000), ESP can be used to set up communication with a host, configure various firmware settings, and control verification processes.
If installing from a Microscan Tools CD: 1. Insert the Microscan Tools CD in your computer’s CD drive. 2. Choose ESP Software from the main menu. 3. Select the Current Version of ESP and follow the file download prompts.
If downloading from the web: 1. Go to http://www.microscan.com/downloadcenter 2. Create a new “myMicroscan” member account or, if you are already a member, enter your user name and password. 3. Click the Download Software link and extract the latest version of ESP to a directory of your choice. Note where your ESP.exe file is stored on your hard drive. 4. At the end of the installation process, the following icon will appear on your desktop:
5. Click the ESP icon to start the program.
System Requirements for ESP • 166 MHz Pentium processor (recommended) • Windows Vista, XP, or 2000 operating system • Internet Explorer 5.0 or higher • 64 MB minimum RAM • 40 MB minimum disk space
1-4
Quadrus Verifier User’s Manual
Page 17
Quick Start
Step 4, Select Model When you start ESP, the following menu will appear:
1. Click the button showing the Quadrus Verifier. 2. Click OK. 3. Note: You can also simply double-click the Quadrus Verifier button to make your selection. 4. Click Yes when the following dialog box appears:
Note: If you need to select another model later, you can find it in Application Mode under Model on the menu toolbar. Quadrus Verifier User’s Manual
1-5
Page 18
Select Communications Protocol
Step 5, Select Communications Protocol When the following dialog appears, make your selection and click Next.
RS-232 1. In the RS-232 dialog, if your communications port is not the default COM1, use the dropdown menu to change your communications port. 2. Click Connect. 3. If the connection fails, click the Autoconnect button, select a different communications port, and try again.
Note: If your host settings cannot be changed to match the Verifier’s settings, check the Force Connect box.
TCP/IP See Chapter 16, Ethernet.
1-6
Quadrus Verifier User’s Manual
Page 19
Quick Start
Step 6, Position Verifier and Symbol When you connect to ESP, the first thing you will see is the Report tab of the Verification view.
This view allows the user to center the symbol before calibrating.
The Quadrus Verifier comes with a reference card that features a Verifier Calibration Symbol and two numbers--the minimum and maximum reflectance values for ISO/IEC 15415 and AIM DPM Reflectance Calibration. Keep this card in a safe place! It is the Verifier’s most critical setup tool.
Use the symbol provided at the top of the Verifier Calibration Symbol card to perform Reflectance Calibration.
Place the symbol at the center of the field of view. Be sure the plane of the symbol is as close to perpendicular as possible relative to the Verifier’s orientation.
Quadrus Verifier User’s Manual
1-7
Page 20
Locate Symbol
Step 7, Locate Symbol After you place the Verifier Calibration Symbol beneath the Verifier’s lighting chamber, click the Locate button. You will see a video representation of the Verifier’s field of view.
This view allows the user to center the symbol before calibrating.
1-8
Quadrus Verifier User’s Manual