Page 131

Scanner Parameters

Additional Information about Gain, Tracking, and Transition Counter Gain The QX-830 Compact Industrial Scanner is an optical device. Optical devices deal with a wide range of brightness, in a way similar to a camera. If the image is too bright, the exposure must be reduced. If the image is too dark, the exposure must be increased. The scanner will only work properly if it is within the range of brightness that can resolve light and dark changes, or “transitions”, in the field of view. If the gain is too low, the entire symbol looks like a dark bar to the scanner. If the gain is too high, the entire symbol looks like a white bar to the scanner. Brightness conditions can be affected by the distance between the symbol and scanner, or by the consistency of the material on which the symbol is printed. The AGC Sampling Mode command has two primary functions: AGC (either Leading Edge or Continuous) and Disabled (Fixed Gain). AGC (Leading Edge or Continuous) AGC self-adjusts the optical-electrical gain to hold brightness to one range for all materials and distances. The Auto Calibration menu shows what AGC is doing to control brightness, as in the example shown here:

The value –35.0 indicates that brightness has been reduced by 35 dB. The symbol used in this example was very close to the scanner, using high laser power. As the laser power is lowered and/or the symbol is moved farther from the scanner, the attenuation (brightness reduction) value becomes lower. The user does nothing to adjust the gain in order to keep the scanner decoding; AGC performs the task of raising gain to accommodate the greater distance. The actual attenuation factor is determined by the return light from the symbol and the “SETPOINT” value that the control system is attempting to hold. (SETPOINT is an abstracted number range that has to do with the feedback system’s dynamic range.) A very low SETPOINT value of 0-20 would tend to keep the symbol amplitude low. A very high SETPOINT value of 150-255 would tend to keep the symbol amplitude high. The gain will change to hold the feedback value equal to the user SETPOINT setting of amplitude (Gain Level) at all times. Disabled (Fixed Gain) Fixed Gain is a gain mode in which there are no changes in the gain system. The system takes the total gain range and divides it into 255 steps. Then the value of the Gain Level setting is converted directly into the gain value. Nothing will change this value unless the user changes the Gain Level setting.

QX-830 Compact Industrial Scanner User Manual

5-65

Page 132

Read Cycle When To Use Fixed Gain vs. AGC The AGC settings (Leading Edge or Continuous) are preferable for most applications, as they provide the best overall scannable area. However, there are times when AGC will have difficulty locking onto the symbol of interest, especially when extraneous non-symbol objects in the field of view appear to be symbols. AGC can only base its measurements on one object. If that object is not the actual symbol, AGC performance and consistency are adversely affected. Leading Edge bases its calculations on the first object it detects, even if the first object is not a symbol. This is helpful if the symbol is always at the leading edge of the decode direction. Continuous is the recommended setting for Microscan scanners. Continuous always samples throughout the entire scan but is most interested in the object with the most sharply-defined bars. This can occasionally mislead the scanner when bar-like objects or specular reflection enter the field of view. Fixed Gain is preferable if AGC repeatedly fails to choose the correct Gain Level setting for the application. In some applications, symbols might be surrounded by a greater number of symbol-like false candidates than actual symbols. The Fixed Gain setting will never change its value, so once it is set, surrounding objects in the field of view will not distract the scanner from decoding actual symbols. Fixed Gain is also useful in applications using symbols that have very low bar counts, or in applications using Symbol Reconstruction on symbols with tilted or skewed placement. The AGC Minimum and AGC Maximum parameters are useful in applications with depth of field limitations that are not conducive to Fixed Gain operation. To determine the appropriate minimum and maximum gain: • Set AGC Sampling Mode to Disabled (Fixed Gain) and find the range of gain levels that yields successful decodes in the application (both the lowest and the highest gain levels). • Enter those values in the AGC Minimum and AGC Maximum fields. This will leave the scanner room for gain changes to increase depth of field (within the limitations of what works for the application). Maximum gain may be too high for symbols on white backgrounds, because it may attempt to process false bars. Especially in applications using symbologies that require more processing, setting gain too high can slow decode performance substantially. Setting a minimum gain can prevent the system from occasionally “under-gaining” because of previous false candidates that appear very bright.

5-66

QX-830 Compact Industrial Scanner User Manual

Page 133

Scanner Parameters

Tracking Tracking is the frequency detection value that determines the threshold between a white bar and a black bar. This is a range of frequencies that have been divided into 127 steps. For most applications, the default tracking value (40) is robust enough to deal with most marginal signals. However, in some instances, symbols are printed on a “noisy” substrate, or symbols are relatively distant from the scanner and the signal-to-noise ratio and optical focus are not ideal. In such cases, using a combination of Gain Level and Tracking can improve decode performance by rejecting “noise” and correctly recognizing the bars of the candidate symbol.

Transition Counter Transition Counter is an integral part of the AGC system. Transition Counter determines the number of black bars that will be counted before an AGC sample is taken. Setting a Transition Counter value lower than the default of 14 gives the scanner less integration time for the candidate symbol, and makes the scanner more susceptible to false objects. A Transition Counter value greater than 14 requires more bars to be present for the symbol even to qualify as an object. When setting the Transition Counter value, it is important to keep in mind what occurs when the symbol is first presented to the scanner - before gain is applied to the symbol. A few bars are usually missed due to distortion at first. Setting Transition Counter to the same value as the number of bars contained in the symbol can cause the scanner not to take an AGC sample at all. The value of 14 is optimal except in the case of low bar count due to: • Symbols that are tilted so that only 10 bars or fewer are available in a single scan line. (See Symbol Reconstruction). • Symbologies with typically low bar counts, such as Pharmacode. Applications using symbols with fewer than 10 bars should use Fixed Gain operation.

QX-830 Compact Industrial Scanner User Manual

5-67

Page 134

Read Cycle

Laser Setup Laser Setup allows the user to configure Laser On/Off, Laser Framing Status, Laser On Position, Laser Off Position, and Laser Power.

Laser On/Off When Enabled, the laser is On only during the read cycle. When Disabled, the laser operates continuously. Note: A serial trigger or external trigger must be enabled for Laser On/Off to take effect.

Laser Framing Status When Enabled, the laser will be off for the duration of the Laser Off time (beginning from the start of the read cycle), and the laser will then be on for the duration of the Laser On time.

Laser On Position The duration of Laser Off time. Laser On Position is a ratio of the total scan width, with increments equal to 1/100th of the total scan width.

5-68

QX-830 Compact Industrial Scanner User Manual

Page 135

Scanner Parameters

Laser Off Position The duration of Laser On time. Laser Off Position is a ratio of the total scan width, with increments equal to 1/100th of the total scan width.

Laser Power Allows the user to select the Laser Power setting as follows: Low: Laser Power = ~0.6mW. Typically only needed for decoding high density symbols located close to the scanner. Medium: Laser Power = ~1.0mW. High: Laser Power = ~1.5mW. Typically used for decoding lower density symbols located far from the scanner.

QX-830 Compact Industrial Scanner User Manual

5-69

Page 136

Symbologies

Symbologies Click the App Mode button and then the Parameters button to display the tree control tabs. Then click the Symbologies tab to display the Symbologies tree control.

To open nested options, single-click the +.

The * indicates that the setting is the default. To change a setting, double-click the setting and use the cursor to scroll through the options.

Note: Symbologies settings can also be sent to the scanner from ESP’s Terminal using Microscan’s K command format. Refer to the Symbologies section of Appendix E. 5-70

QX-830 Compact Industrial Scanner User Manual

Page 137

Scanner Parameters

Code 39 Code 39 is considered the standard for non-retail 1D symbology. An alphanumeric symbology with unique start/stop code patterns, composed of 9 black and white elements per character, of which 3 are wide.

Check Character Status Enables or Disables the check character.

Check Character Output Status Check Character Output Status, when added to the symbol, provides additional data security. When enabled, the check character character is read and compared along with the symbol data. When disabled, symbol data is sent without the check character.

Note: With Check Character Output Status enabled and an External Edge, External Level, or Serial Data trigger option enabled, an invalid check character calculation will cause a No Read Message to be transmitted at the end of the read cycle.

Large Intercharacter Gap Large Intercharacter Gap is helpful for reading symbols that are printed out of specification. When enabled, the scanner can read symbols with gaps between symbol characters that exceed three times (3x) the narrow element width.

Important: Do not use Large Intercharacter Gap with Narrow or Narrow, Enhanced Quiet Zone enabled, because a large intercharacter gap (over 3x) could cause a narrow quiet zone (5x) to be interpreted as an intercharacter gap. QX-830 Compact Industrial Scanner User Manual

5-71

Page 138

Symbologies

Fixed Symbol Length Status When enabled, the scanner will check the symbol length against the symbol length field. If disabled, any length will be considered valid.

Fixed Symbol Length Fixed Symbol Length helps prevent truncations and increases data integrity by ensuring that only one symbol length will be accepted. Specifies the exact number of characters that the scanner will recognize (this does not include start and stop and check character characters). The scanner ignores any symbology that does not match the specified length.

Full ASCII Set Must be enabled when reading characters outside the standard character set (0-9, A-Z, etc.) The user must know in advance whether or not to use the Full ASCII Set option. Since Full ASCII Set requires two code words to encode one character, it is less efficient. Standard Code 39 encodes 43 characters; zero through nine, capital “A” through capital “Z”, minus symbol, plus symbol, forward slash, space, decimal point, dollar sign, and percent symbol. When Full ASCII Set is enabled, the scanner can read the full ASCII character set, from 0 to 255.

5-72

QX-830 Compact Industrial Scanner User Manual

Page 139

Scanner Parameters

Code 128 Code 128 is a smaller symbology useful in applications with limited space and high-security requirements. A very dense alphanumeric symbology. It encodes all 128 ASCII characters, it is continuous, has variable length, and uses multiple element widths measured edge to edge.

Fixed Symbol Length Status When enabled, the scanner will check the symbol length against the symbol length field. If disabled, any length will be considered a valid symbol.

Fixed Symbol Length Fixed Symbol Length helps prevent truncations and increases data integrity by ensuring that only one symbol length will be accepted. This specifies the exact number of characters that the scanner will recognize (this does not include start, stop, and check character characters). The scanner ignores any symbol not having the specified length.

QX-830 Compact Industrial Scanner User Manual

5-73

Page 140

Symbologies

EAN Status When this field is disabled, the scanner will not check any Code 128 labels for conformance to EAN requirements, or perform any special formatting. When enabled, the scanner can read symbols with or without a function 1 character in the first position. If a symbol has a function 1 character in the first position, it must conform to EAN format. Symbols that conform to EAN format will also be subject to the special output formatting options available in this command. If EAN status is required, the scanner will only decode symbols that have a function 1 character in the first position and that conform to EAN format. All symbols read will be subject to the special output formatting options available in this command.

Note: Code 128 status must be enabled for EAN status to be active.

Output Format In Standard, the scanner will not apply special EAN output formatting options. In Application, the scanner will apply the special EAN output formatting options to decoded EAN-conforming symbols.

5-74