The text of a manual · part 7 of 15
MS-710 Scanner User's Manual, pages 61 to 70
Pages 61 to 70 of 145
The original document is the PDF (1.9 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Microscan MS-710
Page 61
Output Menu Partial Output Default:
Disabled
Options:
Disabled, Enabled
When enabled, allows you to pre-select portions of labels to be transmitted by the scanner. In multilabel mode, partial output is performed on each separate label. For example, if Start Position is set to 3 and Partial Length is set to 5, the following labels are transmitted as follows:
2–Menu Config.
1234567890 as “34567” 1234 as
“34”
1234,abcde as “34,cde” (two labels with separator) 12 as
““
Start Position Default:
1
Options:
1 to 63
Allows you to determine the first character from the beginning of a label to transmit.
Partial Length Default:
63
Options:
1 to 63
Allows you to determine the number of characters to be transmitted.
MS-710 Scanner User’s Manual
2-37
Page 62
Chapter 2
Setup Menu
2–Menu Config.
The Setup menu allows you to set specific bar code scanning parameters that affect decode rate, and if equipped with the MS-sensor, allows you to determine why the scanner did not decode a specific label within the triggered read cycle.
Figure 2-10 Setup Menu Structure
Default Clicking the DEFAULT button will reset all parameters on the current menu to defaults. This does not take effect until it is saved to the scanner with the SEND/NO SAVE or SEND/SAVE buttons. Note: Unlike most configuration settings Scanner Type, Scans per Second, and Gain Adjustment are not saved for power-on (but are initialized) when you press Y. To save new settings to NOVRAM for these parameters, enter the <Zp> command. See “Save Scanner Type, SPS, and Gain for Power-on” on “Save Scanner Type, SPS, and Gain for Power-on,” on page 4-12.
2-38
MS-710 Scanner User’s Manual
Page 63
Setup Menu
Factory Settings Scans per Second Default:
50. Corresponds to 500 (scans per second) displayed in the menu.
Options:
30 to 55. Multiply the number to be entered by 10 for actual scans per second.
Allows you to set the number of scans per second by controlling mirror motor speed. The range is 300 to 550 scans per second.
Default:
45 (nominal)
Options:
10 to 255
2–Menu Config.
Gain Adjustment
Note: Gain Adjustment is optimized at the factory before shipment; your default value may not be 45. Allows you to specify a gain value for the analog circuitry. Changes to Gain should typically be done by qualified technicians. Important: Before saving any changes to Gain Adjustment (see “Save Scanner Type, SPS, and Gain for Power-on” on “Save Scanner Type, SPS, and Gain for Power-on,” on page 4-12), first record the optimized factory setting for future reference.
Scanner Type Default:
Standard if the unit is configured for low density; High Density if the unit is configured for high density.
Options:
Standard, High Density
The Scanner Type “default” from the factory indicates the density type of the scanner (with the unit’s optics adjusted accordingly). Do not change this setting unless instructed to do so by a Microscan representative. Important: Entering the <Zp> command (see “Save Scanner Type, SPS, and Gain for Power-on” on page 4-12) will save the current setting for Scanner Type to NOVRAM. If you need to verify the density type of a unit, see your FIS number and contact your Microscan representative.
AGC Sampling Default:
Continuous
Options:
Continuous, Disabled, Leading Edge
Allows you to specify how automatic gain control sampling will be done. Continuous mode continually samples 14 transitions and stores the highest value of the samples and adjusts the AGC value at the end of the scan. Leading Edge mode finds the leading edge by looking for a 40 µs quiet zone followed by 14 transitions and stores the highest value of the samples and adjusts the AGC at the end of the scan. Leading Edge is especially effective when labels are skewed. Changes to AGC Sampling should typically be done by qualified technicians. MS-710 Scanner User’s Manual
2-39
Page 64
Chapter 2
MS-sensor Parameters1 Default:
Disabled
Options:
Disabled, Enabled
2–Menu Config.
Note: See “MS-Sensor” on page A-24. The specular sensor - once you adjust it to your environment - determines why the scanner fails to decode a label within the triggered read cycle. It then outputs the appropriate message to you. The following three conditions are possible: 1. the object’s bar code is defective (you receive a bad bar code message) 2. there is no bar code on the object (you receive a no bar code message) 3. there is no object (you receive a no object message) The sensor is adjusted by the values you assign to Minimum Label Transitions and Reflectance Threshold. These settings are derived from the extended decode rate test, explained on “Extended Mode” on page 4-4.
Minimum Label Transitions Default:
14
Options:
3 to 255
Allows you to define the minimum number of label space transitions (bars and spaces) at which a label is detected. The number you enter is derived by testing several objects with labels while in Extended Mode (see “Extended Mode,” on page 4-4). Select a number that is one or two numbers less than the smallest value displayed for number of label space transitions.
Reflectance Threshold Default:
40
Options:
0 to 255
Allows you to set the reflectance threshold - the level at which an object is detected. The number is derived by testing the reflectivity of empty space. Select a number that is one or two numbers greater than the value posted when nothing is in the scanner’s beam. See “MS-Sensor” on page A-24.2
1. Applies only to FIS options that include the MS-sensor. 2. Reflectance can also be modified by sending the <KQ offset> test command which samples 50 scans for the highest reflectance value and adds a user-defined offset to this value.
2-40
MS-710 Scanner User’s Manual
Page 65
Setup Menu No Barcode Status Default:
Disabled
Options:
Disabled, Enabled
Allows you enable or disable the no label message.
No Barcode Message Default:
NOLABEL
Options:
Up to seven ASCII characters (except NUL)
Bad Barcode Status Default:
Disabled
Options:
Disabled, Enabled
Allows you to enable or disable the bad label message.
Bad Barcode Message Default:
BADCODE
Options:
Up to seven ASCII characters (except NUL)
Allows you to define the message that is sent when a label is defective.
No Object Status Default:
Disabled
Options:
Disabled, Enabled
Allows you to enable or disable the no object message.
No Object Message Default:
MISSING
Options:
Up to seven ASCII characters (except NUL)
Allows you to define the message that is sent when an object is not present.
MS-710 Scanner User’s Manual
2-41
2–Menu Config.
Allows you to define the message that is sent when an object is present without a label.
Page 66
2–Menu Config.
Chapter 2
2-42
MS-710 Scanner User’s Manual
Page 67
Chapter
3
Serial Configuration
Chapter Contents
Most of the configuration changes to the scanner that can be made in the menu can also be accomplished by command strings from the host. Serial command strings are entered from an ASCII terminal. As with menu configuration commands, serial configuration commands relate to the initial scanner setup. To quickly reference the serial configuration commands, see table 3-1 on page 3-2.
MS-710 Scanner User’s Manual
3-1
3–Serial Config.
About Serial Configuration .................................................................. 3-3 Communications Port Commands ...................................................... 3-5 Protocol Commands ........................................................................... 3-6 Operations Commands ..................................................................... 3-10 Code Types Commands ................................................................... 3-13 Scanner Output Commands ............................................................. 3-15 Scanner Setup Commands ............................................................... 3-18
Page 68
Chapter 3 Table 3-1 Summary of Serial Configuration Commands Function Comm. Port
Protocol
3–Serial Config.
Operations
Code Types
Scanner Output
Scanner Setup
All
3-2
K
Command
Format
Ka Ky KT Kf Kd Ke Kc KA KB Kg Kh Ki Kj Km
Host Port Aux Port Communications Status Request Communications Protocol Preamble Postamble LRC Response Timeout Intercharacter Delay Triggering Mode End of Read Cycle Serial Trigger Character External Trigger Level Good Decode Reads
<Kabaud,parity,stop bits,data bits> <Kymode,baud,parity,stop bits,data bits,daisy chain enable,daisy chain id> <KT?> <Kfprotocol,address> <Kdstatus,ASCII characters> <Kestatus,ASCII characters> <Kcstatus> <KAtimeout setting> <KBnumber> <Kgmode,filter time> <Khmode,time> <Kicharacter> <Kjmode> <Kmnumber> <Kntype,sequential matching,match start position,match length,wild card character,sequence on noread,sequence on every mismatch> <Kzstatus> <KLnumber of labels,multilabel separator> <KV?> <Kpstatus,check digit status,check digit output status,large intercharacter gap,fixed code length status,code length> <Kqstatus,start & stop match status,start & stop output status,large intercharacter gap,fixed code length status,code length,check digit type,check digit output> <Krstatus,check digit,check digit output,length 1,length 2> <Ksstatus,EAN status,supplementals status,separator status,separator char.> <Ktstatus,fixed length,length> <Konarrow margins status,symbology ID status> <P> <KW?> <Kvoutput on,polarity,pulse width,# before output> <Kwoutput on,polarity,pulse width,# before output> <Klstatus,when to output> <Kkstatus,output> <Kustatus,volume> <KCstatus> <KSserial command echo status,serial command beep status,control/hex output> <KYstatus,start position,length> <KX?> <KPdensity> <KEspeed> <KDgain adjustment,AGC sampling> <KHstatus,minimum transitions,reflectance> <KQoffset> <KMstatus,message> <KNstatus,message> <KOstatus,message> <KU?> <K?>
Kn
Match Code Type
Kz KL KV
New Master Pin Number of Labels Operations Status Request
Kp
Code 39
Kq
Codabar
Kr
Interleaved 2 of 5
Ks
UPC/EAN
Kt Ko P KW Kv Kw Kl Kk Ku KC
Code 128 Narrow Margins/Symbology ID Autodiscriminate All Codes Code Types Status Request Relay-1 Relay-2 Bar Code Output Noread Message Beeper Laser On/Off
KS
Serial Verification
KY KX KP KE KD KH KQ KM KN KO KU K?
Partial Output Scanner Output Status Request Scanner Type Scans per Second Auto Gain Control MS-sensor Setup Reflectance Setup Test Bad Barcode Message No Barcode Message No Object Output Message Scanner Setup Status Request Return all K Commands
MS-710 Scanner User’s Manual
Page 69
About Serial Configuration
About Serial Configuration Serial Configuration Command Format The format for a serial configuration command is, <Kparameterdata,data,...etc.><initializing command>
Where: • Less than < and greater than > symbols are included as part of the commands. • “parameter,” as used here, are those character(s) that precede the data. • The “initializing command” <A> or <Z> is sent after configuration is complete. <Z> resets and saves for power up. <A> initializes the change to RAM.1 (For more information, see “Reset,” on page 4-11.) For example, the following command enables UPC and saves the change for power-on: <Ks1><Z>.
• Parameters and data are “case sensitive.” That is, characters must be entered as upper or lower case, as specified. • All data fields (except the last) must be followed by a comma (without a space). • If there is no change in a given field, then commas can be entered alone, or with the existing data (for example, <Ka,,,0> or <Ka4,1,0,0>). • All fields preceding the modified field must be included. For example, in Host Port, to change Data Bits to Eight without changing any other field, enter either: <Ka,,,1> or <Ka4,1,0,1>. • All fields following the modified field can be left out. For example, in the RS-232 port, to change Baud Rate to 4800, enter <Ka3>. (See examples on following pages.)
1. See “Initializing serial configuration commands” on page 3-4 for definitions and examples. MS-710 Scanner User’s Manual
3-3
3–Serial Config.
When using serial configuration commands, note also the following conventions:
Page 70
Chapter 3 Initializing serial configuration commands To ensure that a serial configuration command will take effect, you need to follow it with one of the operational commands below: <A> To reset but not save changes for power-on <Z> To reset and save changes for power-on For example, to change Baud Rate and reset without saving changes for power-up, enter <Ka3><A>. To change Baud Rate and reset, saving the changes to NOVRAM, enter <Ka3><Z>. To ensure that a command was received and accepted, you can enter the Show Scanner Status command, <?>.
Concatenating Serial Commands
3–Serial Config.
Commands can be concatenated (added together) to a maximum of 64 characters in a single string or data block. Additional data blocks of 64 or less characters can be sent provided there is at least a 10 ms pause between blocks. For example, <Kc1><Kh1><Ko1><A> enables LRC, sets End of Read Cycle mode to New Trigger, enables Narrow Margins and resets the data buffers (without saving the changes for power-on).
Serial Command Status Request The status of serial commands can be requested by entering the command followed by a question mark. For example, enter <Ke?> to request the status of Postamble. Commands KT?, KV?, KW?, KX?, and KU? are used to request the status of groups of serial commands (see table 3-1 on page 3-2).
Loss of Communications Making changes to communications parameters such as Baud Rate, Parity, Stop Bits, LRC, etc. without corresponding changes in linked device(s) can result in the loss of menu access. If this should occur, default the scanner with the <Zd> Restore/Save Default Configuration for Power-on command as described on page 4-12. CAUTION: Defaulting the scanner will reset all standard configuration parameters except Scanner Type, Scans per Second, and Gain Adjustment to their original default values. Note: Power must be available to the scanner during the default procedure.
3-4
MS-710 Scanner User’s Manual