Page 21
Test Decode Rate with Static Label a) Position the label in front of a functioning scanner, click on the Utilities button, and in the Read Rate dialog either click the Percent or Decodes/Sec radio button and click Start, or enter the <Cp> or <C> command from the host terminal to launch test. b) Observe rate. (The percent or decode rate is added to the front of the data string.) c) End the test by clicking the Stop button or the <J> command. Note: If, after making adjustments, you have changed the read range or another parameter used in the number of scans formula, you should recalculate the number of scans (see “Number of Scans” on page A-11). Note: Variations between labels are common. For this reason, the greater number of sample labels you test, the more likely you are to achieve optimum decode rates.
MS-710 Scanner User’s Manual
1-9
1–Quick Start
Step 8
Page 22
1–Quick Start
Chapter 1
Step 9
Test Decode Rate with Moving Label
Once you’ve mounted the scanner and established a good read rate with a stationary label, you can repeat the test with a moving label. To ensure reliable scanning, you need to apply a minimum of five scans to each label. This means that all variables, label speed, read range, etc. are significant. To calculate in advance the number of scans that the label will receive, use the formulas described in Appendix F, “Number of Scans,” on page A-11.
1-10
MS-710 Scanner User’s Manual
Page 23
1–Quick Start
Step 10 Increase the Number of Scans Where possible, you can increase the number of scans a label receives by: • Increasing label height • Decreasing label length • Increasing scan rate (spinning mirror speed) • Slowing label speed See Appendix F, “Number of Scans,” on page A-11.
MS-710 Scanner User’s Manual
1-11
Page 24
1–Quick Start
Chapter 1
1-12
MS-710 Scanner User’s Manual
Page 25
Chapter
2
Menu Configuration 2–Menu Config.
Chapter Contents About the Menu Configuration Program .................................................. 2-2 Communications Menu............................................................................ 2-5 Protocol Menu ....................................................................................... 2-12 Operation Menu..................................................................................... 2-16 Code Type Menu ................................................................................... 2-25 Output Menu ......................................................................................... 2-32 Setup Menu ........................................................................................... 2-38
With Microscan’s ESP™ (Easy Setup Program) configuration changes can be made in the ESP menus and sent and saved to your scanner. You can also send serial commands to the scanner via the ESP’s Terminal window. ESP can receive configuration settings from a scanner and post these in the Current Menu Settings window. Multiple configuration settings can be saved to host computer can later be recalled and loaded into any MS-710 scanner.
MS-710 Scanner User’s Manual
2-1
Page 26
Chapter 2
About the Menu Configuration Program Pull down menus Toolbar
2–Menu Config.
An “Untitled...” file loads with the original default settings Access configuration menu changes here. Emulation window displays echoed commands and data Also titled “Current Menu/Scanner Settings” following a save or upload You can also access the configuration menus by clicking any item here. Allows you to perform tests and access operational information
Allows you to download/upload between the scanner and the host Figure 2-1 Configuration Program - Main Menu
Using ESP to Configure the Scanner This is the window you will see when you launch the Easy Setup Program. To access the configuration menus, click on the MENU button. If using a mouse, single click on menu items and enter text as appropriate. If using the keyboard, use the Alt key plus underlined letters and the tab key to move to the desired field, use arrow keys or the space bar to scroll through selections, and the Enter key to accept changes.
2-2
MS-710 Scanner User’s Manual
Page 27
About the Menu Configuration Program
Saving Settings to the Scanner You can access the SEND/RECV box by clicking it directly from the Current Settings window, or from within any of the configuration menus, click Apply. To save changes to the scanner, select Apply in the menu pages. (Clicking OK on these menu pages will only update the items in the Current Settings window.) From within the SEND/ RECV box, click SEND/NO SAVE to initialize, send data, and reset counters only, or SEND/SAVE to initialize, reset counters, and save these settings to the scanner’s nonvolatile memory.
To update Current Settings in ESP, click SEND/RECV and then RECEIVE FROM SCANNER. Note: If receiving scan data and you are in Continuous Read, labels being scanned may interfere with your data request.
Saving Settings to the Host Computer When saving a file by selecting SAVE AS (or SAVE) from the File pull-down menu, the settings are saved only on the computer’s hard drive or other storage medium. This is particularly useful for saving special sets of settings that you may want to recall later and use for a different application.
Creating a New Configuration File To create a new file of configuration settings, first be sure you’ve saved the current configuration file to the computer’s hard drive or other storage device. Next pull down the File menu in the ESP and select New. A new Untitled window will open with a listing of the default settings. Change these settings to suit your requirements and save this file to the computer’s hard drive under a new name. These settings can also be downloaded to the scanner at any time under a name of your choice.
Restoring Default Settings Note: Be certain that you’ve saved any other configuration files to the host computer that you may want to reload later. To restore defaults for individual menus in ESP, click the DEFAULT button on the specific menu. To save those settings to the scanner, click Apply. To restore defaults to the scanner (for ALL menus), click the DEFAULT button on the Interface tab of the Utilities menu, or send an <Ad> serial command. You can also default by hardware command.
MS-710 Scanner User’s Manual
2-3
2–Menu Config.
Receiving Settings from the Scanner
Page 28
Chapter 2 Loss of Communications Making changes to communications parameters such as Baud Rate, Parity, Stop Bits, LRC, etc. without corresponding changes in linked device(s) can result in the loss of menu access. If this should occur, try ESP Auto Baud command or restore defaults. Caution: Defaulting the scanner will reset all standard configuration parameters except Scanner Type, Scans per Second, and Gain Adjustment to their original default values.
2–Menu Config.
Note: Power must be available to the scanner during the default procedure.
2-4
MS-710 Scanner User’s Manual
Page 29
Communications Menu
Communications Menu
2–Menu Config.
Figure 2-2 Communications Menu Structure
Default Clicking the DEFAULT button will reset all parameters on the current menu to defaults. This does not take effect until it is saved to the scanner with the SEND/NO SAVE or SEND/SAVE buttons. Note: When you make and save menu communications changes to the scanner, those changes will automatically be matched in the Serial Port pull-down options.
Host Port Note: Changes in Communications parameters or assigning an address to the scanner can cause loss of communications with the configuration terminal when you exit the menu program (whether or not changes are saved for power-on).
MS-710 Scanner User’s Manual
2-5
Page 30
Chapter 2 Baud Rate Default:
9600
Options:
300, 600, 1200, 2400, 4800, 9600, 19.2K, and 38.4K
Allows you to set the number of bits transmitted per second.
2–Menu Config.
Parity Default:
Even
Options:
Even, Odd, None
Allows you to select an error detection routine in which one data bit in each character is set to 1 or 0 so that the total number of 1 bits in the data field is even or odd.
Stop Bits Default:
One
Options:
One, Two
Allows you to select the last one or two bits in each character to indicate the end of the character.
Data Bits Default:
Seven
Options:
Seven, Eight
Allows you to establish the total number of bits in each character.
Aux Port Parameters Options for Baud Rate, Parity, Stop Bits, and Data Bits are identical to the host port parameters on page 2-5 and are changed in the same manner.
Mode Default:
Disabled
Options:
Disabled, Transparent, Half Duplex, Full Duplex, and Daisy Chain
Note: Aux Port is only available through Microscan’s IB-105. Allows you to set communications parameters between the scanner and an auxiliary port. An auxiliary port can be used to configure the menus, send data to the host, and display data transmissions originating from the host or scanner. Note: Aux Port cannot be used in conjunction with RS-485 or RTS/CTS on the host port. Note: The Aux Port baud rate should never exceed Host Port baud rate or auxiliary port data could be lost.
2-6
MS-710 Scanner User’s Manual