Page 31

Multiple Concentrator Network

Multiple Poll Sequence First Poll Sequence. Concentrator 02 in figure 2-5 polls its decoders like a concentrator in a typical, single concentrator network. It receives a hex poll character (1C) and Decoder 01’s data, shown in figure 2-6. Start of Sequence

RES

1C R E Q

D ecoder 01

AC K 1C S TX D A TA E T X LR C

RES

Figure 2-6 Concentrator to Decoder Polling Sequence

Second Poll Sequence. When Concentrator 02 is polled by Concentrator X, Concentrator 02 responds by transmitting its hex poll character (1E), an STX, the decoder’s multidrop address (01), then the decoder's data, shown in figure 2-7. Start of Sequence

Concentrator X Concentrator 02

R ES

1E

End of Sequence

R EQ

AC K 1E STX 01D ATA ETX LR C

RES

Figure 2-7 Concentrator to Concentrator Polling Sequence

When Concentrator X relays this information to the host, it substitutes the polled concentrator’s multidrop address (02) in place of 1E hex and transmits: 0201data

“02” is Concentrator 02’s address; “01” is Decoder 01’s address.

MS-5000 Multidrop Concentrator User’s Manual

2-9

2-Multidrop Comm.

C onc entrator 02

End of Sequence

Page 32

Chapter 2

Multidrop Communications

Multiple Select Sequence When sending a select command (figure 2-5 on page 2-8) to Decoder 01 through Concentrator X and Concentrator 02, for example the Trigger Count request <T>, the host transmits: 0201<T>CR

2-Multidrop Comm.

It does not address Concentrator X because it is communicating with Concentrator X in an unpolled mode. First Select Sequence. When concentrator X receives the command, it strips away the 02, substitutes 1F hex as the select address, and forwards the remaining command, 01<T>, to Concentrator 02, shown in figure 2-8. Start of Sequence

Concentrator X

End of Sequence

RES 1F REQ

Concentrator 02

STX 01<T> ETX LRC 1F ACK

RES 1F ACK

Figure 2-8 Concentrator to Concentrator Select Sequence

Second Select Sequence. Concentrator 02 then relays the command to Decoder 01 in a typical select sequence as shown earlier in figure 2-3 on page 2-4. It strips off the 01 from the command, substitutes 1D hex as the select address, and transmits the remaining data, <T>, to Decoder 01, shown in figure 2-9. Start of Sequence

C oncentrator X C oncentrator 02

R ES 1F R E Q

End of Sequence

S TX 01< T> E T X LR C 1F AC K

RES 1F AC K

Figure 2-9 Concentrator to Decoder Select Sequence

2-10

MS-5000 Multidrop Concentrator User’s Manual

Page 33

Multiple Concentrator Network Select Commands to a Lower Level Multidrop Concentrator. Commands to lower level concentrators in multiple concentrator networks (such as Concentrator 01 or 02 in figure 2-5) include both the concentrator’s address and a 51. For example, to send <Ke1> (Postamble Enable) to Concentrator 02, the host transmits 0251<Ke1>CR in the first sequence. Concentrator X strips away the 02, substitutes 1F as the select address, and forwards the remaining command string to Concentrator 02 as shown in figure 2-10.

Concentrator X Concentrator 02

RES 1F R EQ

End of Sequence

STX 51<Ke1> ETX LR C 1F ACK

RES

1F ACK

Figure 2-10 Concentrator to Lower Level Concentrator Select Sequence

Host-to-Concentrator Polled Network If the host application’s software (or a programmable logic controller (PLC)) Host Polled Protocol can mimic a concentrator, the MSMultidrop or User 5000 concentrator itself can be configDefined Multidrop RS-485 Interface ured in Multidrop (or User Defined Multidrop) and assigned an address. Series (See figure 2-11.) 5000 Concentrator This type of network uses an RS-485 01 Multidrop ONLY (polled) interface between the conRS-485 Interface centrator and the host. (LAN) Select Command to a Multidrop Concentrator. When the host sends Decoder 02 a select command to a concentrator Decoder 03 that’s configured in Multidrop, it transmits the select address (1D hex for Decoder 04 multidrop concentrator address 01 as shown in figure 2-11). It also prefaces Figure 2-11 Special Multidrop Network the command with a 51 just as was done in figure 2-10.

MS-5000 Multidrop Concentrator User’s Manual

2-11

2-Multidrop Comm.

Start of Sequence

Page 34

Multidrop Communications

2-Multidrop Comm.

Chapter 2

2-12

MS-5000 Multidrop Concentrator User’s Manual

Page 35

Chapter

3

Menu Configuration Menu Configuration

Using the Menu Configuration Program............................................... 3-2 Communications Menu ........................................................................ 3-5 Host Protocol Menu ............................................................................. 3-9 Polled Device Status Menu................................................................ 3-15 Time Keeping Menu........................................................................... 3-15 Exiting and Saving Menu Changes.................................................... 3-17

Note: Menu configuration in a multidrop network, as described in this chapter, applies only to the MS-5000 concentrator. Configuration changes to decoders and scanners on a multidrop network can be made with serial (“K”) commands (including device address), or through the menus with an auxiliary monitor or profile card.

MS-5000 Multidrop Concentrator User’s Manual

3-1

3–Serial Config.

Chapter Contents

Page 36

Chapter 3

Menu Configuration

Using the Menu Configuration Program Entering the Menu Configuration Program1

3–Serial Config.

To see the Main menu (figure 1-1), from a host or terminal connected to the host port of the concentrator, enter 51<D>CR.2

Figure 3-1 Main Menu

• 51 is the code which every concentrator recognizes as its own address • <D> is the command to enter the Main menu • CR (carriage return) terminates the command Note: The carriage return is enabled as the default ETX character. If ETX is undefined (nulled), then a time delay exceeding the intercharacter timeout must be inserted before transmitting another command to a different address. Note: Caps Lock must be on during menu configuration.

1. See Table A-1, “ASCII Table with Control Characters,” on page A-3. 2. This command must be sent as a data block from the host or as a string from a softkey because of the intercharacter timeout (default =12 ms).

3-2

MS-5000 Multidrop Concentrator User’s Manual

Page 37

Using the Menu Configuration Program

Navigating the Menu Configuration Program The bottom line on the screen is called the command line. The command line identifies your place in the menu program, shows current status, and allows you to review and change options. Use the following keys to scroll to and select the parameter you wish to change: SP (space bar key) or N to scroll ahead, B to scroll back, CR (carriage return key) to select, and M to return to the previous higher level menu. To return to the Main menu at any time, press ESC (escape key) and M. For example, to change Turnaround Timeout to 10 (see Figure 3-2, “Communications Menu Structure,” on page 3-5), use the following command line path: This is the first prompt displayed in the Main Menu. Press CR to access the Communications menu.

Communications, > Host Port Communications, > LAN Port Communications, > Aux Port

Since Turnaround Timeout is a subtopic of Host Port, press CR to access the Host Port parameters.

Host Port, > Baud Rate = 9600 Host Port, > Parity = Even Host Port, > Stop Bits = One Host Port, > Data Bits = Seven Host Port, > RS-422 = Disabled Host Port, > Turnaround Timeout = 15 Host Port, > Intercharacter Timeout = 15

Baud Rate is the first parameter under Host Port. Press SP until you reach Turnaround Timeout.Press CR to change.

Host Port, > Turnaround Timeout = 15 ––>

Type in 10, and press CR.

To view Turnaround Timeout’s new status in the menu, press M to refresh the screen. To return to the Main menu, press M again. To return to the Main menu at any time, press ESC (escape key) and M.

MS-5000 Multidrop Concentrator User’s Manual

3-3

3–Serial Config.

Main, > Communications Main, > Host Protocol Main, > Device Status Main, > Time Keeping

Page 38

Chapter 3

Menu Configuration

Saving Menu Changes Press ESC (escape key) to see the following on the command line: Press M to return to the Main menu, or press E to exit the Menu Configuration program. If E is pressed, the following question appears: Press N to exit without saving changes, or press Y to retain the current settings for power-on. If Y is selected, a second beep indicates that the save has been carried out.

Loss of Communications

3–Serial Config.

Making changes to communications parameters such as Baud Rate, Parity, and Stop Bits without corresponding changes in linked device(s) can result in the loss of menu access. If this should occur, default the concentrator (see Appendix C, “Defaulting the MS-5000”). Note: Power must be available to the concentrator during the default procedure.

Defining Special Characters To define any control character from the ASCII table: Press SP once, then enter the control character by holding down the control key and simultaneously pressing the desired character. For example to define a line feed, press SP, then Control and J simultaneously. It is displayed as ^J on the command line and as <LF> in the menu when the screen is refreshed. To define CR as a character: Press SP, then CR. It is displayed as ^M on the command line and as <CR> in the menu when the screen is refreshed. To define a space as a character: Press SP twice. It is displayed as a blank space in the menu when the screen is refreshed. While it appears that nothing has been assigned, the hex value 20 will be sent during data transmission. To select NUL as the character: Press SP, then a 0 (zero). It is displayed as <NUL> in the menu when the screen is refreshed.

3-4

MS-5000 Multidrop Concentrator User’s Manual

Page 39

Communications Menu

Communications Menu (For Host, LAN, and Auxiliary Ports) Communications

Bold text represents default settings.

Host Port Stop Bits

Data Bits

– Even

– One

– Seven

– Disabled

– 19.2K – 2400

– Odd

– Two

– Eight

– Enabled

– 38.4K – 4800

– None

Baud Rate – 9600

Parity

– 1200

RS-422

– 600

Intercharacter Poll Response Timeout Timeout

Turnaround Timeout LAN Port Baud Rate

– 15 ms

– 15

– 15

– 1 to 250

– 1 to 250

– 1 to 250

Data Bits

– One

– Seven

– 19.2K – 2400

– Odd

– Two

– Eight

– 38.4K – 4800

– None

Parity

– 1200

Number of Devices – 50

RS-422 – Disabled – Enabled

– 1 to 50

– 600

Aux Port

– 9600

Stop Bits

Data Bits

Aux Mode

– Even

– One

– Seven

– Enabled

– Disabled

– Odd

– Two

– Eight

– Disabled

– Enabled

Parity

Baud Rate – 2400

– 19.2K – 4800

RS-422

– None

Figure 3-2 Communications Menu Structure

Note: Since the procedure for changing communications parameters is identical for each port (Host, LAN, and Aux), this section shows command line examples for Host Port only. Some parameters are not applicable (N/A) to every port; thus each port will have a unique order of command lines. The differences are noted in the following text.

Baud Rate is the first parameter for Host and LAN Port. Aux Mode is the first parameter for Aux Port.

MS-5000 Multidrop Concentrator User’s Manual

3-5

3–Serial Config.

Stop Bits

– Even

– 9600

Page 40

Chapter 3

Menu Configuration Baud Rate Default: 9600 baud for all three ports. Options: Host and LAN: 600, 1200, 2400, 4800, 9600, 19.2K, and 38.4K. Aux Port options: 2400, 4800, 9600, and 19.2K. Number of bits transmitted per second. Note: Due to timing considerations, polled modes require 2400 baud or faster.

3–Serial Config.

Parity Default: Even Options: Even, Odd, None. Error detection routine in which one data bit in each character is set to 1 or 0 so that the total number of 1 bits in the data field is even or odd. Stop Bits Default: One Options: One, Two Last one or two bits in each character to indicate the end of the character. Data Bits Default: Seven Options: Seven, Eight Number of bits in each character. RS-422 Default: RS-232 Options: RS-232, RS-422 Used only in Point-to-Point protocol, and not with RTS or CTS. RS-422 is not used with Aux Port. (See appendix F, “Interface Standards,” for an explanation of RS-422.) When RS-422 is disabled, RS-232 is enabled. An exception to this is that whenever Multidrop or User Defined Multidrop is selected, the functioning standard is RS-485 regardless of the displayed status of RS-422 in the menus. Aux Mode Default: Enabled Options: Enabled, Disabled (Aux Port only) Used only when connected to the Aux Port (labeled “MONITOR” on the concentrator). Aux Mode can be used for menu configuration (see serial commands), device status monitoring, and for displaying host messages.

3-6

MS-5000 Multidrop Concentrator User’s Manual