Page 41

Using ESP

Send/Receive Options To access Receive, Save and Default options, click the Send/Recv button. You can also access these options by right-clicking in any of the configuration views.

Receiving

From the Send/Recv menu, select Receive Reader Settings. Caution: Do not select this option if you do not want to upload the Imager’s settings. For example, if your ESP file has a number of custom settings that you want to maintain and download into the Imager, these settings would be lost by choosing Yes. This is useful if you want to receive (upload) the Imager’s settings and save them as a file for future use. For example, if your Imager has settings that you do not want to change, choosing Yes would allow you to load those settings to ESP and save them in an ESP file for later retrieval. Receiving the Imager’s settings will also assure that you will not be subsequently saving any unwanted changes that you or someone else has made previously in ESP.

Saving

1. Send, No Save (<A>) This saves ESP settings to current memory. 2. Send and Save (<Z>) This activates all changes in current memory and saves to the Imager for power-on.

3. Send and Save as Customer Defaults (<Zc>) Use this option to save your own default settings for quick retrieval with a <Zrc> command. This option will be visible only if you had checked Enable ‘Send and Save as Customer Defaults’ in the ESP Preferences dialog.

MS-4 Imager User Manual

2-15

Page 42

Send/Receive Options

Defaulting

When you select Default Current Menu Settings or Default all ESP Settings, you are only defaulting the ESP settings.

Advanced Options Send Current View

This is the same as Save to Reader > Send No Save except that only the commands in the current configuration tree are sent.

Send Current Command

This is the same as Send Current View , except that it saves only the command that is currently selected.

Add / Remove Exception

After you perform a Receive Reader Settings command1 and you click on the Add Exception option, you may see a list of serial commands. These are commands that may be in your Imager’s firmware, but not included in, or different from, your current version of ESP. You can edit these commands by double-clicking on them and changing them as needed. It is important to note that these commands will be saved to your Imager whenever you send a Save to Reader command, or an <A> or a <Z> command. Also, if there is a corresponding ESP menu item, the ESP Value column for that item will be blank following a Receive Reader Settings command.

1. From the Send/Recv button or by right-clicking from within the configuration trees.

2-16

MS-4 Imager User Manual

Page 43

Using ESP

Using EZ Trax For detailed information about using EZ Trax, refer to the Help menu in EZ Trax software, or the EZ Trax Quick Start Guide, available on the Microscan Tools Drive. For information about how to configure the MS-4 for use with EZ Trax, see Setting Up the Imager for EZ Trax in Chapter 5, Read Cycle.

MS-4 Imager User Manual

2-17

Page 44

Using EZ Trax

2-18

MS-4 Imager User Manual

Page 45

3 Communications Contents Communications by ESP .............................................................................................................. 3-2 Communications Serial Commands.............................................................................................. 3-3 Host Port Connections .................................................................................................................. 3-4 Host Port Protocol......................................................................................................................... 3-5 ACK/NAK Options......................................................................................................................... 3-7 Polling Mode Options.................................................................................................................... 3-8 RS-422 Status............................................................................................................................... 3-9 Auxiliary Port Connections.......................................................................................................... 3-10 Auxiliary Port System Data Status .............................................................................................. 3-18 Daisy Chain Autoconfigure ......................................................................................................... 3-19 Response Timeout...................................................................................................................... 3-20 LRC Status.................................................................................................................................. 3-21 Protocol Configuration Examples................................................................................................ 3-22 USB HID Interface ...................................................................................................................... 3-23 ASCII Character Entry Modifier................................................................................................... 3-32 Preamble..................................................................................................................................... 3-33 Postamble ................................................................................................................................... 3-34

This section explains how to set up communications between the Imager and a host. With Microscan’s ESP (Easy Setup Program), configuration changes can be made in the ESP menus and then sent and saved to your Imager. The user can also send serial commands to the Imager via ESP’s Terminal window.

MS-4 Imager User Manual

3-1

Page 46

Communications by ESP

Communications by ESP

Click this button to bring up the Applications menu.

Click this button to bring up the Communication menu.

To open nested options, single click the +.

3-2

To change a setting, double click the setting and use your cursor to scroll through the options.

MS-4 Imager User Manual

Page 47

Communications

Communications Serial Commands Host Port Connections Auxiliary Port Connections RS-422 Status Host Port Protocol Preamble Postamble Response Timeout LRC Auxiliary Port System Data Status ACK/NAK Options Polling Mode Options USB HID Reports Autoconfiguration Daisy Chain

MS-4 Imager User Manual

<K100,baud rate,parity,stop bits,data bits> <K101,aux port mode,baud rate,parity,stop bits,data bits,daisy chain ID status,daisy chain ID> <K102,status> <K140,protocol,address> <K141,status,preamble characters> <K142,status,postamble characters> <K143,response timeout> <K145,status> <K146,aux port system data status> <K147,RES,REQ,STX,ETX,ACK,NAK> <K148,RES,REQ,STX,ETX,ACK,NAK> <K149,Microscan HID IN Report status,Keyboard IN Report status> <K150DAISY>

3-3

Page 48

Host Port Connections

Host Port Connections The host port can be configured with RS-232 and RS-422 connections. The following settings define the basic transmission speeds and digital standards that ensure common formatting.

Baud Rate, Host Port Usage: Definition: Serial Cmd: Default: Options:

Can be used to transfer data faster or to match host port settings. The rate at which the reader and host transfer data back and forth. <K100,baud rate,parity,stop bits,data bits> 115.2K 0 = 600 1 = 1200 2 = 2400 3 = 4800 4 = 9600 5 = 19.2K 6 = 38.4K 7 = 57.6K 8 = 115.2K 9 = 230K

Parity, Host Port Usage: Definition: Serial Cmd: Default: Options:

Only changed if necessary to match host setting. An error detection routine in which one data bit per character is set to 1 or 0 so that the total number of bits in the data field is either even or odd. <K100,baud rate,parity,stop bits,data bits> None 0 = None 1 = Even 2 = Odd

Stop Bits, Host Port Usage: Definition: Serial Cmd: Default: Options:

Only changed if necessary to match host setting. One or two bits added to the end of each character to indicate the end of the character. <K100,baud rate,parity,stop bits,data bits> One 0 = One 1 = Two

Data Bits, Host Port Usage: Definition: Serial Cmd: Default: Options:

Only changed if necessary to match host setting. One or two bits added to the end of each character to indicate the end of the character. <K100,baud rate,parity,stop bits,data bits> Eight 0 = Seven 1 = Eight

Note: See USB HID Interface on page 3-23 for protocol information relating to USB. 3-4

MS-4 Imager User Manual

Page 49

Communications

Host Port Protocol Usage: Definition: Serial Cmd: Default: Options:

In general, the point-to-point protocols will work well in most applications. They require no address and use RS-232 or RS-422 communications standards. Protocols define the sequence and format in which information is transferred between the reader and the host, or in the case of Multidrop, between the readers and a concentrator. <K140,protocol,address> Point-to-Point 0 = Point-to-Point 1 = Point-to-Point with RTS/CTS 2 = Point-to-Point with XON/XOFF 3 = Point-to-Point with RTS/CTS and XON/XOFF 4 = ACK/NAK 5 = Polling Mode

Note: In all protocol modes, the preamble <K141> and postamble <K142> character strings can be used to frame the decode data, and both are included in calculating the LRC Status (Longitudinal Redundancy Check).

Point-to-Point (Standard) Usage: Definition:

Used only with RS-232 or RS-422. Standard Point-to-Point requires no address and sends the data to the host whenever it is available, without a request or handshake from the host.

Serial Cmd:

<K140,0>

Point-to-Point with RTS/CTS Usage:

Definition: Serial Cmd:

A reader initiates a data transfer with an RTS (request-to-send) transmission. The host, when ready, responds with a CTS (clear-to-send) and the data is transmitted. RTS and CTS signals are transmitted over two dedicated wires as defined in the RS-232 standard. Used only with RS-232. Point-to-Point with RTS/CTS (request-to-send/clear-to-send) is a simple hardware handshake protocol that allows a reader to initiate data transfers to the host. <K140,1>

Point-to-Point with XON/XOFF Usage: Definition: Serial Cmd:

If an XOFF has been received from the host, data will not be sent to the host until the host sends an XON. During the XOFF phase, the host is free to carry on other chores and accept data from other devices.Used only with RS-232. This option enables the host to send the XON and XOFF command as a single byte transmission command of start (^Q) or stop (^S). <K140,2>

MS-4 Imager User Manual

3-5

Page 50

Host Port Protocol

Point-to-Point with RTS/CTS and XON/XOFF Usage: Definition:

Used only with RS-232. This option is a combination of Point-to-Point with RTS/CTS and Pointto-Point with XON/XOFF.

Serial Cmd:

<K140,3>

ACK/NAK Definition:

See the ACK / NAK Options command <K147> on page 3-7.

Serial Cmd:

<K140,4>

Polling Mode Definition: Serial Cmd:

See the Polling Mode Options command <K148> on page 3-8. <K140,5>

Poll Address Serial Cmd: Default: Options:

<K140,protocol,address> 1 1 to 50 1 = Poll address 0x1C, Select address 0x1D 2 = Poll address 0x1E, Select address 0x1F ... 50 = Poll address 0x7E, Select address 0x7F

Note: See USB HID Interface on page 3-23 for protocol information relating to USB.

3-6