Page 281
Configuration Database
Database Mode Switch Mode Selects the event that causes the Imager to load the next database entry to current, active settings. When Frame Count/Time expires and Image Process Looping is enabled, the next database entry with modifications to camera settings will be used. Note: The image capture event always occurs when the first database entry is used. Note: The Switch Mode setting has no effect on Rapid Capture Mode, which always operates in Number of Image Frames mode with a frame count of 1. Serial Cmd: <K256,switch mode,frame count/time,image process looping,image dimensions> Default: 1 Options: 0 = Time 1 = Number of Image Frames Definition:
Time When Switch Mode is set to Time, the Imager will load the next database entry to current, active settings after a predefined time interval. The timer will start upon use of a database entry. If the timer expires during an image capture event, the timer will not start again until that database entry has been incremented and the new database entry has been loaded to current, active settings.
Number of Image Frames When Switch Mode is set to Number of Image Frames, the database entry is incremented after the predetermined number of image capture events has occurred.
Frame Count/Time Indicates the Number of Image Frames that must be captured or the amount of Time that must transpire before the Imager will load the next database index entry. Serial Cmd: <K256,switch mode,frame count/time,image process looping,image dimensions> Default: 1 (frames/mS) Options: 1 to 65535 Definition:
MS-4 Imager User Manual
11-9
Page 282
Database Mode
Image Process Looping Useful in applications where it is necessary to process a single captured image multiple times using different IP and decode parameters. Serial Cmd: <K256,switch mode,frame count/time,image process looping,image dimensions> Default: Disabled Options: 0 = Disabled 1 = Enabled Usage:
Disabled When Image Process Looping is set to Disabled, an image is captured for every database configuration, whether or not any camera settings have been modified.
Enabled When Image Process Looping is set to Enabled, the last captured image frame is re-processed with the new IP and decode parameters. If camera settings have not been changed from the last capture event, and when a database configuration is loaded to current, active settings, no new image is captured. Note: An exception to this is the first database index: a new image is always captured when the first database index is used. When a camera setting has been modified from one database setting to the next, a new image is captured. For example, if all database entries contained the same camera setting values but had different IP and decode parameters, an image frame would only be captured when the first database configuration was used.
11-10
MS-4 Imager User Manual
Page 283
Configuration Database
Image Dimensions Definition: Determines how the image dimension parameters will be implemented. Serial Cmd: <K256,switch mode,frame count/time,image process looping,image dimensions> Default: 0 Options: 0 = Window of Interest 1 = Region of Interest
Window of Interest (WOI) by ESP When Image Dimensions is set to Window of Interest, the database image dimension parameters are camera settings, and they determine the size of the image to be captured.
Camera settings (Capture Index, Shutter Speed, Gain, Sub-Sampling) located in the left half of the highlighted area. Image capture dimensions (Row Pointer, Column Pointer, Row Depth, Column Width) located in the right half of the highlighted area.
MS-4 Imager User Manual
11-11
Page 284
Database Mode
Region of Interest (ROI) When Image Dimensions is set to Region of Interest, the database image dimension parameters are IP and decode settings, and they determine the region or area of the captured image to be processed. ROI coordinate data is based on the full image size. The WOI of the captured image will be configured to cover all the ROI settings in the current, active database entry. In the following example, there are three active database settings, each with a different ROI configuration. Their coordinates are based at point 0,0 of the full scale image (640 x 480). In this example, DB1 and DB3 determine the size of the captured image WOI, while DB2 has no impact. The image WOI is not configurable. It is automatically set up by the database according to the ROI settings. row pointer column pointer
DB 1 IP ROI row size
DB 2 IP ROI column size
DB 3 Camera WOI
IP ROI
Full Image Size (640 x 480) Note: Since the ROI parameters are not a camera setting, a change in ROI parameters from one index to another does not indicate an image capture event. This feature is intended to be used in conjunction with Image Process Looping, to allow different regions of a captured image to be processed using different IP and decode settings. Also, if this feature is used in conjunction with Output Filtering, multiple decoded symbols in a captured image frame can be output according to their location in the field of view. 11-12
MS-4 Imager User Manual
Page 285
Configuration Database
Region of Interest (ROI) by ESP When Image Dimensions is set to Region of Interest, the database image dimension parameters are IP and decode settings, and they determine the region or area of the captured image to be processed.
Threshold Mode, Fixed Threshold Value, Processing Mode, Narrow Margins, Background, and Code Type are all Processing Settings values.
MS-4 Imager User Manual
11-13
Page 286
Save Current Settings to Configuration Database
Save Current Settings to Configuration Database Usage: Serial Cmd:
Allows current, active configuration settings to be saved to a selected database index. <K255+,index>
Example: <K255+,5> This command phrase would save the Imager’s current, active configuration settings to database index 5.
By ESP
Click the Load Current to Index button to save the Imager’s current configuration parameters to the selected database index.
11-14
MS-4 Imager User Manual
Page 287
Configuration Database
Load Current Settings from Configuration Database Definition: Serial Cmd:
Allows the configuration settings contained in a selected database index to be loaded to current, active configuration settings. <K255-,index>
Example: <K255-,5> This command phrase would load the configuration settings contained in database index 5 to current, active configuration settings.
Load Current Settings from Configuration Database by ESP
Click the Load Index to Current button to load configuration settings from the selected index to current, active settings.
Notes on Symbol Type • The current DataBar Expanded status does not change if it is configured as Enabled (Stacked) and the database DataBar Expanded status is Enabled. • The current DataBar-14 status does not change if it is configured as Enabled (Stacked) and the database DataBar Expanded status is Enabled. • Data Matrix ECC level is determined by the current settings and not by database settings. Therefore, the database does not know which ECC level to enable, and has no effect on current Data Matrix ECC settings.
MS-4 Imager User Manual
11-15
Page 288
Request Selected Index Settings
Request Selected Index Settings Definition: Serial Cmd:
Returns configuration settings for the selected database index. <K255?,index>
Example: <K255?,5> This command phrase would return the configuration settings for database index 5.
11-16
MS-4 Imager User Manual
Page 289
Configuration Database
Request All Configuration Database Settings Definition: Serial Cmd:
Returns configuration settings for all indexes in the configuration database. <K255?>
Example: <K255?> This command returns the configuration settings for all 10 database indexes.
MS-4 Imager User Manual
11-17
Page 290
Request All Configuration Database Settings
11-18
MS-4 Imager User Manual