Page 361

Appendices

Polling Mode Data Flow Examples Setup 1 Address RES REQ STX ETX ACK NAK LRC

0x01 (translates to) Poll Req @ ‘0x1C’, Unit Select @ ‘0x1D’ 0x04 0x05 0x02 0x03 0x06 0x15 disabled

Transfer 1 HOST_TX IMAGER_TX HOST_TX IMAGER_TX HOST_TX

‘RES’ ‘0x1D’ ‘REQ’ (Select Unit 1 to receive data) ‘0x1D’ ‘ACK’ (Unit responds with its address) ‘STX’ <T> ‘ETX’ ‘0x1D’ ‘ACK’ (Unit responds with its address) ‘RES’ (Terminate Transfer 2)

Transfer 2 HOST_TX IMAGER_TX HOST_TX IMAGER_TX

‘RES’ ‘0x1C’ ‘REQ’ (Poll Unit 1 for data) ‘0x1C’ ‘STX’ <T/00000> ‘ETX’ ‘ACK’ ‘RES’ (Terminate Transfer 1)

Starting with a ‘RES’ ensures a clean transaction, without “leftovers” from the previous transaction.

Error Condition 1 HOST_TX ‘RES’ ‘0x1C’ ‘REQ’ (Poll Unit 1 for data) IMAGER_TX ‘0x1C’ ‘STX’ <T/00000> ‘ETX’ HOST_TX ‘Nothing’ (Host should ‘ACK’ here) timeout reached... IMAGER_TX ‘REQ’ (Unit requests an ‘ACK’ again) timeout reached... IMAGER_TX ‘REQ’ (Unit requests an ‘ACK’ again) timeout reached... IMAGER_TX ‘REQ’ (Unit requests an ‘ACK’ again) timeout reached... IMAGER_TX ‘RES’ (Terminate Transfer 1, data is flushed)

MINI Hawk High Performance Imager User Manual

A-21

Page 362

Communications Protocol

Error Condition 2 HOST_TX ‘RES’ ‘0x1C’ ‘REQ’ (Poll Unit 1 for data) IMAGER_TX ‘0x1C’ ‘STX’ <T/00000> ‘ETX’ HOST_TX ‘Nothing’ (Host should ‘ACK’ here) timeout reached... IMAGER_TX ‘REQ’ (Unit requests an ‘ACK’ again) HOST_TX ‘NAK’ (Host rejects data frame) (Retry Event) IMAGER_TX ‘0x1C’ ‘STX’ <T/00000> ‘ETX’ (Unit sends again) HOST_TX ‘ACK’ (Host receives data) IMAGER_TX ‘RES’ (Terminate Transfer 1)

The protocol makes 3 retry attempts before data is flushed and transfer is aborted.

A-22

MINI Hawk High Performance Imager User Manual

Page 363

Appendices

Setup 2 Address RES REQ STX ETX ACK NAK LRC

0x01 (translates to) Poll Req @ ‘0x1C’, Unit Select @ ‘0x1D’ 0x04 0x05 0x02 0x03 0x06 0x15 enabled

Transfer 1 HOST_TX IMAGER_TX HOST_TX IMAGER_TX HOST_TX

‘RES’ ‘0x1D’ ‘REQ’ (Select Unit 1 to receive data) ‘0x1D’ ‘ACK’ (Unit responds with its address) ‘STX’ <T> ‘ETX’ ‘LRC’ ‘0x1D’ ‘ACK’ (Unit responds with its address) ‘RES’ (Terminate Transfer 2)

Transfer 2 HOST_TX IMAGER_TX IMAGER_TX HOST_TX

‘RES’ ‘0x1C’ ‘REQ’ (Poll Unit 1 for data) ‘0x1C’ ‘STX’ <T/00000> ‘ETX’ ‘LRC’ ‘ACK’ ‘RES’ (Terminate Transfer 1)

Starting with a ‘RES’ ensures a clean transaction, without “leftovers” from the previous transaction.

Error Condition 1 HOST_TX ‘RES’ ‘0x1C’ ‘REQ’ (Poll Unit 1 for data) IMAGER_TX ‘0x1C’ ‘STX’ <T/00000> ‘ETX’ ‘LRC’ HOST_TX ‘Nothing’ (Host should ‘ACK’ here) timeout reached... IMAGER_TX ‘REQ’ (Unit requests an ‘ACK’ again) timeout reached... IMAGER_TX ‘REQ’ (Unit requests an ‘ACK’ again) timeout reached... IMAGER_TX ‘REQ’ (Unit requests an ‘ACK’ again) timeout reached... IMAGER_TX ‘RES’ (Terminate Transfer 1, data is flushed)

MINI Hawk High Performance Imager User Manual

A-23

Page 364

Communications Protocol

Error Condition 2 HOST_TX ‘RES’ ‘0x1C’ ‘REQ’ (Poll Unit 1 for data) IMAGER_TX ‘0x1C’ ‘STX’ <T/00000> ‘ETX’ ‘LRC’ HOST_TX ‘Nothing’ (Host should ‘ACK’ here) timeout reached... IMAGER_TX ‘REQ’ (Unit requests an ‘ACK’ again) HOST_TX ‘NAK’ (Host rejects data frame) (Retry Event) IMAGER_TX ‘0x1C’ ‘STX’ <T/00000> ‘ETX’ ‘LRC’ (Unit sends again) HOST_TX ‘ACK’ (Host receives data) IMAGER_TX ‘RES’ (Terminate Transfer 1)

Error Condition 3 HOST_TX IMAGER_TX HOST_TX (Retry Event) IMAGER_TX HOST_TX IMAGER_TX

‘RES’ ‘0x1C’ ‘REQ’ (Poll Unit 1 for data) ‘0x1C’ ‘STX’ <T/00000> ‘ETX’ ‘BAD LRC’ ‘NAK’ (Host rejects bad LRC data) ‘0x1C’ ‘STX’ <T/00000> ‘ETX’ ‘GOOD LRC’ (Unit sends again) ‘ACK’ (Host receives data) ‘RES’ Terminate Transfer 1)

The protocol makes 3 retry attempts before data is flushed and transfer is aborted.

A-24

MINI Hawk High Performance Imager User Manual

Page 365

Appendices

Appendix E, ASCII Table Dec 00 01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31

Hex 00 01 02 03 04 05 06 07 08 09 0A 0B 0C 0D 0E 0F 10 11 12 13 14 15 16 17 18 19 1A 1B 1C 1D 1E 1F

Mne NUL SOH STX ETX EOT ENQ ACK BEL BS HT LF VT FF CR SO SI DLE DC1 DC2 DC3 DC4 NAK SYN ETB CAN EM SUB ESC FS GS RS US

Ctrl ^@ ^A ^B ^C ^D ^E ^F ^G ^H ^I ^J ^K ^L ^M ^N ^O ^P ^Q ^R ^S ^T ^U ^V ^W ^X ^Y ^Z ^[ ^\ ^] ^^ ^_

Dec 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63

Hex 20 21 22 23 24 25 26 27 28 29 2A 2B 2C 2D 2E 2F 30 31 32 33 34 35 36 37 38 39 3A 3B 3C 3D 3E 3F

MINI Hawk High Performance Imager User Manual

Ch SP ! “ # $ % & ' ( ) * + , . / 0 1 2 3 4 5 6 7 8 9 : ; < = > ?

Dec 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95

Hex 40 41 42 43 44 45 46 47 48 49 4A 4B 4C 4D 4E 4F 50 51 52 53 54 55 56 57 58 59 5A 5B 5C 5D 5E 5F

Ch @ A B C D E F G H I J K L M N O P Q R S T U V W X Y Z [ \ ] ^ _

Dec 96 97 98 99 100 101 102 103 104 105 106 107 108 109 110 111 112 113 114 115 116 117 118 119 120 121 122 123 124 125 126 127

Hex 60 61 62 63 64 65 66 67 68 69 6A 6B 6C 6D 6E 6F 70 71 72 73 74 75 76 77 78 79 7A 7B 7C 7D 7E 7F

Ch ` a b c d e f g h i j k l m n o p q r s t u v w x y z { | } ~ D

A-25

Page 366

Interface Standards

Appendix F, Interface Standards Interface Standards, established by the Electronic Industries Association (EIA), specify such things as the signaling voltage levels, maximum cable lengths, and number of drivers. In the MINI Hawk, selection of interface is made by pin assignment and, in the case of host communications, by software switching between RS-232 and RS-422/RS-485.

RS-232 RS-232 defines an interface between two devices such as, for example, the reader and host. It differs from the other interfaces by dedicating individual pins to specific functions and by requiring both devices to share a common ground line. Since both device chassis are connected to a common ground, a ground loop potential and the possibility of noise interference exists. Therefore cable lengths are limited to a maximum of 50 feet (19.7m). Despite being the most limited, this interface is used frequently because of the large installed base of RS-232 equipment.

RS-422/RS-485 RS-422, unlike RS-232, measures signals deferentially; that is, the receiver looks at the potentials between the two receive (or transmit) wires rather than the potential between signal and ground. As a result, cables, if shielded, can be up to 4000 feet (1219m) in length. Like RS-232, RS-422 communication is designed for only two devices on a single line and must have a common ground. It can be used wherever RS-232 is used.

USB The Universal Serial Bus can connect up to 127 devices to a host at a maximum of 6 megabytes per second (Mbps) of bandwidth.

A-26

MINI Hawk High Performance Imager User Manual

Page 367

Appendices

Appendix G, Operational Tips Cleaning The MINI Hawk has a hard-coated window that should only be cleaned with alcohol (100% isopropyl).

Mounting When mounting the MINI Hawk, do not insulate the stand mount. The bottom panel is the hottest part of the imager and metal-to-metal contact is necessary for heat dissipation.

MINI Hawk High Performance Imager User Manual

A-27

Page 368

Using an External Trigger

Appendix H, Using an External Trigger Two types of inputs are used on Omron Microscan readers. 5 volt devices have NPN trigger inputs. 10-28 volt devices have optoisolated inputs and outputs.

5 Volt Devices To trigger a 5 volt device, the input has to be pulled down to “signal ground” (NPN). This can be done by relays or by a transistor-sourced signal. If a 5 volt device is used in combination with an IC-332, it must be connected as a 10-28 volt device. Reader

Reader

Sensor

Sensor

10-28 Volt Devices To trigger a 10-28 volt device, a small amount of current needs to flow from “Trigger +” to “Trigger –”. If a PNP source is used, “Trigger –” must be connected to “signal ground”. If an NPN source is used, “Trigger +” must be connected to the supply voltage. Reader

Reader

Sensor

Reader

Sensor

Sensor

PNP

A-28

NPN

MINI Hawk High Performance Imager User Manual

Page 369

Appendices

Appendix I, USB-to-Serial Virtual COM Port Driver (For use with Windows® 7 (32-bit), Vista (32-bit), XP (32-bit), and 2000 Operating Systems) Omron Microscan’s USB-to-Serial Virtual COM Port Driver allows you to send serial data using a USB MINI Hawk.

Installing the USB-to-Serial Virtual COM Port Driver 1. Download the driver files from the Omron Microscan website or Omron Microscan Tools Drive. If the driver files are in a .zip file, extract them to a folder of your choice before running the installer. 2. Double-click on DPInst.exe to launch the Omron Microscan USB-to-Serial Virtual COM Port Driver Installer. Select DPInst.exe to launch the driver installer.

3. Click Next on the driver installer’s welcome screen.

MINI Hawk High Performance Imager User Manual

A-29

Page 370

USB-to-Serial Virtual COM Port Driver 4. Read the License Agreement shown below. If you accept the terms of the agreement, click the radio button next to the statement “I accept this agreement”, and click Next.

Important: After accepting the license agreement and clicking Next, you may see a Windows Security warning that states “Windows can’t verify the publisher of this driver software”. If you receive this warning, choose the “Install this driver software anyway” option. 5. The driver files will begin installing. This may take several seconds. Do not interfere with the installer during this process.

6. Once the driver is installed successfully, you will receive the message shown below. Click Finish.

A-30