Page 301

Terminal

Macros Macros can be stored in a macro selection bar, edited in a separate window, and executed by clicking on the macro name.

Click on Next Row to see the next row of macros

Click on the first arrow here to see Add Macro or Default Macros. When you default macros, the entire macros set is restored to their original macro commands.

Click on subsequent arrows to edit macros.

When you click on the macro name, the macro is executed in the Terminal window. If this is a command, it is sent to the reader at the same time that it is displayed.

Editing a Macro When you click the arrow next to any macro and select Edit, the following appears:

You can edit an existing macro or type in the Macro Name text field and define it in the Macro Value text field. Click OK.

MINI Hawk High Performance Imager User Manual

12-5

Page 302

Terminal Window Menus

Terminal Window Menus Right-click on the Terminal window to display the following menu: • Copy selected text to clipboard. • Paste from Terminal or other text. • Clear all text in Terminal window. • Select All text in the Terminal window. • Save... brings up a Save As dialog. • Change Font... of text in Terminal; brings up a Font dialog. • Change Echo Font... to change typed text; brings up a Font dialog. • Enable Echo enables Echo text (typed by user). • Change Background Color of Terminal window. • Non-Printable Characters allows you to hide non-printable characters, or to show them in Standard or Enhanced format. • Default Settings returns all the above settings to default. • Keyboard Macros allows you to create new keyboard macro commands that can be sent from function keys (F2, F4, F5, etc.)

Terminal Dropdown Menu The dropdown Terminal menu has Capture Text, Save Current Text, Send File, Find Next, and Find Previous functions, as well as the same functions defined above. • Capture Text... lets you append data in real time to a text file of your choice. While in operation, the text file cannot be opened. You can select Pause to interrupt the capture flow or Stop to end the flow and open the file. • Save Current Text... saves all text in the Terminal window to a text file. • Send File allows you to browse for specific files and send them to the reader. • Find Next searches for a user-defined section of text in the Terminal. • Find Previous operates in the same way as Find Next, but searches backward through Terminal text.

12-6

MINI Hawk High Performance Imager User Manual

Page 303

13 Utilities Contents Serial Utility Commands ............................................................................................................. 13-2 Read Rate .................................................................................................................................. 13-4 Counters ..................................................................................................................................... 13-5 Device Control ............................................................................................................................ 13-7 Differences from Default............................................................................................................. 13-8 Master Database ........................................................................................................................ 13-9 Firmware................................................................................................................................... 13-15 Default/Reset/Save................................................................................................................... 13-18 Imager Status Requests ........................................................................................................... 13-20 Learn Operations...................................................................................................................... 13-22 Other Operational Serial Commands ....................................................................................... 13-23

Utility commands are generally commands that are performed during imager operations to check or determine read rates, or to perform miscellaneous operations on imager hardware. Serial utility commands are not prefaced with a “K” and a numeric code, nor do they require an initialization command (<A> and <Z>). They can be entered from within any terminal program or from ESP’s Terminal or Utilities windows.

MINI Hawk High Performance Imager User Manual

13-1

Page 304

Serial Utility Commands

Serial Utility Commands Note: For a list of all K commands, see Serial Configuration Commands on page A-10. Type Read Rate

Counter Request and Clear

Master Database

Part Number / Checksum

Device Control

Default/Reset/Save

Command <C> <Cp> <J> <N> <O> <T> <U> <V> <W> <X> <Y> <G> <Gn> <NEWM> <#> <#a> <#b> <#p> <!> <!a> <!b> <!p> <L1> <L2> <L3> <I1> <I0> <I> <H> <a1> <A> <Ard> <ArdWB> <Arp> <Arc>

13-2

Name Enter Decode Rate Test Enter Percent Rate Test Exit Decode Rate and Percent Rate Test No Read Counter No Read Counter Reset Trigger Counter Trigger Counter Reset Good Read/Match Counter Good Read/Match Counter Reset Mismatch Counter Mismatch Counter Reset Store Next Symbol Read to Database Index 1 Store Next Symbol Read to Database Index n New Master Load Status Display All Firmware Part Numbers Display Application Code Part Number Display Boot Code Part Number Display Profile Module Part Numbers Display All Available Firmware Checksums Display Application Code Code Checksum Display Boot Code Checksum Display Profile Module Checksum Programmable Output 1 Programmable Output 2 Programmable Output 3 Targeting System On Targeting System Off Disable Imager Enable Imager Include PDF Information in Preamble (PDF417) Reset (does not save for power-on) Reset and Recall Omron Microscan Defaults Reset and Recall Omron Microscan White Balance Defaults Reset and Recall Power-On Parameters Reset and Recall Customer Default Parameters MINI Hawk High Performance Imager User Manual

Page 305

Utilities <Z> <Zc> Default/Reset/Save

<Zrd> <ZrdWB> <Zrc>

Imager Status Request Commands

Learn

Other Operational Commands

<?> <K?> <K??> <K?#> <Knnn?> <Knnn??> <Knnn?#> <LEARN> <UNLEARN> <LEARN?> <uy,path,filename> <dy,path,filename> <op,9> <@CAL> <BCCFG> <VAL> <@WB>

MINI Hawk High Performance Imager User Manual

Save Current Settings for Power-On Save Current Settings as Customer Default Parameters for Power-On Recall Omron Microscan Default Parameters and Save for Power-On Reset Omron Microscan Default White Balance Parameters and Save for Power-On Recall Customer Default Parameters and Save for Power-On Imager Status Byte All Status Request All Descriptor Status Request All Range Status Request Single Status Request Single Descriptor Status Request Single Range Status Request Initiate Learn Operation Initiate Unlearn Operation Display Learn Status Y-Modem Upload Transfer Options Y-Modem Download Transfer Options Image Library Request Autocalibrate Enter Bar Code Configuration Static Validation White Balance Calibration

13-3

Page 306

Read Rate

Read Rate Click this button to bring up the Utilities view.

Read Rate by ESP To see the number of decodes per second, click the Decodes/sec radio button on the Read Rate tab in the Utilities view and click the Start button. To see the percentage of decodes, click the Percent radio button and then the Start button. To end a Read Rate test, click the Stop button (the Start button becomes a “Stop” button during an active Read Rate test.) Important: When Read Rate is enabled, the Configuration Database is not active, regardless of how the imager is configured.

Read Rate by Serial Command Enter Decodes/Second Test Sending <C> instructs the imager to transmit the decodes per second and symbol data (if any). The decode rate can vary dramatically due to the angle and location of the symbol in relation to the field of view. This test is very useful in aligning and positioning the imager during setup.

Enter Percent Test Sending <Cp> instructs the imager to transmit the percentage of decodes and any decoded symbol data.

End Read Rate Test Sending <J> ends both the Percent test and the Decodes/Second test.

13-4

MINI Hawk High Performance Imager User Manual

Page 307

Utilities

Counters Counter commands can be a numeric value from 00000 to 65535. After reaching the maximum numeric limit of 65535, an error message will be displayed and the counter will automatically roll over and start counting again at 00000. To obtain the cumulative total of counts after the rollover has occurred, add 65536 per each rollover (the imager does not keep track of the number of rollovers) to the current count. Note: All counter values will be lost if power is recycled to the imager, or if the imager receives a Reset or Save command.

Counters by ESP You can access Counters from ESP’s Utilities menu. Click the Request button to display the appropriate count or Clear to set the counter to zero.

MINI Hawk High Performance Imager User Manual

13-5

Page 308

Counters by Serial Command No Read Counter Sending <N> displays the total number of No Reads that have occurred since the last reset.

No Read Counter Reset Sending <O> sets the No Read Counter to 00000.

Trigger Counter Sending <T> displays the total number of triggers since the last reset.

Trigger Counter Reset Sending <U> sets the trigger counter to 00000.

Good Read/Match Counter (or Good Read Counter) Sending <V> displays the total number of good reads matching the master symbol, or, if Master Symbol is not enabled, the number of good reads since the last reset. This counter is always enabled, but will only work as a match count when Master Symbol is enabled. If Master Symbol is not enabled, this counter records the number of good reads. This count can be requested at any time.

Good Read/Match Counter Reset Sending <W> sets the Match Counter to 00000.

Mismatch Counter Sending <X> displays the number of decoded symbols since the last reset that do not match the master symbol.

Mismatch Counter Reset Sending <Y> sets the Mismatch Counter to zero.

13-6

MINI Hawk High Performance Imager User Manual

Page 309

Utilities

Device Control Device Control by ESP

Device Control by Serial Command Output 1 Pulse Sending <L1> activates the link between Output 1 (+) and Output 1 (–) of the host connector (regardless of Master Symbol or Output 1 status).

Output 2 Pulse Sending <L2> activates the link between Output 2 (+) and Output 2 (–) of the host connector (regardless of Master Symbol or Output 2 status).

Output 3 Pulse Sending <L3> activates the link between Output 3 (+) and Output 3 (–) of the host connector (regardless of Master Symbol or Output 3 status).

Disable Reader Sending <I> will turn the imager OFF, end the current read cycle, and will not allow the imager to enter another read cycle until turned ON. This feature is useful during extended periods of time when no symbols are being decoded, or the imager is being configured. Disabling the imager will not affect any commands that have already been downloaded.

Enable Reader Sending <H> will turn the imager ON and allow it to enter read cycles.

Enable PDF Information Sending <a1> will cause PDF417 data to be prefaced with information consisting of ECC level (ECC level n), number of rows (n rows), number of columns (n columns), number of informative code words (n info code words), and the number of data characters (n data bits). Send <a1> again to disable PDF Information. MINI Hawk High Performance Imager User Manual

13-7

Page 310

Differences from Default

Differences from Default Clicking the Differences from Default button will cause ESP to check all stored configuration settings and compare them to default settings. All settings that are different than default will appear in the left column (shown below), and descriptions of those settings will appear in the right column. Click this button for a list of ESP configuration settings that are different than default settings.

Click Generate Barcode to bring up the Bar Code Dialog. Then create configuration symbols containing the configuration commands of your choice.

Send configuration settings to the imager without saving by clicking Send to Reader.

Click Save As to save the report as plain text or a tab-delimited text file.

Send configuration settings to the imager and save in ESP by clicking Send and Save.

• To create a symbol containing any of the command settings in the table, click Generate Barcode. This will bring up the Bar Code Dialog. • To save the Differences from Default report, either as plain text or as a tab-delimited text file, click Save As. • Click Send and Save to send the settings to the imager and save them, or Send to Reader to send the settings without saving them. Important: The use the Differences from Default feature, you must connect to the imager and Receive Reader Settings via the Send/Recv button on the toolbar.

13-8

MINI Hawk High Performance Imager User Manual