I/O Parameters
Configurable Output 1
Usage: Definition: Serial Cmd: Default: Options:
This option provides discrete signalling to host software to control external devices such as PLCs and relays. It is useful for routing, sorting, and to prevent mis-packaging and mis-routing. Sets the discrete output functions for specific user-selected conditions. Allows the user to set the conditions under which an output (or outputs) will be activated. <K810,output on,output state,pulse width,output mode> Mismatch or No Read 0 = Mismatch or No Read 1 = Match (or good read) 2 = Mismatch 3 = No Read 4 = Trend Analysis 5 = Symbol Quality 6 = Diagnostic Warning 7 = In Read Cycle
Note: If Output On is set to Match or Mismatch, a transition (switching) will not occur unless Matchcode Type is enabled and a master symbol is loaded into memory.
Mismatch or No Read Activates discrete output when the data does not match that of the master symbol or the symbol has not been decoded before the end of the read cycle.
Match or Good Read Activates a discrete output when the symbol data matches the master symbol. Note: If you want to output for a good read and Matchcode is not enabled, you can enable any output for Match.
Mismatch Activates a discrete output whenever the symbol data does not match that of the master symbol.
No Read Activates a discrete output whenever the symbol data is not decoded before the end of the read cycle. Quadrus MINI Imager User’s Manual
7-31
Page 212
Configurable Output 1
Trend Analysis Usage: Definition:
Typically used when successful decodes are occurring but a discrete output is needed to flag a trend in quality issues. Activates discrete output when a trend analysis condition is met, depending on the trend analysis option enabled.
Symbol Quality Usage: Definition:
Typically used when a discrete indication is needed to flag a symbol quality condition. Activates discrete output when a symbol quality condition is met, depending on the symbol quality option enabled.
Diagnostic Warning Usage: Definition:
Typically used when a discrete indication of a diagnostic condition is needed. Activates discrete output when a diagnostic warning condition is met, depending on the diagnostic option enabled.
In Read Cycle Definition:
7-32
Activates a discrete output when the imager is in a read cycle.
Quadrus MINI Imager User’s Manual
Page 213
I/O Parameters
Output State Definition: Serial Cmd: Default: Options:
Sets the active electrical state of the discrete output. <K810,output on,output state,pulse width,output mode> Negative 0 = Negative 1 = Positive
Pulse Width Definition: Serial Cmd: Default: Options:
Sets the time in 10 ms increments that the discrete output remains active. <K810,output on,output state,pulse width,output mode> 5 (Corresponds to 50 ms) 0 to 255 (0 to 2.55 seconds). Divide the number entered on the command line by 100 for time in seconds.
Output Mode Definition: Serial Cmd: Default: Options:
Sets the condition in which the discrete output is de-activated. <K810,output on,output state,pulse width,output mode> Pulse 0 = Pulse 2 = Latch Mode 2 3 = Latch Mode 3
Pulse This is the default mode of operation in which the programmable output is activated when the Output On condition has been met and held active for the duration of the selected pulse width.
Latch Mode 2 (Unlatch Opposite Condition) The programmable output is activated when the Output On condition has been met and held active until the opposite condition selected under Output On has been met. For example, if No Read is enabled under Output On, the programmable output will go active on a No Read and remain active until the opposite condition, a good read, occurs.
Latch Mode 3 (Unlatch Re-enter Read Cycle) The programmable output is active when the Output On condition has been met and is held active until a new read cycle begins. Note: All of the Output On modes are inhibited when any Output on Warning is active for Output 1.
Quadrus MINI Imager User’s Manual
7-33
Page 214
Configurable Output 1
Trend Analysis Output 1 Note: Output On under Output 1 Parameters must be set to Trend Analysis for this output to function.
Under Output 1 Parameters, expand the Trend Analysis tree:
Usage: Definition:
Example:
7-34
Useful in cases where the user doesn’t want to shut down for one condition but wants to monitor quality and read conditions. Applies Trend Analysis settings to Output 1. With Trend Analysis, the user can track the occurrences and frequency of mismatches, No Reads, and the number of reads per trigger, and output the results to any of three outputs. Trend Analysis Mode = No Read Trigger Evaluation Period= 25 triggers (read cycles) Number to Output On = 4 In this example, the imager will activate an output when 4 No Reads occur within a period of 25 triggers (read cycles).
Quadrus MINI Imager User’s Manual
Page 215
I/O Parameters
Trend Analysis Mode Definition:
Sets the trend condition (Mismatch, No Read, or Decodes/Trigger) that will activate the output.
Serial Cmd:
<K780,trend analysis mode,trigger evaluation period,number to output on, decodes/trigger threshold> No Read 0 = Mismatch 1 = No Read 2 = Decodes/Trigger 3 = Bad Symbol 4 = No Symbol
Default: Options:
Mismatch Output will be activated when the number of mismatches equals the value entered for Number to Output On within the trigger window selected in Number of Triggers.
No Read Output will be activated when the number of No Reads equals the value entered for Number to Output On within the trigger window selected in Number of Triggers.
Decodes/Trigger Output will be activated when the number of decodes equals the value entered for Number to Output On within the trigger window selected in Number of Triggers.
Bad Symbol Output will be activated when the number of Bad Symbol occurrences equals the value entered for Number to Output On within the trigger window selected in Number of Triggers. Note: Bad Symbol output is not supported when Q-Mode is enabled. This also applies to Trend Analysis Output 2 and Trend Analysis Output 3.
No Symbol Output will be activated when the number of No Symbol occurrences equals the value entered for Number to Output On within the trigger window selected in Number of Triggers. Note: No Symbol output is not supported when Q-Mode is enabled. This also applies to Trend Analysis Output 2 and Trend Analysis Output 3. Quadrus MINI Imager User’s Manual
7-35
Page 216
Configurable Output 1
Trigger Evaluation Period Definition: Serial Cmd: Default: Options:
The number of triggers to examine for the trend analysis condition. <K780,trend analysis mode,trigger evaluation period,number to output on, decodes/trigger threshold> 0 0 to 255
Number to Output On Usage:
Definition:
Serial Cmd: Default: Options:
Example: If Number to Output On is set to 3 and Trend Analysis Mode is set to No Read, then the output will not be activated until 3 No Reads have occurred. Sets the number of Trend Analysis Mode events (mismatches, No Reads, or reads/trigger as configured by Trend Analysis Mode) to occur within the trigger evaluation period before activating the associated output. <K780,trend analysis mode,trigger evaluation period,number to output on, decodes/trigger threshold> 0 0 to 255
Decodes/Trigger Threshold Definition:
Serial Cmd: Default: Options:
7-36
When set to this mode, and when the appropriate output is set to output on trend analysis, the imager will function in a Decodes per Trigger mode during the read cycle and the trend analysis operation. Output will be activated based on whether or not the symbol decode count at the end of the read cycle crosses the decodes per trigger threshold. Note: Although this setup causes the imager to function in a Decodes per Trigger mode, the decode count will only be appended to the symbol data if the status of the Decodes per Trigger command is enabled. <K780,trend analysis mode,trigger evaluation period,number to output on, decodes/trigger threshold> 0 0 to 65535
Quadrus MINI Imager User’s Manual
Page 217
I/O Parameters
ISO/IEC 16022 Symbol Quality Output 1 Note: Output On under Output 1 Parameters must be set to Symbol Quality for this output to function. Note: ISO/IEC 16022 Symbol Quality Output 1 is not supported when Q-Mode is enabled.
Under Output 1 Parameters, expand the Symbol Quality (ISO/IEC 16022) tree.
Output on Symbol Contrast Usage: Definition: Serial Cmd:
Default: Options:
Lets the user know if symbol quality is less than acceptable. If enabled, toggles Output 1 to an active state when Symbol Contrast Threshold is met. <K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial non-uniformity threshold,output on UEC,UEC threshold> Disabled 0 = Disabled 1 = Enabled
Output on Print Growth Usage: Definition: Serial Cmd:
Default: Options:
Lets the user know if symbol quality is less than acceptable. If enabled, toggles Output 1 to an active state when Print Growth Threshold is met. <K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial non-uniformity threshold,output on UEC,UEC threshold> Disabled 0 = Disabled 1 = Enabled
Quadrus MINI Imager User’s Manual
7-37
Page 218
ISO/IEC 16022 Symbol Quality Output 1
Output on Axial Non-Uniformity Usage: Definition: Serial Cmd:
Default: Options:
Lets the user know if symbol quality is less than acceptable. If enabled, toggles Output 1 to an active state when Axial Non-Uniformity Threshold is met. <K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial nonuniformity threshold,output on UEC,UEC threshold> Disabled 0 = Disabled 1 = Enabled
Output on Unused Error Correction Usage: Definition: Serial Cmd:
Default: Options:
Lets the user know if symbol quality is less than acceptable. If enabled, toggles Output 1 to an active state when UEC Threshold is met. <K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial nonuniformity threshold,output on UEC,UEC threshold> Disabled 0 = Disabled 1 = Enabled
Symbol Contrast Threshold Usage: Definition: Serial Cmd:
Default: Options:
7-38
Lets the user set the acceptable level of symbol quality. Conforms to ISO/IEC 16022 symbol quality grading (A,B,C,D). <K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial non-uniformity threshold,output on UEC,UEC threshold> Grade C 0 = Grade A 1 = Grade B 2 = Grade C 3 = Grade D
Quadrus MINI Imager User’s Manual
Page 219
I/O Parameters
Print Growth Threshold Usage: Definition: Serial Cmd:
Default: Options:
Lets the user set the acceptable level of symbol quality. Conforms to ISO/IEC 16022 symbol quality grading (A,B,C,D). <K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial nonuniformity threshold,output on UEC,UEC threshold> Grade C 0 = Grade A 1 = Grade B 2 = Grade C 3 = Grade D
Axial Non-Uniformity Threshold Usage: Definition: Serial Cmd:
Default: Options:
Lets the user set the acceptable level of symbol quality. Conforms to ISO/IEC 16022 symbol quality grading (A,B,C,D). <K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial nonuniformity threshold,output on UEC,UEC threshold> Grade C 0 = Grade A 1 = Grade B 2 = Grade C 3 = Grade D
Unused Error Correction Threshold Usage: Definition: Serial Cmd:
Default: Options:
Lets the user set the acceptable level of symbol quality. Conforms to ISO/IEC 16022 symbol quality grading (A,B,C,D). <K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial nonuniformity threshold,output on UEC,UEC threshold> Grade C 0 = Grade A 1 = Grade B 2 = Grade C 3 = Grade D
Quadrus MINI Imager User’s Manual
7-39
Page 220
Diagnostics Output 1
Diagnostics Output 1 Note: Output On under Output 1 Parameters must be set to Diagnostic Warning for this output to function.
Under Output 1 Parameters, expand the Diagnostics tree.
Serial Cmd: Default: Options:
<K790,unused 1,service unit> 0 0 = Disabled
Service Unit When Diagnostic Warning is enabled, the Output On configuration has no effect. The output will remain active as long as one of the diagnostic warning conditions is met. The output will become inactive once it detects that there are no diagnostic warning conditions. Definition:
Serial Cmd: Default: Options:
7-40
Allows the user to set up the output to toggle to active when the service timer has expired. This condition will only be held for one service timer click. Note: This feature cannot be used if the imager is in a Continuous Read mode. <K790,unused 1,service unit> 0 0 = Disabled 1 = Enabled