Page 91

Calibration

Initiating Calibration Calibration can be initiated by serial command, the EZ button, or from the Calibration interface in ESP. In ESP, the Calibrate button starts the calibration routine.

To initiate calibration by serial command, enter the serial command <@CAL> from a terminal, such as ESP’s Terminal interface. <@CAL> is a calibration routine that emphasizes contrast and readability.

See Calibrate by EZ Button on page 1-11 for the EZ button calibration procedure.

Quadrus MINI Imager User’s Manual

4-13

Page 92

Initiating Calibration

Calibration Progress Indicators During the calibration process, the imager has a number of ways to communicate its progress to the user. The following three types of feedback are the primary ways that the imager communicates calibration progress.

Read Rate LEDs When calibration is initiated through a serial command or the EZ Button, the imager’s Read Rate LEDs indicate calibration progress as follows: LEDs 20% 20%, 40% 20%, 40%, 60% 20%, 40%, 60%, 80% 20%, 40%, 60%, 80%, 100%

Calibration Progress Search Pass in progress. Focus Pass in progress. Medium Pass in progress. Fine-Tune Pass in progress. Calibrated Successfully (holds for green flash duration).

Calibrated Successfully Fine-Tune Pass Medium Pass Focus Pass Search Pass

Beeper/Green Flash The beeper and green LED flash are used to indicate calibration status as shown in the table below: Indicator 2 quick beeps 5 quick beeps 1 long beep 1 sec. Green Flash

4-14

Status Calibration has started. Calibration has failed. Calibration has been completed successfully. Calibration has been completed successfully.

Quadrus MINI Imager User’s Manual

Page 93

Calibration

Calibration Progress Output to Terminal When initiated by serial command or EZ button, the autocalibration operation will output progress and process information to a terminal if the system is configured to do so.

Terminal Depending upon how the calibration options are configured, the autocalibration process can have up to 5 steps (Quick Focus, Search Pass, Focus Pass, Medium Pass, Fine-Tune Pass). Calibration data is shown in a simple table format under 8 category headings, each of which are defined below. Heading Prog Rating Decode Quality Focus Shutter Gain Locate

Definition Progress indicator with a range of 0 - 100. Indicates percent complete. Rating indicator; higher ratings are more favorable. Number of successful decodes for the pass. Quality indicator; higher ratings are more favorable. Focus position under operation. Shutter setting under operation. Gain setting under operation. Time required to locate symbol, expressed in milliseconds.

Calibration Progress Output Examples (Terminal) Quick Focus This mode is only run if Quick Focus calibration mode is enabled (see Focus Position on page 4-5). Quick Focus calibration mode is designed to quickly locate the focus setting for an object at the center of the imager’s field of view. There is no terminal output for this operation.

Search Pass The purpose of Search Pass is to decode a symbol in as few configurations as possible during calibration. This process is designed to quickly identify symbology type, and any image processing parameters that are already calibrated, as well as determining focus position. Running Search Pass Prog

Rating

Decode

Quality

Focus

Shutter

Gain

Locate

12 14 16 18

0 0 0 2174

0 0 0 1

0 0 0 72

313 313 313 313

250 250 250 250

0 9 18 27

0 0 0 24

Quadrus MINI Imager User’s Manual

4-15

Page 94

Initiating Calibration

Focus Pass The purpose of Focus Pass is to determine the optimal focal position of the optical system. This is accomplished by determining the inside and outside focus positions and then recording the center position as the desired focal position. Running Focus Pass Prog

Rating

Decode

Quality

Focus

Shutter

Gain

Locate

38 38 38 38 38 38 38 38 38 38 38 38 38 38 38 38 38 38 38 38 42 42 42 42 42 42 42 42 42 42 46

0 0 4256 0 4236 0 4236 0 4232 0 2128 0 0 2144 0 0 0 0 0 0 4280 4280 4280 4280 4280 4280 4280 4280 4280 4280 0

0 0 2 0 2 0 2 0 2 0 1 0 0 1 0 0 0 0 0 0 2 2 2 2 2 2 2 2 2 2 0

0 0 48 0 28 0 28 0 24 0 24 0 0 40 0 0 0 0 0 0 72 72 72 72 72 72 72 72 72 72 0

353 353 353 409 409 447 447 492 492 536 536 572 572 572 601 601 601 601 601 601 297 283 270 258 247 236 227 217 209 200 601

250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250

0 9 18 0 9 0 9 0 9 0 9 0 9 18 27 36 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45

0 0 5 0 5 0 4 0 4 0 5 0 0 4 0 0 0 0 0 0 6 5 5 5 5 6 5 5 6 6 0

Inside Focus = 200; Outside Focus = 572; Center Focus = 386

4-16

Quadrus MINI Imager User’s Manual

Page 95

Calibration

Medium Pass The theory behind Medium Pass is to develop a readable range from which to dial in a decoded symbol. Running Medium Pass Prog

Rating

Decode

Quality

Focus

Shutter

Gain

Locate

51 53 55 57 59 61 63 65 67 69 71 73 75

4216 8436 6336 8448 8456 8464 8472 8488 8504 8512 8528 8584 8644

2 4 3 4 4 4 4 4 4 4 4 4 4

8 20 24 32 40 48 56 72 88 96 112 168 228

386 386 386 386 386 386 386 386 386 386 386 386 386

250 250 250 250 250 250 250 250 250 250 250 250 250

0 4 8 12 16 20 24 28 32 36 40 44 48

5 4 4 4 5 4 5 5 5 5 6 5 5

Rating List (Best): Rating = 8606; Focus = 386; Shutter = 250; Gain = 48 Rating = 8581; Focus = 386; Shutter = 250; Gain = 44 Rating = 8553; Focus = 386; Shutter = 250; Gain = 40 Rating = 8527; Focus = 386; Shutter = 250; Gain = 36 Rating = 8505; Focus = 386; Shutter = 250; Gain = 32 Rating = 8488; Focus = 386; Shutter = 250; Gain = 28 Rating = 8476; Focus = 386; Shutter = 250; Gain = 24

Quadrus MINI Imager User’s Manual

4-17

Page 96

Initiating Calibration

Fine-Tune Pass The theory behind Fine-Tune Pass is to dial in a decoded symbol based on the range supplied by the Medium Pass. Running Fine Tune Pass Prog

Rating

Decode

Quality

Focus

Shutter

Gain

Locate

77 79 81 84 86 88 90 93 95 97 100

6416 6424 6424 6432 6448 6464 6476 6488 6506 6524 6538

3 3 3 3 3 3 3 3 3 3 3

104 112 112 120 136 152 164 176 194 212 226

386 386 386 386 386 386 386 386 386 386 386

250 250 250 250 250 250 250 250 250 250 250

38 39 40 41 42 43 44 45 46 47 48

7 6 6 6 5 5 5 5 6 5 6

Rating List (Best): Rating = 6527; Focus = 386; Shutter = 250; Gain = 48 Rating = 6517; Focus = 386; Shutter = 250; Gain = 47 Rating = 6505; Focus = 386; Shutter = 250; Gain = 46 Rating = 6491; Focus = 386; Shutter = 250; Gain = 45 Rating = 6476; Focus = 386; Shutter = 250; Gain = 44 Rating = 6462; Focus = 386; Shutter = 250; Gain = 43 Rating = 6449; Focus = 386; Shutter = 250; Gain = 42 Calibration PASSED.

4-18

Quadrus MINI Imager User’s Manual

Page 97

Calibration

Additional Notes about Calibration The following conditions apply to the imager’s calibration process. Some of these items are noted at various points throughout this section, or in other sections of the imager’s documentation. 1. The Mirrored Image parameter <K514> is not part of the calibration process. 2. Background Color <K451> is calibrated. 3. If Window of Interest Framing is enabled, the WOI will be set to full frame when calibration begins. If WOI Framing is disabled, the current WOI configuration will be used for the Search Pass. 4. IP Mode <K527> will not be altered during calibration. 5. Threshold Mode <K512> will be set to Adaptive during calibration. At the end of a successful calibration, Threshold Mode will remain set to Adaptive regardless of its previous setting. Otherwise, the value will revert to its original state. 6. If Symbol Type is enabled for calibration (Autodiscriminate), then Interleaved 2 of 5 Range Mode Status (Interleaved 2 of 5) <K472> will be enabled. This allows variable-length Interleaved 2 of 5 symbols to be decoded. If an Interleaved 2 of 5 symbol is decoded during calibration, then code length # 1 will be set to the decoded symbol length at the end of calibration. Otherwise, the symbol lengths will be restored to their original configurations. 7. Pharmacode is not calibrated. 8. The symbol size, dimension, and orientation parameters of the No Symbol Message and Bad Symbol Message, and 2D Symbology Qualification command <K718> will be updated if calibration is completed successfully. 9. All symbol types that were enabled before calibration will still be enabled after calibration. For example, if Data Matrix ECC 200 was enabled before calibration, and the calibration routine was performed on a Code 128 symbol, then after calibration is completed both Data Matrix ECC 200 and Code 128 will be enabled. 10. The calibration process does not modify the global Composite status <K453>. The global Composite status must be configured properly before calibration. 11. If the user requires calibration of a stacked symbology <K482>, <K483>, or <K484>, that symbology must be enabled and configured appropriately before calibration. 12. UPC Supplementals will be disabled during calibration. 13. The Search process will use the configurable Window of Interest for image captures. However, when the search process is completed, the WOI will be reduced to include only the symbol of interest and some additional boundary area.

Quadrus MINI Imager User’s Manual

4-19

Page 98

Additional Notes about Calibration

4-20

Quadrus MINI Imager User’s Manual

Page 99

5 Read Cycle Contents Read Cycle by ESP ...................................................................................................................... 5-2 Read Cycle Serial Commands...................................................................................................... 5-3 Read Cycle Setup......................................................................................................................... 5-4 Multisymbol................................................................................................................................... 5-5 Trigger Mode and Filter Duration.................................................................................................. 5-6 External Trigger Polarity ............................................................................................................. 5-11 Serial Trigger .............................................................................................................................. 5-12 Start Trigger Character (Non-Delimited)..................................................................................... 5-13 Stop Trigger Character (Non-Delimited) ..................................................................................... 5-14 End of Read Cycle...................................................................................................................... 5-15 Capture Mode ............................................................................................................................. 5-17 Capture Timing ........................................................................................................................... 5-22 Image Processing Timeout ......................................................................................................... 5-24 Image Storage ............................................................................................................................ 5-25 Minimum Good Reads ................................................................................................................ 5-27 Compare Mode ........................................................................................................................... 5-28 Setting Up the Imager for EZ Trax.............................................................................................. 5-29

After you’ve established communications and completed basic read rate testing, you will need to address the spatial and timing parameters associated with your application. This section explains those parameters.

Quadrus MINI Imager User’s Manual

5-1

Page 100

Read Cycle by ESP

Read Cycle by ESP

Click this button to bring up the App Mode view.

Click this button to bring up the Read Cycle tree control. To open nested options, single-click the +.

5-2

To change a setting, double-click the setting and use your cursor to scroll through the options.

Quadrus MINI Imager User’s Manual