The text of a manual · part 2 of 3
LVS-9570 Operating Instructions, pages 10 to 18
Pages 10 to 18 of 27
The original document is the PDF (1.9 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Microscan LVS-9570 Verifier
Page 10
LVS-9570 Operating Instructions 4. Slowly pull down the LVS-9570 approximately 3” (76 mm) to 4” (101.6 mm). The EAN-13 and UPC-A Master Grade barcodes appear in the field of view on the Calibration tab.
EAN-13 and UPC-A Master Grade barcodes appearing in the field of view.
LVS-9570 Operating Instructions
Page 10 of 27
Page 11
LVS-9570 Operating Instructions 5. Using the mouse, click one time in the “PASS” portion of the EAN-13 or UPC-A Master Grade barcode. The blue line will move to the “PASS” portion of the barcode.
Click anywhere in the “PASS” portion of the EAN-13 or UPC-A Master Grade barcode. The blue line will move to the “PASS” portion of the code. 6. Click the “Calibrate” button.
Successful calibration is indicated by a green “Calibration OK” message.
Failed calibration is indicated by a “Calibration Needed” message and the calibration button area turns red. If calibration fails: o Re-scan the EAN-13 and UPC-A Master Grade barcodes and follow the above steps to calibrate. It may take 2 or 3 attempts before calibration is complete. o If calibration continues to fail, contact Microscan or your Microscan representative for further instructions. IMPORTANT: The calibration score will hardly ever match exactly; this is normal and acceptable as long as the scores are within +/- 5 percent.
7. When calibration is complete, click the “Grading” tab to grade the desired barcodes. To grade a barcode, scan a barcode image, draw a box around the PASS portion of the barcode, ensuring enough space for the quiet zone (see the “Quiet Zone” section for more information). The verification results display immediately. Refer to the “Grading tab” section in the “LVS-95XX Series Barcode Quality
LVS-9570 Operating Instructions
Page 11 of 27
Page 12
LVS-9570 Operating Instructions Station Operations Manual” for detailed steps on grading barcodes; this manual is located on the installation media supplied with your system. After drawing a box around the PASS portion of the barcode, the LVS-9570 indicator light turns one of the following colors: Color
Indicator light
Description
Yellow
Indicates the LVS-9570 is processing the barcode sector.
Green
Indicates all barcode parameters receive a passing grade of A, B, C, or D.
Red
Indicates if any of the barcode parameters receive a failing grade of F or if a barcode was not detected.
Note: If using the “Pass/Fail” option to grade barcodes, the LVS-9570 does not report the barcode parameters listed above. Instead, a green “Pass” message appears for passing barcodes and a red “Fail” message appears for failed barcodes. If using the “Pass/Fail” option, the indicator light turns red for failed barcodes and green for passed barcodes. Refer to the “ISO Grading” section in the “LVS-95XX Series Barcode Quality Station Operations Manual” for more information.
LVS-9570 Operating Instructions
Page 12 of 27
Page 13
LVS-9570 Operating Instructions
Hardware Overview The LVS-9570 is comprised of two major hardware components: the sensor (protected by a glass covering) and a rubber roller. Both of these components must make contact with the surface of the label. The sensor detects the image and the roller activates an encoder that allows the LVS-95XX software to collect the image. It is important that the roller rolls contiguously across the surface along with the sensor. Failure to do so results in a distorted image. A 6-foot (1.8 m) USB cable is provided with each LVS-9570 system. Some users require a longer length USB cable, which is allowable when used with a powered USB hub. USB hubs are available for purchase at local Electronics stores. Additional LVS-9570 hardware images are below.
Front of System
Sensor protected by glass covering
Rubber Roller
Handle
Left and right arrows indicate sensor location Arrow indicates center of sensor
Back of System
USB connector External power supply (not required for use with the LVS-9570) Indicator light
LVS-9570 Operating Instructions
Page 13 of 27
Page 14
LVS-9570 Operating Instructions
Scanning Direction, Speed, and Position The LVS-9570 has a scanning width of 5.4” (137.16 mm) (including the quiet zone) for barcodes in a picket fence orientation and a scanning length of 12.0” (305 mm) (including the quiet zone) for barcodes in a ladder orientation. See the “Quiet Zone” section for more information on quiet zones.
PICKET FENCE ORIENTATION
LADDER ORIENTATION
The LVS-9570 has a scanning width of 5.4” (137.16 mm) for barcodes in a picket fence orientation
The LVS-9570 has a scanning length of 12.0” (305 mm) for barcodes in a ladder orientation
Scanning Direction
Scanning Direction
5.4” (137.16 mm) Scanning Width
12.0” (305 mm) Scanning Length
LVS-9570 Operating Instructions
Page 14 of 27
Page 15
LVS-9570 Operating Instructions
Quiet Zone The Quite Zone is a clear space preceding the start character of a barcode symbol and follows the stop character. When scanning a barcode symbol, you must allow enough space for the Quiet zone. The required Quiet Zone space for each barcode varies by symbology. An error message will appear on the monitor if not enough space has been allowed for the Quiet Zone.
Linear (1D) Barcode Quiet Zones
Quiet Zone
Quiet Zone Start Character
Stop Character
Two-Dimensional (2D) Quiet Zones
Quiet Zone
LVS-9570 Operating Instructions
Page 15 of 27
Page 16
LVS-9570 Operating Instructions
Scanning Direction The LVS-9570 reads barcode labels in one direction only and must be “pulled” through a barcode label; it does not detect anything when “pushed” through a barcode label. IMPORTANT: An image is collected only when the roller is rotating while the sensor passes over the barcode image.
CORRECT
INCORRECT
“Pull” the LVS-9570 through a barcode label.
DO NOT “push” the LVS-9570 through a barcode label.
Back of system
Back of system
LVS-9570 Operating Instructions
Page 16 of 27
Page 17
LVS-9570 Operating Instructions
Scanning Speed You must slowly pull the LVS-9570 through a barcode image. If the LVS-9570 is pulled too quickly through an image, an “Overspeed Problem” message appears on the computer monitor and the image will not be captured.
LVS-9570 Operating Instructions
Page 17 of 27
Page 18
LVS-9570 Operating Instructions
Scanning Positions It is important to maintain a perpendicular orientation to the barcode image. Failure to do so may result in a distorted image. If the image displayed on the monitor appears distorted, scan the image again making sure the LVS-9570 is using a perpendicular orientation to the barcode and is pulled slowly. See the “Correct and Incorrect Images” section for examples of correctly and incorrectly scanned barcode images.
Ladder Orientation If scanning a barcode in a ladder orientation, place the LVS-9570 at the top of the barcode with the back of the system facing the operator (see “A” below). You may also scan from the bottom of the barcode as long as you are leading with the back side of the system (see “B” below). When scanning, be sure to include space for the Quiet Zone.
A
B Back of system
Begin scanning with the back side of the system
Picket Fence Orientation If scanning a barcode in a picket fence orientation, place the LVS-9570 at the top of the barcode and scan the barcode leading with the back of the system, ensuring enough space for the Quiet Zone.
LVS-9570 Operating Instructions
Page 18 of 27