Page 61

Tube/Cap Inspection

Tube/Cap Information by Serial Command Usage: Definition:

Serial Cmd:

Tube/Cap Information is useful in more complex applications that require information about the spatial orientation of test tubes and cap characteristics in addition to presence/absence information. In addition to Tube/Cap Message output, Tube/Cap Information settings can be configured to show a user-defined Separator, a detected tube or cap’s specific Tube/Cap Library Index, a detected test tube’s Tube Angle, and a detected test tube’s Tube/Cap Tip Position. <K850,separator,tube/cap library index,tube angle,tube/cap tip position,cap color,match status,tube diameter>

The following example shows a Tube/Cap Information output string: 001_+001_0001_06.RED.05_c001.d001_0001_CP Tube/Cap Tube Tube/Cap Match Sep. Sep. Sep. Cap Color Sep. Library Angle Tip Pos. Status

Sep.

Tube Dia.

Sep.

Tube/Cap Message

3 digits

1 (+/–) 3 1 4 digits char. digits char.

1 mag. color 1 color.dim 1 4 digits char. category char. char.

1 Up to 15 char. char.

“001”

“_”

“_”

“_”

“+001” “_”

“0001”

06.RED.05 “_” cxxx.dxxx “_”

“0001”

“_CP”

Separator Definition Serial Cmd: Default: Options:

Specifies the ASCII character used to separate tube/cap information. Note: Entering NUL (00 in hex) disables separator. <K850,separator,tube/cap library index,tube angle,tube/cap tip position,cap color,match status,tube diameter> _ (underscore) Any ASCII character

Tube/Cap Library Index Definition Serial Cmd: Default: Options:

When Tube/Cap Library Index is enabled, the three-digit tube/cap library index value will be appended to tube/cap detection output as a preamble. <K850,separator,tube/cap library index,tube angle,tube/cap tip position, cap color,match status,tube diameter> Disabled 0 = Disabled 1 = Enabled Notes: • This value is output in Read Rate when enabled. When enabled, the library index values are required to match in order to be part of the same output. For example, if a Cap Present that matches Tube/Cap Library Index 1 is processed 80% of the time, and a Cap Present that matches Tube/Cap Library Index 2 is processed 20% of the time, the output would be: 80% 001_CP_20% 002_CP_ • The Tube/Cap Library Index value is not required to match when requiring more than one tube/cap to qualify for an output.

EZ Match User Manual

3-17

Page 62

Tube/Cap Information

Tube Angle Definition

Serial Cmd: Default: Options:

When Tube Angle is enabled, a polarity sign (+/–) and a three-digit value representing the angle of the tube in respect to the field of view will be appended to tube/cap detection output as a preamble. A + value indicates that the tube is angled away from the center and toward the bottom of the field of view, while a – indicates that the tube is angled away from the center and toward the top of the field of view. <K850,separator,tube/cap library index,tube angle,tube/cap tip position,cap color,match status,tube diameter> Disabled 0 = Disabled 1 = Enabled Notes: • This value is not output in Read Rate results. • The Tube Angle value is not required to match when requiring more than one tube/cap to qualify for an output.

Tube/Cap Tip Position Definition

Serial Cmd: Default: Options:

When Tube/Cap Tip Position is enabled, a four-digit value will be appended to tube/cap detection output as a preamble. This value represents the distance on the X-axis from the left of the captured image to the tip of the tube or cap, in millimeters x10. Example: 0835 represents 83.5 mm. <K850,separator,tube/cap library index,tube angle,tube/cap tip position, cap color,match status,tube diameter> Disabled 0 = Disabled 1 = Enabled

Cap Color (EZ Match Color Only) Definition Serial Cmd: Default: Options:

3-18

This feature is only available for the color version of the EZ Match Imager. <K850,separator,tube/cap library index,tube angle,tube/cap tip position,cap color,match status,tube diameter> Disabled 0 = Disabled 1 = Color 2 = HSV

EZ Match User Manual

Page 63

Tube/Cap Inspection

Color (Cap Color)

Cap Color is output in three parts (magnitude-color-category), separated by a “.” character. Notes: • When Cap Color is enabled, it is output in read rate. The color output string is required to match to be part of the same output. For example, a RED cap with two different Category values will be output as follows: 80% 08.RED.12_CP_ 20% 08.RED.11_CP_ • If the cap is undetermined, the color of the cap may sometimes be undetermined as well. In such a case, a CNA (Color Not Available) message will supersede the color code. The table below shows the available output options. Magnitude

Color

1-30 1-30 1-30 1-30 1-30 1-30 1-30 1-30 1-30 1-30 1-30 1-30 0 0 0

RED ORANGE YELLOW Y_GRN GREEN B_GRN CYAN G_BLUE BLUE P_BLUE MAGENTA P_RED WHITE BLACK GRAY

C 0 0 0 0 0 0 0 0 0 0 0 0

Category Color Match Resolution H HS 0 1-6 0 1-6 0 1-6 0 1-6 0 1-6 0 1-6 0 1-6 0 1-6 0 1-6 0 1-6 0 1-6 0 1-6 0 0 1-4

HSV 1-30 1-30 1-30 1-30 1-30 1-30 1-30 1-30 1-30 1-30 1-30 1-30

Magnitude Magnitude is a value between 1 and 30 which defines the magnitude of the hue (further divides the (H)ue value to allow more resolution). This is useful in providing a distinction between different shades of a color. For example, the color PINK will have a Color value of RED, but to define PINK further, you would look at the magnitude of the Hue to describe the shade of RED. Color Color refers to the detected color of the cap. Category Category further defines the color into categories based on saturation and brightness.

EZ Match User Manual

3-19

Page 64

Tube/Cap Information

HSV (Cap Color)

When HSV is enabled, cap color values will be output in the following format: (H)ue-(S)aturation-(V)alue Example: 310-450-325 Notes: • This value is not output in read rate. • This value is not required to match when requiring more than one tube/cap to qualify for an output.

Match Status (EZ Match Color Only) Definition

Serial Cmd: Default: Options:

This feature is only available for the color version of the EZ Match Imager. Match Status is output in the following format: Format: cxxx.dxxx where “c” = cap color library match; “d” = cap dimensions library match; “xxx” = tube/cap library index (When “xxx” = “000”, no match has occurred.) While the cap library output mode will output the “best” match, the value output from this mode will offer the last matched cap library index for color and dimension. If both a color and dimension match has occurred, the last cap library index that both color and dimension matched will be output. Notes: • This value is not output in read rate. • This value is not required to match when requiring more than one tube/cap to qualify for an output. <K850,separator,tube/cap library index,tube angle,tube/cap tip position,cap color,match status,tube diameter> Disabled 0 = Disabled 1 = Enabled

Tube Diameter Definition

Serial Cmd: Default: Options: 3-20

When Tube Diameter is enabled, a four-digit value will be appended to tube/cap detection output as a preamble. This value represents the diameter of the tube in millimeters x10. Example: 0835 represents 83.5 mm. <K850,separator,tube/cap library index,tube angle,tube/cap tip position,cap color,match status,tube diameter> Disabled 0 = Disabled 1 = Enabled EZ Match User Manual

Page 65

Tube/Cap Inspection

Tube/Cap Message Formatting by Serial Command Cap Presence Message Definition: Serial Cmd: Default: Options:

Tube/Cap Detection message database index that will be used for a Cap Present condition. <K851,cap presence message index,cap absence message index,cap unknown message index,tube presence message index,tube absence message index> 1 1 to 10

Cap Absence Message Definition: Serial Cmd: Default: Options:

Tube/Cap Detection message database index that will be used for a Cap Absent condition. <K851,cap presence message index,cap absence message index,cap unknown message index,tube presence message index,tube absence message index> 2 1 to 10

Cap Unknown Message Definition: Serial Cmd: Default: Options:

Tube/Cap Detection message database index that will be used for a Cap Unknown condition. <K851,cap presence message index,cap absence message index,cap unknown message index,tube presence message index,tube absence message index> 3 1 to 10

Tube Presence Message Definition: Serial Cmd: Default: Options:

Tube/Cap Detection message database index that will be used for a Tube Present condition. <K851,cap presence message index,cap absence message index,cap unknown message index,tube presence message index,tube absence message index> 4 1 to 10

EZ Match User Manual

3-21

Page 66

Tube/Cap Information

Tube Absence Message Definition: Serial Cmd: Default: Options:

3-22

Tube/Cap Detection message database index that will be used for a Tube Absent condition. <K851,cap presence message index,cap absence message index,cap unknown message index,tube presence message index,tube absence message index> 5 1 to 10

EZ Match User Manual

Page 67

Tube/Cap Inspection

Tube/Cap Message Editing

All Tube/Cap Detection messages can be modified to meet the requirements of the specific application. Note: Tube and cap detection messages can also be edited using the <K852> command. Click the Edit Messages button to bring up the following dialog:

Place your cursor in the text field adjacent to the number of the Tube/Cap Detection message you wish to change. Type the ASCII characters of the new message and click Save.

Click Revert to return to the most recent message settings. Click Default to return to the original default message settings.

The new message will appear on the corresponding dropdown menu in the Messages view.

Hint: To enter Control characters as Tube/Cap Detection messages, right-click in any of the text fields in the Edit Messages dialog and click on Expand. To return to ASCII characters, right-click again and click on Contract. EZ Match User Manual

3-23

Page 68

Tube/Cap Information

Tube/Cap Message Editing by Serial Command

Default Tube/Cap Detection messages are shown in the table below. All Tube/Cap Detection messages can be modified to meet the requirements of the specific application. Serial Cmd: Options: Index 1 2 3 4 5 6 7 8 9 10

<K852,index,message> Any ASCII string up to 15 characters. Messages CP_ CA_ CU_ TP_ TA_ MSG6_ MSG7_ MSG8_ MSG9_ MSG10_

Note: A read cycle or read operation will format the No Read, CA_, and TA_ messages in the following order of precedence: TA_ (all), CA_ (all), No Read (all)

3-24

EZ Match User Manual

Page 69

4 Communications Contents Communications by ESP .............................................................................................................. 4-2 Communications Serial Commands.............................................................................................. 4-3 Host Port Connections .................................................................................................................. 4-4 Host Protocol ................................................................................................................................ 4-5 Poll Address.................................................................................................................................. 4-7 ACK/NAK Options......................................................................................................................... 4-8 Polling Mode Options.................................................................................................................... 4-9 RS-422 Status............................................................................................................................. 4-10 Auxiliary Port Connections.......................................................................................................... 4-11 Auxiliary Port System Data Status .............................................................................................. 4-19 Daisy Chain Autoconfigure ......................................................................................................... 4-20 Response Timeout...................................................................................................................... 4-21 LRC Status.................................................................................................................................. 4-22 Protocol Configuration Examples................................................................................................ 4-23 USB HID Interface ...................................................................................................................... 4-24 ASCII Character Entry Modifier................................................................................................... 4-33 Preamble..................................................................................................................................... 4-34 Postamble ................................................................................................................................... 4-35

This section tells how to set up communications parameters with the host and an auxiliary terminal. With Microscan’s ESP (Easy Setup Program), configuration changes can be made in the ESP menus and then sent and saved to your imager. The user can also send serial commands to the imager via ESP’s Terminal window.

EZ Match User Manual

4-1

Page 70

Communications by ESP

Communications by ESP

Click the Parameters button and then the Communication tab.

To open nested options, single-click the +. To change a setting, double-click the setting and use your cursor to scroll through the options.

4-2

EZ Match User Manual