Page 41
Using ESP
Navigating in ESP To change imager settings, or to access the Utilities, Camera, Terminal, or Output Format views, click the App Mode button.
To return to EZ Mode, click the EZ Mode button.
To make changes to configuration settings in the tree controls: 1. Left click on the + to expand menu items.
The X denotes the default option setting.
2. Double-click the desired parameter and single-click in the selection box to view options. 3. Place your cursor in the selection box, scroll down to the setting you want to change, and single-click the setting. 4. Left click again on the open screen to complete the selection. 5. Right click on the open screen and select Save to Reader to implement the command in the imager. You can send the command without saving it, or you can send and save the command simultaneously.
EZ Match User Manual
2-15
Page 42
Send/Receive Options
Send/Receive Options To access Receive, Save and Default options, click the Send/Recv button. You can also access these options by right-clicking in any of the configuration views.
Receiving
From the Send/Recv menu, select Receive Reader Settings. Caution: Do not select this option if you do not want to upload the imager’s settings. For example, if your ESP file has a number of custom settings that you want to maintain and download into the imager, these settings would be lost by choosing Yes. This is useful if you want to receive (upload) the imager’s settings and save them as a file for future use. For example, if your imager has settings that you do not want to change, choosing Yes would allow you to load those settings to ESP and save them in an ESP file for later retrieval. Receiving the imager’s settings will also assure that you will not be subsequently saving any unwanted changes that you or someone else has made previously in ESP.
Saving Send, No Save (<A>) Saves ESP settings to current memory.
Send and Save (<Z>) Activates all changes in current memory and saves to the imager for power-on.
Send and Save as Customer Defaults (<Zc>) Saves your default settings for quick retrieval. This option will be visible only if you have checked Enable ‘Send and Save as Customer Defaults’ in ESP Preferences.
2-16
EZ Match User Manual
Page 43
Using ESP
Defaulting
When you select Default Current Menu Settings or Default all ESP Settings, you are only defaulting the ESP settings.
Advanced Options Send Current View This is the same as Save to Reader > Send No Save except that only the commands in the current configuration tree are sent.
Send Current Command This is the same as Send Current View , except that it saves only the command that is currently selected.
Add / Remove Exception After you perform a Receive Reader Settings command and you click on the Add Exception option, you may see a list of serial commands. These are commands that may be in your imager’s firmware, but not included in, or different from, your current version of ESP. You can edit these commands by double-clicking on them and changing them as needed. It is important to note that these commands will be saved to your imager whenever you send a Save to Reader command, or an <A> or a <Z> command. Also, if there is a corresponding ESP menu item, the ESP Value column for that item will be blank following a Receive Reader Settings command.
EZ Match User Manual
2-17
Page 44
Send/Receive Options
2-18
EZ Match User Manual
Page 45
3 Tube/Cap Inspection Contents Tube/Cap Inspection Serial Commands ....................................................................................... 3-2 Capture and Test .......................................................................................................................... 3-3 Multi-Operation ............................................................................................................................. 3-4 Multi-Operation by Serial Command............................................................................................. 3-5 Tube/Cap Library .......................................................................................................................... 3-8 Tube/Cap Information ................................................................................................................. 3-16
This section explains how to configure the EZ Match Imager to identify specific tube and cap characteristics, and how to manage the data that is acquired during tube/cap detection and symbol decoding operations.
EZ Match User Manual
3-1
Page 46
Tube/Cap Inspection Serial Commands
Tube/Cap Inspection Serial Commands Active Tube/Cap Library Indexes Tube/Cap Dimensions Multi-Operation Tube/Cap Information Tube/Cap Message Formatting Tube/Cap Message Editing
3-2
<K257,number of active indexes> <K258,index,tube diameter,cap base length,cap tip length,cap base diameter,cap tip diameter,hue,saturation, value> <K260,index,operation,frame count,timeout,database min,database max> <K850,separator,library index,tube angle,tube/cap tip position,cap color,match status,tube diameter> <K851,cap presence message index,cap absence message index,cap unknown message index,tube presence message index,tube absence message index> <K852,index,message>
EZ Match User Manual
Page 47
Tube/Cap Inspection
Capture and Test Click the EZ Match button and then the Test tab to display the view shown below.
Captures an image of the tube, cap, and/or symbol in the field of view and tests Multi-Operation settings (“Bar Code Reader”, “Tube Detection”, and “Cap Detection”).
When enabled, shows the image in higher contrast than the default contrast value.
When enabled, shows the reverse of the original image capture.
Tube or cap detection results are shown in this field, as well as symbol data when applicable.
EZ Match User Manual
3-3
Page 48
Multi-Operation
Multi-Operation Click the EZ Match button and then the Multi-Operation tab to display the following view. Note: Default Multi-Operation configuration shown below.
Each operation can be assigned a range of IP Database indexes. Click Go to IP Database to configure capture settings and processing settings for each index.
Note: All functions in the above view can be set using the Multi-Operation serial command <K260>. 3-4
EZ Match User Manual
Page 49
Tube/Cap Inspection
Multi-Operation by Serial Command Multi-Operation allows users to customize triggered read cycle operations, and to perform those customized operations in a specific sequence. A specific range of IP Database indexes (from 1 to 10) can be assigned to each operation. Switching from one operation to the next can be configured to occur on an image frame count (from 1 to 6) or on a user-defined timeout (1 ms to 65535 ms). The Multi-Operation read cycle is defined to work with the same features and configuration settings as a normal read cycle. Important: Multi-Operation is only active when the imager is in a triggered read cycle. Default Multi-Operation Configuration Index
Frame Count (Rapid Capture Mode only)
Operation
Database Range
Timeout
Min
Max
1
Bar Code Reader
1
0
1000
1
1
2
Tube/Cap Detection
4
0
1500
2
2
3
Disabled
0
0
1000
3
3
Index (Multi-Operation) Serial Cmd: Options:
<K260,index,operation,frame count,timeout,database min,database max> 1 to 3
EZ Match User Manual
3-5
Page 50
Multi-Operation by Serial Command
Operation (Multi-Operation) Definition:
Serial Cmd: Options:
Specifies the operational flow of the read cycle, and allows up to 3 different operations to be sequenced in a single read cycle. The operation field is a bit field, and multiple processes can be defined for a single operation. The examples below demonstrate how to configure a tube and cap detection process to be defined in a single operation. Tube and Cap Detection Operation = 2 + 4 = 6 <K260,index,operation,frame count,timeout,database min,database max> Disabled = 0 Bar Code Reader = Add 1 Tube Detection = Add 2 Cap Detection = Add 4
Disabled When set to Disabled, no processing is specified for the operation.
Bar Code Reader Read cycle operation for the duration of the specified time and number of frames will include processing of image frames for enabled symbologies.
Tube Detection Read cycle operation for the duration of the specified time and number of frames will include processing of image frames for presence or absence of test tubes.
Cap Detection Read cycle operation for the duration of the specified time and number of frames will include processing of image frames for presence or absence of test tube caps.
Frame Count (Multi-Operation) Definition: Serial Cmd: Default: Options:
3-6
Frame Count only applies if the imager is configured in a Rapid Capture mode of operation. The frame count value indicates the number of image frames that will occur before the read cycle will switch to the next operation. <K260,index,operation,frame count,timeout,database min,database max> 0 = Disabled 1 to 6
EZ Match User Manual