Browse
(440) 299-7778
Menu
(440) 299-7778
One-year parts and labor warranty30-day money-back guaranteeEnglishEspañolContact

The text of a manual · part 2 of 13

MS-3000 Single Head Decoder User's Manual, pages 11 to 20

Pages 11 to 20 of 126

The original document is the PDF (730 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Microscan MS-3000

Page 11

Attach Cabling

Under ideal conditions, maximum cable lengths can match the distances shown in table 1-1. However, since cable lengths and sizes are dictated by local conditions such as wire size, wire shape (flat or round), shielding, grounding, extraneous signal noise, etc., maximum cable distances will vary.

1–Setup and Inst.

2 Attach Cabling

Table 1-1 Cable Distances

Maximum Distance

Cabling

RS-232 Decoder to Host 50 ft. (15.2 m) RS-422 Decoder to Host 4000 ft. (1219 m) Decoder to Scan Head

15 ft. (4.57 m)

RS-485 Multidrop Trunk

4000 ft. (1219 m)

RS-485 Multidrop Drop

10 ft. (3 m)

Front Panel Connectors The MS-3000 decoder has six I/O connectors. On the front panel (figure 1-2) there are the following two connectors: a. Trigger (Microscan or other object detector) (6-pin DIN socket) b. Scan head (modular RJ-45)

PW R TRIG

SC AN

IND2

IND1

Figure 1-2 Front Panel

MS-3000 Single Head Decoder User’s Manual

1-3

Page 12 read from a scan by OCR; it may hold reading errors

Chapter 1 Setup and Installation

Trigger Connector

The trigger connector (TRIG) is a 6-pin DIN socket (figure 1-3) that mates with a 240 degree 6-pin DIN plug. Pin assignments are shown in table 1-2.

Table 1-2 Trigger Connector Pin Assignments os) a = xo) c © Q. S Sond ® a =

Pin Function

Trigger Input +3 VDC to + 24 VDC

TTL Relay driver output signal 5 VDC

+5 VDC +12 VDC

Ground

Figure 1-3 Trigger Connector Socket oO }o |; |wo

Not used

Pin 1 is the input from the object detector. When operating the decoder in external trigger mode, a toggle at this pin causes the decoder to begin a read cycle.

Pin 2 of the trigger connector is a programmable relay driver. The MS-3000 software can be programmed to set this pin high or low upon a good read, a no read, a good match, or a mismatch. This pin can source or sink 4 mA (maximum) and can be used to drive a small relay to operate an alarm, diverter, etc.

Microscan offers an optical object detector (P/N 99-440001-03) that plugs directly into this connector and a user-customized trigger port connector (P/ N 20-600090-02). '

Figure 1-4 shows examples of positive and negative external trigger inputs that could be applied to the trigger connector. (Shown in untriggered state.)

1. Trigger sources other than the Microscan object detector can be used. Mechanical switches, relays, etc. - which tend to be slow and bouncy and produce multiple trigger signals - are not recommended unless equipped with optical sensors or filtered transitions (optical, Hall effect, or DC solid state relays).

1-4 MS-3000 Single Head Decoder User’s Manual

Page 13

Attach Cabling

+0 VDC to +24 VDC

1–Setup and Inst.

Positive Trigger

Negative Trigger

J2 1

Trigger input

2

Aux. relay driver

+0 VDC to +24 VDC

J2 1

Trigger input

2

Aux. relay driver

N.C.

N.O.

Solid state switching device

3

+5 VDC

3

+5 VDC

4

+12 VDC

4

+12 VDC

5

Ground

5

Ground

6

Not used

6

Not used

Solid state switching device

Decoder

Decoder

Figure 1-4 Trigger Connector Wiring Diagram (untriggered)

Scan Head Connector To prevent voltage loss, cables between decoder and scan head should not exceed 15 feet (4.57 m) unless wire sizes exceed the minimum 26 AWG. Figure 1-5 shows a MS-3000 decoder connected to a Microscan MS-500 scan head. M S -520 S can Head

MS-3000 Decoder

PW R TRIG

SC AN

IND2

IND1

Figure 1-5 Decoder to Scan Head

MS-3000 Single Head Decoder User’s Manual

1-5

Page 14

1–Setup and Inst.

Chapter 1

Setup and Installation

Back Panel Connectors On the rear panel (figure 1-6) there are the following four connectors: a. Power (5-pin DIN socket) b. LAN (modular RJ-11) c. Host (25-pin D-subminiature socket) d. Monitor (9-pin D-subminiature socket)

HOST

POWER

MONITOR

LAN

Figure 1-6 Rear Panel of MS-3000 Decoder

Power Connector The power connector (figure 1-7) has a 5-pin DIN socket with pin assignments shown in table 1-3. You may also supply the required DC voltages yourself. A mating connector (Switchcraft #05BL5M plug) is required. Caution: Switching power supplies for Microscan equipment with a switching noise of 20 mV or greater with ±12 VDC are not recommended due to excessive ripple characteristics. Table 1-3 Power Connector Pin Assignments

1

Pin

3 5

POWER

2

4

Figure 1-7 Power Connector Socket

1-6

Function

1

DC Ground

2

Chassis Ground

3

+5 VDC

4

–12 VDC

5

+12 VDC

MS-3000 Single Head Decoder User’s Manual

Page 15

Attach Cabling The Local Area Network (LAN) connector allows the MS-3000 decoder to communicate with a multidrop concentrator via a 4-wire cable (preferably shielded) with a 6-pin, 6-wire, RJ-11 type connector. Pin assignments are as shown in table 1-4. The LAN connector is used when a MS-3000 decoder is configured for RS-485 multidrop communications. 1

Table 1-4 LAN Connector Pin Assignments

6

Pin

Figure 1-8 LAN Connector

Function

1

Ground

2

Receive Data (+)

3

Receive Data (–)

4

Transmit Data (+)

5

Transmit Data (–)

For proper operation of RS-485 multidrop communications, the main cable should not exceed 4000 feet, with each drop no more than 10 feet. The Microscan MS-5000 multidrop concentrator can support up to 50 decoders or other multidrop devices on one RS-485 line. Note that the last device must be terminated correctly. RS-485 pinouts are also available at the host connector.

Host Connector The host connector (figure 1-9 on page 1-9), a 25-pin D-subminiature socket, allows the decoder to be connected to a host, a concentrator, or other communications device such as a PLC (programmable logic controller), a monitor, a PC, a relay, a diverter, an alarm, etc. Note: All Microscan products are configured as DTE at the host connector when in RS-232 operation. Caution: Do not use a host cable with more wires connected than are required for the application. The host connector of the decoder has many outputs that could cause damage or interfere with normal operation if connected and improperly used.

MS-3000 Single Head Decoder User’s Manual

1-7

1–Setup and Inst.

LAN Connector

Page 16

1–Setup and Inst.

Chapter 1

Setup and Installation Table 1-5 Host Connector Pin Assignmentsa

Pin

Function

1 2

Chassis ground Transmit data

3

Receive data

4

Request-to-send

5

Clear-to-send

6 7 8 9 10

TTL 5V signal Signal Ground TTL 5V signal +5 VDC Trigger input

11

Default (reset)

12 13 14 16 19

Aux Input RXD + TXD – RXD – TXD +

20

+12 VDC

21 22 23 24 25

Profile Card Input Signal Ground –12 VDC +12 VDC Match Code

Comment connected to dc ground +12v data output from the decoder ±12v signal indicating data from the host to the decoder ±12v signal asserted high by the decoder when it has data to send to the host ±12v signal asserted low by the host to stop the decoder from sending data to it (data transmission will resume when the signal is asserted high.) indicates a good read connected to chassis ground indicates a noread auxiliary supply +0 VDC to +24 VDC resets decoder to default configuration (future use) RS-422/485 RS-422/485 RS-422/485 RS-422/485 data terminal ready (asserted high on power-up to indicate decoder is on)

next label read as master label, if enabled a. The default communications mode does not support pin 4 (RTS). If RTS is not required by the host port, pin 4 should not be wired as the results will be unpredictable.

A good read will cause a 5V signal (high or low) to be output to pin 6. A noread will cause a 5V signal (high or low) to be output to pin 8. Caution: There must be ±12V present before the decoder will transmit data to the host. However, the decoder can receive commands without the presence of either voltage.

1-8

MS-3000 Single Head Decoder User’s Manual

Page 17

Attach Cabling 1–Setup and Inst.

DB-25 DTE to DB-25 DTE Connection

1

Decoder

HOST

Host

Transmit 2

2 Transmit

Receive

3 Receive

3

Signal Grnd 7

7 Signal Grnd

DB-25 DTE to DE-9 DCE Connection

25

Figure 1-9 Host Connector

Decoder

Host

Transmit 2

2 Receive

Receive 3

3 Transmit

Signal Grnd 7

5 Signal Grnd

Figure 1-10 DTE and DCE Host Connections

Monitor Connector The monitor connector (auxiliary terminal) connects the decoder to an auxiliary monitor via a 9-pin D-subminiature socket and cable. Table 1-6 shows auxiliary terminal pin assignments. Communication at this connector is RS-232 only. 1 MONITOR

9

Table 1-6 Monitor Connector Pin Assignments

Pin

Function

2

Receive Data (in)

3

Transmit Data (out)

5

Signal Ground

Figure 1-11 Monitor Connector

MS-3000 Single Head Decoder User’s Manual

1-9

Page 18

Chapter 1

Setup and Installation

1–Setup and Inst.

3 Configure Decoder

Settings for Communications, Operations, Code Types, and User Outputs and Raster Setup are stored in non-volatile memory and can be configured by menu, serial command, or profile card commands. For explanations of configuration settings, see Chapter 2, “ Menu Configuration.” To establish communications you will need to match the host’s or auxiliary terminal’s communication settings with your decoder’s settings (see Communications Menu” on page 2-5). Also make certain that the code type enabled in the decoder matches that of the label being used (see Code Types Menu” on page 2-22). Hint: Sending the <P> command will allow your scan head to read all of the listed code types.

Communicating with and ASCII Terminal The MS-3000 decoder communicates in full duplex, terminal mode with no handshake. It also recognizes carriage returns and line feeds. The host or ASCII terminal must match the following default settings before any communication can take place: 9600 Baud Rate, Seven Data Bits, Even Parity, and One Stop Bit. A PC or Macintosh computer can be used as a dumb terminal if connected as shown in table 1-5 on page 1-8 and running a communications program set to the above defaults. See your computer user’s manual for communication’s port pinouts.

4 Position Scan Head and Label

Before testing, you will need to position the scan head and label in a manner that matches as nearly as possible the actual conditions of your application. Consult your scan head user’s manual for important setup specifications.

1-10

MS-3000 Single Head Decoder User’s Manual

Page 19

Do Read Rate Test

This test calculates the percentage of scans decoded. It is useful during setup procedures to help optimize scan head-to-label orientation. a. Place the label used in your application in front of the scan head and within the desired read range. Note: Ensure that the label being scanned is of a code type enabled in the decoder’s software. Hint: Read rates are easier to read on the screen if Postamble is enabled. b. Select from the ESP Utilities menu or send a <C> command to the decoder to start testing (see the “Enter Read Rate Test” command on page 5-5). A percentage number from 0 to 100 will be displayed on the monitor in the read rate test indicating the ratio of good reads per total number of scans. c. Find the correct read rate area by moving the label in and out and right and left while observing read rate on the monitor.

Read Rate Percentage Label Data

Avoid the specular reflection area (see scan head user’s manual). d. Record the range area measurements and file the test results for future reference. If the results are not satisfactory, reposition or re-configure the decoder or choose a different narrow-bar-width label size. e. Select from the ESP Utilities menu or send a <J> command to the decoder when testing is complete (see the “Exit Read Rate Test” command on page 5-5).

MS-3000 Single Head Decoder User’s Manual

1-11

1–Setup and Inst.

5 Do Read Rate Test

Page 20

Chapter 1

Setup and Installation

1–Setup and Inst.

6 Install Decoder

The MS-3000 decoder can be mounted temporarily using the four rubber pads, or permanently, as follows: a. Position the decoder in a dry place. b. Before mounting, ensure that there is at least a 3 inch (76.2 mm) clearance at the rear and front of the decoder for the connectors and cables being used. c. Use the measurements provided in figure 1-12 to locate centers of mounting holes and drill four 5/32 inch or four 4 mm holes. d. Secure decoder with four 6-32 screws of the appropriate length. Caution: Make certain that mounting screws do not penetrate into the decoder case more than .175 in. (4.4 mm) or damage to the decoder could result. Measurements in inches (millimeters)

1.85" (46.99)

6.2" (157.48)

Fr ont

Optional foot pads (4 places)

2.5" (63.5)

6/32 threads (4 places)

0.5" (12.7)

5.6" (142.24)

Figure 1-12 Bottom Mounting Diagram (not full size)

1-12

MS-3000 Single Head Decoder User’s Manual