Page 1 of Keyence LC Series Ultra-High Accuracy Laser Displacement Meters Spec Sheet

Spec sheet · Keyence

Keyence LC Series Ultra-High Accuracy Laser Displacement Meters Spec Sheet

Spec sheet4 pagesPDF 93 KB

All 4 pages are below, each with its text.

Download the PDF93 KB

Filed with · 2 models

Also filed with these models on the old site

From the sheet · read page by page

Key facts from this sheet

Each line names its page in the PDF.

A four-page catalog section on the Keyence LC Series ultra-high accuracy laser displacement meters. It covers the LC-2400W controller and the LC-2420, LC-2430, LC-2440 and LC-2450 sensor heads; it does not name the LC-2100 or LC-2400A. It gives the features, specifications, functions, interfaces, hints on correct use and dimensions.

Device
LC Series laser displacement meter using triangulation: a controller with a separate sensor head p. 1
Measuring range
Specular-reflective heads up to 30 mm ±0.5 mm; diffuse-reflective heads up to 50 mm ±8 mm p. 1
Resolution
0.01 µm, 0.02 µm, 0.2 µm or 0.5 µm depending on the sensor head p. 2
Speed
50 kHz sampling; 100 µs response time; 20 kHz response frequency p. 2
Laser
Red semiconductor laser, 670 nm; linearity ±0.05% of F.S. p. 2
Analog outputs
Displacement ±10 V in 6 ranges; intensity 0 to 5 V p. 2
Control I/O
Inputs: HOLD timing, AUTO-ZERO ON/OFF, program selection, LASER REMOTE; outputs: upper/lower limit, intensity and AREA OVER alarms, NPN 100 mA max. (30V) p. 2
RS-232C
Displacement data output and control input, 75 to 19200 bps p. 2
Power
100 to 240 VAC ±10%, 50/60 Hz; 70 VA max. p. 2
Ambient temperature
0 to +40°C p. 2
Controller weight
Approx. 6.8 kg p. 2
Averaging
1 to 131072 measurements, 18 selectable settings p. 2
Functions
NORMAL, PEAK TO PEAK, PEAK and BOTTOM modes; AUTO GAIN; AUTO ZERO; output hold; five programs stored in EEPROM p. 3
Centre indicator
LED orange near the centre of range: LC-2420 ±0.02 mm, LC-2430 ±0.05 mm, LC-2440 ±0.3 mm, LC-2450 ±0.8 mm p. 3

What this sheet walks you through

  • Choosing a data processing mode p. 3
  • Setting light intensity limits and the analog output range p. 3
  • Finding the reference distance when mounting the sensor head p. 3
  • Reducing noise interference in wiring and grounding p. 4
  • Sensor head and controller dimensions p. 4

Good to know

  • Controller and sensor head are calibrated in pairs: combine a controller and head with the same serial number. p. 4
  • After a 10°C change in ambient temperature, the LC-2400W takes about 60 minutes to even out. p. 4
  • Keep the wiring away from high-voltage and power lines. p. 4
  • The sensor heads emit laser light; the label warns not to stare into the beam. p. 4

Questions this sheet answers

What is the measuring range of the Keyence LC Series?

30 mm ±0.5 mm for specular-reflective heads and 50 mm ±8 mm for diffuse-reflective heads. p. 1

What power does the Keyence LC-2400W controller need?

100 to 240 VAC ±10%, 50/60 Hz, 70 VA max. p. 2

Can I swap sensor heads between LC Series controllers?

The sheet says controller and head are calibrated in pairs, so combine ones with the same serial number to meet the specifications. p. 4

Does the Keyence LC Series have RS-232C?

Yes: it carries displacement data output and control input at 75 to 19200 bps. p. 2, 3

The file

Size and text

Pages4
PDF93 KB
TextText from the PDF itself.

Every page · 4

Read it here

Each page as it is printed, and its words beside it, so a phone can search and copy them.

Page 1 of Keyence LC Series Ultra-High Accuracy Laser Displacement Meters Spec Sheet

The text of page 1, from the PDF itself

Ultra-high Accuracy Laser Displacement Meters

LC Series

Description Advanced optics and triangulation measurement system The LC Series uses a triangulation measurement system making it the world’s most accurate laser displacement metre.

Features

  • Resolution of 0.01 µm
  • Linearity: ±0.05% of F.S.
  • 12 µm diameter beam spot
  • 50 kHz sampling rate

Visible laser beam spot

Measuring Range

50 kHz sampling rate

Specular-reflective – 30 mm ±0.5 mm

Utilising a high-speed processing circuit developed by KEYENCE, you can more accurately measure the eccentricity of a motor shaft rotating at high speeds.

Diffuse-reflective – 50 mm ±8 mm

The beam spot is only 12 µm in diameter. Even when measuring minute targets, you can easily position the sensor head with the beam spot.

Superior operating distance The LC-2430 has an operating distance of 30 mm, a distance unmatched in specular-reflective sensor heads. The LC Series can be used in situations where the sensor head cannot be mounted in close proximity to the target.

30 mm

Target surface measurement The LC Series can measure irregularities or groove depth on a surface that cannot be measured with conventional displacement metres.

Conventional meter

Measurement of irregularities and grooves on cutter

LC-2400W Irregularities on cutter

Groove on cutter

342

Page 2 of Keyence LC Series Ultra-High Accuracy Laser Displacement Meters Spec Sheet

The text of page 2, from the PDF itself

Ultra-high Accuracy Laser Displacement Meters

LC

Measurement of solder paste height on PC boards

Measurement of silicon wafer thickness

Proximity Sensors

Measurement of pin pitch and alignment

SpecificationsPressure Sensors±0.5 mm±3 mm±8 mm30 mm30 mm50 mmLight source

Red semiconductor laser Wavelength

Red semiconductor laser

670 nm670 nm670 nm670 nmIECClass 2Class 2Class 2Class 2FDAClass IIClass IIClass IIClass IIMinimum spot diameter20 x 12 µm30 x 20 µm35 x 20 µm45 x 20 µmResolution 1.0.01 µm0.02 µm0.2 µm0.5 µm1.±0.05% of F.S.±0.05% of F.S.±0.05% of F.S.±0.05% of F.S.Sampling frequency50 kHz50 kHz50 kHz50 kHzClassLinearityResponse frequency

20 kHz (-3dB, Averaging measurements: 1)

Response time100 µs100 µs100 µs100 µsAveraging measurements

1 to 131072 (18 selectable settings)

1 to 131072 (18 selectable settings)

OFFSET range±199.99 µm±2.9998 µm

Gain adjustment Analog output

Displacement data output Intensity data output Control input

Control I/O Control output Interface

RS-232C

Measurement stability(20 ±5°C)2.

±499.98 µm

±7.9995 µm

AUTO/MANUAL (4 settings)

AUTO/MANUAL (4 settings)

±10 V, Output impedance: 0 Ω, 6 settings

±10 V, Output impedance: 0 Ω, 6 settings

0 to 5 V, Output impedance: 0 Ω

0 to 5 V, Output impedance: 0 Ω

HOLD timing, AUTO-ZERO ON/OFF, Program selection, LASER REMOTE Non-voltage (contact, solid state)

Non-voltage (contact, solid state)

Upper/lower limit, Intensity alarm, AREA OVER alarm NPN: 100 mA max. (30V)

NPN: 100 mA max. (30V)

Displacement data output and control input (baud rate: 75 to 19200 bps selectable) ±0.2% of F.S.

±0.2% of F.S.±0.03% of F.S.±0.2% of F.S.Power supply

100 to 240 VAC ±10%, 50/60 Hz (when using the control unit separately: ±15 VDC ±3%, 0.5 A, +5 V ±5%, 3A)

Power consumption70 VA max.70 VA max.70 VA max.70 VA max.Ambient temperature0 to +40°C0 to +40°C0 to +40°C0 to +40°CWeightSensor headApprox. 500 gApprox. 500 gApprox. 250 gApprox. 250 gControllerApprox. 6.8 kgApprox. 6.8 kgApprox. 6.8 kgApprox. 6.8 kg

1. Using a standard specular-reflective object (LC-2420/LC- 2430) or white diffuse-reflective object (LC-2440/LC-2450) as a target with the averaging measurements set to 512. 2. Data was obtained when the standard target and the sensor head were fixed to a steel plate (LC-2420/LC-2430) or an aluminum plate (LC-2440/LC-2450) at a temperature of 20°C.

5 6High Precision Sensors10 mm7Vision SystemsOperating distanceLC-2400W8Bar Code Readers±0.2 mmController4910

Thrubeam Measuring Displacement Sensors Instruments

Measuring range

Counters, Control Units

LC-2450311 12

Analog Sensor Controllers

LC-2430213Video MicroscopeLC-2420

Diffuse-reflective LC-2440

Programmable Logic Controllers

Specular-reflectiveSensor headModelTechnical Guide

Measurement of CD pickup travel

1Safety Light CurtainsPhotoelectric SensorsApplications14343

Page 3 of Keyence LC Series Ultra-High Accuracy Laser Displacement Meters Spec Sheet

The text of page 3, from the PDF itself

LC Ultra-high Accuracy Laser Displacement Meters Functions Data processing modes

Light intensity limits

Each of the four modes can be set by simply pressing a key.

If the measured value is inaccurate because of excessive or insufficient reflected light or ambient light, setting upper and lower limits for received light intensity will eliminate this problem.

Data processing mode NORMAL PEAK TO PEAK (P-P) PEAK BOTTOM

Function Measures displacement from reference position. Measures displacement between maximum and minimum values. Measures maximum value. Measures minimum value.

Analog voltage output range selectable You can select from six analog voltage output ranges. By selecting the analog voltage output range based on the measuring range, a minute change in measured values can be monitored with high accuracy without being affected by noise interference. Output range of LC-2420 sensor head (µm/ V) 2.5

5102550100

Analog voltage output (intensity) Analog voltage (0 to 5 V) proportional to the intensity level is continuously output. The analog voltage output is useful for monitoring changes in intensity level over a period and for setting upper/lower intensity limits.

AUTO GAIN The controller can be set to automatically switch between 4 gain levels. Receiving sensitivity will automatically adjust to different target colours and materials.

Five program storage capability

Displacement Sensors

10

Up to five sets of parameters can be stored in EEPROM. All of the settings, including preset tolerance and calibration, can be easily switched depending on the workpiece.

AUTO ZERO Pressing this button sets the current measured value to “0”, regardless of target position. This simplifies offset target positioning.

Output hold The displayed value and output value are retained by simply pressing a key.

Easy confirmation of reference distance When the target is in the centre area of the measuring range, the LED indicator lights orange. When it is outside the centre area but in the measuring range, the LED indicator lights green. This allows you to easily find the reference distance (position where the laser beam spot is smallest) and mount or adjust the sensor head. Sensor head

Range the LED lights orange

LC-2420±0.02 mmLC-2430±0.05 mmLC-2440±0.3 mmLC-2450±0.8 mm

Error message The LC displays an error message to indicate that a problem has occurred. Since the displayed error message reveals the cause of the problem, you can act to prevent further errors from occurring.

Interface Analog voltage output

Control output

Analog voltage is output in proportion to the measured value, enabling quick data processing and analysis. The output range can be changed to 6 steps.

This displacement metre is equipped with the HIGH and LOW comparator outputs as the standard function. In addition, the LIGHT INTENSITY alarm output (DARK, BRIGHT) and the AREA OVER alarm output (FAR, NEAR) are provided.

RS-232C The RS-232C interface enables communication with a computer. Thus, data transfer and remote operation for changing settings are possible.

344

Page 4 of Keyence LC Series Ultra-High Accuracy Laser Displacement Meters Spec Sheet

The text of page 4, from the PDF itself

Ultra-high Accuracy Laser Displacement Meters

LC

Measures to reduce noise interference

Keep the ambient temperature at a constant level. Fluctuations in ambient temperature may cause measurement error.

  • Keep the wiring or connection cable away from highvoltage lines or power lines to prevent the LC-2400W from malfunctioning due to noise. Be sure to keep wiring separate.

[Note] If the ambient temperature changes by 10°C, it takes approximately 60 minutes for the temperature distribution in the LC-2400W to become uniform.

  • Be sure to earth-ground the LC-2400W through the earth ground terminal. Insulating the sensor head is also effective in reducing noise.

Compatibility

Proximity Sensors

With the LC Series, the controller and sensor head are calibrated in pairs. Therefore, to meet specifications, be sure to combine a controller and sensor head having the same serial number.

Class 2FDAClass IICAUTION

DO NOT STARE INTO BEAM Maximum output Pulse duration Emitted wavelength

AVOID EXPOSURE

1.9mW 10µs 670nm

LASER RADIATIONDO NOT STARE INTO BEAM SEMICONDUCTOR LASER 670nm MAXIMUM OUTPUT 1.9mW (AVERAGE OUT PUT 950µW) PULSE DURATION 10µSEC CLASS II LASER PRODUCT

CLASS 2 LASER PRODUCT in conformity to IEC825-1 11.1993

LASER RADIATION IS EMITTED FROM THIS APERTURE.

Counters, Control Units

IECLASER RADIATIONLC-2450

Dimensions LC-2440/LC-2450

LC-2420/LC-243058.558.568 45.5 22.525ø7 Cable length: 20 cm1.5ø1055.5700.5 5.5 X (X-0.5)θ4.560

0.5 3 x ø4.5 (mounting hole)

22.511X (X-0.5)35.5 36604θ521512.515.517.80.50.5 228θXModelθXLC-242063°10.5LC-244030°30.5LC-243045°30.5LC-245022°50.5910 11 12 13Video MicroscopeModel

3 x ø4.5 (Mounting hole)

6 719617012.53125ø7 Cable length: 20cm1.5Unit: mm1818Sensor head5High Precision SensorsLC-24404Vision SystemsLC-2430Bar Code ReadersLC-2420FDA Class II3

Thrubeam Measuring Displacement Sensors Instruments

ModelIEC Class 22

Analog Sensor Controllers

The LC Series conforms to IEC and FDA standards as follows:

Programmable Logic Controllers

Pressure SensorsWarningø101Safety Light CurtainsOperating environmentPhotoelectric SensorsHints on Correct UseControllerTechnical GuideLC-2400W 110 123234 320(47) 23232 2971422345

More documents · 1

For the same models