Trakker Antares 24XX Terminal User's Manual Addendum: Chapter 9 - Configuration Command Reference, pages 380 to 389

Pages 380 to 389 of 501, from Trakker Antares 24XX Terminal User's Manual Addendum

The original document is the PDF (2.2 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Intermec 2455

Page 380

Configuration Command Reference

Or:

9

To set EOM to one or two ASCII characters for one serial port: Note: To scan these bar code labels, you must configure the T2455 to use Code 39 in Full ASCII mode. For help, see “Code 39” earlier in this chapter. 1. Scan this bar code: Enter Accumulate Mode / Set EOM

*/K/O/D/KYZ* *+/$+YZ*

2. Scan the serial port from these bar codes: COM1

COM4

*1/N*

*4/N*

*1.*

*4.*

3. Scan one or two bar codes for data from the “Full ASCII Bar Code Chart” in Appendix B. 4. Scan this bar code: Exit Accumulate Mode

*/M/O* *-/*

5. Repeat Steps 1 through 4 to set the EOM for another serial port. Notes:

For COM1 only, you can use either command syntax listed next: Syntax: or:

YZ1.data PFdata where data is one or two ASCII characters. For COM4, you must use the YZ configuration command.

9-51

Page 381

TRAKKER Antares 2455 Vehicle-Mount Terminal User’s Manual

Flash Memory Configuration Purpose:

Configures the use of the optional 2MB flash memory. If you have a terminal with the 4MB flash memory option, you can configure the extra 2MB of flash memory either as a drive D or as a storage space for double-byte fonts. You can store up to 32 files on drive D or store a 2MB double-byte font set. If you configure the flash memory as drive D, use this drive to store large lookup tables and data files. Note: You can only configure 2MB (of the total 4MB) of flash memory as a drive or to store fonts. You cannot use the space for both. If you configure drive D, you cannot store a font in flash memory.

Syntax:

FFdata Acceptable values for data are: 0 32

Default: Scan:

Configure flash memory for double-byte fonts Configure flash memory as a 2MB drive D

0 (flash memory for double-byte fonts) 1. One of these bar codes to configure flash memory: Configure Flash Memory For Double-Byte Fonts

*$+FF0* *$+FF0* Configure Flash Memory as Drive D

*$+FF32* *$+FF32*

Note: A value of 32 configures drive D as a 2MB drive. 2. To save the changes in flash memory, scan this bar code: Save Configuration in Flash Memory

*.+1* *.+1*

3. Scan this bar code to boot the terminal and use the 2MB flash memory to store double-byte fonts or as drive D: Reset Firmware

*-.* *-.*

9-52

Page 382

Configuration Command Reference

9

Note: When you boot or reset the terminal, any fonts or files in flash memory on Drive D are erased. 4. If you configured flash memory to store double-byte fonts, use the TRAKKER Antares Font Editor to download the double-byte font set to the terminal. See your local Intermec sales representative for information about the TRAKKER Antares Font Editor.

Flow Control Purpose:

Regulates the data transmission through the serial port. The terminal is a DTE (Data Terminal Equipment) device. There are several flow control options: •

CTS checking where Clear To Send (CTS) is a hardware signal flow control. The terminal sets and clears Request To Send (RTS) when it is ready to receive. The terminal checks CTS when it tries to transmit data.

•

XON/XOFF response causes the terminal to respond to XON/XOFF characters received while transmitting.

•

XON/XOFF control specifies that the terminal transmits XON/XOFF characters to control the incoming data flow.

•

XON/XOFF response and control specify that the terminal responds to and transmits XON/XOFF characters.

•

CTS/RTS for DTE-DTE allows for peer connections. Each terminal’s RTS output pin is connected to the CTS input pin of the other terminal. Each terminal sets RTS whenever it is ready to receive data and waits for CTS to be set before sending data.

Note: For TRAKKER Antares firmware version 4.X and earlier, you must disable the EOM command (use Character mode) to enable Flow Control with the XON/XOFF options. Syntax:

YLn.data where n is: 1 4

COM1 serial port COM4 serial port

Acceptable values for data are: 0 1 2 3

No flow control CTS checking XON/XOFF response XON/XOFF control

9-53

Page 383

TRAKKER Antares 2455 Vehicle-Mount Terminal User’s Manual

Flow Control (continued) 4 5 Default: Scan:

XON/XOFF response and control CTS/RTS for DTE-DTE

No flow control To set the default flow control for each serial port, scan these bar codes: No Flow Control for COM1

*$+YL1.0* *$+YL1.0* No Flow Control for COM4

*$+YL4.0* *$+YL4.0*

Or:

To set the flow control for one serial port: 1. Scan this bar code: Enter Accumulate Mode / Set Flow Control

*+/$+YL* *+/$+YL*

2. Scan the serial port from these bar codes: COM1

COM4

*1.*

*4.*

*1.*

*4.*

3. Scan the flow control setting from these bar codes:

9-54

No Flow Control

CTS Checking

*0*

*1*

*0*

*1*

XON/XOFF Response

XON/XOFF Control

*2*

*3*

*2*

*3*

XON/XOFF Response and Control

CTS/RTS for DTE-DTE

*4*

*5*

*4*

*5*

Page 384

Configuration Command Reference

9

4. Scan this bar code: Exit Accumulate Mode

*-/* *-/*

5. Repeat Steps 1 through 4 to set the flow control for another serial port. Notes:

For COM1 only, you can use either command syntax listed next: Syntax: or:

YL1.data ILdata where data is a numeric value from 0 to 5. For COM4, you must use the YL configuration command.

Handshake Purpose:

Enables or disables the handshake event that is an affirmative acknowledge to a message received through the serial port. Note: Before you can enable Handshake, you must configure the EOM command. Handshake is also referred to as AFF (affirmative acknowledge) on other Intermec data collection devices.

Syntax:

YJn.data where n is: 1 4

COM1 serial port COM4 serial port

Acceptable values for data are: Disabled Enabled Default: Scan:

No ACK or NAK transmitted Set to ACK (Acknowledge, hexadecimal value \x06)

Disabled To disable handshake, scan one of these bar codes: Disable Handshake for COM1

*$+YJ1.* *$+YJ1.*

9-55

Page 385

TRAKKER Antares 2455 Vehicle-Mount Terminal User’s Manual

Handshake (continued) Disable Handshake for COM4

*$+YJ4.* *$+YJ4.*

Or:

To enable handshake, scan one of these bar codes: Note: To scan these bar code labels, you must configure the T2455 to use Code 39 in Full ASCII mode. For help, see “Code 39” earlier in this chapter. Enable Handshake (Set to ACK) for COM1

*/D/KYJ1/N/DF* *$+YJ1.$F* Enable Handshake (Set to ACK) for COM4

*/D/KYJ4/N/DF* *$+YJ4.$F*

Note: ACK is encoded as $F in a Code 39 bar code label. Notes:

For COM1 only, you can use either command syntax listed next: Syntax:

YJ1.data or:

PGdata where data is disabled or enabled (set to ACK). For COM4, you must use the YJ configuration command.

9-56

Page 386

Configuration Command Reference

9

Host IP Address Purpose:

Defines the IP address assigned to the host computer in your 2.4 GHz RF network. An IP address is a unique network level address you assign to each device in a TCP/IP direct connect network. The host IP address you set on the T2455 must match the address that is set on the host.

Syntax:

NCn.n.n.n where each n address segment is a number from 0 to 255. The host IP address field consists of four separate numbers, each separated by a period. Note: The RF network cannot be activated if the first address segment in the IP address is set to 0, 127, or a number greater than 223.

Default: Scan:

0.0.0.0 To set the default host IP address, scan this bar code: Default Host IP Address

*$+NC0.0.0.0* *$+NC0.0.0.0*

Or:

To set the host IP address: 1. Scan this bar code: Enter Accumulate Mode / Set Host IP Address

*+/$+NC* *+/$+NC*

2. Scan a numeric value from 0 to 255 to set an n field of the host IP address from these bar codes.

*0*

*1*

*0*

*1*

*2*

*3*

*2*

*3*

*4*

*5*

*4*

*5*

9-57

Page 387

TRAKKER Antares 2455 Vehicle-Mount Terminal User’s Manual

Host IP Address (continued)

*6*

*7*

*6*

*7*

*8*

*9*

*8*

*9*

3. Scan this bar code: . (Period)

*.* *.*

4. Repeat Steps 2 and 3 to set the next three numbers in the host IP address field. After you scan the last address segment, go to Step 5. Do not scan the period after the last address segment. 5. Scan this bar code: Exit Accumulate Mode

*-/* *-/*

9-58

Page 388

Configuration Command Reference

9

Interleaved 2 of 5 Purpose:

Enables or disables decoding of Interleaved 2 of 5 (I 2 of 5) symbology. I 2 of 5 is a high-density, self-checking, continuous numeric symbology. It is mainly used in inventory distribution and the automobile industry. Enabling I 2 of 5 automatically disables Code 2 of 5.

Syntax:

CAdata Acceptable values for data are: 0 2-32 97 98 99

Disabled Fixed length (even number only) Variable length without a check digit Case code (6 or 14) with a check digit Variable length with a check digit

Caution Using the variable length without a check digit configuration option can cause substitution errors. Conseil Des erreurs de substitution peuvent survenir si vous utilisez la longueur variable sans option de vérification de configuration de chiffres. Default: Scan:

Disabled One of these bar codes: Disable Interleaved 2 of 5

*$+CA0* *$+CA0* Enable Variable Length Without a Check Digit

*$+CA97* *$+CA97* Enable Variable Length With a Check Digit

*$+CA99* *$+CA99*

9-59

Page 389

TRAKKER Antares 2455 Vehicle-Mount Terminal User’s Manual

Interleaved 2 of 5 (continued) Enable Case Code With a Check Digit

*$+CA98* *$+CA98*

Or:

To set Interleaved 2 of 5 to a fixed length: 1. Scan this bar code: Enter Accumulate Mode / Set Fixed Length

*+/$+CA* *+/$+CA*

2. Scan a numeric value for data from these bar codes. (Use even numbers 2-32 only.)

*0*

*1*

*0*

*1*

*2*

*3*

*2*

*3*

*4*

*6*

*4*

*6*

*8* *8*

3. Scan this bar code: Exit Accumulate Mode

*-/* *-/*

9-60