TC1100 Reference Manual: 3 Software Configuration Strings

Pages 28 to 41 of 131, from TC1100 Reference Manual

The original document is the PDF (979 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Datalogic TC1100

Page 28

TC1100

3

3

SOFTWARE CONFIGURATION STRINGS

3.1

DEFAULT CONFIGURATION

RS232 INTERFACE 9600 baud, parity disabled, 8 data bits, 1 stop bit, handshaking disabled, ACK/NACK from Host disabled, 100 ms RX timeout, FIFO enabled, inter-character delay disabled

WEDGE INTERFACE USA keyboard, caps lock off, delays disabled, control character emulation = ctrl+shift+key;

PEN Emulation INTERFACE interpret mode, conversion to code 39 disabled, output level normal, idle level normal, minimum output pulse 600 µs, overflow medium, inter-block delay disabled

DATA FORMAT Code identifier disabled, field adjustment disabled, code length TX disabled, character replacement disabled, header = <STX>, terminator = <CR> <LF>, no read character = <CAN> POWER SAVE 270 read/sec, sleep disabled, standby disabled READING PARAMETERS On-Line mode, beeper intensity = high intensity, beeper tone = tone 2, beeper type = monotone, beeper length = short DECODING PARAMETERS Ink spread enabled, overflow control enabled, interdigit control enabled, Puzzle Solver™ disabled, decoding safety = 1 read

20

Page 29

SOFTWARE CONFIGURATION STRINGS

3

CODE SELECTION enabled codes •

Interleaved 2 of 5 Check digit transmitted, variable length code; 4-99 characters

•

EAN8/EAN13UPC A/UPC B without ADD ON Check digit transmitted, no conversions

•

Code 39 variable length code; 1-99 characters

•

Code 128 Check digit not transmitted, variable length code; 1-99 characters PDF417 (TC1100-x200 only)

• disabled codes EAN 128, ISBT128,Code 93, Codabar, pharmaceutical codes, Delta IBM, Code 16K, Code 49, (MSI/Plessey, Telepen, Pharmacode only TC1100-x100), (Codablock-A, Codablock-F Standard and EAN only TC1100-x200)

21

Page 30

TC1100

3

3.2

TC1100 CONFIGURATION

TC1100 is factory set for RS232 Interface applications. The easiest way to configure TC1100 is by using the DL Sm@rtSet™ utility program installed from the CD included with this manual. Otherwise you can configure the scanner using the serial string which follow: To configure TC1100 by using the configuration strings: 1)

Connect your TC1100 to a PC RS232 port according to the information in chapter 2. Set the PC serial port to the TC1100 default RS232 communication parameters.

2)

Using a Terminal Emulation Program, send the Restore Default string to the reader using the syntax described on the next page.

3)

Send all the necessary command strings according to your application's requirements.

CAUTION

The WEDGE or PEN Emulation interface selections must be sent last since they will cause the reader to no longer accept RS232 communications.

If working with the Wedge or Pen emulation interfaces, it is possible to return to the RS232 connection by reading the code given below: Return to RS232 Interface i i i

$1 $1 $1

For WEDGE or PEN Emulation Users

22

Page 31

SOFTWARE CONFIGURATION STRINGS

3

Command Syntax The command syntax is given below with an example:

Command(s)

$+

$-

<CR>

The new setting will be definitive (stored in FLASH EPROM) Example: Multiple command programming sequence: $+

BK0 CD7

AC410132

$-

CR

Carriage return character (0D Hex.) Exit and save new configuration Matrix 2/5 3 bars: no check digit, variable length code from 1 to 32 characters Baud rate = 19200 Serial On-Line operating mode Enter configuration environment

Each configuration parameter setting removes the condition previously active for that parameter.

NOTE

The device buffer can contain about 400 characters. If your programming string goes over this value, you must split it into separate groups and send each group after a delay of at least 3 seconds to give the reader time to empty the buffer and interpret the commands.

23

Page 32

TC1100

3 SERIAL CONFIGURATION STRINGS ENTER/EXIT CONFIGURATION COMMANDS DESCRIPTION Enter Configuration Exit and Save Configuration Restore Default Transmit Software Release (not for PEN emulation) Get Configuration from Reader (not for PEN emulation)

STRING $+ $$+$*<CR> $+$!<CR> $+$&<CR>

These commands do not require $-.

INTERFACE SELECTION DESCRIPTION RS232 WEDGE for IBM AT for IBM Terminals: 31xx, 32xx, 34xx, 37xx; make-break keyboard for IBM Terminals: 31xx, 32xx, 34xx, 37xx; make-only keyboard Keyboard Type for IBM Terminals 31xx, typewriter 32xx, 34xx, 37xx advanced for IBM XT for IBM Terminal 3153 for IBM PC Notebook for IBM SURE1 for IBM AT - ALT mode for IBM PC Notebook - ALT mode for Wyse Terminal - ANSI Keyboard for Wyse Terminal - PC Keyboard for Wyse Terminal - ASCII Keyboard for Wyse Terminal - VT220 style Keyboard for Digital Terminals VT2xx/3xx/4xx for Apple ADB Bus PEN EMULATION

24

STRING CP0 CP500 CP501 CP502 FK0 FK1 CP503 CP504 CP505 CP506 CP507 CP508 CP509 CP510 CP511 CP514 CP512 CP513 CP6

Page 33

SOFTWARE CONFIGURATION STRINGS

3 RS232

DESCRIPTION Baud rate

STRING CD0 CD1 CD2 CD3 CD4 CD5 CD6 CD7 CD8 CC0 CC1 CC2 CA0 CA1 CA2 CB0 CB1 CE0 CE1 CE2 CE3 CF0 CF3 EC0 EC1 CK00 - CK99 CL00 - CL99

150 300 600 1200 2400 4800 9600 19200 38400 Parity none even odd Data bits 7 8 9 Stop bits 1 2 Handshaking disable RTS/CTS XON/XOFF RTS always On ACK/NACK Protocol disable enable FIFO disable enable Inter-character delay (ms) RX Timeout (100 ms)

WEDGE DESCRIPTION Keyboard nationality

Belgian English French German Italian Spanish Swedish USA

STRING FJ7 FJ4 FJ2 FJ3 FJ1 FJ6 FJ5 FJ0

25

Page 34

TC1100

3 WEDGE (continued) DESCRIPTION Caps Lock Num Lock Delays Control Character Emulation caps Lock ON caps Lock OFF Toggle Num Lock Num Lock Unchanged Inter-Character (ms) Inter-Code (s) Ctrl + Shift + Key Ctrl + Key

STRING FE1 FE0 FL1 FL0 CK00 - CK99 FG00 - FG99 FO0 FO1

PEN DESCRIPTION Operating mode Minimum output pulse

Conversion to Code 39 Output level Idle level Overflow

Inter-Block delay (100 ms)

26 interpret (does not require $+ or $-) transparent (does not require $+ or $-) 200µs 400µs 600µs 800µs 1 ms 1.2 ms disable conversion to Code 39 enable conversion to Code 39 normal inverted normal inverted narrow overflow medium overflow wide overflow

STRING $] $[ DG0 DG1 DG2 DG3 DG4 DG5 DA0 DA1 DD0 DD1 DE0 DE1 DH0 DH1 DH2 CK00-CK99

Page 35

SOFTWARE CONFIGURATION STRINGS

3

DATA FORMAT NOT FOR PEN EMULATION INTERFACES DESCRIPTION Code Identifier disable Datalogic standard AIM standard Custom

Custom Code Identifier Headers

Terminators

No Read Character no header one character two characters three characters four characters five characters six characters seven characters eight characters no terminator one character two characters three characters four characters five characters six characters seven characters eight characters disable enable

STRING EB0 EB1 EB2 EB3 EHabc EA00 EA01x EA02xx EA03xxx EA04xxxx EA05xxxxx EA06xxxxxx EA07xxxxxxx EA08xxxxxxxx EA10 EA11x EA12xx EA13xxx EA14xxxx EA15xxxxx EA16xxxxxx EA17xxxxxxx EA18xxxxxxxx EP0 EP1x a = ASCII character. b, c, x = HEX values representing an ASCII character. a = ASCII character of the DATALOGIC STANDARD Code Identifier from the table on page 60. b = Hex value of the first Custom Code Identifier character from 00 to FE; FF = disable Code Identifier c = Hex value of the second Custom Code Identifier character from 00 to FE; FF = disable second character of Custom Code Identifier x = Hex value from 00 to FE

27

Page 36

TC1100

3 DATA FORMAT (continued) NOT FOR PEN EMULATION INTERFACES DESCRIPTION Code Length Tx not transmitted transmitted in variable-digit format transmitted in fixed 4-digit format disable right addition left addition right deletion left deletion

Field Adjustment

Field Adjustment Character Character Replacement disable character replacement first character replacement second character replacement third character replacement

STRING EE0 EE1 EE2 EF0 EFa0d EFa1d EFa2d EFa3d EGe EO0 EO1afg EO2afg EO3afg a = ASCII character. d = a number from the Hex/Numeric Table e, f, g = HEX values representing an ASCII character a = ASCII character of the DATALOGIC STANDARD Code Identifier from the table on page 60. d = a number in the range 01-32 from the Hex/Numeric Table e = Hex value from 00 to FE f = Hex value of the character to be replaced from 00 to FE g = Hex value of the new character to insert from 00 to FE FF = replace with no new character (remove character)

POWER SAVE DESCRIPTION Read Rate

Sleep State Enter Sleep Timeout (100 ms) Standby

28

67 reads per sec. 135 reads per sec. 270 reads per sec. disable enable enable disable

STRING BT0 BT1 BT2 BQ0 BQ1 BR00-BR99 BM0 BM1

Page 37

SOFTWARE CONFIGURATION STRINGS

3

READING PARAMETERS DESCRIPTION Operating Modes serial on-line on-line See par. 5.5.1 for details automatic automatic/lighting system standby Test Mode disable enable Beeper Intensity very low intensity low intensity medium intensity high intensity Beeper Tone Tone 1 Tone 2 Tone 3 Tone 4 Beeper Type monotone bitonal Beeper Length long short PDF Decoding Recognition low Intensity high

STRING BK0 BK1 BK3 BK2 #+DStat0<CR> #+DStat1<CR> BG0 BG1 BG2 BG3 BH0 BH1 BH2 BH3 BJ0 BJ1 BI0 BI1 BW0 BW1

DECODING PARAMETERS DESCRIPTION Ink-Spread Overflow Control Interdigit Control Puzzle SolverTM Decoding Safety disable enable disable enable disable enable disable enable one read two reads three reads four reads

CODE SELECTION DESCRIPTION DISABLE ALL FAMILY CODES EAN/UPC Disable EAN/UPC family EAN 8/EAN 13/UPC A/UPC E without ADD ON with ADD ON with and without ADD ON

STRING AX0 AX1 AW1 AW0 AV0 AV1 AU0 AU1 ED0 ED1 ED2 ED3

STRING AZ0 AA0 AA1 AA5 AA8

29

Page 38

TC1100

3 CODE SELECTION (continued) DESCRIPTION EAN 8/EAN 13

UPC A/UPC E

EAN 8 check digit transmission EAN 13 check digit transmission UPC A check digit transmission UPC E check digit transmission Conversions

ISBN Conversion codes

Code 39 without ADD ON with ADD ON 2 ONLY with ADD ON 5 ONLY with ADD ON 2 AND 5 without ADD ON with ADD ON 2 ONLY with ADD ON 5 ONLY with ADD ON 2 AND 5 disable enable disable enable disable enable disable enable UPC E to UPC A UPC E to EAN 13 UPC A to EAN 13 EAN 8 to EAN 13 enable ISBN enable ISSN enable ISBN and ISSN disable ISBN and ISSN disable Code 39 family Standard no check digit control check digit control and transmission check digit control without transmission Full ASCII no check digit control check digit control and transmission check digit control without transmission CIP 39 Code 32 code length xxxx = ASCII numbers that define the code length where: • First 2 digits = minimum acceptable code length. • Second 2 digits = maximum acceptable code length. The minimum code length must always be less than or equal to the maximum. The maximum code length for all codes is 99 characters.

Examples: 0132 = variable length from 1 to 32 digits in the code. 1010 = 10 digit code length only.

30

STRING AA3 AAK AAL AA6 AA4 AAM AAN AA7 AAG0 AAG1 AAH0 AAH1 AAI0 AAI1 AAJ0 AAJ1 AAA AAB AAC AAD AP1 AP2 AP3 AP0 AB0 AB11 AB12 AB13 AB21 AB22 AB23 AB3 AB4 AB*xxxx

Page 39

SOFTWARE CONFIGURATION STRINGS

CODE SELECTION (continued) DESCRIPTION 2/5 disable Code 2/5 family Interleaved 2/5 no check digit control check digit control and transmission check digit control without transmission Normal 2/5 5 bars no check digit control check digit control and transmission check digit control without transmission Industrial 2/5 (IATA) no check digit control check digit control and transmission check digit control without transmission Matrix 2/5 3 bars no check digit control check digit control and transmission check digit control without transmission CIP/HR Codabar disable Codabar family Standard no start/stop character equality control nor transmission no start/stop character equality control but transmission start/stop character equality control but no transmission start/stop character equality control and transmission ABC Codabar no start/stop character equality control but transmission Codabar ABC forced concatenation code length start/stop character case in transmission lower case upper case

3

STRING AC0 AC11xxxx AC12xxxx AC13xxxx AC21xxxx AC22xxxx AC23xxxx AC31xxxx AC32xxxx AC33xxxx AC41xxxx AC42xxxx AC43xxxx AC5 AD0 AD111 AD112 AD121 AD122 AD212 AD232 AD*xxxx ADA0 ADA1 xxxx = ASCII numbers that define the code length where: • •

First 2 digits = minimum acceptable code length. Second 2 digits = maximum acceptable code length.

The minimum code length must always be less than or equal to the maximum. The maximum code length for all codes is 99 characters:

EXAMPLES: 0132 = variable length from 1 to 32 digits in the code. 1010 = 10 digit code length only.

31

Page 40

TC1100

3 CODE SELECTION (continued) DESCRIPTION Code 128

Code 93

Codablock-A (TC1100-x200 only)

Codablock-F (TC1100-x200 only)

MSI (TC1100-x100 only)

Plessey (TC1100-x100 only)

Telepen (TC1100-x100 only)

disable Code 128 family enable Code 128 control without transmission of check digit enable EAN 128 control without transmission of check digit code length add GS before Code disable enable ISBT 128 enable ISBT 128 disable Code 93 family enable Code 93 control without transmission of check digit disable enable disable the family enable Standard enable EAN disable the family no check MOD10 no tx MOD10 with tx MOD11-MOD10 no tx MOD11-MOD10 with tx MOD10-MOD10 no tx MOD10-MOD10 with tx disable the family Standard no check Standard check - with tx Standard check - no tx Anker no check Anker check - with tx Anker check - no tx disable the family Numeric no check Numeric check - with tx Numeric check - no tx Alpha no check Alpha check - with tx Alpha check - no tx xx =

ASCII numbers that define the minimum acceptable code length;

yy =

ASCII numbers that define the maximum acceptable code length;

32

STRING AI0 AI11 AI21 AILxxyy EQ0 EQ1 AI31 AK0 AK1 AO0 AO1 AN0 AN1 AN2 AE0 AE1 AE2 AE3 AE4 AE5 AE6 AE7 AF0 AF11 AF12 AF13 AF21 AF22 AF23 AL0 AL11 AL12 AL13 AL21 AL22 AL23

Page 41

SOFTWARE CONFIGURATION STRINGS

3

CODE SELECTION (continued) DESCRIPTION Delta IBM

Code 16K Code 49 PDF417 (TC1100-x200 only)

Pharmacode (TC1100-x100 only)

disable the family no check Type 1 check Type 2 check disable enable disable enable disable enable disable forward code TX direction reverse code TX direction overflow ratio xx =

ASCII numbers that define the minimum acceptable code length;

yy =

ASCII numbers that define the maximum acceptable code length;

STRING AH0 AH1 AH2 AH3 AJ0 AJ1 AM0 AM1 AR0 AR1 AQ0 AQ1xxyy AQ2xxyy AQAzz

The minimum code length is 2 characters. The maximum code length is 16 characters. zz =

ASCII numbers that define the overflow ratio value in the range 02-10.

33