Modify > Serial Communications > Serial Port n (Message Definitions Tab) Selection Serial Port n Framing
Definition Header – Character string specified for the header. Sent at beginning of every message. Trailer – Character string specified for the trailer. Use to terminate every message. Sequence – Select a sequence specification from the dropdown list. When enabled, a sequence number consisting of the specified number of digits is inserted just after the header. Check Digits – Select BCC or CRC from the drop-down list. When enabled, a two-character hexadecimal check character is inserted just prior to the trailer. Include All Bar Codes In One Message – Affects how the reader handles multiple bar code transmissions. If enabled, the bar codes are sent in one message (between one header and trailer) and each bar code is separated by the specified “separator.” If disabled, all bar codes are always sent as a complete message (with a header and trailer). Bar Code Separator – Enter a string to use to separate multiple bar codes. Power Up/Connect Message – Enter a string that will always be sent on reader initialization. Note that for a Network connection, this string is always sent when the socket connection is made.
Phone: 800.894.0412 - Fax: 888.723.4773 - Web: www.ctiautomation.net - Email: info@ctiautomation.net
Page 172
Serial Port n Messages
Read – Select None, Send Bar Code, or User Defined from the drop-down list. • None: Nothing is sent for a valid read. • Send Bar Code: The bar code is sent. • User Defined: A user specified message. No Read – Select None, Repeat Character, or User Defined from the drop-down list. • None: Nothing is sent for a no-read • Repeat Character: Tthe specified “repeated character” is sent. • User Defined: A user specified message is sent. Multiple Bar Code – Select None, Repeat Character, or User Defined from the drop-down list. • None: A multiple bar code condition will be reported as a “Read” of the first bar code reported. • Repeat Character: The specified “repeated character” is sent. • User Defined: A user specified message. Match - Select None, Send Bar Code, or User Defined from the drop-down list. • None: If the bar code matches any of the match codes, no message is sent. • Send Bar Code: The bar code is sent. • User Defined: A user specified message is sent. No Match – Select None, Repeat Character, or User Defined from the drop-down list. • None: If the bar code does not match any of the match codes (or was a no-read), no message is sent. • Repeat Character: The specified “repeated character” is sent. • User Defined: A user specified message. No Match Read – Select None, Send Bar Code, or User Defined from the drop-down list. • None: if a valid bar code was read but it does not match any of the match codes, no message is sent. • Send Bar Code: The bar code is sent. • User Defined: A user specified message is sent. Append QQ (code quality) to message – When enabled, a two digit “code quality” indication (number from 00 to 99) is appended to the bar code message. The code quality value is separated from the message with the specified separator. Append Distance to message – When enabled, the approximate distance to the bar code can be appended to the bar code message. The distance data is separated from the message with the specified separator.
Phone: 800.894.0412 - Fax: 888.723.4773 - Web: www.ctiautomation.net - Email: info@ctiautomation.net
Page 173
Heart Beat Message – When enabled, if no other message is sent for the time period specified in Heart Beat Time Out, the specified heart beat message string is sent. A time-out of zero (0) disables the heart beat message. Heart Beat Time Out (0-30,000 ms) – 0 = Disable Heart Beat Message, 1-30,000 = Delay before sending Heart Beat Message
Phone: 800.894.0412 - Fax: 888.723.4773 - Web: www.ctiautomation.net - Email: info@ctiautomation.net
Page 174
Modify Settings > Network
Selection IP Address
Definition An IP address specifies a unique host ID for the reader on a particular TCP/IP network. Consult your network administrator for a valid IP address for the reader.
Net Mask
A netmask modifies a standard IP address into subnets. A default class C net mask should be valid in most cases (either 255.255.255.0 or 255.255.240.0).Consult your network administrator for the net mask value.
Gateway IP Address
A gateway allows the reader to reach other systems not on the local network and remote systems to reach the reader. If the reader requires external connectivity, consult your network administrator for a valid gateway address.
Net Host Port n
Select a connection types from the drop-down list: • TCP/IP Server: As a TCP/IP server, a Port # must be specified. In this case it is the TCP/IP client’s responsibility to initiate an reader connection. • TCP/IP Client: As a TCP/IP client, aside from the Port #, the Remote IP Address of the TCP/IP server which is to receive the bar code messages must be specified. The Reconnect Time specifies the interval (in seconds) at which the reader attempts to establish a connection. • UDP Messaging: In some instances, it may be desirable to use a connectionless UDP message to send bar code data. In this case, only the Port # and Remote IP Address of the receiver need to be specified.
Phone: 800.894.0412 - Fax: 888.723.4773 - Web: www.ctiautomation.net - Email: info@ctiautomation.net
Page 175
• EtherNet/IP (Net Host Port 1 only): This enables use of the EtherNet/IP protocol on the reader. Message Format – Select one of four message format options from the drop-down list: • Custom Message • Standard Message • User Defined Message • MUX (ADP) Message. Controller /Client Configuration
Number of Client Readers – The number of client readers in a multi-reader network must be defined on the reader acting as the Controller. Select the number of client readers from the drop-down list (up to 14). Group Number – All readers in a Controller/Client group (both the Controller and ALL Clients) must be set to the same controller group number. The default of zero (0) disables a reader from listening for messages to become a client. Redundant Controller Configuration – Select None, Primary Controller, or Backup Controller from the drop-down list. • None: This reader will act only as a client • Primary Controller: This reader will act as the primary controller • Backup Controller: This reader will act as a backup controller if the primary controller fails
Ethernet Link
Select AutoDetect, 100 Mbps/Full Duplex, 100 Mbps/Half Duplex, 10 Mbps/Full Duplex, or 10 Mbps/Half Duplex from the drop-down list.
Controller Multicast IP Address Telnet Port
Enter an IP address for multicast. Enter a Telnet port ID.
Phone: 800.894.0412 - Fax: 888.723.4773 - Web: www.ctiautomation.net - Email: info@ctiautomation.net
Page 176
Modify Settings > Network (Message Definition Tab)
Selection Network 1 Framing
Definition Header - Character string specified for the header. Sent at beginning of every message. Trailer - Character string specified for the trailer. Use to terminate every message. Sequence - When enabled, a sequence number consisting of the specified number of digits is inserted just after the header. Check Digits - When enabled, a two-character hexadecimal check character is inserted just prior to the trailer. Include All Bar Codes In One Message (checkbox) - Affects how the reader handles multiple bar code transmissions. If enabled, the bar codes are sent in one message (between one header and trailer) and each bar code is separated by the specified “separator.” If disabled, all bar codes are always sent as a complete message (with a header and trailer). Bar Code Separator - String used to separate multiple bar codes (see above). Power Up Message - String that is always sent on reader initialization. Note that for a Network connection, this string is always sent when the socket connection is made.
Network 1 Messages
Read – Select None, Send Bar Code, or User Defined from the drop-down list. • None: Nothing is sent for a valid read. • Send Bar Code: The bar code is sent. User Defined: A user specified message.
No Read – Select None, Repeat Character, or User Defined from the drop-down list.
Phone: 800.894.0412 - Fax: 888.723.4773 - Web: www.ctiautomation.net - Email: info@ctiautomation.net
Page 177
• None: Nothing is sent for a no-read • Repeat Character: Tthe specified “repeated character” is sent. • User Defined: A user specified message is sent. Multiple Bar Code – Select None, Repeat Character, or User Defined from the drop-down list. • None: A multiple bar code condition will be reported as a “Read” of the first bar code reported. • Repeat Character: The specified “repeated character” is sent. • User Defined: A user specified message. Match - Select None, Send Bar Code, or User Defined from the drop-down list. • None: If the bar code matches any of the match codes, no message is sent. • Send Bar Code: The bar code is sent. • User Defined: A user specified message is sent. No Match – Select None, Repeat Character, or User Defined from the drop-down list. • None: If the bar code does not match any of the match codes (or was a no-read), no message is sent. • Repeat Character: The specified “repeated character” is sent. • User Defined: A user specified message. No Match Read – Select None, Send Bar Code, or User Defined from the drop-down list. • None: if a valid bar code was read but it does not match any of the match codes, no message is sent. • Send Bar Code: The bar code is sent. • User Defined: A user specified message is sent. Append QQ (code quality) to message – When enabled, a two digit “code quality” indication (number from 00 to 99) is appended to the bar code message. The code quality value is separated from the message with the specified separator. Append Distance to message – When enabled, the approximate distance to the bar code can be appended to the bar code message. The distance data is separated from the message with the specified separator. Heart Beat Message – When enabled, if no other message is sent for the time period specified in Heart Beat Time Out, the specified heart beat message string is sent. A time-out of zero (0) disables the heart beat message. Heart Beat Time Out (0-30,000 ms) – 0 = Disable Heart Beat Message, 1-30,000 = Delay before sending Heart Beat Message.
Modify Settings > Trigger There are three options for trigger source:
Phone: 800.894.0412 - Fax: 888.723.4773 - Web: www.ctiautomation.net - Email: info@ctiautomation.net
Page 178
• Hardware Trigger • Software Trigger • Continuous Trigger
Trigger Source = Hardware Trigger Selection Hardware Trigger
Definition Select Active Low or Active High from the drop-down list. • Active Low: The trigger is active when the input signal is low. • Active High: The trigger is active when the input signal is high. Debounce Duration – Defines the minimum time required for detection of a change in the hardware trigger status.
Trigger Source = Software Trigger
Selection Software Trigger
T r i g g
Definition Select a Host n or Net n trigger source from the drop-down list. The software trigger can be any “host port” either serial or network based. Start Char – Define the ASCII characters that, when received, will start trigger for the reader. End Char – Define the ASCII characters that, when received, will end trigger for the reader.
Phone: 800.894.0412 - Fax: 888.723.4773 - Web: www.ctiautomation.net - Email: info@ctiautomation.net
Page 179
Trigger Source = Continuous Trigger
Selection Continuous Trigger Type
Definition Select a Continuous, Timed, or Once from the drop-down list. • Continuous: Reader is always in trigger and transmits any bar codes that are seen. • Timed: Whenever a code is transmitted, trigger is removed for the period of time specified in Trigger OffTime. • Once: Same as continuous, but the same bar code will not be sent a second time until the Bar Code Timeout value has expired. Bar Code Time-Out – When Continuous Trigger Type is set to Once, the same bar code will not be sent a second time until the Bar Code Timeout value has expired.
Phone: 800.894.0412 - Fax: 888.723.4773 - Web: www.ctiautomation.net - Email: info@ctiautomation.net
Page 180
Modify Settings > Relay
I/O 1 Configuration = Used as input Selection Custom Input
Definition Select this for use with applications involving custom software to select a customized input option.
Learn Match Code
When this input is selected, the AL5010 is placed into Test Mode in order to "learn" a match code. When the line is deactivated, Test Mode is exited. When this option is enabled, the last code read is stored in the first entry of the Match Table.
End Trigger
Active as end of trigger signal.
Height Tracking Photoeye
NOTE: Available only at I/O#1 if the Trigger Source is set to Hardware Start/End Trigger. Used for applications that require Height Tracking Photoeyes. Up to three photoeyes may be used to divide the reader's depth of field into up to four zones. The Distance from reader (in inches) must also be set. If an error message is displayed when saving or retrieving parameters, be sure that the proper dimensions are entered in the Reader Mounting (near/far) and I/O Configuration (distance from scanner). For installation details, reference the AL5010 Product Manual, Chapter 3, Mounting the Height Tracking Photoeyes.
Active Low/ Active High
Select Active Low or Active High from the drop-down list. • Active Low: The trigger is active when the input signal is low. • Active High: The trigger is active when the input signal is high.
Phone: 800.894.0412 - Fax: 888.723.4773 - Web: www.ctiautomation.net - Email: info@ctiautomation.net