The text of a manual · part 3 of 8

Apex II Portable Data Collection Terminal Operations & Programming Manual, Part 1: User's Manual, pages 19 to 27

Pages 19 to 27 of 71

The original document is the PDF (342 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Compsee Apex II

Page 19

Step 5.

Action/Result Testing the Scanner • Apex II with tethered Wand. To test using a bar code wand, depress <Scan>. Place the tip of the wand either before or after the bar code symbol. Move the wand across the entire bar code at an even speed. A good read is indicated by an audible tone. If a no read condition occurs, scan the symbol again. Make sure the wand intersects all the symbol elements. Please observe 1/4" Quiet Zone on each side. If the symbol still does not read, review the troubleshooting procedures in Section 5 or call your Compsee dealer for assistance. The wand powers down when the trigger of the input device or the Apex II <SCAN> key is released, unless it is programmed to perform differently (programmer designates that when a certain key is pressed, the wand is activated until the key is pressed again). • Apex II with Integrated Laser. Depress <Scan>. Make sure to orient the laser scanning unit so that the red aiming beam intersects all bar code elements. A good read is indicated by an audible tone accompanied by a steady green LED indication. If a no read occurs, scan the symbol again. Make sure to orient the laser scanning unit so that the red aiming beam intersects all bar code elements. If the scanner still does not read the bar code, review the troubleshooting procedures in Section 5 or call your Compsee dealer for assistance. Note DO NOT aim the unit's laser perpendicular at the bar code symbol. When reading symbols printed on a glossy surface, this causes a reflection that interferes with the scanning action. Point the scanner so that the laser is angled and NOT pointing directly at the bar code.

6.

Observe Automatic Power Down. If the unit is powered up and left inactive for a period of 90 seconds, the unit will automatically power down. The 90 second power down feature is a default setting which can be modified. Refer to the SYSPARMS statement in the COMPSEE BASIC Interpreter Programmer's Manual. Reactivate the unit by pressing the <POWER> Key.

7.

Observe Buffer Full Indication (all terminal models). Unless an application program has been loaded into the Apex II terminal, the scan buffer becomes full after several scans. The Apex II's buffer is defined as the assigned part of memory used as a temporary holding area. To observe the buffer full indication, scan additional bar code symbols. After several reads, the terminal will emit ten (10) audible tones, indicating that the buffer is full. Press <ALT> and <Z> to clear the buffer. Alternatively, you can clear the buffer by rebooting the terminal as follows. In addition to clearing the scan buffer, rebooting the terminal erases any programs and data that are in RAM.

Compsee, Inc. Apex-0698-Part 1: User’s Manual

19 Part 1

Page 20

Step 8.

Action/Result Reboot the Terminal. 1.

Power down the unit.

2.

Simultaneously hold down the <ALT> and <.> keys.

3.

While still holding down the keys, press <POWER>.

4.

The screen will clear and then the Compsee Inc. copyright screen will appear.

CAUTION Rebooting an Apex II terminal erases your application program and all data files. In addition, loading a new operating system version will cause the Apex II to reboot.

Apex II STANDARD PARTS AND ACCESSORIES After unpacking the Apex II, inspecting for damage, and performing the Checkout procedure, take time now to learn the functions of the various Apex II terminal component parts and accessories shown in Figure 1-2.

NO.

Part

Description

1.

Power Key

Turns unit on/off.

2.

Scan Key

Activates the scanner on the Apex II.

3.

LCD Display

4-lines by 16 characters per line. The liquid crystal display provides the operator with visual feedback (prompts and warning messages).

4.

Scanner Port

Standard -DB 9 Connector Industry Standard scanner connection for wands, lasers, CCD's, etc. Accepts wand or laser digital emulation signals. Integrated Laser - Located on top of the Apex II. Emits the light which scans the bar code.

5.-7.

Keypad

This rubber-like keypad enables the operator to respond to prompts and/or to enter data manually. Specific keys are color-coded to provide quick key recognition.

8.

RS-232-C Port

A 10-pin communications port which allows the Apex II terminal to communicate with any computer host, serial printer, modem, point-of-sale terminal, and many other devices that support RS-232-C connections and ASCII data formats. This port is used primarily for loading COMPSEE BASIC programs and data files into the terminal memory and for unloading any data collected by the Apex II. See Section 10 for additional information on data communications and pin assignments for this port.

9.

Battery Case

(Located on the back of the Apex II). Accepts either four Alkaline batteries or a Nickel Metal Hydride Battery Cell. It provides power for approximately 15 hours of operation at an average of 30 scans per minute.

10.

Good Read LED Indicator

Integrated Laser Units Only. Will flash a steady green light when a bar code is successfully scanned.

Apex II KEYPAD DEFINITIONS The following is a description of the various keys on the terminal and their functions and values at as they exist when the terminal is shipped to you (prior to loading of your applications program).

Compsee, Inc. Apex-0698-Part 1: User’s Manual

20 Part 1

Page 21

Note Some of the keys on the Apex II have specific values at boot- up time (starting values are set at the Compsee factory), but all of them can be redefined by the programmer. Consult the Keypad Table topic in the COMPSEE BASIC Interpreter Programmer's Manual for instructions.

Key

Function

POWER

Used to turn the Apex II on and off. Red Key

<ALT>

Considered a control key and is used in conjunction with other keys to perform certain functions. (See Section 8.)

Arrow keys

The arrow keys do not have a specific function. They're defined by the programmer as a way of paging up or down to display information on the Apex II's display.

F1/F2/F3/F4

These keys do not have a designated function at boot-up, but can be defined by the programmer to be used within the application program.

A-Z

Used for keypad entry of data or responses within a program. Some are used in conjunction with the <ALT> key to perform other functions. Refer to Section 8 for a listing of these functions.

0-9

Used for keypad entry of numeric data.

<SPC>

Functions like the Spacebar on a typewriter or PC.

<BKSP>

Functions like the Backspace key on a typewriter or PC.

<.>

The period key can be defined to be used as a decimal point.

<ENTER>

Used to accept keypad entry as input into the program and is defined as such at boot-up time.

ADDITIONAL REQUIREMENTS The following equipment and software (not supplied with your Apex II terminal) is required for a complete data collection system.

•

Intelligent source/host (PC, VAX, etc.) to load COMPSEE BASIC programs and data files into the scanner and to extract data collected from the scanner. Details and instructions for these procedures are provided in Section 10.

•

Communications software to enable a two-way communication between the Apex II terminal and an intelligent source or host. The demonstration software that we have supplied with the manual has built-in communications capabilities. The user can write and modify their own program(s) or use a third party package such as PROCOMM+ or CrossTalk for communication purposes. Compsee sells the Apex II Program Generator software tool which can interface an Apex II terminal to an IBM or compatible PC. The Apex II Program Generator is designed to allow anyone, regardless of technical expertise, to design a COMPSEE BASIC program using the program generator. Programs that have been written using the program generator can be easily modified with it. The Apex II Program Generator contains built-in communications capability. (See Section 12 for product part numbers.)

Compsee, Inc. Apex-0698-Part 1: User’s Manual

21 Part 1

Page 22

•

An application program to actually perform the data collection. (The pre-loaded Apex-Trak programs have been provided for immediate usage.) See PART 2 for details on how to use these programs.

Once you know how to operate the Apex II terminal, you can write a customized program for your application. Consult Section 4, Building Your Application, for details on how to do this.

2

OPERATING SYSTEMS MODES

To accommodate the various functions and operations that the Apex II is capable of performing, three system modes are provided: IMMEDIATE, RUN and DIAGNOSTIC. Each of these modes represent a different state of functionality. Consequently, it is required that the operator know which commands and which functions can be performed in each of these modes. The following is an extensive explanation of the modes and the operations which can be performed while in those modes.

IMMEDIATE MODE In IMMEDIATE mode, the operating system is looking at the RS-232-C port for direction. This is the mode the Apex II terminal should be in if you plan to send information such as a program, an operating system update, or a file from the host to the Apex II, or from the Apex II to the host. When an Apex II terminal is in IMMEDIATE mode, it is very much like a PC waiting at the C:> prompt. The C:> indicates that the application program is not running and the operating system is waiting for instructions. While in this mode, you can use your PC to send commands such as DIR, CLS, and BEEP to the Apex II. When COMPSEE BASIC is in IMMEDIATE mode, a statement can be entered and executed immediately, or it can be entered and stored away for later execution in RUN mode. The statement must be terminated by a carriage return (ASCII code 13) and a linefeed (ASCII code 10). If the statement is to be stored for later execution, it must be preceded by a line number. If no line number precedes, the statement executes immediately. For example, the following statement executes immediately, since the statement "PRINT 5" is not preceded by a line number:

PRINT 5 5 > In this case, "PRINT 5" is preceded by a line number. Before this statement executes, COMPSEE BASIC must be placed in RUN mode (using the RUN command). When the Apex II terminal is in RUN mode, the application program stored in memory is executed.

10 >RUN 5 >

PRINT 5

It is necessary to be in IMMEDIATE mode to : • Receive a BASIC program, or • Change/edit a BASIC program in Apex II memory via a PC, or • Receive a new operating system, or • Interact with the operating system (via RS-232-C). Terminals in the IMMEDIATE mode are those which: • Are re-booted. • The word QUIT (diagnostic command) has been typed on an Apex II that was in DIAGNOSTIC mode. This causes the Apex II to drop into this mode.

Compsee, Inc. Apex-0698-Part 1: User’s Manual

22 Part 1

Page 23

To determine if the terminal is in this mode: • Press <ALT><V>. This should produce a message on the Apex II's screen containing the operating system version, or • Press <ALT><X>. This should produce the MODE setting control screen. Either one of the above will indicate that a terminal is in the IMMEDIATE mode. If the Apex II LCD doesn't display one of these messages, then you may be in the RUN or DIAGNOSTIC modes.

RUN MODE The RUN mode is where the Apex II terminal is being controlled by the instructions of a BASIC program. The operating system is ultimately controlling the terminal; however, all direction is coming from the user's program. While in this mode, keyboard inputs, RS-232-C communications, file manipulation, and other functions are the responsibility of the BASIC program running within the terminal. All input to the terminal (keyboard/scanner/ RS-232-C) are accepted, processed, or ignored by the BASIC program. For example, if an inventory program is running on the Apex II terminal and the PC sends the DIR command through the RS-232-C port to the Apex II, then the operating system will not receive it since it is now considered data by the BASIC program. You must be in RUN mode to: • Execute the instructions of a BASIC program. You can get into RUN mode by: • Sending the RUN command to the Apex II via the RS-232-C port, or • Pressing <ALT> <R>, or • Typing G and the program line number will return you to the RUN mode. You can determine if the terminal is in the RUN mode by : • Pressing <ALT><C> will produce a PRESS ENTER message on the Apex II. This key sequence accesses the DIAGNOSTIC mode and only produces this message in the RUN mode.

DIAGNOSTIC MODE In DIAGNOSTIC mode, the operating system is looking at the Apex II's keyboard for direction. This mode is used for debugging a BASIC application program, single stepping through a program, input values for variables, printing the values of variables, setting break points, etc. A programmer would use this mode to aid in debugging a BASIC program running within the Apex II. Once a unit is placed in DIAGNOSTIC mode, the operating system is once again in control of the Apex II terminal. Diagnostic mode supports a number of options that allow the user to directly manipulate limited functions of the operating system. This includes checking the contents of variables, changing the value of variables, changing the flow of the program, and tracing the flow of the program. You must be in Diagnostic mode to: • use diagnostic functions (as defined in this COMPSEE BASIC Interpreter Programmer's Manual), or • to break into the IMMEDIATE mode from the keypad. You can get into Diagnostic mode by: • Pressing and holding <ALT>; then pressing<C>. When the >ENTER prompt is displayed, press and hold <ALT>; then press <D>, or • Having a STOP statement in the BASIC program, or • Setting a BREAK point (diagnostic mode) in the program, or • Running a BASIC program within the terminal that encounters an un-trapped error. This will cause the scanner to break into this mode.

Compsee, Inc. Apex-0698-Part 1: User’s Manual

23 Part 1

Page 24

You can determine if the terminal is in this mode by: • viewing the Apex II's display. You should find the Apex II prompt (>) on the second line. Additionally, when you press keys on the Apex II keypad, the corresponding character will be displayed on the Apex II's screen. The following diagram represents mode flow. Each of these modes represent a different state of functionality. Terminal rebooted

Diagnostic Mode - type “Quit”

IMMEDIATE

RUN

With a program loaded, send RUN command via RS-232C or With a program loaded, press <ALT><R> <ALT><C> and <ALT><D> on keypad Encounter an error Encounter a STOP statement Encounter a Break Point

DIAGNOSTIC Enter QUIT from the keypad

KEYBOARD COMMANDS While in IMMEDIATE mode, COMPSEE BASIC recognizes certain ASCII characters received from the keyboard as special keyboard commands. When an Apex II terminal is first powered up, the default keys are assigned the proper ASCII values to invoke the keyboard commands. The user is free to reassign these ASCII values to other keys. See Section 8, Keypad Table, for further information on reassigning key values.

Compsee, Inc. Apex-0698-Part 1: User’s Manual

24 Part 1

Page 25

The following is a quick reference guide for Compsee keyboard commands: FUNCTION

KEYBOARD COMMAND

Reboot Apex II - Program remains in FLASH

Hold down ALT•(period); Then turn Apex II ON

Reboot Apex II - Program is erased from FLASH

Hold down ALT SPC; Then turn Apex II ON

Check Version

ALT

V

Display communication parameters

ALT

X

Set communication parameters when they are displayed

<X> to change first number <Y> to change second number Press <ENTER> to exits

Clear Buffers

ALT

Z

Run the Program

ALT

R

ALT

C (followed by) ALT D

Contrast Level Up

ALT

I

Contrast Level Down

ALT

J

Save program to FLASH

ALT

S

Stop a Program that is Running (unit is then in debug mode)

Keyboard Command: CONTRAST LEVEL UP/DOWN <ALT><I> or <ALT><J> Keystrokes assigned by the operating system to control contrast. <ALT><I> increases the contrast, while<ALT> <J> decreases the contrast.

Keyboard Command: RUN <ALT><R> Executes the application program stored in memory.

Keyboard Command: VERSION <ALT> <V> Displays the COMPSEE BASIC version level of the operating system stored in RAM for three seconds. The content of the screen is saved before the version is displayed. Three seconds later, the content of the screen is restored.

Compsee, Inc. Apex-0698-Part 1: User’s Manual

25 Part 1

Page 26

Keyboard Command: MODE <ALT> <X> Change the communications parameters of the RS-232-C port. The content of the screen is saved and the following message is displayed on the first line: MODE X,Y where X denotes the baud rate, and Y is the status. The data mode is always set to binary. Change the communications parameters on the Apex II's RS 232 port as follows: 1.

Press <ALT> <X> to view the current parameters. A message similar to the one below will appear. A table of mode settings and their characteristics is provided at the end of this MODE Command topic. MODE 5,3

2.

Ensure that the cursor is flat (the cursor is shown next to the second number), which means it must appear as an underscore, not a rectangle. Press <ALT> until the flat cursor appears.

3.

Change the first number by pressing <X> until the correct baud rate value appears on the screen. The following is a chart of all the possible "X" values and the baud rate that they represent:

4.

"X" Value

Baud Rate

0 1 2 3 4 5 6

300 baud 600 baud 1,200 baud 2,400 baud 4,800 baud 9,600 baud 19,200 baud

Change the second number by pressing <Y> until the desired number is reached.

Compsee, Inc. Apex-0698-Part 1: User’s Manual

26 Part 1

Page 27

The following is a chart of all the possible "Y" values and what they represent:

5.

Status Byte "Y" Value

Flow Control

Data Bits

Parity

Stop Bits

0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

RTS/CTS XON/XOFF RTS/CTS XON/XOFF RTS XON/XOFF RTS/CTS XON/XOFF RTS/CTS XON/XOFF RTS/CTS XON/XOFF RTS/CTS XON/XOFF RTS/CTS XON/XOFF

8 8 7 7 8 8 7 7 8 8 7 7 8 8 7 7

N N E E O O O O N N E E O O O O

1 1 1 1 1 1 1 1 2 2 2 2 2 2 2 2

Press <Enter> to save the new settings.

NOTES 1.

2.

3.

When modifying the communications parameters, DO NOT power down the Apex II terminal until you have completed all of the steps listed above. If you power down prior to completing the procedure and then power the unit on again, the last command entered will still be displayed. In some cases, after you have changed the communication parameters, the screen buffer may contain the previous communications parameters. This may lead you to believe the terminal did not recognize the new settings. To verify that the new settings are stored, press <ALT> <X>. The new settings will appear. Press <Enter> to return to the previous screen. While the keyboard is executing the MODE command, <ALT> <X> steps through the baud rate values and <ALT> <Y> (ASCII value 25) steps through the status values. The <Enter> key (ASCII value 13) terminates the MODE command, sets the RS-232-C port, and restores the screen. The keyboard remains in ALT mode until the <ALT> key is pressed.

Keyboard Command: CLEAR BUFFERS <ALT> <Z> Clears the four COMPSEE BASIC buffers: 1. 2. 3. 4.

RS-232-C receive RS-232-C transmit Keyboard Bar Code

Compsee, Inc. Apex-0698-Part 1: User’s Manual

27 Part 1