Hand-Held Decoded Bar Code Scanners (Cat. No. 2755-HDG-4) User Manual: 5 - Troubleshooting and Maintenance

Pages 86 to 89 of 115, from Hand-Held Decoded Bar Code Scanners (Cat. No. 2755-HDG-4) User Manual

The original document is the PDF (622 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Allen-Bradley 2755-HDG-4

Page 86

Chapter

5

Troubleshooting and Maintenance This chapter describes how to troubleshoot and maintain the scanner.

• troubleshooting chart • cleaning the scan window • Allen-Bradley Global Technical Support

Publication 2755-6.2

Page 87

5-2

Troubleshooting and Maintenance

Troubleshooting the Scanner This section provides a list of the most common operating problems, the probable causes, and suggested corrective actions. Problem

Probable Cause(s) 1. Power supply not connected to Synapse cable.

1. Make sure power supply is connected and plugged into a power source.

2. Scan beam has timed out.

2. Normal operation. Release trigger completely and scan again.

3. Defective interface cable.

3. If possible, try another cable.

4. Defective scanner.

4. If possible, try another scanner using the same connections. Contact Allen-Bradley Global Support (GTS).

1. Defective scanner.

1. Contact Allen-Bradley Global Technical Support.

1. Scanner not configured for correct wedge cable.

1. Make sure scanner is configured for correct wedge cable.

2. Caps lock is on.

2. Turn Caps lock off.

3. Incorrect interface cable.

3. Check that you are using the proper cable.

4. Decoder options not properly configured.

4. Check all options such as UPC-E to UPC-A conversion.

5. Scanned data still appears incorrectly.

5. Contact Allen-Bradley Global Technical Support.

No Scan Beam.

Scan Beam Appears as a Dot Not a line

Scanned data appears incorrectly on a display terminal when using a keyboard wedge cable.

Publication 2755-6.2

Corrective Action

Page 88

Troubleshooting and Maintenance

Problem

Scan Beam Present, Symbols Not Read

Data is decoded but not transmitted to host.

5-3

Probable Cause(s)

Corrective Action

1. Scanner not set to decode symbol type scanned.

1. Verify the scanner decoder settings.

2. Scanner not held at slight angle to symbol.

2. Hold scanner at an angle to the symbol.

3. Scan beam not crossing entire symbol.

3. Make sure scan beam crosses every bar and space on the symbol.

4. Poor quality bar code symbols.

4. Use the symbols provided on the inside back cover for testing.

5. Loose cable connections.

5. Check for loose cable connections.

6. Incorrect interface cable.

6. Check that you are using the proper cable.

1. Scanner is not configured for the proper cable.

1. Make sure the scanner is configured for the correct cable (standard RS-232 or smart cable).

2. Loose connections

2. Check all cable connections.

3. Scanner communication settings do not match host requirements.

3. Verify cable communication settings such as baud rate.

Publication 2755-6.2

Page 89

5-4

Troubleshooting and Maintenance

Cleaning the Scan Window You may need to clean the window of the scanner. Carefully clean the window by first removing loose particles of dirt with clean air. Then use an optical quality cloth moistened with an optical quality cleaning fluid for plastic lenses and wipe the window in a single direction (don’t wipe cloth back and forth across window). Do not leave streaks.

!

!

ATTENTION: Do not use abrasive materials such as disposable wipes and facial tissue. Do not use solvents like alcohol or acetone. These materials will damage the window or the finish on the scanner.

ATTENTION: The scanner has no serviceable parts. Do not open the housing of the scanner.

A-B Technical Support If you should require assistance or need additional information on operating the scanner or Allen-Bradley decoder, contact Allen-Bradley Global Technical Services (GTS) at (216) 646-6800.

Publication 2755-6.2