Two-page Thornton spec sheet from 1998 for the 1800 Series Smart Calibrators, headed as the 1855 Calibration Set. It explains how the calibrators plug into a Thornton Series 7 instrument in place of a sensor to verify and calibrate it, and lists specifications for the resistance/temperature calibrators and the voltage calibrator.
- Device
- Smart Calibrator set that stands in for a process sensor to verify and calibrate a Thornton Series 7 instrument p. 1
- Accuracy
- Within ±0.8% of value, NIST traceable p. 1
- Calibration period
- Valid for one year; ambient temperature range 20-30°C p. 2
- How it works
- Memory (NVRAM) in each calibrator holds its ID, calibration value and last factory calibration date; the instrument enters calibration mode on its own p. 1
- Calibration time
- Verifying and calibrating an instrument takes under 2 minutes p. 1
- Included
- Certificate of NIST traceability, blank verifier labels and instructions with every calibrator; sets come with a carrying case p. 2
- Voltage calibrator
- 410 mV nominal, ±0.17 mV after factory calibration, ±0.09 mV drift over the period; for pH, pressure and tank level inputs p. 2
- Resistance/temperature calibrators
- For resistivity, conductivity, TDS, % concentration and pH (temperature only); temperature accuracy ±0.3°C, drift ±0.5°C p. 2
- Resistance calibrator values
- 1854-02 at 1 Mohm-cm or 1 µS/cm; 1854-03 at 0.1 Mohm-cm or 10 µS/cm; 1854-04 at 10 kohm-cm or 100 µS/cm (0.1/cm cell constant) p. 2
- Patent
- U.S. Patent 5,248,933 p. 2
What this sheet walks you through
- Calibrating an installed instrument channel by channel with the Smart Calibrators p. 1
- How the smart calibration technology works p. 1
- Calibrator specifications p. 2
- Ordering the calibration sets p. 2
Good to know
- Calibration values are valid for one year; they can then be verified with suitable test equipment or sent to Thornton for recertification. p. 1
- A calibration is called for when the displayed and reference values differ by more than the 0.3% instrument accuracy. p. 1
- Specifications assume an ambient temperature of 20-30°C. p. 2
Questions this sheet answers
How accurate are the Thornton Smart Calibrators?
Their values are certified within ±0.8% of reading and are traceable to NIST. p. 1
How long is a Thornton Smart Calibrator calibration valid?
One year; the value is printed on each calibrator's label. p. 1
How long does it take to calibrate an instrument with the 1855 set?
The sheet gives under 2 minutes in total to verify and calibrate, with each adjustment done within 3 seconds. p. 1
What does the Thornton voltage calibrator output?
A nominal 410 mV with ±0.17 mV accuracy after factory calibration. p. 2
The file
Size and text
Every page · 2
Read it here
Each page as it is printed, and its words beside it, so a phone can search and copy them.
Page1of 2
Open in the PDF
The text of page 1, from the PDF itself
SMART CALIBRATORS
...innovations in instrumentation
1800 SERIES
FEATURES
NIST traceable calibration Accurate to ±0.8% of value Fast and easy to use No other equipment required Unique Patented Thornton Technology
APPLICATIONS With the advent of Total Quality Management (TQM) and ISO 9000 programs, there is a growing need for accurate and convenient equipment to verify and calibrate process control instrumentation on a routine basis. In response to this requirement, Thornton has developed a unique patented calibration system for use with the 770PC Instrument/transmitter. The Thornton Smart Calibrator set consists of two calibrators; one for resistance and temperature measurements and the other for millivolt measurements. Each calibrator contains a circuit that develops precise values, certified within ±0.8%, of the reading, traceable to NIST. The calibration value of each parameter is stated on the label of each calibrator and is valid for one year. These values can be verified at established time intervals by a user with access to adequate test instrumentation, or, sent to Thornton for recertification.
SMART CALIBRATORS IN USE The procedure for performing a quality management calibration on an installed 700PC and in-line sensors is as follows:
- One of the two process sensors is disconnected from the Smart connector/patch cable junction, located 5 feet from the sensor.
- One of the Smart Calibrators is attached to the open patch cable, in place of the sensor.
- The 770PC instantly recognizes the calibrator and automatically reconfigures the instrument to be in the calibration mode, verifying the instrument’s functionality. The 700PC also displays its measurement of the precision reference signal(s) from the Smart Calibrator, including the effect of the patch cable.
- The measurement displayed is compared with the precision reference value listed on the calibrator’s label.
- If the two values differ by more than the 0.3% instrument accuracy, the the operator select “Do Calibration”.
- Within 3 seconds, the instrument performs the necessary calculations and circuit adjustments to make the displayed value correspond to the reference value.
- The procedure is repeated with a second Smart Calibrator.
- Replace the process sensor on the patch cable.
- Repeat the procedure on the other channel with both Smart Calibrators.
1855 Calibration Set The second process sensor is then replaced on the patch cable. The 770PC and both sensors are back on-line monitoring the process.
- The total elapsed time to verify and calibrate the instrument is less than 2 minutes.
SMART CALIBRATION SYSTEM TECHNOLOGY Smart Calibrators are key elements in the Thornton Series 7 multiparameter process control system. They use patented technology that includes a non-volatile memory device (NVRAM) in each sensor’s assembly. The calibrator’s identification code and precise calibration value(s), which are NIST traceable and accurate to ±0.8%, are saved in each calibrator’s memory. The date of the last factory calibration is also stored in the calibrator memory. The signal is used by the instrument for three functions.
- When a smart calibrator is connected to a 700PC multiparameter instrument, the instrument recognizes the calibrator and puts the instrument into a calibration mode.
- The measured value of the calibration circuit is displayed for a visual evaluation of the instrument’s condition.
- The precise calibration value is used by the instrument to calibrate its circuitry. The Thornton Smart Calibration System provides unsurpassed convenience and accuracy, while meeting system validation requirements. Unit Under Calibration
Smart Calibrator
Test/Measurement Instrument
Calibrator Circuit
- Calibrator I.D.
- Calibration Value to ±0.8% NIST Traceable
Page2of 2
Open in the PDF
The text of page 2, from the PDF itself
SPECIFICATIONS
1800 Smart Calibrators
Resistance/Temperature Calibrators - for resistivity, conductivity, TDS, % concentration and pH (temperature only) measurements Ambient Temperature Range: 20-30°C Calibration Period: valid for one year
Temperature Accuracy: ±0.3°C after factory calibration Temp. Drift Over Calibration Period: ±0.5°C
Calibrator Model Nominal Calibrator Value (0.1/cm cell constant)
18541854-021854-031854-04
18.2 MΩ-cm, 0.055 µS/cm
1 MΩ-cm, 1 µS/cm0.1 MΩ-cm, 10 µS/cm10 KΩ-cm, 100 µS/cm
Resistivity/Conductivity Accuracy (factory calibration)
±0.0072 MΩ-cm±0.0004 MΩ-cm±0.004 µS/cm±0.04 µS/cm
Drift Over Calibration Period
±0.0029 MΩ-cm±0.00016 MΩ-cm±0.0016 µS/cm±0.016 µS/cmNominal Temp. Value25°C77.6°C25°C25°C
1855 Voltage Calibrator - pH, pressure, tank level, and other voltage measurements Ambient Temperature Range: 20-30°C Calibration Period: valid for one year Nominal Calibration Value: 410 mV Voltage Accuracy: ±0.17 mV after factory calibration Voltage Drift Over Calibration Period: ±0.09 mV All calibrators include a certificate of NIST traceability, blank verifier labels and instructions. Thornton product specifications are subject to change without notice.
DIMENSIONS
ORDERING INFORMATION Part Number 1855
1855-02
Description Smart Calibration Set – with carrying case, includes the following two calibrators: 1853 Voltage Calibrator pH, pressure and tank level measurements; 1854 Resistance/Temperature Calibrator for high resistivity, very low conductivity, TDS and pH (temperature only) measurements Smart Calibration Set – with carrying case, includes the following three calibrators: 1854-02 Smart Resistance Calibrator for resistivity/conductivity, TDS and pH (temperature only) at 1 Mohm-cm or 1 µS/cm with 0.1/cm cell constant 1854-03 Smart Resistance Calibrator for resistivity/conductivity, TDS and pH (temperature only) at 0.1 Mohm-cm or 10 µS/cm with 0.1/cm cell constant 1854-04 Smart Resistance Calibrator for resistivity/conductivity, TDS, % concentration and pH (temperature only) at 10 kohm-cm or 100 µS/cm with 0.1/cm cell constant
THORNTON ASSOCIATES, INC. 1432 MAIN ST., WALTHAM, MA 02451 TEL. (781) 890-3399 Smart Calibrators U.S. Patent 5,248,933
Smartline: 800-642-4418 Faxline: 781-890-5507 E-mail: info@thorntoninc.com www.thorntoninc.com ML0006 Rev. C 11/98 Printed in U.S.A.