Spec sheet · Sick
SICK LUT 1-4 Luminescence Scanners Technical Data
Spec sheet2 pagesPDF 238 KB
Both pages are below, each with its text.
Covers · 1 model
The models it covers
From the sheet · read page by page
Key facts from this sheet
Each line names its page in the PDF.
Two SICK catalog pages on the LUT 1-4 luminescence scanners, variants LUT 1-400 through LUT 1-450. They give the dimensional drawing, connection pins, an output truth table, technical data and order numbers. A reader can identify a variant by its part number and find its colour filters and spare lamp.
- Device
- Luminescence scanner with a UV light source using a high-pressure mercury vapour lamp p. 1
- Part numbers
- LUT 1-400: 1 007 626; LUT 1-410: 1 005 935; LUT 1-420: 1 005 936; LUT 1-430: 1 005 937; LUT 1-440: 1 005 938; LUT 1-450: 1 005 939 p. 2
- Switching outputs
- QL (signal output) and QB (operational readiness), PNP, light- or dark-switching selectable p. 2
- Connection pins
- 7-pin + PE: L+, M, QL, QB, QA, NC, NC, PE p. 1
- Indicators
- Readiness indicator (red) and status indicator (green) p. 1
- Optics
- Scanning distance set by interchangeable lenses; fibre-optic connection; additional optical filter p. 1
- Ambient temperature
- Operation -10 °C to +45 °C; storage -25 °C to +75 °C p. 2
- Colour filters
- OBF-OG 570-4, OBF-RG 610-4, OBF-RG 630-4, OBF-RG 665-4 p. 2
- Spare lamp
- SLA-LUT 1-4 p. 2
What this sheet walks you through
- Choosing a LUT 1-4 variant by its part number p. 2
- Reading the output and indicator states for light- and dark-switching p. 1
- Mounting dimensions of the housing p. 1
Good to know
- A flashing state in the truth table means the lamp power is still sufficient for operation. p. 1
- Scanning distances are measured from the front edge of the lens. p. 2
Questions this sheet answers
What is the part number of the LUT 1-410?
The order table lists the LUT 1-410 as part number 1 005 935. p. 2
What is the operating temperature of the LUT 1-4 scanner?
-10 °C to +45 °C in operation; storage is -25 °C to +75 °C. p. 2
Which spare lamp fits the LUT 1-4?
The accessory list names the spare lamp SLA-LUT 1-4. p. 2
What outputs does the LUT 1-4 have?
PNP switching outputs QL (signal) and QB (operational readiness), light- or dark-switching selectable, plus an analogue output QA. p. 2
The file
Size and text
Every page · 2
Read it here
Each page as it is printed, and its words beside it, so a phone can search and copy them.
Page1of 2
Open in the PDF
The text of page 1, from the PDF itself
LUT 1-4 Luminescence scanners Dimensional drawing
253138460.550"X"A"X"12 0°A-B3030°R 3.360° 12930M5Bø48,5 ø621012535371
- UV semi-conductor light source
with high-pressure mercury vapour lamp
- Scanning distance selectable by using interchangeable lenses
- Fibre-optic connection
- Additional optical filter
6425LUT 1-410 LUT 1-420LUT 1-430 LUT 1-440LUT 1-450131012Adjustments possible5050All types50200Luminescence scannersø50 88
Scanning distance 10 . . . 90 mm
OUTPUT NOT READY /ALARM
1 2 3t on/off [ms]2.54 50.1E max6SENSITIVITY[E]
Connection type Bridge LUT 1 - 4
C V
Accessories Cable receptacles Fibre-optic cable
page 496 5527-pin + PE red1Lenses556blu2grn3brn4wht5blk6 7yel422SENSICK CATALOGUE
Lens, see dimensional drawing Readiness indicator Status indicator Sensitivity adjustment Digital intensity signal Threaded mounting hole, 12 mm deep
Truth table LUT 1 - 4 Internal jumper
oaobo
oaobo
Switching type Sender lamp
Delivered state Light-switching started off
Dark-switching started off
Luminescence Output QB (PNP) Readiness indicator (red) Output QL (PNP) Status indicator (green)
– LOW on LOW off
yes no HIGH off/flashing*) HIGH on
LOW off
– LOW an LOW off
yes no HIGH off/flashing*) LOW on
- Flashing: lamp power still sufficient for operation
L+ M QL QB QA NC NC PEa bHIGH off
Page2of 2
Open in the PDF
The text of page 2, from the PDF itself
LUT 1-4 Technical Data
LUT 1-
Scanning distance1)/ light spot diameter
With fibre-optic cable, no lens With fibre-optic cable, with 144 lens With fibre-optic cable, with 144 lens and diaphragm
410420430440
High-pressure mercury vapour lamp UV, 365 nm 4000 h 18 . . . 30 V DC 2) 2V 700 mA
Switching outputs QL5) and QB6)
Light-/dark-switching, selectable PNP: HIGH = VS – < 2 V/LOW = 0 V 200 mA 5 kHz 200 Hz 0.1 ms 2.5 ms 0 . . .1.5 V Plug
VDE protection class 8) Circuit protection 9) Enclosure rating
V
Ambient temperature
Operation – 10 °C . . . + 45 °C Storage – 25 °C . . . + 75 °C 2.5 kg Die-cast metal
A, B, C IP 63
Weight Housing material/surface 1) From front edge of lens 2) Limit values 3) May not exceed or fall short of VS 4) Without load
5) QL = signal output 6) QB = operational readiness
7) With light/dark ratio 1:1 8) Reference voltage DC 50 V
9) A = VS connections reverse-polarity protected B = Outputs QL, QB and QLU short-circuit protected C = Interference pulse suppression
Scanning distance 1 2 3 4 5
200 1 100 502Rel. sensitivity in %
10 mm 20 mm 50 mm 125 mm 300 mm
5 5
20
Order information Type Part no. LUT 1-400 1 007 626 LUT 1-410 1 005 935 LUT 1-420
1 005 936LUT 1-4301 005 937LUT 1-4401 005 938LUT 1-4501 005 939
Special accessories Colour filter
4105 10
Scanning distance Scanning distance Scanning distance Scanning distance Scanning distance
320(mm) 2450
10 mm/l 3 mm 20 mm/l 4 mm 50 mm/l 8 mm 125 mm/l 15 mm 300 mm/l 40 mm 8 mm/l 10 mm 15 mm/l 6 mm 15 mm/l 3 x 6 mm
Light source Light type, wavelength Average service life Supply voltage VS Ripple 3) Current consumption 4)
Output current lA max. Max. switching frequency 7) At max. sensitivity Response time At max. sensitivity Analogue output QA Connection type
40050 100 200 500 1000TypeWavelengthPart no.
OBF-OG 570-4 OBF-RG 610-4
570 . . . 750 nm 610 . . . 750 nm
4 005 810
OBF-RG 630-4 OBF-RG 665-4 Spare lamp
630 . . . 750 nm 665 . . . 750 nm
SLA-LUT 1-4
4 012 735 4 014 153 4 014 154 1 002 262
SENSICK CATALOGUE
423