The text of a manual · part 16 of 19
Falcon DOS Portable Data Terminals Advanced User's Guide, pages 151 to 160
Pages 151 to 160 of 184
The original document is the PDF (7.1 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the PSC 335, PSC 340, PSC 345, Datalogic 335, Datalogic 340 and Datalogic 345
Page 151
C Programming Parameters
APPENDIX CONTENTS Code 39 .................................................................. B-3 Interleaved 2 of 5 .................................................... B-3 Matrix 2 of 5........................................................... B-3 Standard 2 of 5 ....................................................... B-3 Code 11 .................................................................. B-4 Codabar/Ames ........................................................ B-4 MSI......................................................................... B-4 Code 93 .................................................................. B-4 Universal Product Code-A (UPC-A) ....................... B-4 Universal Product Code-E (UPC-E) ....................... B-5 European Article Numbering (EAN) / Japan Article Numbering (JAN) ............................................ B-5 UPC, EAN, JAN Extensions.................................... B-5 Code 128 ................................................................ B-5 Labelcode 4/5 ......................................................... B-5 Other Controls ....................................................... B-5
Page 152
Programming Parameters
Overview This appendix contains information about programmable settings for the Falcon. Use the Falcon Configuration Utility in Chapter 2 or the bar codes provided in Appendix D to program the Falcon. Table l on page C-2 provides the following information: z
Code Parameter is the “human” name for the programming option.
z
I.D. # is the “decoder” name for the programming option. For example, z z z to set a Code 39 minimum label length, use I.D. #01. Programming I.D. numbers given in this appendix can be used with all programming methods. Type tells what kind of setting to use for each code parameter. z On/Off is a toggle. 1 turns the parameter on, and 0 turns it off. z Value requires a two-character entry (e.g., 02 for two beeps after each good read). Acceptable Input gives the settings or range of settings for each code parameter. Defaults tells how the parameter is set when the predefined default D0, D1, or D2 is selected. z Predefined Default 0 (D0) turns every on/off parameter off and sets all minimum and maximum lengths to the lowest values. z Predefined Default 1 (D1) turns every on/off parameter on, sets all minimum lengths to the lowest values, and sets all maximum lengths to the highest values. This default set is normally used only for troubleshooting. It gives the best chance of reading an unknown bar code symbology and also identifies the symbology of each bar code scanned. z Predefined Default 2 (D2) is the default set that was installed in the Falcon at the factory. This default set will work for most applications. To reset the Falcon to the original defaults, scan the D2 bar code on page D-2.
Table C-1: Parameters, Settings and Defaults
C-2
Falcon® DOS Portable Terminals
Page 153
Code 39
Code Parameter
I.D. #
Type
Acceptable Input
Code 39
Enable
Defaults D0
D1
D2
Enter 1 for On and 0 for Off.
00
On/Off
On or Off
Off
On
On
Minimum length
01
Value
00–50
00
00
00
Maximum length
02
Value
01–50
01
50
20
Enable checksum
03
On/Off
On or Off
Off
Off
Off
Send checksum
04
On/Off
On or Off
Off
Off
Off
Full ASCII mode
05
On/Off
On or Off
Off
On
On
08
On/Off
On or Off
Interleaved 2 of 5
Enable
Enter 1 for On and 0 for Off.
Off
On
On
Minimum length
09
Value
02–50
02
02
06
Maximum length
0A
Value
02–50
02
50
10
Enable checksum
0B
On/Off
On or Off
Off
Off
Off
Send checksum
0C
On/Off
On or Off
Off
Off
Off
Use lengths 6 and 14 only (case code)
0D
On/Off
On or Off
Off
Off
Off
Matrix 2 of 5
Enter 1 for On and 0 for Off.
Enable
10
On/Off
On or Off
Off
On
Off
Minimum length
11
Value
01–50
01
01
06
Maximum length
12
Value
01–50
01
50
10
Enable checksum
13
On/Off
On or Off
Off
Off
Off
Send checksum
14
On/Off
On or Off
Off
Off
Off
15
On/Off
On or Off
Standard 2 of 5
Enable
Enter 1 for On and 0 for Off.
Off
On
Off
Minimum length
16
Value
01–50
01
01
06
Maximum length
17
Value
01–50
01
50
10
Enable checksum
18
On/Off
On or Off
Off
Off
Off
Send checksum
19
On/Off
On or Off
Off
Off
Off
Use 2-bar start/stop
1A
On/Off
On or Off
Off
Off
Off
Advanced User’s Guide
C-3
Page 154
Programming Parameters
Code Parameter
I.D. #
Type
Acceptable Input
Code 11
Defaults D0
D1
D2
Enter 1 for On and 0 for Off.
Enable
1B
On/Off
On or Off
Off
On
Off
Minimum length
1C
Value
01–50
01
01
04
Maximum length
1D
Value
01–50
01
50
10
Require 2 check digits
1E
On/Off
On or Off
Off
Off
Off
Send check digit(s)
1F
On/Off
On or Off
Off
On
Off
51
On/Off
On or Off
Off
On
Off
Codabar Enable
20
On/Off
On or Off
Off
On
On
Minimum length
21
Value
01–50
01
01
04
Maximum length
22
Value
01–50
01
50
20
Send start/stop
23
On/Off
On or Off
Off
On
Off
Codabar/Ames
Ames Enable
Enter 1 for On and 0 for Off.
Codabar-to-CLSI conversion
24
On/Off
On or Off
Off
Off
Off
Wide intercharacter gaps allowed
25
On/Off
On or Off
Off
Off
On
MSI
Enter 1 for On and 0 for Off.
Enable
26
On/Off
On or Off
Off
On
Off
Minimum length
27
Value
01–14
01
01
04
Maximum length
28
Value
01–14
01
14
10
Require 2 check digits
29
On/Off
On or Off
Off
Off
Off
2nd check digit Mod 11
2A
On/Off
On or Off
Off
Off
Off
Send check digit(s)
2B
On/Off
On or Off
Off
On
Off
Code 93
Enter 1 for On and 0 for Off.
Enable
2C
On/Off
On or Off
Off
On
Off
Minimum length
2D
Value
01–50
01
01
02
Maximum length
2E
Value
01–50
01
50
20
30
On/Off
On or Off
Off
On
On
Universal Product Code-A (UPC-A)
Enable
Enter 1 for On and 0 for Off.
Send system digit
31
On/Off
On or Off
Off
On
On
Send check digit
32
On/Off
On or Off
Off
On
Off
Convert UPC-A to EAN-13
33
On/Off
On or Off
Off
On
Off
C-4
Falcon® DOS Portable Terminals
Page 155
Universal Product Code-E (UPC-E)
Code Parameter
I.D. #
Type
Acceptable Input
Universal Product Code-E (UPC-E)
Defaults D0
D1
D2
Enter 1 for On and 0 for Off.
Use system digit 0
34
On/Off
On or Off
Off
On
On
Use system digit 1
35
On/Off
On or Off
Off
On
On
Convert UPC-E to UPC-A
36
On/Off
On or Off
Off
On
Off
Send system digit
37
On/Off
On or Off
Off
On
Off
Send check digit
38
On/Off
On or Off
Off
On
Off
European Article Numbering (EAN) / Japan Article Numbering (JAN)
Enter 1 for On and 0 for Off.
Enable EAN-8/JAN-8
39
On/Off
On or Off
Off
On
On
Enable EAN-13/JAN-13
3A
On/Off
On or Off
Off
On
On
Convert EAN-13 to ISBN
3B
On/Off
On or Off
Off
Off
Off
Send EAN/JAN checksum
3F
On/Off
On or Off
Off
Off
Off
UPC, EAN, JAN Extensions
Enter 1 for On and 0 for Off.
Allow 2-digit extensions
3C
On/Off
On or Off
Off
On
On
Allow 5-digit extensions
3D
On/Off
On or Off
Off
On
On
Require extensions
3E
On/Off
On or Off
Off
Off
Off
40
On/Off
On or Off
Code 128
Enable
Enter 1 for On and 0 for Off.
Off
On
On
Minimum length
41
Value
01–50
01
01
02
Maximum length
42
Value
01–50
01
50
20
Enable UCC/EAN 128
43
On/Off
On or Off
Off
Off
Off
Enable
52
On/Off
On or Off
Off
On
Off
Convert
53
On/Off
On or Off
Off
Off
Off
Laser programming enable
B0
On/Off
On or Off
On
On
On
Autoterminator
B1
Value
Any single ASCII charact er (00 = Off)
(CR)
(CR)
(CR)
Auto-off timer
B2
Value
01–99 (in 4sec. increments) (00 = Off)
75
75
75
Send assigned symbology identifiers*
B5
On/Off
On or Off
Off
On
Off
Labelcode 4/5
Enter 1 for On and 0 for Off.
Other Controls
Advanced User’s Guide
Enter 1 for On and 0 for Off.
C-5
Page 156
Programming Parameters
Code Parameter *A = UPC-A B = I 2 of 5 C = Code 39
D = M 2 of 5 E = UPC-E F = S 2 of 5
Good-read beep tone
I.D. #
Type
G = EAN-8 H = MSI I = Codabar
Acceptable Input J = Code 11 K = Code 128 L = Code 93
B8
Value
Number of good-read beeps
B9
Good-read beep duration
BA
Beeper volume
Defaults D0
D1
D2
M = EAN-13 N = Labelcode 4/5 O = Ames
00 = 2400 Hz 01 = 2600 Hz 02 = 2800 Hz 03 = 3000 Hz 04 = 3200 Hz 05 = 3400 Hz 06 = 3600 Hz 07 = 3800 Hz
00
00
00
Value
01–04
01
01
01
Value
00 = 0.07 sec. 01 = 0.13 sec. 02 = 0.18 sec. 03 = 0.36 sec.
00
00
00
BC
Value
00 = Off 01–09 (01 = Lowest; 07 = Highest; 08 = Higher; 09 = Lower)
07
07
07
Error beep tone
BD
Value
00 = 2400 Hz 01 = 2580 Hz 02 = 2770 Hz 03 = 3000 Hz 04 = 3270 Hz 05 = 3600 Hz 06 = 4000 Hz 07 = 4520 Hz
01
01
01
Long-range trigger mode (long-range scanner models only)
D4
On/Off
On = Release Scan Off = Spot Timeout
Off
Off
Off
C-6
Falcon® DOS Portable Terminals
Page 157
Other Controls
Code Parameter
I.D. #
Type
Acceptable Input
Defaults D0
D1
D2
Spot beam timeout (long-range scanner models only)
D5
Value
00 = 0.25 sec. 01 = 0.5 sec. 02 = 1 sec. 03 = 1.5 sec. 04 = 2 sec.
01
01
01
Release scan timeout (long-range scanner models only)
D6
Value
01–30 (seconds)
02
02
02
Spotting Beam Enable (long-range scanner models only)
D7
On/Off
On or Off
On
On
On
Keypress sound
DD
Value
00 = Off 01 = Click 02 = Beep
01
01
01
Enable CTL-ALT-DEL reboot
E0
On/Off
On or Off
Off
Off
On
Enable trigger programmability (Note that for the Falcon 33x, this will effect Enter key programmability.)
E1
On/Off
On or Off
On
On
On
Backlight auto-off timeout
E2
Value
01–99 (seconds) (00 = Off)
15
15
15
Double Action Key Mode (33x models only)
E3
Value
01 = Double Strike 02 = Press & Wait
01
01
01
Double Action Key Timeout (33x models only)
E4
Value
00 = 165 ms. 01 = 220 ms. 02 = 275 ms. 03 = 330 ms. 04 = 385 ms. 05 = 440 ms. 06 = 495 ms. 07 = 550 ms. 08 = 605 ms. 09 = 660 ms.
03
03
03
Advanced User’s Guide
C-7
Page 158
Programming Parameters
NOTES
C-8
Falcon® DOS Portable Terminals
Page 159
D Bar Codes This appendix provides bar codes for common setup parameters for programming the Falcon. Factory default settings are underlined. To make settings that aren’t provided in this chapter, design custom bar codes or use the Falcon Configuration Utility to program the Falcon (refer to Chapter 2).
APPENDIX CONTENTS Predefined Defaults................................................ D-2 Code 39.................................................................. D-2 Interleaved 2 of 5 .................................................. D-3 Matrix 2 of 5 ......................................................... D-3 Standard 2 of 5 ...................................................... D-4 Code 11 ................................................................. D-5 Codabar/Ames ...................................................... D-6 MSI ....................................................................... D-7 Code 93 ................................................................. D-8 Code 128 ............................................................... D-8 Labelcode 4/5 ........................................................ D-9 UPC-A (Universal Product Code-A) ........................... D-9 UPC-E (Universal Product Code-E) .......................... D-10 EAN (European Article Numbering)/ JAN (Japan Article Numbering).......................... D-10 UPC/EAN/JAN Extentions ................................. D-11 Other Controls .................................................... D-11
Page 160
Bar Codes
Predefined Defaults
MAXIMUM LENGTH 1
D1 *
$ + $
- D 1 E E *
D2 *
$ + $
- D 2 E E *
*
$ + $
-
0 2 0 1 E E *
*
$ + $
-
0 2 1 0 E E *
*
$ + $
-
0 2 2 0 E E *
*
$ + $
-
0 2 3 0 E E *
*
$ + $
-
0 2 4 0 E E *
*
$ + $
-
0 2 5 0 E E *
10 20 30
Code 39
40
ENABLE 50 On *
$ + $ 0 0 1 E E *
*
$ + $ 0 0 0 E E *
Off
ENABLE CHECKSUM On
MINIMUM LENGTH
*
$ + $
-
0 3 1 E E *
*
$ + $
-
0 3 0 E E *
Off 0 *
$ + $
-
0 1 0 0 E E *
*
$ + $
-
0 1 1 0 E E *
10 20 *
$ + $
-
0 1 2 0 E E *
*
$ + $
-
0 1 3 0 E E *
30
On *
$ + $
-
0 4 1 E E *
*
$ + $
-
0 4 0 E E *
Off FULL ASCII MODE On
40 *
$ + $
-
0 1 4 0 E E *
*
$ + $
-
0 5 1 E E *
*
$ + $
-
0 5 0 E E *
Off
50 *
D-2
SEND CHECKSUM
$ + $
-
0 1 5 0 E E *
Falcon® DOS Portable Terminals