Page 181

I/O Parameters

EZ Button The EZ button has four positions, selectable by the length of time the button is held down, and indicated by one, two, three, and four beeps in succession. Each position can be programmed for any of eight options. Definition: Serial Cmd: Default: Options:

Serves as a master switch to enable/disable the EZ button status. <K770,global status,default on power-on,load configuration database, save for power-on> Enabled 0 = Disabled 1 = Enabled 2 = Trigger 3 = Unlatch Outputs 4 = Parameter Switch

Disabled When set to Disabled, the EZ button does not function.

Enabled When selected, the EZ button is enabled and the function of each button position is selected by the EZ Button Mode command.

Trigger When selected, the EZ button acts as a trigger for the Imager to start and end read cycles. All other button operations are inactive. In External Level: In External Edge:

The read cycle endures for as long as the EZ button is pressed, unless a timeout occurs and Timeout is enabled for End of Read Cycle. As with Level, Edge allows a read cycle to be initiated by pressing the EZ button, but unlike Level, the read cycle ends with a good read output, a timeout, or a new trigger.

Unlatch Outputs In this configuration, the EZ button will unlatch any logic outputs that have been latched.

Parameter Switch The parameter switch toggles between custom defaults and power-on settings. The condition is the same as that achieved by sending the <Arc> and <Arp> commands consecutively.

MS-4 Imager User Manual

7-25

Page 182

EZ Button

Default on Power-On Definition: Serial Cmd: Default: Options:

When enabled, if the EZ button is held down on power-on the Imager will default to customer defaults and save for power-on. This is the same as sending a <Zrc> command. <K770,global status,default on power-on,load configuration database,save for power-on> Enabled 0 = Disabled 1 = Enabled

Load Configuration Database Definition:

Serial Cmd: Default: Options:

Allows the user to load the Configuration Database with calibration results. When the user performs a calibration using the EZ button, all the database entries are moved down one index and the results of the calibration are saved to index 0. The results will be saved as current settings as well. <K770,global status,default on power-on,load configuration database,save for power-on> Enabled 0 = Disabled 1 = Enabled

Save for Power-On Definition: Serial Cmd: Default: Options:

7-26

If enabled, after calibration is complete, all parameters will be saved for power-on. <K770,global status,default on power-on,load configuration database, save for power-on> Disabled 0 = Disabled 1 = Enabled

MS-4 Imager User Manual

Page 183

I/O Parameters

EZ Button Modes Usage: Definition: Serial Cmd: Options:

Useful for performing multiple, repetitive tasks at the work site. Allows the user to program each of the EZ button’s 4 positions from a selection of 8 modes. <K771,position 1 mode,position 2 mode,position 3 mode,position 4 mode> 0 = Disabled 1 = Enabled

Single Beep 0 = Disabled 1 = Read Rate 2 = Calibrate 3 = Save for Power-On 4 = Unused 5 = Load New Master 6 = Unused 7 = Target System 8 = Bar Code Config.

Two Beeps 0 = Disabled 1 = Read Rate 2 = Calibrate 3 = Save for Power-On 4 = Unused 5 = Load New Master 6 = Unused 7 = Target System 8 = Bar Code Config.

Three Beeps 0 = Disabled 1 = Read Rate 2 = Calibrate 3 = Save for Power-On 4 = Unused 5 = Load New Master 6 = Unused 7 = Target System 8 = Bar Code Config.

Four Beeps 0 = Disabled 1 = Read Rate 2 = Calibrate 3 = Save for Power-On 4 = Unused 5 = Load New Master 6 = Unused 7 = Target System 8 = Bar Code Config.

Disabled When set to disabled, the associated button position will have no function associated with it, and the position will be skipped over.

Read Rate Read Rate will be initiated when the associated button position is selected. Read Rate will perform decodes/second and is the same as sending a <C> from the Terminal. To exit Read Rate Mode quickly press and release the EZ button.

Calibrate Calibration will be initiated when the associated button position is selected. To abort calibration, quickly press and release the EZ button.

Save for Power-On All Imager settings will be saved to non-volatile memory to be recalled on power-on whenever the associated button position is selected. This is the same as sending a <Z> from the terminal.

Load New Master Functions in the same way as new master pin input whenever the associated button position is selected. The new master pin’s Consecutive Decode requirement holds true for this function. MS-4 Imager User Manual

7-27

Page 184

EZ Button Modes

Target System Turns on the target pattern whenever the associated button position is selected. To disable, quickly press and release the EZ button. Note: This mode is the only one that does not require that the button be released before taking effect. Thus, as soon as the Imager beeps the appropriate number of times for the position, it will take effect. If it is necessary to have the target system on before another operation such as calibration or read rate is performed, ensure that the target system mode is assigned a lower position so that it will be activated first.

Bar Code Configuration Enables bar code configuration mode whenever the associated button position is selected. When enabled, the Imager can accept configuration commands from bar codes without first reading a special, coded Data Matrix ISO/IEC 16022 symbol. To disable, quickly press and release the EZ button.

EZ Button Operation To execute EZ button commands,

Single Beep (Position #1) Hold down button until a single beep is heard (and the 20% LED illuminates).

Two Beeps (Position #2) Hold down button until two quick beeps are heard (and the 20% and 40% LEDs illuminate.)

Three Beeps (Position #3) Hold down button until three quick beeps are heard (and the 20%, 40%, and 60% LEDs illuminate).

Four Beeps (Position #4) Hold down button until four quick beeps are heard (and the 20%, 40%, 60%, and 80% LEDs illuminate).

7-28

MS-4 Imager User Manual

Page 185

I/O Parameters

Configurable Output 1

Usage: Definition: Serial Cmd: Default: Options:

This option provides discrete signalling to host software to control external devices such as PLCs and relays. It is useful for routing, sorting, and to prevent mis-packaging and mis-routing. Sets the discrete output functions for specific user-selected conditions. Allows the user to set the conditions under which an output (or outputs) will be activated. <K810,output on,output state,pulse width,output mode> Mismatch or No Read 0 = Mismatch or No Read 1 = Match (or good read) 3 = No Read 4 = Trend Analysis 6 = Diagnostic Warning 7 = In Read Cycle

2 = Mismatch 5 = Symbol Quality

Note: If Output On is set to Match or Mismatch, a transition (switching) will not occur unless Matchcode Type is enabled and a master symbol is loaded into memory.

Mismatch or No Read Activates discrete output when the data does not match that of the master symbol or the symbol has not been decoded before the end of the read cycle.

Match or Good Read Activates a discrete output when the symbol data matches the master symbol. Note: If you want to output for a good read and Matchcode is not enabled, you can enable any output for Match.

Mismatch Activates a discrete output whenever the symbol data does not match that of the master symbol.

No Read Activates a discrete output whenever the symbol data is not decoded before the end of the read cycle. MS-4 Imager User Manual

7-29

Page 186

Configurable Output 1

Trend Analysis Usage: Definition:

Typically used when successful decodes are occurring but a discrete output is needed to flag a trend in quality issues. Activates discrete output when a trend analysis condition is met, depending on the trend analysis option enabled.

Symbol Quality Usage: Definition:

Typically used when a discrete indication is needed to flag a symbol quality condition. Activates discrete output when a symbol quality condition is met, depending on the symbol quality option enabled.

Diagnostic Warning Usage: Definition:

Typically used when a discrete indication of a diagnostic condition is needed. Activates discrete output when a diagnostic warning condition is met, depending on the diagnostic option enabled.

In Read Cycle Definition:

7-30

Activates a discrete output when the Imager is in a read cycle.

MS-4 Imager User Manual

Page 187

I/O Parameters

Output State Definition: Serial Cmd: Default:

Sets the active electrical state of the discrete output. <K810,output on,output state,pulse width,output mode> Negative

Options:

0 = Negative

1 = Positive

Pulse Width Definition: Serial Cmd: Default: Options:

Sets the time in 10mS increments that the discrete output remains active. <K810,output on,output state,pulse width,output mode> 5 (Corresponds to 50 mS) 0 to 255 (0 to 2.55 seconds). Divide the number entered on the command line by 100 for time in seconds.

Output Mode Definition: Serial Cmd: Default:

Sets the condition in which the discrete output is de-activated. <K810,output on,output state,pulse width,output mode> Pulse

Options:

0 = Pulse 2 = Latch Mode 2 3 = Latch Mode 3

Pulse This is the default mode of operation in which the programmable output is activated when the Output On condition has been met and held active for the duration of the selected pulse width.

Latch Mode 2 (Unlatch Opposite Condition)

The programmable output is activated when the Output On condition has been met and held active until the opposite condition selected under Output On has been met. For example, if No Read is enabled under Output On, the programmable output will go active on a No Read and remain active until the opposite condition, a good read, occurs.

Latch Mode 3 (Unlatch Re-enter Read Cycle) The programmable output is active when the Output On condition has been met and is held active until a new read cycle begins. Note: All of the Output On modes are inhibited when any Output on Warning is active for Output 1.

MS-4 Imager User Manual

7-31

Page 188

Trend Analysis Output 1

Trend Analysis Output 1 Note: Output On under Output 1 Parameters must be set to Trend Analysis for this output to function.

Under Output 1 Parameters, expand the Trend Analysis tree:

Usage: Definition:

Example:

7-32

Useful in cases where the user doesn’t want to shut down for one condition but wants to monitor quality and read conditions. Applies Trend Analysis settings to Output 1. With Trend Analysis, the user can track the occurrences and frequency of mismatches, No Reads, and the number of reads per trigger, and output the results to any of three outputs. Trend Analysis Mode = No Read Trigger Evaluation Period= 25 triggers (read cycles) Number to Output On = 4 In this example, the Imager will activate an output when 4 No Reads occur within a period of 25 triggers (read cycles).

MS-4 Imager User Manual

Page 189

I/O Parameters

Trend Analysis Mode Definition:

Sets the trend condition (Mismatch, No Read, or Decodes/Trigger) that will activate the output.

Serial Cmd:

<K780,trend analysis mode,trigger evaluation period,number to output on, decodes/trigger threshold> No Read

Default: Options:

0 = Mismatch 3 = Bad Symbol

1 = No Read 4 = No Symbol

2 = Decodes/Trigger

Mismatch Output will be activated when the number of mismatches equals the value entered for Number to Output On within the trigger window selected in Number of Triggers.

No Read Output will be activated when the number of No Reads equals the value entered for Number to Output On within the trigger window selected in Number of Triggers.

Decodes/Trigger Output will be activated when the number of decodes equals the value entered for Number to Output On within the trigger window selected in Number of Triggers.

Bad Symbol Output will be activated when the number of Bad Symbol occurrences equals the value entered for Number to Output On within the trigger window selected in Number of Triggers.

No Symbol Output will be activated when the number of No Symbol occurrences equals the value entered for Number to Output On within the trigger window selected in Number of Triggers.

MS-4 Imager User Manual

7-33

Page 190

Trend Analysis Output 1

Trigger Evaluation Period

Default:

The number of triggers to examine for the trend analysis condition. <K780,trend analysis mode,trigger evaluation period,number to output on, decodes/trigger threshold> 0

Options:

0 to 255

Definition: Serial Cmd:

Number to Output On

Default:

Example: If Number to Output On is set to 3 and Trend Analysis Mode is set to No Read, then the output will not be activated until 3 No Reads have occurred. Sets the number of Trend Analysis Mode events (mismatches, No Reads, or reads/trigger as configured by Trend Analysis Mode) to occur within the trigger evaluation period before activating the associated output. <K780,trend analysis mode,trigger evaluation period,number to output on, decodes/trigger threshold> 0

Options:

0 to 255

Usage: Definition: Serial Cmd:

Decodes/Trigger Threshold

Default:

When set to this mode, and when the appropriate output is set to output on trend analysis, the Imager will function in a Decodes per Trigger mode during the read cycle and the trend analysis operation. Output will be activated based on whether or not the symbol decode count at the end of the read cycle crosses the decodes per trigger threshold. Note: Although this setup causes the Imager to function in a Decodes per Trigger mode, the decode count will only be appended to the symbol data if the status of the Decodes per Trigger command is enabled. <K780,trend analysis mode,trigger evaluation period,number to output on, decodes/trigger threshold> 0

Options:

0 to 65535

Definition:

Serial Cmd:

7-34

MS-4 Imager User Manual