Page 81
Calibration
Calibration Options This command specifies the operation of the imager’s calibration feature. The default configuration is set up to perform calibration on gain, focus, and symbol type. The default shutter speed setting is disabled, because not all applications require shutter speed, and when it is required, the application-specific settings tend to vary widely. Note: Background Color (page 6-51) is part of the calibration process regardless of the calibration setup. After successful completion of the calibration routine, the background color will be set to the appropriate value, depending on the symbol in the field of view. The Mirrored Image parameter (page 10-28) is not part of the calibration process, and it must be configured appropriately prior to calibration.
Gain Definition:
Serial Cmd: Default: Options:
When enabled, gain will be calibrated to provide the best available image quality and performance. When disabled, gain is fixed and is not part of the calibration process. <K529,gain,shutter speed,focus position,symbol type,WOI framing,WOI margin,line scan height,processing> Enabled 0 = Disabled 1 = Enabled
Shutter Speed Unless an application is a static setup, the shutter speed setting should be configured by the user based on the application. For dynamic applications, the user should configure the shutter speed setting so the calibration process can optimize the gain for that setting. Serial Cmd: <K529,gain,shutter speed,focus position,symbol type,WOI framing,WOI margin,line scan height,processing> Default: Disabled Options: 0 = Disabled 1 = Enabled 2 = Fast Shutter Note: The following table shows general guidelines for shutter speed configurations based on various line speeds. The configurations shown depend on the imager’s optical configuration, and on symbol element size. Definition:
Shutter Speed 0 - 250 250 - 750 750 - 1500 1500 - 2000 2000 - 2500
Line Speed static 5” / sec 10” / sec 15” / sec 20” / sec
Quadrus MINI Velocity User’s Manual
4-3
Page 82
Calibration Options
Disabled (Shutter Speed) When this feature is disabled, the shutter speed is fixed and is not part of the calibration process.
Enabled (Shutter Speed) When this feature is enabled, shutter speed will be calibrated to provide the best possible image quality and performance.
Fast Shutter The calibration process will concentrate on achieving the fastest possible shutter setting that will still provide good performance. The image quality or contrast may not be as good as what would be achieved with the Enabled setting. The calibration process is not designed to choose the fastest shutter speed that can decode a symbol, but rather to optimize for the fastest shutter speed that still provides good image quality.
4-4
Quadrus MINI Velocity User’s Manual
Page 83
Calibration
Focus Position Definition:
Serial Cmd: Default: Options:
Since focus can be configured by entering the target distance from the imager’s parting line, the user will normally configure the focus setting without needing to calibrate. However, if it is necessary to calibrate the focus distance, there are two methods for doing so: the standard “Search Method”, or a “Quick Focus” method, both of which are defined below. <K529,gain,shutter speed,focus position,symbol type,WOI framing, WOI margin,line scan height,processing> Quick Focus 0 = Disabled 1 = Enabled (Search Method) 2 = Quick Focus
Disabled When this feature is disabled, the focus position is fixed and is not part of the calibration process.
Enabled (Search Method) When this feature is enabled, the focus position will be calibrated to provide the best possible image quality and performance. This method is essentially a search method that attempts to identify the desired focus using a search algorithm that cycles through focus settings to try and locate the desired focus as quickly as possible. This is accomplished when the given symbol in the field of view is decoded. Once a focus distance that will decode the given symbol has been located, the search algorithm will fine-tune the search to locate the “inside” and “outside” focal distances. The final focus distance will be between the “inside” and “outside” values. Note: This method can be time-consuming if the focus is not found on the first pass.
Quick Focus Quick Focus is designed to quickly locate the focus setting for an object in the field of view. This is accomplished with minimal processing of the image by analyzing a histogram of the image frame. Before performing the focus calibration, a value for shutter speed and gain is determined and then the calibration process will step through the focus range of the system, capturing the required number of image frames for each focal position. A histogram is performed on each image frame and the histogram results for each focus position are averaged. When this process is completed, the data is analyzed and the optimal focus position is determined. Note: Quick Focus may not work well for all applications. It is not recommended for applications that use very small symbols.
Quadrus MINI Velocity User’s Manual
4-5
Page 84
Calibration Options
Symbol Type Serial Cmd: Default: Options:
<K529,gain,shutter speed,focus position,symbol type,WOI framing,WOI margin,line scan height,processing> Enabled 0 = Disabled 1 = Enabled
Disabled When this feature is disabled, only the currently-enabled symbologies will be considered during the calibration process.
Enabled When this feature is enabled, autodiscrimination will be in effect during the calibration process. All symbologies that are readable by the imager except PDF417 and Pharmacode will be considered during calibration. Any new symbologies successfully decoded during calibration will remain enabled at the end of the process. All enabled symbologies will remain enabled. For example, assume that only Code 39 is enabled at the beginning of calibration. If a Code 128 symbol is read during calibration, then Code 128, as well as Code 39, will be enabled after the calibration process is complete.
4-6
Quadrus MINI Velocity User’s Manual
Page 85
Calibration
Window of Interest (WOI) Framing Definition:
Serial Cmd: Default: Options:
Configured WOI
If a WOI framing mode is enabled, it will set the camera WOI to a full-size image when calibration begins. Once a symbol is decoded, the camera WOI will be zoomed-in both vertically and horizontally (regardless of which WOI mode is enabled) to include the symbol plus an additional margin area. This is done to accelerate the calibration process. When a successful calibration process is completed, the camera WOI will be adjusted according to the mode enabled. Otherwise, the original WOI configuration will be retained. <K529,gain,shutter speed,focus position,symbol type,WOI framing,WOI margin,line scan height,processing> Disabled 0 = Disabled 1 = Row and Column 2 = Row 3 = Column 4 = Straight Line 5 = Straight Line Framed
Starting WOI
After Decode
Calibration Passed
Calibration Failed
If WOI framing is not enabled, the current WOI configuration will be used until a symbol has been decoded. After a symbol is decoded, the WOI is framed exactly as it would be if a WOI framing mode was enabled. When the calibration process is complete, the original WOI configuration is restored. Refer to the diagram below:
Configured WOI
After Decode
Quadrus MINI Velocity User’s Manual
Calibration Passed
Calibration Failed
4-7
Page 86
Calibration Options
Disabled When this feature is disabled, the Window of Interest is not modified after the calibration process is complete.
Row and Column If the calibration process is successful, the Window of Interest will be modified to frame the symbol as well as an additional margin area around the symbol, determined by the WOI Margin parameter. Refer to the diagram below:
Column If the calibration process is successful, the Window of Interest columns will be modified to vertically frame the symbol, plus an additional margin area around the symbol, determined by the WOI Margin parameter. Refer to the diagram below:
Row If the calibration process is successful, the Window of Interest rows will be modified to horizontally frame the symbol, plus an additional margin area around the symbol, determined by the WOI Margin parameter. Refer to the diagram below:
4-8
Quadrus MINI Velocity User’s Manual
Page 87
Calibration
Straight Line This feature is intended for use with linear symbologies. If the calibration process is successful, the orientation of the symbol is determined and the Window of Interest is modified according to the symbol orientation. The scan line orientation is determined to be vertical if the symbol tilt is between 225° and 315°, or between 45° and 135°. Otherwise, the scan line will be horizontal. If the symbol is vertical, the image column size will be set by the scan height parameter, and will be configured for full row resolution. If the symbol is horizontal, the image row size will be set by the scan height parameter, and will be configured for full column resolution. The scan line will be centered in the middle of the symbol. If the symbol is tilted such that the scan line will not pass completely through the symbol, the scan width will be adjusted to include the entire symbol. Refer to the diagram below:
Straight Line Framed This parameter is the same as “Straight Line”, except that the Window of Interest will also frame the scan line on the symbol length as well. The scan line includes the symbol plus an additional margin area determined by the WOI Margin parameter.
Quadrus MINI Velocity User’s Manual
4-9
Page 88
Calibration Options
Window of Interest (WOI) Margin Definition: Serial Cmd: Default: Options:
Sets the margin size that is applied to the calibrated symbol. This parameter is expressed in number of pixels. If the margin causes the image to exceed the maximum image size, it will be reduced accordingly. <K529,gain,shutter speed,focus position,symbol type,WOI framing,WOI margin,line scan height,processing> 75 (pixels) 20 to 752
Line Scan Height Definition: Serial Cmd: Default: Options:
This parameter is only used with the Straight Line modes. It sets the scan height of the straight-line image, and it is expressed in number of pixels. <K529,gain,shutter speed,focus position,symbol type,WOI framing,WOI margin,line scan height,processing> 64 (pixels) 3 to 480
Processing Definition: Serial Cmd: Default: Options:
This setting defines the amount of time and effort the imager will spend attempting to decode a symbol for each parameter configuration. <K529,gain,shutter speed,focus position,symbol type,WOI framing,WOI margin,line scan height,processing> Medium 0 = Low 1 = Medium 2 = High 3 = Definable
Low The imager will spend a low amount of effort attempting to decode the given symbol for each parameter configuration.
Medium The imager will spend a medium amount of effort attempting to decode the given symbol for each parameter configuration.
High The imager will spend a high amount of effort attempting to decode the given symbol for each parameter configuration.
Definable The processing time for each image frame is defined by the Image Processing Timeout parameter <K245>. 4-10
Quadrus MINI Velocity User’s Manual
Page 89
Calibration
Calibration by ESP ESP’s Calibration view incorporates all the functionality of the <K529> (Calibration Options) command in a single easy-to-use, intuitive interface. This calibration process allows the user much finer control of individual parameters than the calibration routine in EZ Mode or on the Video tab.
Before/During Calibration Before and After tabs give the user an authentic representation of the difference in what the imager “sees” before and after the calibration routine.
Focus, Shutter, Gain, and Image Quality values are all tracked in real-time during the calibration process. Gain, Shutter Speed, and Focal Distance (Capture Settings) are all configurable in the Calibration interface.
Determines the amount of time and effort the imager will spend attempting to decode a symbol for each parameter configuration. (Low, Medium, High, Definable).
Dropdown menu of 1D and 2D symbologies, and a check box for enabling or disabling Composite symbologies.
Initiates the calibration routine. Cancels the calibration process, if necessary.
The Window of Interest section of the Calibration interface allows the user to make precise adjustments to WOI Framing, WOI Margin (in pixels), and the scan height of the straight-line image (in pixels).
Quadrus MINI Velocity User’s Manual
4-11
Page 90
Calibration by ESP
After Calibration Notice the improved image resolution after the calibration process is complete. The Focus, Shutter, Gain, and Quality indicator bars remain stationary at the end of calibration to provide a visual reference for the final values of these four parameters.
The calibration progress bar indicates that the process has been completed.
Saves the optimal configuration parameters determined by calibration results.
4-12
Quadrus MINI Velocity User’s Manual