Page 251

Symbol Quality

Microscan Symbol Quality Output Percent Cell Damage (Data Matrix Only) Note: Percent Cell Damage is not supported when Q-Mode is enabled. Definition: Serial Cmd: Default: Options:

When this feature is enabled, the cell damage percentage is appended to data output. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled

Total Read Time Definition:

Serial Cmd: Default: Options:

The time that transpires between the image capture and the output of the decoded data, including locate time. When enabled, the total read time is appended to the symbol data. <K710,percent cell damage,total read timea,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled a. Total Read Time applies globally to all three Symbol Quality standards.

Capture Time Definition:

Serial Cmd: Default: Options:

Capture time (in milliseconds) is a fixed “overhead” that includes the time of capture and transfer of the image. When enabled, the capture time is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled

Locate Time Definition:

Serial Cmd: Default: Options:

The time in milliseconds from the start of image processing until the symbol has been located and is ready to be decoded. When enabled, the locate time is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled

Quadrus MINI Velocity User’s Manual

8-13

Page 252

Microscan Symbol Quality Output

Decode Time Definition: Serial Cmd: Default: Options:

The time in milliseconds required to decode a symbol. When enabled, the decode time is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled

Pixels Per Element (Data Matrix Only) Note: Pixels Per Element is not supported when Q-Mode is enabled. Definition:

Serial Cmd: Default: Options:

The number of pixels for each element, either dark or light for both x and y directions. When enabled, the pixels per element value is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled

Error Correction Level (Data Matrix Only) Note: Error Correction Level is not supported when Q-Mode is enabled. Definition: Serial Cmd: Default: Options:

Outputs the Data Matrix ECC level. When enabled, the ECC level is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled

Matrix Size (Data Matrix Only) Note: Matrix Size is not supported when Q-Mode is enabled. Definition:

Serial Cmd: Default: Options: 8-14

Defines the symbol matrix size, in number of pixels in both the x and y axis. When enabled, the matrix size value is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled Quadrus MINI Velocity User’s Manual

Page 253

Symbol Quality

Quiet Zone (Data Matrix Only) Definition: Serial Cmd:

Default: Options:

When this feature is enabled, the size of the quiet zone is evaluated and a PASS or FAIL message is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled

Symbol Angle (Data Matrix Only) Definition: Serial Cmd:

Default: Options:

When this feature is enabled, the symbol orientation is appended to data output as a degree value representing the angle of the Data Matrix symbol’s L-shaped finder pattern relative to the imager. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC level,matrix size,quiet zone,symbol angle> Disabled 0 = Disabled 1 = Enabled

Quadrus MINI Velocity User’s Manual

8-15

Page 254

Microscan Symbol Quality Output by ESP

Microscan Symbol Quality Output by ESP ESP’s Symbol Quality interface allows you to evaluate Data Matrix symbols for compliance with a rigorous set of standards, such as Locate Time, Capture Time, and Decode Time. First, determine which parameters you want to evaluate, using the Symbol Quality tree control.

Once you have set these Data Matrix evaluation parameters, go to the Preferences Tab and set your Data Matrix Grading Report output preferences.

After your output preferences are set, click the Data Matrix Grading button.

There will be a wait of a few seconds, and then the evaluation results will appear in the Symbol Quality view, in a format similar to the one shown below.

If you want to view the results in a report format, click the Save Report button. The report will be output in the format you chose in the Preferences dialog.

8-16

Quadrus MINI Velocity User’s Manual

Page 255

9 Matchcode Contents Matchcode by ESP ....................................................................................................................... 9-2 Matchcode Serial Commands ...................................................................................................... 9-3 Overview of Matchcode ................................................................................................................ 9-4 Matchcode Type ........................................................................................................................... 9-5 Match Replace............................................................................................................................ 9-10 Mismatch Replace ...................................................................................................................... 9-11 New Master Pin .......................................................................................................................... 9-12

This section explains the Quadrus MINI Velocity’s matchcode output functions and the master symbol database setup.

Quadrus MINI Velocity User’s Manual

9-1

Page 256

Matchcode by ESP

Matchcode by ESP

Click this button to bring up the App Mode view.

Click this button to bring up the Matchcode tree control. To open nested options, single-click the +.

9-2

To change a setting, double-click the setting and use your cursor to scroll through the options.

Quadrus MINI Velocity User’s Manual

Page 257

Matchcode

Matchcode Serial Commands Matchcode Type Number of Master Symbols New Master Pin Enter Master Symbol Data Request Master Symbol Data Delete Master Symbol Data Match Replace Mismatch Replace

Quadrus MINI Velocity User’s Manual

<K223,matchcode type,sequential matching,match start position,match length,wild card character,sequence on No Read,sequence on mismatch> <K224,number of master symbols> <K225,status> <K231,master symbol number,master symbol data> <K231?,>[for all] or <K231?,master symbol number> <K231,master symbol number,> <K735,status,match replacement string> <K736,status,mismatch replacement string>

9-3

Page 258

Overview of Matchcode

Overview of Matchcode Definition:

Usage:

Matchcode allows the user to store master symbol data in the imager’s memory, compare that data against other symbol data, and define how symbol data and/or discrete signal output will be directed. A master symbol database can be set up for up to 10 master symbols. Note: Matchcode will function with multiple symbols; however, if Matchcode Type is set to Sequential or if Triggering Mode is set to Continuous Read 1 Output, the imager will behave as if Number of Symbols were set to 1, regardless of the user-defined configuration. Matchcode is used in applications to sort, route, or verify data based on matching the specific symbol in a variety of ways as defined in this section. For example, a manufacturer might sort a product based on dates that are embedded in the symbol.

Steps for Entering and Using Master Symbols 1. Set Triggering Mode to External or Serial. 2. Choose the method of symbol comparison that fits your application. 3. Define the output you want to achieve with your matchcode setup: a. Symbol data output b. Discrete output 4. Select the number of master symbols you want to create. 5. Decide which of 4 ways you want to enter your master symbol(s): a. Use ESP to type master symbol data directly. b. Send a serial command with symbol data in the form of <M231,master symbol#, data>. c. Send a <G> (Read Next Symbol as Master Symbol) command. d. Enable the New Master Pin command and activate the discrete input to store the next symbol read as the master symbol.

9-4

Quadrus MINI Velocity User’s Manual

Page 259

Matchcode

Matchcode Type Definition:

Allows the user to choose the way that master symbols will be compared with subsequently read symbols. Note: First set Triggering Mode to External or Serial.

ESP:

Serial Cmd:

Default: Options:

Disabled: Enabled: Wild Card: Sequential:

<K223,matchcode type,sequential matching,match start position, match length,wild card character,sequence on No Read,sequence on mismatch> Disabled 0 = Disabled 1 = Enabled 2 = Wild Card 3 = Sequential Has no effect on operations. Instructs the imager to compare symbols or portions of symbols with the master symbol. Allows the user to enter user-defined wild card characters in the master symbol. Instructs the imager to sequence after each match (numeric only) and compare symbols or portions of symbols for sequential numbers. Note: If Matchcode Type is set to Sequential, the imager will behave as if Number of Symbols were set to 1, regardless of the user-defined configuration.

Quadrus MINI Velocity User’s Manual

9-5

Page 260

Matchcode Type

Sequential Matching Usage: Definition: Serial Cmd: Default: Options:

Useful in tracking product serial numbers that increment or decrement sequentially. With Sequential enabled, Sequential Matching determines if a count is in ascending (incremental) or descending (decremental) order. <K223,matchcode type,sequential matching,match start position,match length,wild card character,sequence on No Read,sequence on mismatch> Increment 0 = Increment 1 = Decrement

Match Start Position Usage:

Match Start Position is useful in defining specific portions of a symbol for comparison. For example, if a symbol contains a part number, manufacturing date, and lot code info, but you are only interested in the part number information, you can set the imager to sort only the part number and ignore the other characters. Definition: Match Start Position determines the portions of symbols that will be matched by defining the first character in the symbol (from left to right) that will be compared with those of the master symbol, when Matchcode Type is set to Enabled or Sequential. Function: For example, if Match Start Position is set to 3, the first 2 characters read in the symbol will be ignored and only the 3rd and subsequent characters to the right will be compared, up to the number of characters specified by Match Length. Serial Cmd: <K223,matchcode type,sequential matching,match start position, match length,wild card character,sequence on No Read,sequence on mismatch> Default: 0 Options: 0 to 3000 Note: Match Start Position must be set to 1 or greater to enable this feature. A 0 setting will disable this feature.

Match Length Usage: Definition: Serial Cmd: Default: Options: 9-6

Example: If Match Length is set to 6 in a 10-character symbol, and Match Start Position is set to 2, only the 2nd through 7th characters (from left to right) will be compared. Defines the length of the character string that will be compared with that of the master symbol when Match Start Position is set to 1 or greater. When Match Start Position is set to 0, no comparison will occur. <K223,matchcode type,sequential matching,match start position,match length,wild card character,sequence on No Read,sequence on mismatch> 1 1 to 3000 Quadrus MINI Velocity User’s Manual