For the Leuze BCL-40

Page 78

Leuze electronic

Communicating with the Device Manual definition of the reference code Command

’RS’

Description

This command can be used to define a new reference code in the BCL 40 by means of direct entry via the serial interface. The data are saved in the parameter set according to their input under reference code 1 or 2 and stored in the working buffer for further processing.

Parameter

’RSyxxzzzzzzzz’ y, x and z are place holders (variables) for the actual input. y: def. reference code No ’1’ (Code 1) ’2’ (Code 2) … xx def. code type (see command "’CA’" on page 75) z def. code information (1 … 31 characters)

Acknowledgeno ment Example input = ’RS10678654331’ (Code 1 (1), UPC (06), 78654331)

Teach-In Command

’RT’

Description

This command enables a reference code to be defined quickly by reading an example label.

Parameter

’RTy’ y: Function ’1’ defines reference code 1 ’2’ defines reference code 2 ’-’ exit the Teach-In process

Acknowledgezzzzzz Contents of the decoded label ment Notice! With this function, only code types are recognised that are identified using the "autoConfig" function or which were set in the set-up. © After each reading via an ’RTx’ command, explicitly switch off the function again using the ’RT-’ command since failure to do so will interfere with other commands as well as prevent execution of a new ’RTx’ command.

76

BCL 40

Leuze electronic

Page 79

Leuze electronic

Communicating with the Device

Detect code Command

’CC’

Description

Reads the code type and contents of an unknown label without storing it in Setup.

Parameter no

Acknowledge- xxyyzzzzzz ment The acknowledgement is the same as that of the "autoConfig" function.

9.3.2

"Online" Commands for System Control Activate sensor input 1 Command

’+’

Description

Activates decoding (sensor input 1). A signal can be simulated on sensor input 1 with this command.

Parameter no

Acknowledgeno ment

Command

’-’

Description

Deactivates decoding (sensor input 1). A signal can be simulated on sensor input 1 with this command.

Parameter no

Acknowledgeno ment

Leuze electronic

BCL 40

77

TNT 35/7-24V

Deactivate sensor input 1

Page 80

Leuze electronic

Communicating with the Device Activate sensor input 2 Command

’,’

Description

Activates the definition of reference code 1 (sensor input 2). A signal can be simulated on sensor input 2 with this command.

Parameter no

Acknowledgeno ment

Deactivate sensor input 2 Command

’.’

Description

Deactivates the definition of reference code 1 (sensor input 2). A signal can be simulated on sensor input 2 with this command.

Parameter no

Acknowledgeno ment

9.3.3

"Online" Commands for System Checking Diagnosis laser Command

’DL’

Description

The command controls or tests the laser diode.

Parameter

’+’ ’-’ ’C’ ’T’ switches on the laser diode switches off the laser diode outputs the laser current outputs the laser temperature.

no ’DS03’: invalid parameter for the command ’DL’ Acknowledge’DS04’: invalid command length ment ’DTxxx.x’ ’DCyyy’ xxx.x is the laser temperature in °C yyy is the laser current in mA

78

BCL 40

Leuze electronic

Page 81

Leuze electronic

Communicating with the Device

Diagnosis motor Command

’DMS’

Description

Returns the rotary frequency of the polygon mirror wheel drive motor.

Parameter no

Acknowledge’DMSxxxxx’ ment Description xxxxx is the current rotary frequency of the motor in revolutions per minute.

Supply voltage Command

’DUV’

Description

Returns the supply voltage of the BCL 40.

Parameter no

Acknowledge’DUVxx.x’ ment xx.x is the current supply voltage in V.

TNT 35/7-24V

Description

Leuze electronic

BCL 40

79

Page 82

Leuze electronic

Communicating with the Device 9.3.4

"Online" Commands for Querying Statistical Data Query device statistics Command

’SR’

Description

Returns the current contents of the given field of the statistical device data memory.

Parameter

’00’ to ’19’

Acknowledge’ST00XXXXX’ to ’ST19XXXXX’ ment

Description

80

ST00XXXXX’: XXXXX is the length of time in seconds that the laser diode has been switched on since the last initialisation ’ST01XXXXX’: XXXXX is the length of time in minutes that the polygon wheel motor has been switched on since the last initialisation ’ST05XXXXX’: XXXXX the number of correctly received commands via the interface ’ST06XXXXX’: XXXXX the number of incorrectly received commands via the interface ’ST07XXXXX’: XXXXX the number of sent commands via the interface that have received positive acknowledgement ’ST08XXXXX’: XXXXX the number of sent commands via the interface that have received negative acknowledgement ’ST09XXXXX’: XXXXX the number of correctly received commands via the host ’ST10XXXXX’: XXXXX the number of incorrectly received commands via the host ’ST11XXXXX’: XXXXX the number of sent commands via the host that have received positive acknowledgement ’ST12XXXXX’: XXXXX the number of sent commands via the host that have received negative acknowledgement ’ST13XXXXX’: XXXXX is the number of scans since the last read enable ’ST14XXXXX’: XXXXX is the number of read scans since the last reset ’ST15XXXXX’: XXXXX is the number of read scans since the last read enable ’ST16XXXXX’: XXXXX the number of read labels from the last scan ’ST17XXXXX’: XXXXX is the number of read enables since the last reset ’ST18XXXXX’: XXXXX the number of correctly decoded labels since the last reset ’ST19XXXXX’: XXXXX the number of incorrectly decoded labels since the last reset

BCL 40

Leuze electronic

Page 83

Leuze electronic 9.3.5

Communicating with the Device

"Online" Commands for Parameter Set Operations Test parameter set Command

’PA’

Description

Tests the given parameter set for validity using a checksum.

Parameter

’0’ ’1’ tests the current parameter set tests the customer-specific parameter set (MA 10 only)

TNT 35/7-24V

’Psx’ x: Status '0' valid parameter set '1' invalid message '2' invalid message length '3' invalid block check type Acknowledge'4' invalid block check checksum ment '5' invalid data length '6' invalid message data '7' invalid start address '8' invalid parameter set '9' invalid parameter type

Leuze electronic

BCL 40

81

Page 84

Leuze electronic

Communicating with the Device Copy parameter set Command

’PC’

Description

Copies the given source parameter set into the given target parameter set

Parameter

BCL 40 Stand alone: ’0’ Current parameter set. ’2’ parameter set with factory default settings.

Example

’PC20’: Reset to factory default settings BCL 40 with MA 10: ’0’ Current parameter set. ’1’ customer-specific parameter set. ’2’ parameter set with factory default settings.

Example

’PC20’: Reset to factory default settings ’PC01’: The current parameter set is copied in the customer-specific parameter set (backup) ’PC10’: Reset to customer-specific settings

’PSx’ x: Status '0' valid parameter set '1' invalid message '2' invalid message length '3' invalid block check type Acknowledge'4' invalid block check checksum ment '5' invalid data length '6' invalid message data '7' invalid start address '8' invalid parameter set '9' invalid parameter type

Notice! A valid parameter set operation must be acknowledged with ’PS0’. Should an error occur, the command should be repeated. Repeated error acknowledgements indicate an error in the parameter set. If in doubt, contact your Leuze sales or service office (see back page for addresses).

82

BCL 40