Page 109
Power Supply BL-U1 5.4 Details of Command Communication All RS-232C and RS-422A serial communication commands can be used.
Communication Procedure
ID•••1 S TX
010
02
LON
7E
E
S TX
010
02
LON
7E
E
S TX
009
02
OK
3F
E
T X
ID•••2 T X
-Timing input turns ON.
T X
ID•••3
ID•••4
1. Send a command from the host computer to the BL series you wish to operate. [STX] mmm nn
Command cc
[ETX]
2. Data, including the same command, loops back to the host computer. [STX] mmm nn
Command cc
[ETX]
3. The specified BL series receives the command and performs the required operation. 4. The BL series sends back the following data in response to the command. [STX] mmm nn
Response cc
[ETX]
Response to the direct control command or the setting change command: OK Response to the setting confirmation command: Current settings If an error occurs with the specified command, the BL series sends back the following response. [STX] mmm nn
ERR** cc
[ETX]
** indicates an error code (see p. 70).
If incorrect characters are specified for the data communication commands (P, C), the BL series sends back ERR**.
97
Page 110
Power Supply BL-U1 Example:
When the command to shift to the setting mode (SSET) is sent to the BL series of ID No. 3:
1. From the host computer: [STX]
011
03
SSET
22
[ETX]
011
03
SSET
22
[ETX]
009
03
OK
3E
[ETX]
2. Loopback data: [STX]
3. From the BL series: [STX]
CAUTION
If an incorrect ID number is specified, an incorrect format is used to send a command, or a data transmission error occurs (checksum is incorrect), the BL series sends back no response. You will need to design a program so that the same command is sent again (Retry) if the BL series sends back no response within a specified period (Timeout). With RS-232C or RS-422A communication, the BL series sends back no data in response to some of the direct control commands (e.g. LOG<CR>). With the multidrop link, however, the BL series sends back the following data in response to those commands. During normal processing: OK When error occurs: ERR** ** indicates an error code (see p. 70).
98
Page 111
Power Supply BL-U1
6
Precautions for Programming • Since the multidrop link requires communication of many commands, the host computer must have high processing ability. As the number of connected BL series increases, the host computer needs higher data processing speed. When the host computer’s processing speed is insufficient, the BL series’ transmission buffer will become full, and the BL series will send back the following response. [STX]
011
03
OVER
22
[ETX]
The BL series clears all data stored in the transmission buffer. To prevent this, replace the host computer with a higher performance unit (including processing of the programming language), or reduce the number of the BL series connected to the host computer. • If an incorrect ID number is specified, an incorrect format is used to send a command, or a data transmission error occurs (checksum is incorrect), the BL series sends back no response. You will need to design a program so that the same command is sent again (Retry) if the BL series sends back no response within a specified period (Timeout).
99
Page 112
Power Supply BL-U1
BL-U1 Specifications Model
BL-U1
Power supply for bar code reader
5 VDC±5% (1.5 A)
Power supply for sensor
12 V ±10% (300 mA)
OK/NG output Trigger input
7
Input rating
8.5 to 30 VDC, 10 mA max.
Max. OFF current
1.0 mA
Output type
NPN open-collector
Rated load
30 VDC, 100 mA
Leakage current (at OFF)
0.1 mA max.
Residual voltage (at ON)
Environment
Interface Ambient temperature
0 to 50°C (32 to 122°F), No freezing
Relative humidity
35 to 85%, No condensation
Ambient atmosphere
No dust, no corrosive gas
Vibration
10 to 55 Hz, 1.5 mm double amplitude in X, Y and Z directions, two hours respectively
Noise immunity Insulation resistance
Rating
Insulation withstand voltage
1000 V p-p, 1 µs (Power line) At least 100 M across earth ground and AC power supply terminals (by 500 VDC Megger) 1500 VAC across earth ground and AC power supply terminals, for 1 minute
Power supply voltage
100 to 240 VAC (50/60 Hz)
Current consumption
40 VA (100 VAC), 50 VA (240 VAC)
Weight
100
1 V max. RS-232C, RS-422A, RS-485 multidrop (Up to 31 units can be connected. Max. cable extension: 1.2 km)
Approx. 615 g (including cable)
Page 113
Power Supply BL-U1 BL-U1
Approx. 40
100 to 240 VAC (with 3-core plug)
90
98 110
ø6.5, Cable length: 2 m
4 x ø5 mounting hole 150 186
4
4
120°
91 max.
51
101
Page 114
Power Supply BL-U1 When using the BL-U1
For a PC with a 9-pin D-sub RS-232C connector If the RS-232C interface of your personal computer uses a D-sub 9-pin connector, prepare a cable as described below. BL-U1 FG 1 SD 2 RD 3 RS 4 CS 5 SG 7 DR 6 ER 20 D-sub 25-pin (male) M 2.6 screw
PC – Connector case 2 RD 3 SD 4 ER 6 DR 5 SG 7 RS 8 CS D-sub 9-pin (female) #4-40 screw
Be sure to use a metal connector case for the 9-pin D-sub connector. Connect the connector case to the shield wire.
The BL-U1 can also be connected to a personal computer using a 25-pin to 9-pin adapter and a 25-pin to 25-pin null modem cable as described below. The optional OP-22149 null modem cable (1.5 m) is available from KEYENCE. 25-pin to 25-pin null modem cable OP-22149 (1.5 m)
BL-U1
102
25- to 9-pin adapter OP-25057
IBM-PC/AT compatible
Page 115
WARRANTIES AND DISCLAIMERS: (1) KEYENCE warrants the Products to be free of defects in materials and workmanship for a period of one (1) year from the date of shipment. If any models or samples were shown to Buyer, such models or samples were used merely to illustrate the general type and quality of the Products and not to represent that the Products would necessarily conform to said models or samples. Any Products found to be defective must be shipped to KEYENCE with all shipping costs paid by Buyer or offered to KEYENCE for inspection and examination. Upon examination by KEYENCE, KEYENCE, at its sole option, will refund the purchase price of, or repair or replace at no charge any Products found to be defective. This warranty does not apply to any defects resulting from any action of Buyer, including but not limited to improper installation, improper interfacing, improper repair, unauthorized modification, misapplication and mishandling, such as exposure to excessive current, heat, coldness, moisture, vibration or outdoors air. Components which wear are not warranted. (2) KEYENCE is pleased to offer suggestions on the use of its various Products. They are only suggestions, and it is Buyer's responsibility to ascertain the fitness of the Products for Buyer's intended use. KEYENCE will not be responsible for any damages that may result from the use of the Products. (3) The Products and any samples ("Products/Samples") supplied to Buyer are not to be used internally in humans, for human transportation, as safety devices or fail-safe systems, unless their written specifications state otherwise. Should any Products/Samples be used in such a manner or misused in any way, KEYENCE assumes no responsibility, and additionally Buyer will indemnify KEYENCE and hold KEYENCE harmless from any liability or damage whatsoever arising out of any misuse of the Products/Samples. (4) OTHER THAN AS STATED HEREIN, THE PRODUCTS/SAMPLES ARE PROVIDED WITH NO OTHER WARRANTIES WHATSOEVER. ALL EXPRESS, IMPLIED, AND STATUTORY WARRANTIES, INCLUDING, WITHOUT LIMITATION, THE WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE, AND NON-INFRINGEMENT OF PROPRIETARY RIGHTS, ARE EXPRESSLY DISCLAIMED. IN NO EVENT SHALL KEYENCE AND ITS AFFILIATED ENTITIES BE LIABLE TO ANY PERSON OR ENTITY FOR ANY DIRECT, INDIRECT, INCIDENTAL, PUNITIVE, SPECIAL OR CONSEQUENTIAL DAMAGES (INCLUDING, WITHOUT LIMITATION, ANY DAMAGES RESULTING FROM LOSS OF USE, BUSINESS INTERRUPTION, LOSS OF INFORMATION, LOSS OR INACCURACY OF DATA, LOSS OF PROFITS, LOSS OF SAVINGS, THE COST OF PROCUREMENT OF SUBSTITUTED GOODS, SERVICES OR TECHNOLOGIES, OR FOR ANY MATTER ARISING OUT OF OR IN CONNECTION WITH THE USE OR INABILITY TO USE THE PRODUCTS, EVEN IF KEYENCE OR ONE OF ITS AFFILIATED ENTITIES WAS ADVISED OF A POSSIBLE THIRD PARTY'S CLAIM FOR DAMAGES OR ANY OTHER CLAIM AGAINST BUYER. In some jurisdictions, some of the foregoing warranty disclaimers or damage limitations may not apply. BUYER'S TRANSFER OBLIGATIONS: If the Products/Samples purchased by Buyer are to be resold or delivered to a third party, Buyer must provide such third party with a copy of this document, all specifications, manuals, catalogs, leaflets and written information provided to Buyer pertaining to the Products/Samples.
103
Page 116
Specifications are subject to change without notice.
KEYENCE CORPORATION 1-3-14, Higashi-Nakajima, Higashi-Yodogawa-ku, Osaka, 533-8555, Japan AUSTRIA Phone: +43-2236-378266-0 BELGIUM Phone: +32 2 716 40 63 CANADA Phone: +1-905-696-9970 CHINA Phone: +86-21-68757500 CZECH REPUBLIC Phone: +420 222 191 483 FRANCE Phone: +33 1 56 37 78 00
PHONE: +81-6-6379-2211
GERMANY Phone: +49-6102-36 89-0 HONG KONG Phone: +852-3104-1010 HUNGARY Phone: +36 14 748 313 ITALY Phone: +39-2-6688220 JAPAN Phone: +81-6-6379-2211 KOREA Phone: +82-31-642-1270
MALAYSIA Phone: +60-3-2092-2211 MEXICO Phone: +52-81-8220-7900 NETHERLANDS Phone: +31 40 20 66 100 POLAND Phone: +48 71 36861 60 SINGAPORE Phone: +65-6392-1011 SLOVAKIA Phone: +421 2 5939 6461
SWITZERLAND Phone: +41 43 455 77 30 TAIWAN Phone: +886-2-2718-8700 THAILAND Phone: +66-2-369-2777 UK & IRELAND Phone: +44-1908-696900 USA Phone: +1-201-930-0100
A4WW1-MAN-0059
Copyright (c) 1995 KEYENCE CORPORATION. All rights reserved. 0103E 0059-2 96M1125 Printed in Japan