Keyence's instruction manual (AP3-IM-1-0398, 1998) for the AP-30 Series two-color digital display pressure sensors: AP-31, AP-32, AP-33 and AP-34, and their zero-shift Z types. It does not name the AP-32K or AP-34K it is filed under. It gives the specifications, wiring, piping and mounting, the four operation modes, setup steps, error indications and dimensions.

Device
Pressure sensor with a two-color digital display and two switching outputs p. 1
Pressure ranges
AP-31: 0 to -101.3 kPa; AP-32: 0 to +100.0 kPa; AP-33: 0 to +1.000 MPa; AP-34: +101.3 to -101.3 kPa p. 4
Fluids
Air or noncorrosive gases; gauge pressure p. 1
Power supply
12 to 24 VDC ±10%, ripple 10% max.; 50 mA at 24 V, 90 mA at 12 V p. 1
Control outputs
2 NPN open-collector outputs, 100 mA max. (40 V max.), N.O./N.C. selectable p. 1
Analog output
1 to 5 V (load impedance 47 kΩ min.); not on Z types p. 1
Zero-shift input
Z types only; non-voltage input, 20 ms or more p. 1
Display
3 1/2-digit, 2-color, 7-segment LED, 11 mm characters p. 1
Response time
2.5, 5, 100 or 500 ms, selectable p. 1
Repeatability
±0.2% of F.S. (5 ms or more) p. 1
Pressure port
Rc (PT) 1/8 internal thread; three selectable port positions p. 2
Wire colors
Brown 12 to 24 VDC, blue 0V, black control output 1, white control output 2 p. 1
Weight
Approx. 120 g including 2 m cable p. 1
Panel mount
Panel mounting holder set OP-31357, for panels 1 to 3.5 mm thick p. 2

What this manual walks you through

  • Wiring the outputs, zero-shift input and analog output p. 1
  • Piping and mounting the sensor with brackets A and B p. 2
  • Choosing auto-tuning, hysteresis, 2-independent or window mode p. 2, 3
  • Setting display units, N.O./N.C., response time and display color p. 3
  • Adjusting the zero point at atmospheric pressure p. 3
  • Using the zero-shift function for a leakage test p. 4
  • Showing and resetting peak-hold and bottom-hold values p. 4
  • Locking the front panel keys p. 4
  • Reading error indications and their remedies p. 4

Good to know

  • The manual states the AP-30 series does not have an explosion-proof structure and must not be used to detect flammable gases. p. 2
  • The hexagonal socket bolt takes 0.3 Nm at most and the port stopper 10 Nm at most. p. 2
  • Block an unused pressure port with the supplied hexagonal-socket port stopper. p. 2
  • Output voltage may drift ±1.0% right after power-up; allow 15 to 30 minutes warm-up for fine measurements. p. 3
  • The zero-shift value is lost when the power is turned off. p. 4

Questions this manual answers

What does FFF or -FFF mean on the AP-30 display?

The applied pressure is outside the display range, -15% to 110% of the rated pressure. p. 4

How do I zero a Keyence AP-30 sensor?

At normal atmospheric pressure, hold the AUTO key for at least 2 seconds in measurement mode; it works within ±5% of F.S. p. 3

What is the pressure range of the AP-32?

0 to +100.0 kPa; the AP-34 covers +101.3 to -101.3 kPa. p. 4

How do I unlock the front panel keys on an AP-30?

Hold down the AUTO key and press the down arrow; unL flashes for 2 seconds. p. 4

The file

Size and text

Pages4
PDF351 KB
TextText from the PDF itself.

Every page · 4

Read it here

Each page as it is printed, and its words beside it, so a phone can search and copy them.

Page 1 of Keyence AP-32K and AP-34K Manual

The text of page 1, from the PDF itself

SPECIFICATIONS Negative pressure AP-31(Z)

Compound pressure Model AP-32(Z) AP-33(Z) AP-34(Z) +101.3 to 0 to -101.3 kPa 0 to 100.0 kPa 0 to 1.000 MPa –101.3 kPa Rated pressure (0 to -760 (0 to 1 (0 to 10 (+760 mmHg mmHg) kgf/cm2) kgf/cm2) to -760 mmHg) 490 kPa 490 kPa 1.47 MPa 490 kPa Proof pressure (5 kgf/cm2) (5 kgf/cm2) (15 kgf/cm2) (5 kgf/cm2) Pressure type Gauge pressure Fluid types Air or noncorrosive gases Display 3 1/2-digit, 2-color, 7-segment LED (Character height: 11 mm) 0.1 kPa, 0.1 kPa, 0.001 Mpa, 0.2 kPa, Display resolution 1 mmHg, 0.001kgf/cm2, 0.01kgf/cm2, 2 mmHg, 0.02 Psi 0.02 Psi 0.2 Psi 0.04 Psi Detectable pressure range -15% to +110% of F.S. Repeatability ±0.2% of F.S. (5 ms or more) Response time (chattering 2.5/5/100/500 ms (selectable) prevention function) Non-voltage input (contact, solid-state), Zero-shift input 2. Input time: 20 ms or more NPN open-collector: 100 mA max. (40 V max.), Residual Control output voltage: 1 V max. 2-output (N.O./N.C. selectable) Analog output 3. 1 to 5 V (Load impedance: 47 kΩ min.) Temperature fluctuation ±2% max. (of F.S.) of detecting pressure at 25°C (0 to 50°C) for analog output Type

Two-color Digital Display Pressure Sensor

AP-30 Series Instruction Manual

1.

Temperature fluctuation for display Current consumption Power supply Ambient temperature Relative humidity

Positive pressure

±1% max. (of F.S.) of detecting pressure at 25°C (0 to 50°C) 50 mA (at 24 V), 90 mA (at 12 V) 12 to 24 VDC±10%, Ripple (p-p): 10% max. 0 to 50°C 35 to 85% 10 to 55 Hz, 1.5 mm double amplitude in X, Y, and Z directions, 2 hours respectively Front housing: Polyamide, Front panel sheet: PET, Rear housing: Polysulfone, Pressure port: Die-cast zinc, Cable: Oil-proof cabtyre cable

Vibration Material Weight (including 2 m cable)

Approx. 120 g

1. The zero-shift type sensor is suffixed with Z after the model name. 2. Z type only. 3. Not provided with Z type

  • ACCESSORIES
  • Instruction manual: 1 • Display unit Mounting Mounting label sheet: 1 bracket A: 1 bracket B: 1
  • Hexagonal-socket port stopper: 1
  • Quick reference sheet: 1

PART NAMES AND FUNCTIONS Output indicator 1 (Red LED)

  • Connections
  • Drive current load

Brown12 to 24 VDCBlack or white LoadA Blue

The AP-30 series enables you to select the display units for pressure. Attach the included display unit label for the desired units at the position in the figure.

Voltage outputBlue0V

  • Input/output circuit Brown

Load12 to 24 VDCLoad• Output circuitHexagonal socket bolt

Rear metal casing (Die-cast zinc)

  • Display unit label

Black or white

Black (Control output 1) White (Control output 2) Blue

0V

AP-31Z/32Z/33Z/34Z (Z type only)

AP-31/32/33/34 (Except for Z type)

Input circuit (Zero-shift input) Zero-shift input resets the display to “0” at the rising edge of the signal.

Analog output circuit12 to 24 VDC3.3 k Ω Pink 11 k ΩBlue

Pressure sensor main circuit

Housing

UP/DOWN key Use these keys to set output modes, or to change preset values or units.

12 to 24 VDC 4.7 k Ω

Over current protection circuit

Display unit label

Brown

Pressure sensor main circuit

Output indicator 2 (Green LED)

0V

  • Input to voltage input equipment

SET key Use this key to display or change preset values.

Pressure sensor main circuit

AUTO key In auto-tuning mode, use this key to detect pressure. In measurement mode, press this key for 2 seconds or more to adjust the zero-point.

CONNECTIONS AND INPUT/OUTPUT CIRCUIT

Protection circuitPinkBlueAnalog output (+)0V

Page 2 of Keyence AP-32K and AP-34K Manual

The text of page 2, from the PDF itself

PIPING/MOUNTING Hexagonal socket bolt

  • You can select from three pressure port positions by selecting or replacing the pressure ports. Select the optimal position based on your location.
  • The arrow shows the pressure port positions that can be selected.

CAUTION

The maximum tightening torque for the hexagonal socket bolt is 0.3 Nm (Approx. 3 kgf•cm). To avoid breakage, do not exceed the specified value.

  • An Rc (PT) 1/8 internal thread is provided on the pressure port of the AP-30 series. Commercially available air-pressure pipe joints or nipples can be used.
  • When connecting a joint or a plug, use a wrench (14 mm) to hold the metallic part as shown in the figure to avoid a large force being applied to the sensor housing (resin part).
  • Special mounting brackets are included with the AP-30 series. Use the type of mounting bracket appropriate for the location where the sensor is mounted. To mount the sensor, remove the hexagonal socket bolts and then retighten them through the mounting bracket.

Hexagonal socket bolts

CAUTION

The maximum tightening torque for the hexagonal socket bolt is 0.3 Nm (Approx. 3 kgf•cm). To avoid breakage, do not exceed the specified value.

SAFETY PRECAUTIONS Be sure to follow the instructions below to avoid malfunctions. CAUTION

Hold this part with a wrench (14 mm).

  • Be sure to block an unused pressure port with the hexagonal-socket port stopper provided. CAUTION

The maximum tightening torque for the hexagonal-socket port stopper is 10 Nm (Approx. 100 kgf•cm). To avoid breakage, do not exceed the specified value.

  • To prevent air leaks, wrap the male screw with sealing tape.
  • To mount the AP-30 series to a panel using the panel mounting holder set (OP-31357), use a panel with 1 to 3.5 mm thickness.
  • Connection
  • When using a commercially available switching regulator, be sure to ground the frame ground terminals.
  • Isolate the sensor’s wiring from power lines or high-voltage lines; otherwise, the sensor may malfunction due to noise interference.
  • Other
  • Do not use the AP-30 series for the detection of corrosive gases or liquid.
  • Do not insert any objects, such as wires, from the pressure port. The pressure-sensing element may break, resulting in malfunctions.
  • Do not press the front panel keys with a pointed object.
  • The AP-30 series does not have an explosion-proof structure. Do not use it for the detection of flammable gases.

OPERATION MODE SELECTION

  • Auto-tuning mode (F-1)
  • 2-independent mode (F-3)

Using the AUTO key, detect the upper limit value (A) and the lower limit value (b). The detection level (C) is automatically set at the midpoint between the two values. (You can finely adjust the preset value C within the range between A and b.)

Set two desired detection points (A and B).

Control output 1: The sensor turns on when the pressure exceeds the preset value C. Control output 2: The sensor turns on when the pressure goes outside the stability levels.

  • The stability levels are automatically set as shown in the following calculations.

A SH Preset value C SL b 0

SH =

Control output 1 ON (OUT 1) OFF Control output 2 ON (OUT 2) OFF

SL =

(A + C) 2 (C + b) 2

  • Hysteresis mode (F-2) Set desired detection level (H) and hysteresis (h) for the detection. Control output 1: The sensor turns on when the pressure exceeds the preset value H. When the pressure falls by the preset value h, the sensor turns off. Control output 2: The sensor turns on when the pressure goes outside the hysteresis width (H - h). Pressure Preset value H

0 Control output 1 ON (OUT 1) OFF Control output 2 ON (OUT 2) OFF

Pressure Preset value A Preset value b

Pressure

H-h

Control output 1: The sensor turns on when the pressure exceeds the preset value A. Control output 2: The sensor turns on when the pressure exceeds the preset value b.

h

h: Hysteresis width of OUT1

  • When h is set to a value close to 0, if pressure fluctuates around the detection point, OUT1 will chatter.

0 Control output 1 ON (OUT 1) OFF Control output 2 ON (OUT 2) OFF

  • Window mode (F-4) Set desired upper limit value (H) and lower limit value (L). Control output 1: The sensor turns off when the pressure goes outside of the range between the upper limit value (H) and lower limit value (L). Control output 2: The sensor turns off when the pressure goes outside of the stability levels. Pressure Preset value H SH SL Preset value L 0 Control output 1 ON (OUT 1) OFF Control output 2 ON (OUT 2) OFF
  • The stability levels are automatically set as shown in the following calculations. SH = H – SL = L +

(H - L) 4 (H - L) 4

Note 1: The above description shows the operation of control outputs 1 and 2 when the output selector switch is set to N.O. When the output selector switch is set to N.C., the operation of control outputs 1 and 2 is inverted. Note 2: Except for OUT1 in hysteresis mode, each control output includes an internal hysteresis of 0.5% of F.S.

Page 3 of Keyence AP-32K and AP-34K Manual

The text of page 3, from the PDF itself

ADJUSTMENT Measurement mode

▲▲▲▲

Press the button for 3 seconds or more.

  • Unit Setting
  • Preset Value Input Mode

Determine the desired units. In measurement mode, press for at least 3 seconds. “- - - -“ appears first, and then the current units are displayed. Use ▲ or completes the unit ▼ to select the desired units. Pressing setting procedure and enters operation mode selection.

Determine the preset values.

Std

AP-31/34: mmHg, AP-32/33: kgf/cm2

PA

AP-31/32/34: kPa, AP-33: MPa

  • When the units are changed, the preset values are automatically converted to appropriate values for the updated units.

psiPress the button once.▲ ■ Operation Mode

Determine the desired operation mode. (➮ Refer to “OPERATION MODE SELECTION” on page 2.) The current operation mode is displayed. Use ▲ or ▼ to select the operation mode. Pressing completes the operation mode setting procedure and enters N.O./N.C. selection.

F-1Auto-tuning modeF-2Hysteresis modeF-3

2-independent output mode

F-4Window modePress the button once.

▲

  • N.O./N.C. Selection Select N.O. (normally open) or N.C. (normally closed). The current selection of “no” (normally open) or “nc” (normally closed) is displayed. Use ▲ or ▼ to select the desired mode. Pressing completes the N.O./N.C. selection procedure and enters the chattering prevention setting.

Press the button once.

▲

  • Chattering Prevention Determine the desired response time. The current response time is displayed. Use ▲ or ▼ to select the response time. Pressing completes the setting procedure and enters the display color selection.

2.5 5 100 500

2.5 ms 5 ms 100 ms 500 ms

Press the button once.

▲

  • Display Color Selection Determine the desired LED color for numerical value display. The current color is displayed. Use ▲ or ▼ to select the color. completes the setting procedure and returns to the Pressing measurement mode.

1-[ 2-[

Press the button once.

  • Auto-tuning mode (F-1) (Measurement mode) 1. In measurement mode with the ❋❋❋❋ current measured value disCurrent value
  • •• played, press . The AP-30 ❋❋❋❋ enters the preset value input Flashes Current upper alternately mode. (lower) limit value A A ••• 2. “A” and the current preset value ❋❋❋❋ The updated value displays for 1 sec. flash alternately. ❋❋❋❋ 3. Position the target at the desired Flashes Current upper alternately upper (lower) limit. (lower) limit
  • A value b 4. Press A to register the value. ❋ ❋ ❋ ❋ The updated value displays for 1 sec. The updated value is displayed ❋❋❋❋ Flashes for 1 second. The set value alternately C displays. 5. “b” and the current preset value
  • •• A+b (C= 2 ) flash alternately. 6. Position the target at the desired lower (upper) limit. 7. Press A to register the value. The updated value is displayed for 1 second. 8. “C” and the calculated preset value C flash alternately. (You can change the C value to any value between A and b using ▲ or ▼ .) 9. Press to register the C value. The setting procedure is completed and the unit returns to measurement mode.
  • To confirm the preset value, press repeatedly.
  • Example of auto-tuning mode setting: Confirmation of work piece pick-up. Set the upper limit (A) to the position where the work piece is taken. Set the lower limit (b) to the position where the nozzle becomes open after releasing the work piece. Press A to register the upper and lower limit values. The C value is automatically set to the midpoint between the upper and lower limit values.
  • Hysteresis Mode (F-2), 2-independent Output Mode (F-3), Window Mode (F-4) [Example: In hysteresis mode] 1. In measurement mode with the (Measurement mode) current measured value displayed, ❋❋❋❋ press . The AP-30 enters the Current value
  • •• preset value input mode. Flashes 1. alternately 2. “H” 1. and the current preset value ❋❋❋❋ Set the desired value flash alternately. using or 3. Use ▲ or ▼ to change the value
  • •• Flashes 2. alternately to to the desired value. Press ❋❋❋❋ Set the desired value register the updated H value. using or 4. “h” 2. and the current preset value
  • •• Flashes 3. alternately flash alternately. ❋❋❋❋ 5. Use ▲ or ▼ to change the value 1. “A” appears in the 2-independent to the desired value. Press to output mode. register the updated h value. 2. “b” appears in the 2-independent 6. “P” 3. and the shift value of the output mode. “L” appears in the window mode. zero-shift adjustment flash 3. Shown with Z type only. alternately. 8. Press to complete the setting procedure and return to measurement mode.
  • To confirm the preset value, press repeatedly. Note 1: In hysteresis mode, (h - F.S.) cannot be set to a value greater than H. Note 2: In window mode, (L + 1% of F.S.) cannot be set to a value greater than H.

Red LED only Red/green LED

  • The setting is saved in the EEPROM.
  • Zero-point adjustment At normal atmospheric pressure (1 atm.), press A for at least 2 seconds in measurement mode. The display changes to “----“, then to “0”. The zero adjustment function can be used when the pressure is within ±5% of F.S.
  • The setting is saved in the EEPROM.

Note 1: When the operation mode is changed, check the preset values in the preset value input mode. Note 2: When units of psi are selected with the AP-34Z, the display range becomes 19.99 to 19.99. Note 3: Perform the zero-shift adjustment periodically. Note 4: The initial output voltage may fluctuate by ±1.0% immediately after the power is turned on. To measure minute differences in pressure, let the sensor warm up for approximately 15 to 30 minutes.

Page 4 of Keyence AP-32K and AP-34K Manual

The text of page 4, from the PDF itself

OTHER FUNCTIONS AND ERROR INDICATION

  • Zero-shift function (Z type only)
  • Peak-hold/bottom-hold display function

The zero-shift function is used to reset the current pressure value to “0” using an external signal input, in order to prevent measurements from being affected by fluctuations in base pressure.

The AP-30 series internally updates the peak-hold and bottom-hold values at all times.

  • To display hold values
  • While ▲ is held down in measurement mode, the peak-hold value is displayed.
  • While ▼ is held down in measurement mode, the bottom-hold value is displayed.

Example: Leakage test Input a zero-shift value after air supply is completed so that air leakage after a specified time is displayed as a negative value. The AP-30’s detection is unaffected by fluctuations in air supply volume. Zero-shift input 1 MPa

Zero-shift input -0.1 MPa is acceptable.

  • To reset the peak-hold and bottom-hold values
  • Hold down ▲ and press ▼ in measurement mode.

-0.6 MPa is not acceptable.

  • The peak-hold and bottom-hold values are also reset using the following procedure.
  • Turn the power off.
  • Press for 3 seconds or more and change any settings.

Atmospheric pressure

When the power is turned off, the value updated after the zero-shift input (zero-shift value) is lost.

Note: The hold values cannot be displayed when the front panel keys are locked with the key protection function. Disable the function before displaying the hold values.

Note 1: The zero-shift function cannot be used in auto-tuning mode. Note 2: The zero-shift input is effective when the current pressure is between -3% of F.S. and F.S. for a shift of 0 (P = 0). Note 3: If the applied pressure is outside the range of -15% to 110% of the rated pressure, “-FFF” or “ FFF” appears.

  • Analog output function (Except for Z type) The voltage value according to the pressure value is output.
  • Key protection function The key protection function is used to lock the front panel key in order to prevent preset values from being accidentally changed. To enable the key protection function, hold down A and press ▲ . “Loc” flashes for 2 seconds and the keys are locked. To disable the key protection function, again hold down A and press ▼ . “unL” flashes for 2 seconds and the keys are unlocked.

Model1 V to 5 VAP-310 to -101.3 kPaAP-320 to +100.0 kPaAP-330 to +1.000 MPaAP-34+101.3 to -101.3 kPa

  • Error indications and remedies Error indication

Using the EEPROM, the AP-30 series can retain the preset values even if the power is turned off.

Remedy

Zero-point adjustment was Perform zero-point adjustment executed at a pressure of at normal atmospheric ±5% or more of F.S. pressure.

Ec

Overcurrent through OUT1 Turn power off and adjust the or 2 load so that the current is within the rated range.

-FFF, FFF

Applied pressure was outside of the display range.

  • Display color selection You can set the color of the LED display either to the two-color mode which displays the numerical value in green or red according to OUT1, or to the single color mode which always shows the value in red. The two-color display allows you to check the output condition at a glance. (Refer to “ADJUSTMENT” on page 3 for the setting procedure.)

Problem

E

Adjust the pressure to within the rated range.

  • N.O./N.C. selection The N.O. or N.C. output can be selected according to the device’s control method. When the output status is changed, the color of the numerical value display LED is inverted.

In two-color mode (Regardless of N.O./N.C. selection)

  • When OUT1 is turned on: Red
  • When OUT1 is turned off: Green
  • Chattering prevention function The chattering prevention function is used to prevent outputs from chattering by changing the response time. The response time can be selected from 4 settings. When the detection (non-detection) state continues for more than a preset response time, the output is produced.

DIMENSIONS AP-30 (Sensor unit)

2031312 x M3 screw148.5

When mounting bracket A is attached [Top view] 2-ø4.5 [Side view] mounting hole

25332024.5616.5 6.5(4)

Cable length: 2 m ø4, 5 x 0.18 mm2 core

45

t=1.6

When mounting bracket B is attached

30 2030 20 14142-ø4.52-ø3.51331311345

KEYENCE FRANCE S.A. PHONE: 01 47 92 76 76 FAX: 01 47 92 76 77 KEYENCE SINGAPORE PTE LTD PHONE: 392-1011 FAX: 392-5055

45.5

KEYENCE (THAILAND) CO., LTD PHONE: 02-934-6772~4 FAX: 02-934-6775

19 8.6 20 45

KEYENCE DEUTSCHLAND GmbH PHONE: 0711-7973710 FAX: 0711-7977799

KEYENCE SENSORS & MEASURING SDN BHD PHONE: 03-252-2211 FAX: 03-252-2131

20

2-ø3.5

KEYENCE CORPORATION OF AMERICA PHONE: 201-930-0100 FAX: 201-930-0099 KEYENCE (UK) LIMITED PHONE: 01908-696900 FAX: 01908-696777

Mounting bracket B (Accessory)

1625 3522

Pressure port: Rc (PT) 1/8

6.5

Mounting bracket A (Accessory) 22

31

16

Worldwide Headquarters KEYENCE CORPORATION 1-3-14, Higashi-Nakajima, Higashi-Yodogawa-ku, Osaka, 533-8555, Japan PHONE: 81-6-379-2211 FAX: 81-6-379-2131

202-ø4.5t=1.5t=1.6 20

KEYENCE KOREA CORPORATION PHONE: 02-563-1270 FAX: 02-563-1271 © KEYENCE CORPORATION, 1998 AP3-IM-1-0398 Printed in Japan

19 t=1.5