Page 46
1. 2. 3. 4. 5. 6. 7. 8. 9.
System will operate in Edge Sync or Envelope sync mode (~LX#). Sync polarity programmable (~LP#) Use ~PR0200 for proper port initialization. Port 1 will pulse active ON for a bar code quality or data error condition. Port 2 will pulse active ON for a NO READ, partial or low % decode error condition. Port 3 will pulse active ON for a non-error condition. Port pulse times programmable. (~PP801xxx) Pulses act like re-triggerable 1-shot logic Pushing the reset button or power re-cycle will place Ports 1, 2 and 3 in their inactive states and reset the # codes per sync counter. 10. Error conditions available are: a. Partial Read (Programmable, ( ~Lp#, LQ#) b. % decode (Programmable, ~LD##) c. Bad Quiet Zone: < 10X (~PB816xxx100) d. ANSI Defects grade (Programmable, ~PB806xxx000) e. ANSI Decodability grade (Programmable, ~PB802xxx100) f. Symbol Contrast grade (Programmable, ~PB804xxx100) g. Overall ANSI Grade (Programmable, ~LAxx) h. No Read (if sync is received) i. Number of codes per sync programmable (~LN##, ~LZ##) j. X dimension Range (Programmable ~PB811xxxyyy) k. Ratio (Programmable ~PB815xxx000) l. Symbology Mod Check data error m. I 2 of 5 optional Mod 10 Check Digit data error (if enabled) n. Data Match Error (~BC – see Operator’s Guide) o. Increment or Decrement data Error (~Br – see Operator’s Guide)
B. LED Operation 1.
LED1 will turn on when Port 1 is active due to ANSI, Symbol Contrast, X dimension, ratio, mod check or quiet zone error conditions. 2. LED1 will flash when Port 1 is active due to a data increment or decrement error condition. 3. LED2 will turn on when Port 2 is active due to a partial, % decode or no read error conditions. 4. LED2 will flash when Port 1 is active due to a Data Match error condition. 5. Flashing on indication will override a solid indication if both types of error conditions occur. Note; Both LEDs can be on in cases where multiple bar codes are analyzed in a sync period.
C. SV Commands Important for Mode 17 This system is operating in a “mode” rather than fully programmable logic. The port activation parameters are programmable via the ~PB8 rather than ~PB1, ~PB2, etc. for each individual port. Other commands, such as mode commands are also available. The commands most useful for this system application are described below. Please see the SV Series Operator’s Guide for additional command description details. ~LV17 This command sets this mode of operation. ~Lp0 This command turns off all partial decode logic at the decoder level ~Lp1 This command turns on partial decode logic at the decoder level
43
Page 47
~LQ1 This command allows partial decodes to activate the output ports ~LQ0 This command disables partial decodes to activate output ports. ~LAxx This command sets the passing threshold for the Overall ANSI method Symbol Grade. For example: ~LA28 causes an Overall Symbol Grade of 2.7 or lower calculated for the code analyzed to set a failure condition for this parameter. ~LDxx This command sets the % decode threshold for a passing condition. If analyzing poorly printed symbols, it is recommended that partial decodes be enabled at the decoder level when this command is enabled. xx = failure threshold. Example ~LD75: if 74 % or less of scans on a code were not fully decoded, this sets a failure for this parameter. ~PP801xxx This command sets the “active” time for the port pulses. The value “xxx” sets the time in .1 second intervals. Any value greater than 100 (10 seconds) creates a latching output. ~PB816xxx100 This command sets the minimum percent of scans on a code which calculate good quiet zones to determine an acceptable quiet zone analysis. The field xxx is the minimum passing threshold. Example: if xxx = 030, a minimum of 30% of all fully decoded scans on a code must calculate a good quiet zone, or a failure condition is set for this parameter. ~PB806xxx000 This command sets the threshold for the ANSI Defects calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 025, a Defects analysis of 26% or higher will cause a failure condition for this parameter. ~PB802xxx100 This command sets the threshold for the ANSI Decodability calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 037, a Decodability analysis of 36% or lower will cause a failure condition for this parameter. ~PB804xxx100 This command sets the threshold for the ANSI Symbol Contrast calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 020, a Symbol Contrast analysis of 19% or lower will cause a failure condition for this parameter. ~PB815xxx000 This command sets the threshold for the ratio calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 025, a ratio analysis of 2.4 or lower will cause a failure condition for this parameter. ~PB811xxxyyy This command sets the threshold for acceptable X dimension ranges. The value for xxx is the narrowest acceptable X dimension. The value yyy is the widest acceptable X dimension. Units are in .001 inch (mils). For example: ~PB811010020 will cause a failure for an X dimension less than 10 mils or greater than 20 mils. ~LN##
44
Page 48
This command sets the minimum number of codes to be read during a sync period. ## = the number of codes. For example: ~LN02 causes a No Read condition to be set if less than 2 bar codes are fully decoded during a sync interval. See the ~LZ## for priority. If the value ## = 00 (and ~LZ00 is also set), a sync input is disregarded, therefore a No Read condition is disabled. ~LZ## This command sets the exact number of codes to be read during a sync period. ## = the number of codes. For example: ~LZ02 causes a No Read condition to be set if exactly 2 bar codes are not fully decoded during a sync interval. This command overrides the ~LN## command if the number of codes set does not = 00. If ~LZ## is set to ~LZ00, the ~LN## command takes precedence. ~Hx This command stores all parameters into non-volatile memory. This command should be the last command sent after parameters are programmed via the above commands.
Mode 18 (~LV18) A. I/O operation 1. System will operate Envelope or Edge sync mode (~LX#) 2. Sync polarity – programmable (~LP#) 3. Use ~PR0210 for proper port initialization. 4. Port 1 will go active OFF (no current) on an error condition. 5. Port 2 will go active ON (sink current) on an error condition. 6. Port 3 will operate in per analysis pulse mode. Active polarity ON (sink current), approximately 500 millisecond duration. Port 3 will activate (pulse) on an error condition.
7. Port 4 will operate in per analysis pulse mode. Active polarity ON (sink current), approximately 50 millisecond duration. Port 4 will activate (pulse) on an error condition. 8. Port 5 will pulse active ON for an acceptable condition (absence of error condition.) 9. Port 5 pulse time programmable (~PP801xxx). Pulses will act like a re-triggerable 1-shot. 10. Pushing the reset button or power re-cycle will place Ports 1- 5 in their inactive states. 11. Error conditions available are: c. Partial Read (Programmable, ( ~Lp#, LQ#) d. % decode (Programmable, ~LD##) e. Bad Quiet Zone: < 10X (~PB816xxx100) f. ANSI Defects grade (Programmable, ~PB806xxx000) g. ANSI Decodability grade (Programmable, ~PB802xxx100) h. Symbol Contrast grade (Programmable, ~PB804xxx100) i. Overall ANSI Grade (Programmable, ~LAxx) j. No Read (if sync is received) k. Number of codes per sync programmable (~LN##, ~LZ##) l. X dimension Range (Programmable ~PB811xxxyyy) m. Ratio (Programmable ~PB815xxx000) n. Symbology Mod Check data error o. I 2 of 5 optional Mod 10 Check Digit data error (if enabled) p. Data Match Error (~BC – see Operator’s Guide) q. Increment or Decrement data error (~Br – see Operator’s Guide)
45
Page 49
B. LED Operation 1.
LED1 will turn on when Ports 1 and 2 are active due to ANSI, Symbol Contrast, X dimension, ratio, mod check or quiet zone error conditions. 2. LED1 will flash when Ports 1 and 2 are active due to a data increment or decrement error condition. 3. LED2 will turn on when Ports 1 and 2 are active due to a partial, % decode or no read error conditions. 4. LED2 will flash when Ports 1 and 2 are active due to a Data Match error condition. 5. Flashing indication will override a solid on indication if both types of error conditions occur. Note; Both LEDs can be on in cases where multiple bar codes are analyzed in a sync period.
C. SV Commands Important to this System This system is operating in a “mode” rather than fully programmable logic. The port activation parameters are programmable via the ~PB8 rather than ~PB1, ~PB2, etc. for each individual port. Other commands, such as mode commands are also available. The commands most useful for this system application are described below. Please see the SV Series Operator’s Guide for additional command description details. ~LV18 This command sets this mode of operation. ~Lp0 This command turns off all partial decode logic at the decoder level ~Lp1 This command turns on partial decode logic at the decoder level ~LQ1 This command allows partial decodes to activate the output ports ~LQ0 This command disables partial decodes to activate output ports. ~LAxx This command sets the passing threshold for the Overall ANSI method Symbol Grade. For example: ~LA28 causes an Overall Symbol Grade of 2.7 or lower calculated for the code analyzed to set a failure condition for this parameter. ~LDxx This command sets the % decode threshold for a passing condition. If analyzing poorly printed symbols, it is recommended that partial decodes be enabled at the decoder level when this command is enabled. xx = failure threshold. Example ~LD75: if 74 % or less of scans on a code were not fully decoded, this sets a failure for this parameter. ~PP801xxx This command sets the “active” time for the port pulses. The value “xxx” sets the time in .1 second intervals. Any value greater than 100 (10 seconds) creates a latching output. ~PB816xxx100 This command sets the minimum percent of scans on a code which calculate good quiet zones to determine an acceptable quiet zone analysis. The field xxx is the minimum passing threshold. Example: if xxx = 030, a minimum of 30% of all fully decoded scans on a code must calculate a good quiet zone, or a failure condition is set for this parameter. ~PB806xxx000
46
Page 50
This command sets the threshold for the ANSI Defects calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 025, a Defects analysis of 26% or higher will cause a failure condition for this parameter. ~PB802xxx100 This command sets the threshold for the ANSI Decodability calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 037, a Decodability analysis of 36% or lower will cause a failure condition for this parameter. ~PB804xxx100 This command sets the threshold for the ANSI Symbol Contrast calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 020, a Symbol Contrast analysis of 19% or lower will cause a failure condition for this parameter. ~PB815xxx000 This command sets the threshold for the ratio calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 025, a ratio analysis of 2.4 or lower will cause a failure condition for this parameter. ~PB811xxxyyy This command sets the threshold for acceptable X dimension ranges. The value for xxx is the narrowest acceptable X dimension. The value yyy is the widest acceptable X dimension. Units are in .001 inch (mils). For example: ~PB811010020 will cause a failure for an X dimension less than 10 mils or greater than 20 mils. ~LN## This command sets the minimum number of codes to be read during a sync period. ## = the number of codes. For example: ~LN02 causes a No Read condition to be set if less than 2 bar codes are fully decoded during a sync interval. See the ~LZ## for priority. If the value ## = 00 (and ~LZ00 is also set), a sync input is disregarded, therefore a No Read condition is disabled. ~LZ## This command sets the exact number of codes to be read during a sync period. ## = the number of codes. For example: ~LZ02 causes a No Read condition to be set if exactly 2 bar codes are not fully decoded during a sync interval. This command overrides the ~LN## command if the number of codes set does not = 00. If ~LZ## is set to ~LZ00, the ~LN## command takes precedence. ~Hx This command stores all parameters into non-volatile memory. This command should be the last command sent after parameters are programmed via the above commands.
Mode 19 (~LV19) A. I/O operation 1. Envelope type sync input on pin 4 2. Port 3: per sync type. The state of this output must be stable within 12.5 milliseconds after the sync input goes low (trailing edge.) Typical time for stability will be less than 7.5 milliseconds if the sync is received when a bar code or complex graphic is not in the beam. The port shall go active (sink current) for the following error conditions; a. Partial Read (Programmable, ( ~Lp#, LQ#) b. % decode (Programmable, ~LD##)
47
Page 51
c. Bad Quiet Zone: < 10X (~PB816xxx100) d. ANSI Defects grade (Programmable, ~PB806xxx000) e. ANSI Decodability grade (Programmable, ~PB802xxx100) f. Symbol Contrast grade (Programmable, ~PB804xxx100) g. Overall ANSI Grade (Programmable, ~LAxx) h. No Read (if sync is received) i. Number of codes per sync programmable (~LN##, ~LZ##) j. X dimension Range (Programmable ~PB811xxxyyy) k. Ratio (Programmable ~PB815xxx000) l. Symbology Mod Check data error m. I 2 of 5 optional Mod 10 Check Digit data error (if enabled) n. Data Match Error (~BC – see Operator’s Guide) o. Increment or Decrement data error (~Br – see Operator’s Guide) 3. 4.
5. 6.
Sync polarity - Leading edge high going; time between trailing (low going) edge to leading edge - 12.5 milliseconds minimum Port 1; acts as strobe (signal ensuring Port 3 condition is stable) in response to trailing edge of sync input. This signal will go “on” (sink current) between 20 and 100 microseconds after Port 3 condition is stable. Ports 1 and 3 will go high (no current) within 12.5 milliseconds after the sync input goes high (leading edge). Use ~PR0200 for proper port initialization.
B. LED Operation 1.
LED1 will turn on when Port 3 is active due to ANSI, Symbol Contrast, X dimension, ratio, mod check or quiet zone error conditions. 2. LED1 will flash when Port 3 is active due to a data increment or decrement error condition. 3. LED2 will turn on when Port 3 is active due to a partial, % decode or no read error conditions. 4. LED2 will flash when Port 3 is active due to a Data Match error condition. 5. Flashing indication will override a solid on indication if both types of error conditions occur. Note; Both LEDs can be on in cases where multiple bar codes are analyzed in a sync period.
C. SV Commands Important to this System This system is operating in a “mode” rather than fully programmable logic. The port activation parameters are programmable via the ~PB8 rather than ~PB1, ~PB2, etc. for each individual port. Other commands, such as mode commands are also available. The commands most useful for this system application are described below. Please see the SV Series Operator’s Guide for additional command description details. ~LV19 This command sets this mode of operation. ~Lp0 This command turns off all partial decode logic at the decoder level ~Lp1 This command turns on partial decode logic at the decoder level ~LQ1 This command allows partial decodes to activate the output ports ~LQ0 This command disables partial decodes to activate output ports. ~LAxx
48
Page 52
This command sets the passing threshold for the Overall ANSI method Symbol Grade. For example: ~LA28 causes an Overall Symbol Grade of 2.7 or lower calculated for the code analyzed to set a failure condition for this parameter. ~LDxx This command sets the % decode threshold for a passing condition. If analyzing poorly printed symbols, it is recommended that partial decodes be enabled at the decoder level when this command is enabled. xx = failure threshold. Example ~LD75: if 74 % or less of scans on a code were not fully decoded, this sets a failure for this parameter. ~PB816xxx100 This command sets the minimum percent of scans on a code which calculate good quiet zones to determine an acceptable quiet zone analysis. The field xxx is the minimum passing threshold. Example: if xxx = 030, a minimum of 30% of all fully decoded scans on a code must calculate a good quiet zone, or a failure condition is set for this parameter. ~PB806xxx000 This command sets the threshold for the ANSI Defects calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 025, a Defects analysis of 26% or higher will cause a failure condition for this parameter. ~PB802xxx100 This command sets the threshold for the ANSI Decodability calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 037, a Decodability analysis of 36% or lower will cause a failure condition for this parameter. ~PB804xxx100 This command sets the threshold for the ANSI Symbol Contrast calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 020, a Symbol Contrast analysis of 19% or lower will cause a failure condition for this parameter. ~PB815xxx000 This command sets the threshold for the ratio calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 025, a ratio analysis of 2.4 or lower will cause a failure condition for this parameter. ~PB811xxxyyy This command sets the threshold for acceptable X dimension ranges. The value for xxx is the narrowest acceptable X dimension. The value yyy is the widest acceptable X dimension. Units are in .001 inch (mils). For example: ~PB811010020 will cause a failure for an X dimension less than 10 mils or greater than 20 mils. ~LN## This command sets the minimum number of codes to be read during a sync period. ## = the number of codes. For example: ~LN02 causes a No Read condition to be set if less than 2 bar codes are fully decoded during a sync interval. See the ~LZ## for priority. If the value ## = 00 (and ~LZ00 is also set), a sync input is disregarded, therefore a No Read condition is disabled.
~LZ## This command sets the exact number of codes to be read during a sync period. ## = the number of codes. For example: ~LZ02 causes a No Read condition to be set if exactly 2 bar codes are not fully decoded during a sync interval.
49
Page 53
This command overrides the ~LN## command if the number of codes set does not = 00. If ~LZ## is set to ~LZ00, the ~LN## command takes precedence. ~Hx This command stores all parameters into non-volatile memory. This command should be the last command sent after parameters are programmed via the above commands.
50
Page 54
ANSI PARAMETER GRADE THRESHOLDS ANSI parameter fail thresholds are set in the SV Series by commands requiring numeric settings that correlate to the parameter calculations. Below are descriptions of how the numeric calculations are divided into letter grades per the ANSI Bar Code Quality Specification Rmin A < .5 x Rmax F > .5 x Rmax Symbol Contrast A > 70% B > 55% C > 40% D > 20% F < 20% Minimum. Edge Contrast A > 15% F < 15% Modulation A < .70% B < .60% C < .50% D < .40% F > .40% Decodability > .62 = A > .50 = B > .37 = C > .25 = D < .25 = F Defects A < .15% B < .20% C < .25% D < .30% F > .30% Conversion of letter scan grade to numeric grade A=4, B=3, C=2, D=1, F=0 Conversion of numeric symbol grade to letter symbol grade (used in averaging) 3.5 < A < 4.0 2.5 < B < 3.5 1.5 < C < 2.5 0.5 < D < 1.5 F < 0.5 Reference Decode A = ANSI method algorithm decoded the symbol F = ANSI method algorithm could not decode the symbol
51