Celeritous' two-page spec sheet for the LPD2001 Laser Position Detector, an all-digital 2D instrument that reads quadrant, duolateral and tetralateral detectors for beam position, centering and alignment work. It lists the features, outputs and power, how channels, amplifier pods and detectors fit together, and the order numbers.

Device
Fully integrated, all-digital 2D laser position instrument with detector electronics, display and analog output p. 1
Detectors
Quadrant, 2D duolateral and 2D tetralateral detectors p. 1
Channels
1 or 4 channels: LPD2001-1 single channel, LPD2001-4 four channel p. 1, 2
Resolution
10 bit, all digital with no analog dividers p. 1
Analog output
Normalized: X and Y ± 1 V, SUM 0-1V p. 1
Pulsed sample rate
800Hz with 1 channel, 200Hz with 4 channels p. 1
Power
Universal AC supply, 85-220Vac 50/60 Hz p. 1
Display
Large backlit LCD showing position in mm; adjustable backlight and contrast p. 1
Gain
Adjustable 1-100 on each channel; X, Y output inversion p. 1
Amplifier pods
LPD2001-Q (quad), LPD2001-T (tetralateral), LPD2001-D (duolateral) p. 2
Detectors sold
LPD2001-10MQ and -3MQ quad cells, LPD2001-10MT and -3MT tetralateral (10mm or 3mm) p. 2
Pod cables
5' cable standard, 10' available; detector to pod 1' cable p. 2
Detector housing
1" x 1" x 2" plastic enclosure, DIN cable, 8-32 threaded stud p. 2
Beam modes
CW or pulsed; background subtraction in pulsed mode p. 1

What this sheet walks you through

  • Matching amplifier pods and detectors to each channel p. 2
  • Choosing the sync edge for pulsed beams p. 2
  • Ordering the instrument, pods and detectors p. 2

Good to know

  • All channels must use the same type and size of detector. p. 2
  • All channels run in the same mode, either CW or pulsed. p. 2
  • The instrument and detectors are sold separately; existing detectors connect through the matching pod. p. 2

Questions this sheet answers

What detectors work with the Celeritous LPD2001?

Quadrant, 2D duolateral and 2D tetralateral detectors, each through an amplifier pod for that detector family. p. 1, 2

How fast does the LPD2001 sample pulsed beams?

800Hz with one channel and 200Hz with four channels. p. 1

Can I use my own detectors with the LPD2001?

Yes. Users may connect existing detectors through the appropriate detector pod. p. 2

What is the difference between the LPD2001-1 and LPD2001-4?

The LPD2001-1 is the single channel instrument and the LPD2001-4 the four channel one. p. 2

The file

Size and text

Pages2
PDF198 KB
TextText from the PDF itself.

Every page · 2

Read it here

Each page as it is printed, and its words beside it, so a phone can search and copy them.

Page 1 of Celeritous LPD2001 Spec Sheet

The text of page 1, from the PDF itself

CELERITOUS DESIGNING THE FUTURE

LPD2001 Laser Position Detector

A fully integrated, all digital 2D Laser position instrument. Uses Include: Precision Beam Position Measurement Beam Parameter measurement Beam Centering / Boresighting Automatic Beam Alignment Surface Deflection Measurements

Compatible with: Quadrant Detectors 2D Duolateral Detectors 2D Tetralateral Detectors

Fully Integrated Detector Electronics, Display, Analog Output, Multi-Channel, CW or Pulsed Beam All-in-one compact unit. Celeritous Technical Services, Corp 5109 82nD St #7 PMB1200 Lubbock, TX 79424 http://www.celeritous.com

Features Include: All Digital – No Analog Dividers 10 Bit resolution 1 or 4 channels Digital position display in mm Normalized Analog Output X - ± 1 V into 50 W Y - ± 1 V into 50 W SUM 0-1V into 50 W CW or Pulsed Beams Background subtraction in Pulsed Mode Pulsed sample rate 1 Channel - 800Hz 4 Channel - 200Hz Large, Backlit LCD Display Adjustable Display Backlight, Contrast X, Y output inversion Adjustable 1-100 gain on each channel Universal AC Supply (85-220Vac 50/60 Hz) 800-687-6510 806-793-0708 FAX 806-793-0710 sales@celeritous.com

Page 2 of Celeritous LPD2001 Spec Sheet

The text of page 2, from the PDF itself

CELERITOUS DESIGNING THE FUTURE

Configuration Notes 1. Currently all channels must use the same type and size of detectors (quad, duolateral or tetralateral). 2. Each detector must have an amplifier pod configured for that detector family. Pods come with a 5’ cable standard. A 10’ cable is also available. Detectors connect to the amplifier pods with a 1’ cable. 3. All channels are either in CW or Pulsed mode. User may select rising or falling edge of Sync Input pulse for pulsed beam presence. Background light is sampled on opposite edge. 4. User may provide their own detectors or purchase detectors. Detectors are housed in a 1”x 1”x 2” Plastic enclosure with DIN cable and 8-32 threaded stud for standard optical posts.

Celeritous Technical Services, Corp 5109 82nD St #7 PMB1200 Lubbock, TX 79424 http://www.celeritous.com

Ordering Information: Instrument Unit and detectors are sold separately. Users may connect to existing detectors with appropriate detector pod. LPD2001-1 Single Channel Instrument LPD2001-4 Four Channel Instrument1 Amplifier Pods LPD2001-Q for Quad Detectors LPD2001-T for 2D Tetralateral Detectors LPD2001-D for 2D Duolateral Cells Detectors LPD2001-10MQ 10mm Quad Cell (± 5mm) LPD2001-3MQ 3mm Quad Cell (± 1.5mm) LPD2001-10MT 10mm Tetralateral (± 5mm) LPD2001-3MT 3mm Tetralateral (± 1.5mm) Other detector types and sizes available upon request.

800-687-6510 806-793-0708 FAX 806-793-0710 sales@celeritous.com